PAN-AMERICA HYPERBARICS HYPERBARIC CHAMBER SYSTEM, MODEL PAH-M3+1
K021690 · Pan-America Hyperbarics, Inc. · CBF · Oct 24, 2002 · Anesthesiology
Device Facts
Record ID
K021690
Device Name
PAN-AMERICA HYPERBARICS HYPERBARIC CHAMBER SYSTEM, MODEL PAH-M3+1
Applicant
Pan-America Hyperbarics, Inc.
Product Code
CBF · Anesthesiology
Decision Date
Oct 24, 2002
Decision
SESE
Submission Type
Abbreviated
Regulation
21 CFR 868.5470
Device Class
Class 2
Attributes
Therapeutic
Intended Use
It is the expressed, intended use of the Pan-America Hyperbarics' PAH-M3+1 Hyperbaric Chamber System to provide therapy to those patients with selected medical conditions that have been determined to respond to the application of hyperbaric oxygen. As a Class II prescriptive device, it is further intended for physician involvement in its procurement and routine use. The UHMS is the professional medical organization chartered with setting the standards of care defining the appropriate use of hyperbaric oxygen. More specifically, the UHMS publishes a listing of medical conditions that have been clearly established as appropriate primary or adjunctive use of hyperbaric oxygen (HBO). The disorders on the list have been scientifically validated and verified through extensive data collection. It should be noted that the list is dynamic. Based on the strength of the scientific data, disorders are both added and removed from the list, depending on the outcomes of scientific pursuit. The conditions listed as appropriate for the use of HBO in the current edition of the Hyperbaric Oxygen Therapy Committee Report (1999) are as follows: - Air or gas embolism 1. - Carbon monoxide poisoning and carbon monoxide poisoning complicated by 2. cyanide poisoning - Clostridial myositis and myonecrosis ന് - 4. Crush injury, compartment syndrome, and other acute traumatic ischemias - 5. Decompression sickness - റ. Enhanced of selected problem wounds - 7. Exceptional blood loss anemia - 8. Necrotizing soft tissue infections - ு. Osteomyelitis (refractory) - 10. Delayed radiation injury (soft tissue and bony necrosis) - 11. Skin grafts and flaps (compromised) - 12. Thermal burns - 13. Intracranial abscess The Pan-America Hyperbarics' PAH-M3+1 Hyperbaric Chamber System is designed to be installed and operated in medical facilities as defined by the NFPA 99, Health Care Facilities, Chapter 19, Hyperbaric Facilities (Chapter 20, 2002 Edition). Further, this system is intended to be operated only by medical personnel specifically trained in the appropriate use of HBO and the safe operations of all related equipment such as the hyperbaric chamber.
Device Story
Multiplace hyperbaric chamber system; treats up to 4 patients; max operating pressure 3 ATA (29.4 psig). Inputs: compressed air for pressurization; 100% oxygen administered via oronasal masks or head tents. Operation: manual pneumatic control by trained medical personnel in clinical facilities. Features: two compartments (main/transfer); water deluge and handline fire suppression; low-voltage intercommunication system; Teledyne TED-191 oxygen analyzer for continuous monitoring; retractable gurney/sliding bunk for patient loading. Output: hyperbaric environment for therapeutic oxygen delivery. Clinical benefit: treatment of conditions responsive to hyperbaric oxygen as defined by UHMS. Physician-prescribed and operated by trained staff.
Clinical Evidence
Bench testing only. No clinical data presented. Device design and safety verified through compliance with ANSI/ASME Boiler and Pressure Vessel Code Section VIII, Division 1, ANSI/ASME-PVHO-1, and NFPA 99 standards.
Technological Characteristics
Class A multiplace hyperbaric chamber; 4500mm length, 1016mm diameter. Materials: pressure vessel construction per ANSI/ASME Boiler and Pressure Vessel Code Section VIII, Div 1 and PVHO-1. Energy: compressed air (pressurization), electrical (intercom/monitoring). Monitoring: Teledyne TED-191 oxygen analyzer. Safety: dual fire suppression (water deluge/handline), safety interlock for door. Connectivity: penetrator plate for user-supplied medical monitoring leads.
Indications for Use
Indicated for patients requiring hyperbaric oxygen therapy for conditions including air/gas embolism, carbon monoxide poisoning, clostridial myositis/myonecrosis, crush injury, compartment syndrome, acute traumatic ischemias, decompression sickness, problem wounds, exceptional blood loss anemia, necrotizing soft tissue infections, refractory osteomyelitis, delayed radiation injury, compromised skin grafts/flaps, thermal burns, and intracranial abscess.
