The Sigma Delta 'T' vaporizer range is designed to attach to the back bar of an anaesthetic machine to provide controlled concentrations of anaesthesia gasses into a fresh gas. The fresh gas is then delivered into a patient breathing system, from where the anaesthesia ventilator and breathing circuit will then deliver this fresh gas mixture to the patient. The Vaporizer is calibrated in Vol % of the fresh gas flow and is compensated for changes in both fresh gas flows and temperature changes within the specified range.
Device Story
The Sigma Delta 'T' is an anesthetic vaporizer designed for attachment to the back bar of an anesthesia machine. It functions by introducing controlled concentrations of anesthetic gas into a fresh gas flow. The device features internal compensation mechanisms to account for variations in fresh gas flow rates and ambient temperature, ensuring consistent output calibrated in volume percentage (Vol %). The device is operated by clinicians in a hospital or surgical setting as part of an anesthesia delivery system. It benefits patients by enabling precise control over the concentration of anesthetic agents delivered via the breathing circuit.
Clinical Evidence
bench testing only
Technological Characteristics
The device is a mechanical anesthetic vaporizer designed for back-bar mounting. It utilizes flow and temperature compensation mechanisms to maintain output accuracy. It is a non-software, non-electronic device.
Indications for Use
Indicated for the delivery of controlled concentrations of anesthetic gases into a fresh gas flow for patient breathing systems during anesthesia administration.
Regulatory Classification
Identification
An anesthetic vaporizer is a device used to vaporize liquid anesthetic and deliver a controlled amount of the vapor to the patient.
K012924 — DATEX-OHMEDA TEC 7 ANESTHESIA VAPORIZER · Datex-Ohmeda, Inc. · Jan 30, 2002
K031027 — DATEX-OHMEDA TEC 7 ANESTHESIA VAPORIZER (ENFLURANE) · Datex-Ohmeda, Inc. · Jun 26, 2003
K172702 — Tec 820, Tec 850 · Datex-Ohmeda, Inc. · Jan 4, 2018
K051666 — GAV - GENERAL ANAESTHETIC VAPORIZER · General Anaesthetic Services Limited · Aug 23, 2005
K980489 — VAPAMASTA 6 · Medical Industrial Equipment · Aug 31, 1998
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image is a circular seal or logo. The seal contains the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter of the circle. Inside the circle is a stylized image of an eagle or bird-like figure, composed of three curved lines that suggest wings and a body. The overall design is simple and monochromatic.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
AUG - 9 2004
Mr. Alan Green Technical Director Penlon Limited Abingdon Science Park Barton Lane Abingdon United Kingdom OX14 3PH
Re: K041076
Trade/Device Name: Anaesthetic Vaporizer Regulation Number: 868.5880 Regulation Name: Anesthetic Vaporizer Regulatory Class: II Product Code: CAD Dated: July 12, 2004 Received: August 3, 2004
Dear Mr. Green:
We have reviewed your Section 510(k) premarket notification of intent to market the device We have reviewed your becaler researce is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device interstate conninetee pror to may 20, 10, 10, 10, 10, 10, 20, 10, 2011, 10, 2019, 10, 10, 10, 10, 10, 10, 10, 10, 10, 10, 10, 10, 10, 10, 10, 10, 10, 10, 10, 1 Alliendinents, on to devices that nave of the that do not require approval of a premarket the Federal Pood, Drug, and Cosments , therefore, market the device, subject to the general approval approvisions of the Act. The general controls provisions of the Act include controls provisions of the ristration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III II your device is classified (500 above) in the major regulations affecting (FMA), it may oc subject to back addin the Code of Federal Regulations, Title 21, Parts 800 to 898. In your device can be found in the Orderning your device in the Federal Register.
{1}------------------------------------------------
Page 2 -Mr. Green
Please be advised that FDA's issuance of a substantial equivalence determination does not Please be auvised that I Dr. 3 issualled or that your device complies with other requirements mean that i DA mas made a detect and regulations administered by other Federal agencies. of the Act of ally I edelar statues and registered to the limited to: registration You must comply with an the Pict 8 rog (21 CFR Part 801); good manufacturing practice allu listing (21 CF R Part 067), abolity systems (QS) regulation (21 CFR Part 820); and if requirements as set form in alle qt radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your Section 510(k) Fins letter will anow yourse of substantial equivalence of your device to a premarket notified.com - The Pre results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), If you desire specific at Compliance at (301) 594-4646. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the may obtain ourer generational and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html
Sincerely yours,
Carl
Chiu Lin, Ph.D.
Chiu Lin, Ph.D. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## ATTACHMENT 10
## Indications for Use
510(k) Number (if known):
K041076
Device Name:
Anaesthetic Vaporizer
Indications For Use:
The Sigma Delta 'T' vaporizer range is designed to attach to the back bar of an anaesthetic The Digma Bolla - 1- Taprolled concentrations of anaesthesia gasses into a fresh gas. The fresh gas is then delivered into a patient breathing system, from where the anaesthesia ventilator and breathing circuit will then deliver this fresh gas mixture to the patient.
The Vaporizer is calibrated in Vol % of the fresh gas flow and is compensated for changes in both fresh gas flows and temperature changes within the specified range.
The indicated use of the Sigma Delta 'T' has therefore not changed from that of the legally marketed device the Sigma Delta.
> . AND/OR Prescription Usc (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Clue O. Sulton
Sign-Off)
Division of Anesthesiology, General Hospital,
Infection Control, Dental Devices
510(k) Number: K041076
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches