MONTGOMERY TRACHEOSTOMY SPEAKING VALVE
Device Facts
| Record ID | K955634 |
|---|---|
| Device Name | MONTGOMERY TRACHEOSTOMY SPEAKING VALVE |
| Applicant | Boston Medical Products, Inc. |
| Product Code | CBP · Anesthesiology |
| Decision Date | Jun 10, 1996 |
| Decision | SESE |
| Submission Type | Traditional |
| Regulation | 21 CFR 868.5870 |
| Device Class | Class 2 |
| Attributes | Therapeutic |
Indications for Use
For use as a one-way speaking valve between a tracheostomy tube and ventilation equipment.
Device Story
Montgomery Tracheostomy Speaking Valve; in-line device between tracheostomy tube and ventilator; features thin silicone diaphragm; opens on inspiration, closes on expiration; allows ventilator-dependent patients to speak during both phases of ventilation cycle without waiting for ventilator cycling; used in clinical settings; operated by clinicians/caregivers; output is restored speech capability for patient.
Clinical Evidence
Bench testing only; visual inspection for defects; testing with ISO 5356-1 test gauges for 15mm and 22mm connections.
Technological Characteristics
Silicone diaphragm; 15mm and 22mm connections; ISO 5356-1 compliant connections; mechanical one-way valve operation.
Indications for Use
Indicated for ventilator-dependent patients requiring a one-way speaking valve to facilitate speech during both inspiration and expiration while connected to ventilation equipment.
Regulatory Classification
Identification
A nonrebreathing valve is a one-way valve that directs breathing gas flow to the patient and vents exhaled gases into the atmosphere.
Predicate Devices
- Tracheostomy Speaking Valve (15mm / 22mm), Product No. 007 (Passy and Passy)
Related Devices
- K041485 — MARPAC MESSENGER MODEL #501 · Marpac, Inc. · Aug 2, 2004
- K043160 — TRACOE PHON ASSIST I - SPEAKING VALVES, REFERENCE MODELS: PHON ASSIST I-650-T/TO AND 650-S/SO · Tracoe Medical GmbH · Feb 18, 2005
- K122932 — SPIRO SPEAKING VALVE WITH OXYGEN PORT SPIRO SPEAKING VALVE WITHOUT OXYGEN PORT · Fogless International AB · Dec 23, 2013
- K100995 — SHIKANI SPEAKING VALVE · Shikani Medical, LLC · Jun 25, 2010
- K982128 — SHIKANI-FRENCH SPEAKING VALVE · Pilling Weck Surgical · Oct 23, 1998