Regulatory Classification
Identification
A hyperbaric chamber is a device that is intended to increase the environmental oxygen pressure to promote the movement of oxygen from the environment to a patient's tissue by means of pressurization that is greater than atmospheric pressure. This device does not include topical oxygen chambers for extremities (§ 878.5650).
K082455 — RECTANGULAR MULTIPLACE HYPERBARIC CHAMBER · Gulf Coast Hyperbarics, Inc. · May 22, 2009
Submission Summary (Full Text)
{0}------------------------------------------------
K021690
OCT 2 4 2002
# 510(k) NOTIFICATION SUMMARY (Per 21 CFR 807.92)
Prepared: 20 May 2002
TRADE NAME: PAH-M3+1 Hyperbaric Chamber System
COMMON NAME OF DEVICE: Multiplace Hyperbaric Chamber
CLASSIFICATION: 73 CBF, 21 CFR 868.5470
ESTABLISHMENT REGISTRATION NUMBER: Pending
### CLAIMED PREDICATE DEVICE(S):
Perry Sigma II Hyperbaric Chamber - K862198 Perry Sigma Plus/II Hyperbaric Chamber - K983648
ADDRESS OF MANUFACTURER:
Pan-America Hyperbarics, Inc. No.9, Lane 12, Guang Feng 1st Street Ba-de City, Tao Yuan County Taiwan, R.O.C. +886-3-3676676
CONTACT PERSON: Cheng, Kuo-Chung
#### EXECUTIVE SUMMARY
The PAH-M3+1 Hyperbaric Chamber System is a Class A multiplace hyperbaric chamber designed to treat up to 4 patients at up to a maximum operating pressure of 3 Atmospheres Absolute (ATA) or 29.4 pounds per square inch gauge (psig). The chamber uses compressed air as the pressurization gas and 100% oxygen as the hyperbaric treatment gas.
The PAH-M3+1 Hyperbaric Chamber System is intended to be procured and used by physicians to treat a variety of medical conditions that respond to hyperbaric oxygen. The Undersea & Hyperbaric Medical Society (UHMS) produces a list of medical conditions that have been identified for the appropriate primary or adjunctive use of hyperbaric oxygen. These approved conditions include: air or gas embolism; carbon monoxide poisoning and smoke inhalation; clostridial myonecrosis (gas gangrene); crush injury, compartment syndrome and other acute traumatic ischemias; decompression sickness; enhanced healing of selected problem wounds: exceptional blood loss anemia: necrotizing soft tissue infections: osteomyelitis (refractory); radiation tissue damage (osteoradionecrosis); compromised skin flaps and grafts; thermal burns; and, intracranial abscess. Aggressive research into the beneficial effects of hyperbaric oxygen, when appropriately applied, will result in additional medical conditions being added to the list of indications by the UHMS.
{1}------------------------------------------------
The PAH-M3+1 Hyperbaric Chamber System is designed and fabricated in accordance with the requirements of the ANSI/ASME Boiler and Pressure Vessel Code, Section VIII, Division 1. Pressure Vessels: ANSI/ASME-PVHO-1 (American Society of Mechanical Engineers-Pressure Vessels for Human Occupancy); and, NFPA 99, Health Care Facilities, Chapter 19, Hyperbaric Facilities (Chapter 20, 2002 Edition). The overall external length of the chamber is 4500mm (approx. 14.76 ft). Its internal diameter is 1016mm (approx. 40 inches). There are two compartments: main compartment and transfer compartment. Three (3) removable seats and one fixed (1) seat are installed in main compartment and transfer compartment respectively. Two independent fire suppression systems, water deluge system and handline system, are installed in accordance with the requirements of NFPA 99, Chapter 19 (Chapter 20, 2002 Edition). Pressurization is provided by compressed air with 100% oxygen administered to the patient by using properly fitting oronasal masks or head tents. A lowvoltage patient intercommunication system designed and installed in accordance with NFPA 99. Chapter 19 (Chapter 20, 2002 Edition) and provides communications between the patients in the chamber and the outside chamber operator. A Teledyne TED-191 oxygen analyzer is installed to monitor the concentration of oxygen inside the chamber continually. The system consists of an operator control console that contains all of the controls and connection points. Single operator chamber pressure control is achieved via a simple manual pneumatic control. A penetrator plate is provided in the vessel wall to allow user supplied medical monitoring leads, etc., to be used as required. The patients are loaded and unloaded by a retractable gurney, which is equipped with a sliding transport chair. When supine position is needed, a sliding bunk will be used instead. The chamber is also equipped with safety switch for pressurization. There is no gas supply before the chamber's door is closed and secured thoroughly.
