The MEDISIZE GOLD Heater and Booster T-Piece is intended to feed additional heat and moisture into the system, through the T-Piece or via the Medisize Gold, during anaesthesia and respiratory care, specifically, when the use of an HME alone is no longer adequate due to the patients moisture deficit.
Device Story
The Medisize Gold Heater and Booster T-Piece is a respiratory gas humidifier system; combines a reusable heater with a disposable single-use T-Piece component. The device is placed in the respiratory gas stream to provide active humidification and heating; supplements HME function when patient moisture deficit exceeds HME capacity. The system is self-regulating; requires minimal clinician intervention. It administers heat and humidity to the anesthesia gas stream at approximately 33°C and 33 mg/L. Used in clinical settings (anesthesia/respiratory care) by healthcare professionals. Benefits include improved respiratory gas conditions, preventing complications associated with inadequate moisture during ventilation.
Clinical Evidence
No clinical testing was performed. Substantial equivalence is supported by bench testing, including electrical safety (IEC 60601-1), electromagnetic compatibility (IEC 60601-1-2), and performance testing per ISO 8185.
Technological Characteristics
Respiratory gas humidifier; reusable heater with disposable T-Piece. Operates at ~33°C and 33 mg/L humidity. Power supply: 100-240V~/50-60Hz input, 9V/1.0A output. Complies with IEC 60601-1 and IEC 60601-1-2. Materials identical to predicate (K070714, K052615).
Indications for Use
Indicated for patients undergoing anesthesia or respiratory care requiring supplemental heat and moisture when a Heat and Moisture Exchanger (HME) alone is inadequate due to moisture deficit.
Regulatory Classification
Identification
A respiratory gas humidifier is a device that is intended to add moisture to, and sometimes to warm, the breathing gases for administration to a patient. Cascade, gas, heated, and prefilled humidifiers are included in this generic type of device.
K990614 — HME-BOOSTER · King Systems Corp. · May 23, 2000
K071958 — THERA-HEAT HEATED HUMIDIFIER AND ADULT DUAL AND SINGLE LIMB HEATED WIRE VENTILATOR BREATHING CIRCUITS · Smiths Medical Asd, Inc. · Nov 21, 2007
K983783 — HEATED HUMIDIFIER · Pegasus Research Corp. · Oct 24, 2000
Submission Summary (Full Text)
{0}------------------------------------------------
K12885
Medisize B.V. Traditional 510(k) Premarket Submission Medisize Gold Heater and Booster T-Piece
# Section 5 - 510(k) Summary for Medisize Gold Heater and Booster T-Piece
- 1. Submission Sponsor
Medisize B.V. Edisonstraat 1 Hillegom 2181 AB The Netherlands Tel: +31 252 576 815 Fax: +31 252 519 825 Contact name: Ms Desire Santiago Title: Regulatory Affairs Manager Email: desire.santiago@phillipsmedisize.com
AUG 2 9 2013
### 2. Submission Correspondent
Emergo Group Prinsessegracht 20 2514 AP, The Hague THE NETHERLANDS Cell Phone: 508.838.9139 Office Phone: +31 70 345 8570 Fax: +31 70 346 7299 Contact: Richard A. VINCINS, Vice President QA Email: PMTeurope@emergogroup.com
#### 3. Date Prepared
29 August 2013
### 4. Device Name
Trade/Proprietary Name: Medisize Gold Heater and Booster T-Piece . Common/Usual Name: Respiratory gas humidifier Classification Name: Respiratory Gas Humidifier Classification Regulation: 21 CFR 868.5450 Classification Panel: Anesthesiology Product Code: BTT Device Class: Class II
FDA Establishment Registration #: 3002598726
{1}------------------------------------------------
Medisize B.V. Traditional 510(k) Premarket Submission Medisize Gold Heater and Booster T-Piece
# 5. Predicate Devices
MEDISIZE GOLD Heater and Booster manufactured by Medisize B.V., 510(k) number: K070714
#### 6. Device Description
The reusable MEDISIZE GOLD Heater in combination with the disposable single use Medisize Gold Booster T-Piece (or Medisize Gold, 510k number K052615) combines the simplicity of a Heat and Moisture Exchanger (HME) with the features of active humidification. Use of the MEDISIZE GOLD Heater in combination with a HME increases the number of options for humidifying and heating respiratory gases, greatly expanding the field of application for HMEs. When used in combination with a HME, the MEDISIZE GOLD Heater improves the absolute humidity, the relative humidity and the temperature of the respiratory gases. Use of the MEDISIZE GOLD Heater makes it possible to administer additional moisture and heat in a simple way, which is self-regulating. As a result, use of the MEDISIZE GOLD Heater is not particularly labor or knowledge intensive. The ease of use makes it a reliable, stable and effective method to administer additional moisture and heat to patients when required.
# 7. Intended Use
The MEDISIZE GOLD Heater and Booster T-Piece is intended to feed additional heat and moisture into the system, through the T-Piece or via the Medisize Gold, during anaesthesia and respiratory care, specifically, when the use of an HME alone is no longer adequate due to the patients moisture deficit.
#### 8. Technological Characteristics and Substantial Equivalence
The following table compares the predicate Medisize Gold Heater and Booster (with T-Piece) to the modified Medisize Gold Heater and Booster T-Piece with respect to intended use, technological characteristics and principles of operation, providing more detailed information regarding the basis for the determination of substantial equivalence.
