K980884 · Siemens Medical Solutions USA, Inc. · CAD · Aug 19, 1998 · Anesthesiology
Device Facts
Record ID
K980884
Device Name
SIEMENS SEVOFLURANE VAPORIZER SV 953
Applicant
Siemens Medical Solutions USA, Inc.
Product Code
CAD · Anesthesiology
Decision Date
Aug 19, 1998
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 868.5880
Device Class
Class 2
Attributes
Therapeutic, Pediatric
Indications for Use
The Sevoflurane Vaporizer SV 953 is designed for use with the Servo Ventilator 900 and for vaporizing the liquid anesthetic agent sevoflurane in conjunction with the controlled administration of anesthetic gas mixtures during surgery. The Sevoflurane Vaporizer SV953 is intended to be used by Healthcare providers, i.e. Physicians, Nurses and Technicians. The Sevoflurane Vaporizer SV953 is intended for general and critical ventilatory to be used on Adult, Pediatric, Infant and Neonatal populations. The Sevoflurane Vaporizer SV953 is intended to be used in the environment where patient care is provided by Healthcare Professionals. The unit is designed to be used at the bedside. It is not intended for transport use in ambulances or helicopters in the U.S. market.
Device Story
The Sevoflurane Vaporizer SV 953 is an anesthetic vaporizer designed for integration with the Siemens Servo Ventilator 900. It functions by vaporizing liquid sevoflurane for controlled delivery of anesthetic gas mixtures to patients during surgery. The device is operated by healthcare professionals (physicians, nurses, technicians) in clinical bedside settings. Modifications from the predicate include a yellow color-coded knob, label, and filling mechanism, a concentration scale ranging from 0.2% to 8%, and an increased capillary tube diameter (0.36 mm) to support the 8% concentration output. The device is not intended for transport in ambulances or helicopters and is not MRI compatible.
Clinical Evidence
Bench testing only. The device design was validated through testing of all settings under the expanded indications, with results meeting or exceeding the criteria applied to the predicate devices.
Technological Characteristics
Anesthetic vaporizer; mechanical operation. Features a concentration scale (0.2% to 8%) and a 0.36 mm diameter capillary tube. Color-coded yellow for sevoflurane identification. Complies with ISO/DIS 8835-1.2 and ISO 5360. Not MRI compatible.
Indications for Use
Indicated for use in adult, pediatric, infant, and neonatal populations requiring general or critical ventilation during surgery for the administration of sevoflurane anesthetic gas mixtures.
Regulatory Classification
Identification
An anesthetic vaporizer is a device used to vaporize liquid anesthetic and deliver a controlled amount of the vapor to the patient.
K980489 — VAPAMASTA 6 · Medical Industrial Equipment · Aug 31, 1998
Submission Summary (Full Text)
{0}------------------------------------------------
AUG / 9 1998 510(k) Notification
Siemens Sevoflurane Vaporizer SV 953
# 510(k) Summary and Certification
# Submitter's Name and Address
Siemens-Elema AB Röntgenvägen 2 S-171 95 Solna Sweden
# Official Correspondent
Mr. David Simard
Telephone 978-907-7737 978-777-3398 Telefax
# Contact Person for this Submission
Mr Wulf R. Trepte
Telephone 011-46 8 730 72 28 011-46 8 98 63 05 Telefax
## Device Name
| Common Name: | Sevoflurane Vaporizer 953 |
|------------------------|---------------------------|
| Classification Name: | Anesthetic Vaporizer |
| Regulation Number: | 21 CFR 868.5580 |
| Classification Number: | 73BSZ<br>73CAD |
| Class: | Class II |
# Establishment Registration Number
The Establishment Registration Number for Siemens-Elema AB is: 8010042
## Manufacturing Facility Address
Siemens-Elema AB S-171 95 Solna Sweden
### Reason for Pre-market Notification
The reason for this pre-market notification is an expanded indication to the existing = ~*** device.
## Company Confidential
Siemens-Elema AB Electromedical Systems Division Röntgenvägen 2 S-171 95 SOLNA SWEDEN
tel: (46) 8 730 7228 fax: (46) 8 98 63 05
198088
{1}------------------------------------------------
#### 510(k) Notification
Siemens Sevoflurane Vaporizer SV 953
## · ------------Predicate Device
The legally marketed device to which equivalence is being claimed is:
Siemens Vaporizers 950/951/952, i.e Halothane Vaporizer 950, Enflurane Vaporizer 951, Isoflurane Vaporizer 952, they were found Substantially Equivalent on June 7, 1984 (Premarket Notification K841157).
### Device Description
The Sevoflurane Vaporizer 953 is an anesthetic vaporizer used together with Siemens Servo Ventilator 900 in anesthetic applications. It is a modification of the Siemens Vaporizers 950/951/952.
The modifications made are:
- At the knob a scale for Sevoflurane is added. The scale indicates concentration . values from 0.2% to 8%.
- The knob scale, label and filling mechanism, are yellow color coded. .
- To reach the concentration values 8% the inside diameter of a capillary tube is . 0.36mm instead of 0.30mm as in the other vaporizers
#### Intended Use
#### Purpose and function of the Sevoflurane Vaporizer SV953:
The Sevoflurane Vaporizer SV 953 is designed for use with the Servo Ventilator 900 and for vaporizing the liquid anesthetic agent sevoflurane in conjunction with the controlled administration of anesthetic gas mixtures during surgery.
#### Intended Operator:
The Sevoflurane Vaporizer SV953 is intended to be used by Healthcare providers, i.e. Physicians, Nurses and Technicians.
#### Intended Patient Populations:
The Sevoflurane Vaporizer SV953 is intended for general and critical ventilatory to be used on Adult, Pediatric, Infant and Neonatal populations.
#### Intended Use Environment:
The Sevoflurane Vaporizer SV953 is intended to be used in the environment where patient care is provided by Healthcare Professionals. The unit is designed to be used at the bedside. It -is not intended for transport use in ambulances or helicopters in the U.S. market.
#### Company Confidential
Siemens-Elema AB Electromedical Systems Division
Röntgenvägen 2 S-171 95 SOLNA SWEDEN
tel: (46) 8 730 7228 fax: (46) 8 98 63 05
{2}------------------------------------------------
#### 510(k) Notification
Siemens Sevoflurane Vaporizer SV 953
## Substantial Equivalence:
The enhanced functionality for the Sevoflurane Vaporizer SV953 is equivalent to the Siemens Halothane Vaporizer HV950.
The Siemens Halothane Vaporizer HV950 was granted pre-market approval under 510(k) file number K841157.
### MRI Compatibility Statement:
The Sevoflurane Vaporizer SV953 is not compatible for use in a MRI magnetic field.
## Biocompatibility
Not applicable
## Sterilization
Not applicable
# Specific Standards & Guidances
Siemens Sevoflurane Vaporizer SV 953 complies with the following standards:
- ISO/DIS 8835-1.2
- ISO 5360
Company Confidential
Siemens-Elema AB Electromedical Systems Division
Röntgenvägen 2 S-171 95 SOLNA SWEDEN
tel: (46) 8 730 7228 fax: (46) 8 98 63 05
{3}------------------------------------------------
## 510(k) Notification Siemens Sevoflurane Vaporizer SV 953
## Comparison of Technological Characteristics
The hardware modification, compared to the predicate device, is:
- At the knob a scale for Sevoflurane is added. The scale indicates concentration . values from 0.2% to 8%.
- The knob scale, label and filling mechanism, are yellow color coded. .
- To reach the concentration values 8% the inside diameter of a capillary tube is 0.36 . mm instead of 0.30 mm as in the other vaporizers
The functionality of the Sevoflurane Vaporizer SV 953 is identical to the functionality of the Siemens Vaporizers 950/951/952; i.e Halothane Vaporizer 950, Enflurane Vaporizer 951, Isoflurane Vaporizer 952.
## Tests Used in Determination of Substantial Equivalence
The design of the Seveflurane Vaporizer SV 953 has been thoroughly validated. All different settings of the new expanded indications has been tested, all test were passed according to criteria that are equal or more stringent than the test criteria which were applied to the predicate device.
## Conclusion
Analysis and tests have shown that the new expanded indications doesn't adversely affect patient safety.
Therefor, we conclude that the requirements specifications and validation testing show that the modified device is as safe and effective, and performs as well or better as the predicate device.
## Company Confidential
Siemens-Elema AB Electromedical Systems Division
Röntgenvägen 2 S-171 95 SOLNA SWEDEN
tel: (46) 8 730 7228 fax: (46) 8 98 63 05
{4}------------------------------------------------
Image /page/4/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features the department's name encircling a symbol. The symbol is a stylized representation of an eagle or bird, with three curved lines forming its body and wings. The overall design is simple and recognizable, representing the department's role in health and human services.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
AUG | 9 1998
Mr. David M. Simard Siemens Medical Systems, Inc. 16 Electronics Avenue Danvers, MA 01923
Re: K980884 Siemens Sevoflurane Vaporizer SV 953 Regulatory Class: II (two) Product Code: 73 CAD June 19, 1998 Dated: Received: July 1, 1998
Dear Mr. Simard:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and ---------prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major requlations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in requlatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.
{5}------------------------------------------------
#### Page 2 - Mr. David M. Simard
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling requlation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4648. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597, or at its internet address "http://www.fda.gov/cdrh/dsma/dsmamain.html".
Sincerely yours,
Thomas J. Collethon
Thomas J. Callahan, Ph.D. Director Division of Cardiovascular, Respiratory, and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
510(k) Sevoflurane Vaporizer SV953
## Indications for Use Statement:
Page1 off 2
پ ۽ پيلي جي جي جي جي جي جي جي جي جي جي جي جي جي جي جي جي جي جي جي جي جي جي جي جي جي جي جي جي جي جي جي جي جي جي جي جي جي جي جي جي جي جي جي جي جي جي جي جي جي جي جي جي جي جي جي
510(k) Number (if known): K980884
Sevoflurane Vaporizer SV953 Device Name:
Indications For Use:
#### Purpose and function of the Sevoflurane Vaporizer SV953:
The Sevoflurane Vaporizer SV 953 is designed for use with the Servo Ventilator 900 and for vaporizing the liguid anesthetic agent sevoflurane in conjunction with the controlled administration of anesthetic gas mixtures during surgery.
#### Intended Operator:
The Sevoflurane Vaporizer SV953 is intended to be used by Healthcare providers, i.e. Physicians, Nurses and Technicians.
### Intended Patient Populations:
The Sevoflurane Vaporizer SV953 is intended for general and critical ventilatory to be used on Adult, Pediatric, Infant and Neonatal populations.
#### Intended Use Environment:
The Sevoflurane Vaporizer SV953 is intended to be used in the environment where patient care is provided by Healthcare Professionals. The unit is designed to be used at the bedside. It is not intended for transport use in ambulances or helicopters in the U.S. market.
#### Substantial Equivalence:
The enhanced functionality for the Sevoflurane Vaporizer SV953 is equivalent to the Siemens Halothane Vaporizer HV950.
The Siemens Halothane Vaporizer HV950 was granted pre-market approval under 510(k) file number K841157.
#### MRI Compatibility Statement:
The Sevoflurane Vaporizer SV953 is not compatible for use in a MRI magnetic lield.
{7}------------------------------------------------
510(k) Sevoflurane Vaporizer SV953
Page2 off 2
. .
# (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NECESSARY)
Concurrence of CDRH, Office of Device Evaluation (ODE)
| | <img alt="Signature" src="signature.png"/> |
|-------------------------------------------------------------------|--------------------------------------------|
| (Division Sign-Off) | |
| Division of Cardiovascular, Respiratory, and Neurological Devices | |
| 510(k) Number | |
| Prescription Use | OR | Over-The-Counter Use |
|------------------|----|----------------------|
|------------------|----|----------------------|
(Per 21 CFR 801.109)Siemens-Elema AB. Electromedical Systems Division.
. . . .
------------------------------------------------------------------------------------------------------------------------------------------------------------------------------
. .
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.