#### Intended Use:
It is the expressed, intended use of the Pan-America Hyperbarics' PAH-M3+1 Hyperbaric Chamber System to provide therapy to those patients with selected medical conditions that have been determined to respond to the application of hyperbaric oxygen. As a Class II prescriptive device, it is further intended for physician involvement in its procurement and routine use.
The UHMS is the professional medical organization chartered with setting the standards of care defining the appropriate use of hyperbaric oxygen. More specifically, the UHMS publishes a listing of medical conditions that have been clearly established as appropriate primary or adjunctive use of hyperbaric oxygen (HBO). The disorders on the list have been scientifically validated and verified through extensive data collection. It should be noted that the list is dynamic. Based on the strength of the scientific data, disorders are both added and removed from the list, depending on the outcomes of scientific pursuit.
The conditions listed as appropriate for the use of HBO in the current edition of the Hyperbaric Oxygen Therapy Committee Report (1999) are as follows:
- Air or gas embolism 1.
- Carbon monoxide poisoning and carbon monoxide poisoning complicated by 2. cyanide poisoning
- Clostridial myositis and myonecrosis ന്
- 4. Crush injury, compartment syndrome, and other acute traumatic ischemias
- 5. Decompression sickness
- റ. Enhanced of selected problem wounds
- 7. Exceptional blood loss anemia
- 8. Necrotizing soft tissue infections
- ு. Osteomyelitis (refractory)
- 10. Delayed radiation injury (soft tissue and bony necrosis)
- 11. Skin grafts and flaps (compromised)
{2}------------------------------------------------
- 12. Thermal burns
- 13. Intracranial abscess
The Pan-America Hyperbarics' PAH-M3+1 Hyperbaric Chamber System is designed to be installed and operated in medical facilities as defined by the NFPA 99, Health Care Facilities, Chapter 19, Hyperbaric Facilities (Chapter 20, 2002 Edition). Further, this system is intended to be operated only by medical personnel specifically trained in the appropriate use of HBO and the safe operations of all related equipment such as the hyperbaric chamber.
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" around the perimeter. Inside the circle is a symbol consisting of three horizontal bars that curve downward, resembling a stylized human figure. The image is black and white and appears to be a scan or photocopy.
OCT 24 2002
d and Drug Administr 9200 Corporate Boulevard Rockville MD 20850
Pan-America Hyperbarics, Incorporated ^C/O Mr. W. T. Workman Workman Hyperbaric Services, Incorporated 18111 Copper Ridge Drive San Antonio, Texas 78259-3612
Re: K021690
Trade/Device Name: Hyperbaric Chamber System, Model PAH-M3+1 Regulation Number: 868.5470 Regulation Name: Hyperbaric Chamber Regulatory Class: II Product Code: CBF Dated: September 5, 2002 Received: September 6, 2002
Dear Mr. Workman:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies.
{4}------------------------------------------------
Page 2 - Mr. Workman
You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 21 CFR Part 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4646. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html
Sincerely yours,
Timothy A. Ullatowski
Timothy Ulatowski Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
## 510(k) Number (if known): KO2ICe 90
Device Name: Pan-America Hyperbarics, Inc., PAH-M3+1 Hyperbaric Chamber System
1
Indications for Use:
. . . . . .
The conditions listed as appropriate for the use of HBO in the current edition of the Hyperbaric Oxygen Therapy Committee Report (1999) are as follows:
- 1. Air or gas embolism
- 2. Carbon monoxide poisoning and carbon monoxide poisoning complicated by cyanide poisoning
- 3. Clostridial myositis and myonecrosis
- 4. Crush injury, compartment syndrome, and other acute traumatic ischemias
- 5. Decompression sickness
- 6. Enhanced healing of selected problem wounds
- 7. Exceptional blood loss anemia
- 8. Necrotizing soft tissue infections
- 9. Osteomyelitis (refractory)
- 10. Delayed radiation injury (soft tissue and bony necrosis)
- 11. Skin grafts and flaps (compromised)
- 12. Thermal burns
- 13. Intracranial abscess
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IS NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)

umber: K021/640
(Optional Format 3-10-98)
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.