MERCURY MEDICAL EXPIRATORY RESISTANCE EXERCISER, RESISTEX
K954492 · Mercury Medical · BWF · Oct 21, 1996 · Anesthesiology
Device Facts
Record ID
K954492
Device Name
MERCURY MEDICAL EXPIRATORY RESISTANCE EXERCISER, RESISTEX
Applicant
Mercury Medical
Product Code
BWF · Anesthesiology
Decision Date
Oct 21, 1996
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 868.5690
Device Class
Class 2
Attributes
Therapeutic
Intended Use
PERMITS RESISTANCE BREATHING DURING AEROSOL TREATMENT
Device Story
Mercury Expiratory Resistance Exerciser (RESISTEX) is a therapeutic spirometer accessory used in-line with aerosol treatment systems. Device features a rotating resistance orifice and one-way valve positioned between a nebulizer and an exhalation resistor. Operates by providing variable positive expiratory pressure (PEP) during aerosol drug delivery. Used in clinical or home settings to facilitate resistance breathing exercises during medication administration. Healthcare providers or patients use the device to combine aerosol therapy with expiratory resistance, potentially improving respiratory muscle function or drug deposition.
Clinical Evidence
No clinical data provided; substantial equivalence is based on descriptive comparison of technological characteristics and intended use.
Technological Characteristics
Materials: Styrene and Low-Density Polyethylene (LDPE). Mode of operation: Rotating resistance orifice with in-line nebulizer. Includes one-way valve between nebulizer and exhalation resistor. Provides variable resistance adjustment for positive expiratory pressure (PEP). Patient connection via LDPE mouthpiece.
Indications for Use
Indicated for patients requiring resistance breathing during aerosol treatment.
Regulatory Classification
Identification
An incentive spirometer is a device that indicates a patient's breathing volume or flow and that provides an incentive to the patient to improve his or her ventilation.
Predicate Devices
Circulaire™ Aerosol Drug Delivery System (Westmed)
Related Devices
K972042 — PARI PEP THERAPY SYSTEM · Pari Holding Co. · Aug 5, 1997
K181660 — Acapella Choice Blue Vibratory PEP Device · Smiths Medical · Oct 24, 2019
K991788 — SALTER LABS PEP DEVICE · Salter Labs · Aug 6, 1999
K983308 — RC CORNET, MODEL #44F50 · Pari Holding Co. · Jun 4, 1999
{0}
DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
Food and Drug Administration
9200 Corporate Boulevard
Rockville MD 20850
OCT 21 1996
Ms. Kathy L. Hann
QA/Regulatory Manager
Mercury Medical®
11300A-49th Street North
Clearwater, Florida 34622-4800
Re: K954492
Trade Name: Mercury Expiratory Resistance Exerciser
Regulatory Class: II
Product Code: 73BWF
Dated: July 9, 1996
Received: July 17, 1996
Dear Ms. Hann:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Good Manufacturing Practice for Medical Devices: General (GMP) regulation (21 CFR Part 820) and that, through periodic GMP inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.
{1}
Page 2 - Ms. Kathy L. Hann
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for *in vitro* diagnostic devices), please contact the Office of Compliance at (301) 594-4648. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or at (301) 443-6597.
Sincerely yours,
Thomas J. Callahan
Thomas J. Callahan, Ph.D.
Director
Division of Cardiovascular, Respiratory, and Neurological Devices
Office of Device Evaluation
Center for Devices and Radiological Health
{2}
OCT-18-1996 08:35 FROM MERCURY MEDICAL
TO 13015943076 P.02
MERCURY MEDICAL®
K954492
# REVISED ENCLOSURE I
510 (k) SUMMARY
OCT 21 1996
- Mercury Expiratory Resistance Exerciser (RESISTEX®)
- Common Name - Expiratory Resistance Exerciser and Adapter
- Classification Name - Spirometer, Therapeutic
This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of the Safe Medical Devices Act 1990 and 21CFR897.92. The Mercury Expiratory Resistance Exerciser, when placed in-line with an aerosol treatment, permits resistance breathing during aerosol treatment in a manner consistent with the Circulaire™ Aerosol Drug Delivery System by Westmed.
## TABLE OF COMPARISON
| | MERCURY
EXPIRATORY RESISTANCE
EXERCISER | WESTMED
CIRCULAIRE™ AEROSOL
DRUG DELIVERY SYSTEM |
| --- | --- | --- |
| INTENDED USE | PERMITS RESISTANCE
BREATHING DURING
AEROSOL TREATMENT | PERMITS RESISTANCE
BREATHING DURING
AEROSOL TREATMENT |
| MATERIALS
MODE OF OPERATION | STYRENE & LDPE
ROTATING RESISTANCE
ORIFICE WITH IN-LINE
NEBULIZER | POLYPROPYLENE
ROTATING RESISTANCE
ORIFICE WITH IN-LINE
NEBULIZER |
| ONE WAY VALVE | BETWEEN NEBULIZER AND
EXHALATION RESISTOR | BETWEEN NEBULIZER AND
EXHALATION RESISTOR |
| PROVIDES POSITIVE
EXPIRATORY PRESSURE
BENEFIT DURING
TREATMENT | HAS VARIABLE
RESISTANCE ADJUSTMENT | HAS VARIABLE RESISTANCE
ADJUSTMENT |
| PATIENT CONNECTION | LDPE MOUTHPIECE | LDPE MOUTHPIECE |
| PACKAGING | SINGLE PACKED 20/BOX | SINGLE PACKED 25/BOX |
Arthur J. Ward
10/18/96
Date
11300A-49th Street North • Clearwater, Florida 34622-4800 • 813-573-0088
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches