CIRQLATOR INTRATHORACIC PRESSURE REGULATOR (ITPR)

K070490 · Advanced Circulatory Systems, Inc. · BWF · Jun 20, 2007 · Anesthesiology

Device Facts

Record IDK070490
Device NameCIRQLATOR INTRATHORACIC PRESSURE REGULATOR (ITPR)
ApplicantAdvanced Circulatory Systems, Inc.
Product CodeBWF · Anesthesiology
Decision DateJun 20, 2007
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 868.5690
Device ClassClass 2
AttributesTherapeutic

Intended Use

The CirQlator™ Intrathoracic Pressure Regulator is indicated for the temporary decrease in intrathoracic pressure to increase blood circulation, as prescribed by a physician, licensed practitioner, or qualified technician. Recommended duration of use is up to four hours.

Device Story

CirQlator ITPR interfaces with patient airway and ventilation source; connects to vacuum source to provide continuous low-level vacuum; vacuum interrupted during positive pressure breaths from manual resuscitator. Device decreases intrathoracic pressure between ventilations; increases vital organ perfusion; decreases intracranial pressure. Used in hospital and pre-hospital (EMS) settings; operated by physicians, licensed practitioners, or qualified technicians. Output is mechanical pressure regulation; healthcare providers use device to manage hemodynamics in shock, cardiac arrest, or low blood flow states. Benefits patient by increasing cardiac preload and improving circulation.

Clinical Evidence

Bench testing compared performance features and functionality against predicate devices. Animal testing was performed to demonstrate the device functions as intended. No human clinical data provided.

Technological Characteristics

Intrathoracic pressure regulator; interfaces with airway and external vacuum/ventilation sources. Operates by applying continuous low-level vacuum, interrupted by positive pressure ventilation. Mechanical device; no software or electronic components described.

Indications for Use

Indicated for temporary decrease in intrathoracic pressure to increase blood circulation in patients needing assisted ventilation suffering from poor circulation, low blood pressure, or insufficient cardiac preload. Contraindicated in patients with pneumothorax, hemothorax, hypertension, or uncontrolled bleeding.

Regulatory Classification

Identification

An incentive spirometer is a device that indicates a patient's breathing volume or flow and that provides an incentive to the patient to improve his or her ventilation.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ | | JUN 20 2007 Non-Confidential Summary of Safety and Effectiveness<br>Page 1 of 2<br>14-Jun-07 | |-----------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------| | Advanced Circulatory Systems, Inc.<br>7615 Golden Triangle Drive, Suite A<br>Eden Prairie, MN 55344 | Tel - 612-986-3917<br>Fax -952-942-8336 | | Official Contact: | Keith Lurie, MD, Chief Medical Officer | | Proprietary or Trade Name: | CirQlator™ Intrathoracic Pressure Regulator | | Common/Usual Name: | Spirometer, therapeutic (incentive) | | Classification Name: | Spirometer, therapeutic (incentive) | | Device: | Intrathoracic Pressure Regulator (ITPR) | | Predicate Devices: | ResQPOD® Circulatory Enhancer - K022906, K033401<br>Bird Mark 8 ventilator – pre-amendment | # 11 188 9 11 2007 Non-Confidential Summary of Safety and Effectiveness #### Device Description: The CirQlator"M is an intrathoracic pressure regulator intended to interface with the airway of a patient and a ventilation source. A vacuum source is connected to the ITPR that provides continuous low-level vacuum except when a positive pressure breath is given by a ventilation source, e.g. manual resuscitator. The applied vacuum decreases the intrathoracic pressure. When interposed between positive pressure ventilations, a decrease in intrathoracic pressure has been shown to increase vital organ perfusion and decrease intracranial pressure during states of shock, cardiac arrest, and other low blood flow states in animal studies. | Indications for Use: | | |----------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Indicated Use - | The CirQlator™ Intrathoracic Pressure Regulator is indicated for the temporary<br>decrease in intrathoracic pressure to increase blood circulation, as prescribed by a<br>physician, licensed practitioner, or qualified technician. Recommended duration of<br>use is up to four hours. | | Patient Population - | The CirQlator™ Intrathoracic Pressure Regulator may benefit people needing<br>assisted ventilation that suffer from states of poor circulation, low blood pressure,<br>or insufficient cardiac preload that may be reflected by low blood pressure. The<br>device is not intended for those patients who would not benefit from an increase in<br>cardiac preload. | | Environment of Use - | Hospital, pre-hospital (including EMS) | | Contraindications - | persons with pneumothorax;<br>persons with hemothorax:<br>persons who are hypertensive;<br>persons with uncontrolled bleeding. | {1}------------------------------------------------ ## Non-Confidential Summary of Safety and Effectiveness Page 2 of 2 14-Jun-()7 ### Performance testing: Comparative bench testing was performed for the proposed device and the predicates and the performance features and functionality where found to be comparable. In addition, animal testing was performed to demonstrate the proposed device performed as intended. ### Comparison to Legally Marketed Predicate Devices: The ITPR is viewed as substantially equivalent to the following predicate devices - - ResQPOD Circulatory Enhancer (K022906, K033401) ● - . Bird Mark 8 respirator (pre-amendment) There are no significant differences that affect the safety or effectiveness of the intended device as compared to the predicate devices. {2}------------------------------------------------ Public Health Service Image /page/2/Picture/2 description: The image shows the seal of the Department of Health & Human Services (HHS) of the United States. The seal features an eagle-like emblem with three stylized legs or supports, symbolizing stability and support. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the emblem, indicating the department's name and national affiliation. Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 Advanced Circulatory Systems, Incorporated C/O Mr. Paul Dryden President ProMedic, Incorporated 3460 Pointe Creek Court # 102 Bonita Springs, Florida 34134-2015 JUN 2 U 2007 Re: K070490 Trade/Device Name: CirQlator" Intrathoracic Pressure Regulator (ITPR) Regulation Number: 21 CFR 868.5690 Regulation Name: Incentive Spirometer Regulatory Class: II Product Code: BWF, JOW Dated: June 14, 2007 Received: June 15, 2007 Dear Mr. Dryden: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {3}------------------------------------------------ #### Page 2 – Mr. Dryden Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510/k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801). please contact the Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html. Sincerely yours, Chiu Lin, Ph.D. Chiu Lin, Ph.D. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {4}------------------------------------------------ ## Indications for Use Statement Page 1 of 1 510(k) Number: KOJO 490 490 (To be assigned) Device Name: CirQlator"M Intrathoracic Pressure Regulator (ITPR) Indications for Use: The CirQlator™ Intrathoracic Pressure Regulator is indicated for the temporary decrease in intrathoracic pressure to increase blood circulation, as prescribed by a physician, licensed practitioner, or qualified technician. Recommended duration of use is up to four hours. Prescription Use XX (Part 21 CFR 801 Subpart D) or Over-the-counter use (21 CFR 807 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) Sion Sign Off) Culsion of Anesthesiology, General Hospital, Imaction Control, Dental Devices K(k) Number. K070490
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...