The Rüsch Ultra Tracheoflex® Tracheostomy Set-Cuffed is a tracheostomy tube with accessories, intended for intubation through a surgical incision into the trachea, for the purpose of providing a patent airway.
Device Story
Sterile, single-use cuffed tracheostomy tube with fixed flange; includes radiopaque tube, obturator, cough cap, 15mm airway connector, inner cannulas, neck band. Used in clinical settings for airway management. Device provides patent airway via surgical tracheal incision. Features spring (Bespack) valve for cuff pressure maintenance and ISO standard screw-luer fitting for secure connection to airway accessories. Physician-operated.
Clinical Evidence
No clinical data; bench testing only.
Technological Characteristics
Radiopaque tracheostomy tube; materials meet ASTM F1242-89. Includes spring (Bespack) valve for cuff inflation and ISO standard screw-luer fitting. Complies with ISO 5366-1 (connectors), ISO 5366-2 (adult tracheostomy tubes), ISO 5356 (conical connectors), and ISO 5361 (cuffs/tubes). Sterile, disposable, single-use.
Indications for Use
Indicated for patients requiring a patent airway via surgical tracheostomy incision.
Regulatory Classification
Identification
A tracheostomy tube and tube cuff is a device intended to be placed into a surgical opening of the trachea to facilitate ventilation to the lungs. The cuff may be a separate or integral part of the tracheostomy tube and is, when inflated, intended to establish a seal between the tracheal wall and the tracheostomy tube. The cuff is used to prevent the patient's aspiration of substances, such as blood or vomit, or to provide a means for positive-pressure ventilation of the patient. This device is made of either stainless steel or plastic.
K031553 — SILICONE FOAM CUFF TRACHEOSTOMY TUBES, SILICONE CUFFLESS NEONATAL AND PEDIATRIC TRACHEOSTOMY TUBES · Arcadia Medical Corporation · Nov 18, 2003
K251313 — Shiley Adult Flexible Tracheostomy Tube XLT with TaperGuard Cuff, Distal with Disposable Inner Cannula; Shiley Adult Flexible Tracheostomy Tube XLT with TaperGuard Cuff, Proximal with Disposable Inner Cannula; Shiley Adult Flexible Tracheostomy Tube XLT Cuffless, Distal with Disposable Inner Cannula ; Shiley Adult Flexible Tracheostomy Tube XLT Cuffless, Proximal with Disposable Inner Cannula; Shiley Adult Flexible Disposable Inner Cannula XLT · Covidien, LLC · Dec 29, 2025
Submission Summary (Full Text)
{0}
RUSCH.
INTERNATIONAL
Group Regulatory Affairs
A Subsidiary of Teleflex Incorporated (USA)
Tall Pines Park
Jaffrey, NH 03452
(603) 532-7706
FAX (603) 532-8211 or 6108
JUN 28 1995
K955564
510(K) SUMMARY
ULTRA TRACHEOFLEX® TRACHEOSTOMY SET - CUFFED
Submitter Information:
Neil R. Armstrong
Group Regulatory Affairs Manager
Rusch International
Tall Pines Park
Jaffrey, NH 03452
Phone Number: (603) 532-7706
Fax Number: (603) 532-8211
Date 510(k) Summary Prepared: October 10, 1995
Name of the Device:
Trade or Proprietary Name: Rusch Ultra Tracheoflex® Tracheostomy Set - Cuffed
Common Name: Tracheostomy Set-Cuffed, Fixed length
Classification Name: Tracheostomy tube and tube cuff, and accessories: 21 CFR 868.5800
Airway Connector: 21 CFR 868.5810
Attachment 1
{1}
Identification of Legally Marketed Device to which the Submitter Claim Equivalence:
The Ultra Tracheoflex® Tracheostomy set cuffed is substantially equivalent to two products marketed by Rüsch prior to May 28, 1976: These are the Safetyflex® Tracheotomy Set available since April, 1974, and the Tracheoflex® Tracheotomy Set, available since April 1971. This device is also similar to the Portex Flexible D.I.C. Tracheostomy Tube found substantially equivalent under K934465.
The subject devices are similar or equivalent to the predicates in significant features, materials, dimensions, and intended use.
## Description of the Subject Devices:
The Ultra Tracheoflex® Tracheostomy Set-Cuffed is a sterile, disposable, single use cuffed tracheostomy tube with a fixed flange. The set includes a radiopaque tube, a tube obturator, a cough cap, a 15 mm airway connector, exchangeable inner cannulas, and a neck band. The inner cannula will be also available as a separate item.
These devices meet applicable portions of the following voluntary standards:
- ASTM F1242-89, Standard Specification for Cuffed and Uncuffed Tracheal Tubes.
- ISO 5366-1, Tracheostomy Tubes, Part 1, Connectors for Tubes for Adults
- ISO 5366.2, Basic requirements for Tracheostomy Tubes for Adults.
- ISO 5356: Anesthesia and Respiration Equipment, Conical Connectors, Cones and Sockets.
- ISO 5361, Tracheal Tubes, Part 5: Requirements and Tests for Cuffs and Tubes.
Attachment 1
{2}
# Intended Use of Subject Device
The Rüsch Ultra Tracheoflex® Tracheostomy Set-Cuffed is a tracheostomy tube with accessories, intended for intubation through a surgical incision into the trachea, for the purpose of providing a patent airway.
# Technological Characteristics of the Subject Devices
The subject device and the predicates have the same technological characteristics as the predicate devices except for:
1. The subject device's tube cuff is filled via a pilot balloon tube that ends in a spring (Bespack) valve. The Rüsch predicate devices have a balloon that ended in a capped luer. This technological improvement was made to aid in the prevention of the loss of cuff pressure due to the improper placement of the cap after cuff inflation.
2. The Rüsch predicate devices had a 15 mm airway connector permanently attached to the tube flange. The subject device has a ISO standard screw-luer fitting permanently attached. This allows a locking connections to a 15 mm airway connector, a cough cap, or other connectors as needed.
No clinical testing was completed.
Attachment 1
Panel 1
/
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.