CUFFED OROPHARYNGEAL AIRWAY (COPA)

K955721 · Mallinckrodt Medical · CAE · Mar 28, 1997 · Anesthesiology

Device Facts

Record IDK955721
Device NameCUFFED OROPHARYNGEAL AIRWAY (COPA)
ApplicantMallinckrodt Medical
Product CodeCAE · Anesthesiology
Decision DateMar 28, 1997
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 868.5110
Device ClassClass 1
AttributesTherapeutic

Device Story

COPA (Cuffed Oropharyngeal Airway) is an airway management device for anesthetized patients. Device consists of an airway tube with an inflatable cuff at the distal end. Inserted into the pharynx; cuff inflation occludes the nasopharynx, displaces the base of the tongue and epiglottis anteriorly, and seals the posterior pharynx. Operated by clinicians (anesthesiologists) during general anesthesia. Provides a patent airway for spontaneous breathing. Compared to face masks and laryngeal mask airways (LMA). Clinical utility involves manual airway maintenance, such as jaw lift or head rotation, to ensure clear airway. Benefits include reduced potential for glottic or tracheal stimulation compared to tracheal tubes and performance characteristics similar to a 'hands-free' face mask/Guedel airway.

Clinical Evidence

Evidence includes multiple clinical studies: 1) IDE G900092 (n=10) feasibility study; 2) European study (n=20) showing ease of use and minor sequelae; 3) Australian study (n=100) evaluating spontaneous breathing performance; 4) Randomized, multi-site study (IDE G960100, n=452) comparing COPA to LMA. Results: COPA provided clear airway in 98-100% of patients. Multi-site study showed COPA and LMA equivalent in safety and efficacy, though LMA required fewer manipulations (p<0.001) and had shorter time to 'hands-free' status (p=0.004). Complication rates (aspiration, hypoxia, laryngospasm) were statistically similar between devices.

Technological Characteristics

Oropharyngeal airway with inflatable cuff. Conforms to ISO 5364 standards for molding tolerances, 15 mm connector dimensions, and resistance to collapse/distortion. Materials provide airway integrity, cuff resting volume/diameter, and bond strength. Mechanical device; no energy source or software.

Indications for Use

Indicated for use as an airway management device in anesthetized patients undergoing spontaneously breathing general anesthesia.