{2}------------------------------------------------
Medisize B.V. Traditional 510(k) Premarket Submission Medisize Gold Heater and Booster T-Piece
# Table 5-A: Device Comparison Table
| Manufacturer | Medisize B.V. | Medisize B.V. |
|--------------------------------|----------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------|
| Trade Name | Medisize Gold Heater and<br>Booster T-Piece | Medisize Gold Heater and<br>Booster (with T-Piece)<br>(Predicate device) |
| 510(k) Number | K122885 | K070714 |
| Product Code | BTT | BTT |
| Regulation<br>Number | 21 CFR 868.5450 | 21 CFR 868.5450 |
| Regulation<br>Name | RESPIRATORY GAS<br>HUMIDIFIER | RESPIRATORY GAS<br>HUMIDIFIER |
| Indications for<br>use: | Feeds additional heat and<br>moisture into the respiratory<br>system when the use of a<br>HME alone is not adequate. | Feeds additional heat and<br>moisture into the respiratory<br>system when the use of a<br>HME alone is not adequate. |
| Overall Design | Re-usable heater used with<br>single use disposable<br>component in gas stream | Re-usable heater used with<br>single use disposable<br>component in gas stream |
| Mode of<br>Operation | Can administer heat and<br>humidity to anaesthesia gas<br>stream at temperature of<br>approximately 33°C, 33 mg/L | Can administer heat and<br>humidity to anaesthesia gas<br>stream at temperature of<br>approximately 33°C, 33 mg/L |
| Complies with<br>IEC 60601-1 | Yes | Yes |
| Complies with<br>IEC 60601-1-2 | Yes | Yes |
| Power Supply | FW7660M/09<br>Input:100-240V~/50-<br>60Hz/250mA<br>Output: 9V --- /1,0A | FW7555M/12<br>Input:100-240~/50-<br>60Hz/400mA<br>Output: 12V --- /1,25A |
| Operating<br>Conditions | See Power Supply | See Power Supply |
| Input Cable | See Power Supply | See Power Supply |
| Input Cable<br>Adapter | See Power Supply | See Power Supply |
{3}------------------------------------------------
# 9. Non-Clinical Testing
The device has been tested for electrical safety according to IEC 60601-1 and IEC 60601-1-2 and passed all the testing. Details of the specific performance standards used are listed below and test results summarized in Table 1:
| Standard | Standard Title | Test Results |
|--------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------|
| IEC 60601-1:2005 | Medical Electrical Equipment – Part 1:<br>General requirements for basic safety and<br>essential performance | Passed |
| IEC 60601-1-2:2007 | Medical Electrical Equipment – Part 1-2:<br>General requirements for basic safety and<br>essential performance – Collateral<br>standard: Electromagnetic Compatibility –<br>Requirements and Tests | Passed |
### Table 5-B: Summary of Non-Clinical Test Results
The device was tested for the performance testing according to the voluntary standard ISO 8185, Respiratory Tract Humidifiers for Medical Use - Particular requirements for respiratory humidification systems. The testing supports the acceptable results of the device according to ISO 8185 requirements.
No biocompatibility testing was performed on the Medisize Gold Heater and Booster T-Piece as part of this submission because the materials are identical to the Medisize Gold Heater and Booster (with T-Piece) cleared under the previous 510(k) clearances K070714 and K052615.
# 10. Clinical Testing
There was no clinical testing required to support the medical device as the indications for use is equivalent to the predicate device. The substantial equivalence of the device is supported by the non-clinical testing. The verification and validation testing of the device electrical safety and EMC testing of the device was found to be acceptable and supports the claims of substantial equivalence.
# 11. Conclusion
By definition, a device is substantially equivalent to a predicate device when the device has the same intended use and the same technological characteristics as the previously cleared predicate device. Or the device has the same intended use and different technological characteristics that can be demonstrated that the device is substantially equivalent to the predicate device, and that the new device does not raise different questions regarding its safety and effectiveness as compared to the predicate device.
The modified Medisize Gold Heater and Booster T-Piece, as designed and manufactured, is determined to be substantially equivalent to the referenced predicate device.
{4}------------------------------------------------
Image /page/4/Picture/1 description: The image shows the logo for the Department of Health & Human Services (HHS). The logo features a stylized symbol resembling a bird or abstract human figure, composed of three curved lines. The symbol is positioned to the right of a circular seal containing the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
August 29, 2013
Medisize BV C/O Mr. Richard Vincins Emergo Group 816 Congress Avenue, Suite 1400 Austin, TX 78701
Re: KI22885
Trade/Device Name: Medisize Gold Heater and Booster T-Piece Regulation Number: 21 CFR 868.5450 Regulation Name: Respiratory Gas Humidifier Regulatory Class: II Product Code: BTT Dated: August 7, 2013 Received: August 8, 2013
Dear Mr. Vincins:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or 10 devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{5}------------------------------------------------
Page 2 - Mr. Vincins
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (2) CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers. International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours.
Kwame 0. Ulmer Ar. And an -5
Kwame Ulmer M.S. Acting Division Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
# 85
Medisize B.V. Traditional 510(k) Premarket Submission Medisize Gold Heater and Booster T-Piece
# Section 4 - Indications for Use Statement
510(k) Number (if known): Not known
Device Name: MEDISIZE GOLD Heater and Booster T-Piece
Indications for Use:
The MEDISIZE GOLD Heater and Booster T-Piece is intended to feed additional heat and moisture into the system, through the T-Plece or via the Medisize Gold, during anaesthesia and respiratory care, specifically, when the use of an HME alone is no longer adequate due to the patients moisture deficit.
X _____ (Part 21 CFR 801 Subpart D) Prescription Use _ AND/OR Over-The-Counter Use _____________________________ (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Anya C Harry
Digitally signed by Anya C. Ham XX: c=US. o=U.S. Government, ou=HHS DA. ou=People, comAnva C. Harr 0.9 2342.19200300.100.1.1=0011315. Date: 2013.08.28 16:48:48 -04'0
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.