Regulatory Classification

Identification

An oropharyngeal airway is a device inserted into a patient's pharynx through the mouth to provide a patent airway.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0} K955721 # Non-Confidential Safety and Effectiveness Summary ## Preliminary Feasibility Studies MAR 28 1997 An earlier design of this device was evaluated in a clinical study under IDE G900092. In that study, ten patients were studied with the device in place during emergence from a short general anesthetic. There was no difference between positive inspiratory pressure necessary to generate adequate ventilation volumes before placement of the device, (face mask only), during placement with the COPA™ cuff deflated, or during placement with the COPA™ cuff inflated. Use of the device did not affect end-tidal carbon dioxide concentrations. No patient had a complication as a result of participating in the study. ## Design Modifications While the preliminary studies demonstrated the feasibility of the COPA™ device, they also implied that a change in the shape of the inflatable cuff would improve that reliability of the device. All later studies were performed on devices with a cuff design as indicated in the attached drawing. ## Functional Testing Bench testing of sterile prototypes verified conformance to ISO 5364 -- Oropharyngeal airways, in its requirements for: 1. molding tolerances 2. dimensions of 15 mm connectors and airway 3. resistance to collapse of the buccal end 4. resistance to distortion Bench testing of sterile prototypes also verified the integrity and consistency of the product in the following areas: 1. airway integrity 2. resting volumes and diameters of cuffs 3. cuff integrity 4. cuff herniation over the distal end of the airway 5. strength of the tail-airway bond 6. disconnect force of the connector from the breathing circuitry The functionality of the COPA™ device was compared to that of predicate devices in cadaver studies. (See attachment 5) The results of these studies show that: 1. The cuff is appropriately placed in the pharynx when the device is placed as directed in the product insert. When inflated, the cuff occludes the nasopharynx and displaces the base of the tongue and epiglottis anteriorly while sealing the posterior pharynx. 2. The cuff seals the oropharynx at lower pressures than both the Sheridan Combi-Tube and the Laryngeal Mask Airway. 3. The COPA™ airway does not extend inferiorly past the vallecula or into the esophagus. Therefore, it does not raise the potential for glottic or tracheal stimulation in an anesthetized patient. 1 of 3 {1} # Clinical Studies ## A. Conducted August 1995. The current design was evaluated in a European study of twenty patients. The results of this study, as summarized in the attached abstracts, (attachment 3) show that: 1. Device placement and maintenance of a patent airway were easy. 2. Anesthetic sequelae (coughing, bucking, and sore throat) were minor and comparable to those occurring with use of tracheal tubes and laryngeal mask airways. 3. Anesthetic doses needed for device placement and maintenance are similar to those needed for the predicate devices. ## B. Conducted July 1996 A two-site Australian study was conducted on 100 patients to evaluate the $\mathrm{COPA}^{\mathrm{TM}}$ device for spontaneously breathing anesthesia. During airway maintenance with the $\mathrm{COPA}^{\mathrm{TM}}$ device, adverse events and interventions were recorded on videotape, and by verbal commentary and hand-written notes. An assessment was made of device positioning (fiberoptically, from within the device), postoperative sore throat, and skill acquisition. The $\mathrm{COPA}^{\mathrm{TM}}$ device provided a clear airway in $98\%$ of the patients during manually assisted breathing (MAB), $100\%$ during spontaneous breathing (SB) and emergence from anesthesia. Useful interventions were jaw lift and rotation of the head. Less useful were size changes, withdrawing the $\mathrm{COPA}^{\mathrm{TM}}$ device slightly, or changing cuff volume. Most interventions occurred during the first three minutes of each phase. Jaw lift (manual or mechanical chinlift) was required for $21.4\%$ of the time during SB. The oxygen saturation $(\mathrm{SpO}_2)$ briefly fell to $87 - 89\%$ on six occasions. During SB the vocal cords were visible (from a vantage point of just inside the distal tip of the device) in $29\%$ and the epiglottis in $90\%$. Emergence characteristics were excellent. Mild sore throat occurred in $4\%$. There was tentative evidence for skill acquisition. The investigators concluded that the cuffed oropharyngeal airway is suitable for spontaneous breathing anesthesia. Most patients require one or more interventions to provide a clear airway. The device has the performance characteristics of a relatively "hands-free" face mask/Guedel airway. It is well tolerated and has a low complication rate. ## C. Completed September 1996 A randomized, controlled, multi-site study was carried out under IDE G960100 to compare the safety, efficacy, and utility of the $\mathrm{COPA}^{\mathrm{TM}}$ device to that of the laryngeal mask airway (LMA) in patients undergoing spontaneously breathing general anesthesia. The results of this study indicate that the $\mathrm{COPA}^{\mathrm{TM}}$ device and the LMA are equivalent devices in terms of 1) ease of use, 2) physiological tolerance and 3) minor and major complications. Although the LMA was associated with fewer airway manipulations, both devices were equivalent in establishing a safe and effective airway for spontaneously breathing anesthetized adults. There were no reports of unanticipated adverse device effects on the health or safety of the patients, or any life threatening problems or deaths caused by or associated with either tested device in this study. The time to hands free airway (seconds) was 229 (310) for the COPA and 137 (121) for LMA. It was significantly shorter for the LMA $(p = 0.004)$. There were significantly more patients who did not achieve a hands free airway with the COPA $(29/302 = 9.6\%)$ than the LMA $(1/150 = 0.66\%)$ $(p &lt; 0.001)$. The time to spontaneous breathing (minutes) was similar at 7.3 (6.8) for COPA and 6.7 (6.4) for LMA $(p = 0.44)$. Airway manipulations post-incision - propofol off. Total (%) | Airway Manipulations | COPA | LMA | | --- | --- | --- | | 0 | 135/295 (45.8) | 144/150 (96) | | 1 | 39/295 (13.2) | 5/150 (3.3) | | 2 | 14/295 (4.8) | 1/150 (0.7) | | >2 or continuous support | 107/295 (36.3) | 0/150 (0) | | Total | 295 (100) | 150 (100) | {2} Major and minor problems reported are summarized in the following table: Problems occurring during use of the device: events/patients (%) | type | COPA | LMA | p value Chi Squared | | --- | --- | --- | --- | | aspiration | 1/302 (0.3) | 0/151 (0) | 0.479 | | regurgitation | 2/302 (0.7) | 1/151 (0.7) | 1.000 | | laryngospasm | 6/302 (2) | 2/151 (1.3) | 0.614 | | succinylcholine given | 5/302 (1.7) | 0/151 (0) | 0.112 | | wheeze | 1/302 (0.3) | 0/151 (0) | 0.479 | | hypoxia (SaO_{2} < 92%) | 30/302 (9.9) | 14/151 (9.3) | 0.822 | | failed use | 14/302 (4.6) | 4/151 (2.6) | 0.291 | | bucking | 2/302 (0.7) | 1/151 (0.7) | 1.000 | | hiccuping | 5/302 (1.7) | 8/151 (5.3) | 0.029 | | shivering | 4/302 (1.3) | 0/151 (0) | 0.155 | | gagging | 4/302 (1.3) | 4/151 (2.7) | 0.313 | | coughing | 28/302 (9.3) | 11/151 (7.3) | 0.477 | | stridor | 9/302 (3.0) | 3/151 (2.0) | 0.535 | | nasogastric tube inserted | 0/302 (0) | 1/151 (0.7) | 0.157 | | blood detected | 17/294 (5.8) | 23/150 (15.3) | 0.001 | | movement | 15/302 (5.0) | 9/151 (6.0) | 0.656 | | continuous chin support | 90/302 (29.8) | 0/151 (0) | 0.001 |
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...