CuffGuard

K243562 · Idmed · BSK · Aug 12, 2025 · Anesthesiology

Device Facts

Record IDK243562
Device NameCuffGuard
ApplicantIdmed
Product CodeBSK · Anesthesiology
Decision DateAug 12, 2025
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 868.5750
Device ClassClass 2
AttributesTherapeutic, Pediatric

Indications for Use

The CuffGuard is an endotracheal tube cuff controller designed to continuously monitor and maintain user-defined endotracheal tube cuff pressure for patients (adults and pediatrics) who require mechanical ventilation and who are intubated with endotracheal tube including cuff in the ICU, OR and for intra-hospital transport.

Device Story

CuffGuard is an endotracheal tube cuff controller for ICU, OR, and intra-hospital transport use. Operated by trained physicians or nurses, it continuously monitors and automatically maintains user-defined cuff pressure. The device utilizes a pressure sensor and air pump to adjust cuff pressure via embedded software. It provides visual and audible alarms for leaks or pressure deviations. Output is displayed to the clinician to ensure cuff pressure remains within a safe, user-defined range (5–40 cmH2O), preventing complications associated with improper cuff inflation. The device is portable, powered by battery or AC, and intended for use on adult and pediatric patients.

Clinical Evidence

No clinical or animal studies were conducted. Substantial equivalence is supported by bench testing, including electrical safety (IEC 60601-1), electromagnetic compatibility (IEC 60601-1-2), and software verification and validation.

Technological Characteristics

Portable, software-controlled pressure controller. Features: pressure sensor, air pump, visual/audible alarms, battery/AC power. Pressure range: 5–40 cmH2O; control accuracy: +/- 2 cmH2O. Compliant with IEC 60601-1, IEC 60601-1-2, IEC 60601-1-6, and IEC 60601-1-8. No patient-contact materials.

Indications for Use

Indicated for adult and pediatric patients requiring mechanical ventilation who are intubated with an endotracheal tube. Contraindicated if the endotracheal tube itself is contraindicated.

Regulatory Classification

Identification

An inflatable tracheal tube cuff is a device used to provide an airtight seal between a tracheal tube and a patient's trachea.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0} FDA U.S. FOOD & DRUG ADMINISTRATION August 12, 2025 Idmed Perrine Moelle Quality & Regulatory Affairs Director 3 rue John Maynard Keynes Marseille, 13013 France Re: K243562 Trade/Device Name: CuffGuard Regulation Number: 21 CFR 868.5750 Regulation Name: Inflatable Tracheal Tube Cuff Regulatory Class: Class II Product Code: BSK Dated: July 7, 2025 Received: July 7, 2025 Dear Perrine Moelle: We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" U.S. Food & Drug Administration 10903 New Hampshire Avenue Silver Spring, MD 20993 www.fda.gov {1} K243562 - Perrine Moelle Page 2 (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download). Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181). Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050. All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems. For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. {2} K243562 - Perrine Moelle Page 3 See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, Bradley Q. Quinn -S Bradley Quinn Assistant Director DHT1C: Division of Anesthesia, Respiratory, and Sleep Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT, and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {3} DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration Indications for Use Form Approved: OMB No. 0910-0120 Expiration Date: 07/31/2026 See PRA Statement below. Submission Number (if known) K243562 Device Name CuffGuard Indications for Use (Describe) The CuffGuard is an endotracheal tube cuff controller designed to continuously monitor and maintain user-defined endotracheal tube cuff pressure for patients (adults and pediatrics) who require mechanical ventilation and who are intubated with endotracheal tube including cuff in the ICU, OR and for intra-hospital transport. Type of Use (Select one or both, as applicable) ☑ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C) ## CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov > "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {4} idmed an eye on your patient # 510 (k) Summary This summary of 510(k) information is being submitted in accordance with the requirements of 21 CFR 807.92. # 1. Submitter Submitter: IDMED Hotel Technoptic 2 rue Marc Donadille 13013 Marseille France +33 (0)974 980 541 Establishment Registration Number: N° DUNS 260233256 Contact person: Perrine MOLLE Quality & Regulatory Affairs Director E-mail : p.molle@idmed.fr Date Summary Prepared: 2025/07/03 # 2. Device Trade Name: CuffGuard (references CG-MU_ID and CG-MU_DR) Common/Usual Name: Tracheal tube cuff Pressure Controller Regulation Name: Inflatable tracheal tube cuff Classification Regulation: 21 CFR 868.5750 Product Code: BSK Regulatory Class: Class II # 3. Predicate device Predicate Device: Intellicuff, manufactured by Hamilton Medical, K150893 # 1.1. Device description The CuffGuard is an endotracheal tube cuff controller designed to continuously monitor and maintain user-defined endotracheal tube cuff pressure for patients (adults and pediatrics) who require mechanical 510(k) - CuffGuard {5} idmed an eye on your patient ventilation and who are intubated with endotracheal tube including cuff in the ICU, OR and for intra-hospital transport. It is used by health professionals (doctor or nurse) specifically trained in this tool. The CuffGuard has a pressure sensor and software that can monitor and automatically adjust the cuff pressure. The accessories associated include: - Connection tube with filter - Power supply - Micro USB connection cable - Fixation clamp and its tilt block) At a high level, the comparison of the subject and predicate devices is based on the following table: | Features | CuffGuard | Intellicuff | | --- | --- | --- | | Product code | BSK | BSK | | Regulation number | 868.5750 | 868.5750 | | Class | II | II | | Indication for use | | | | Intended use | The CuffGuard is an endotracheal tube cuff controller designed to continuously monitor and maintain user-defined endotracheal tube cuff pressure for patients (adults and pediatrics) who require mechanical ventilation and who are intubated with endotracheal tube including cuff. | The IntelliCuff device is intended to continuously measure and automatically maintain the user-set cuff pressure of an endotracheal tube (ETT) or tracheostomy tube (TT) during mechanical ventilation | | Environment of use | In a professional health care setting such as hospital or health care facility: in an Intensive Care Unit, an Operating Room and during intra-hospital transportation. | Intensive care units, operating rooms, and during interhospital transport: - In the intensive care ward or in the recovery room - In the operation room during intubation narcosis - For emergency medical care or primary care - During transport within and outside of the hospital During transfer by rescue vehicles, ship, jet, or helicopter | 510(k) - CuffGuard {6} idmed an eye on your patient | Features | CuffGuard | Intellicuff | | --- | --- | --- | | Product code | BSK | BSK | | Regulation number | 868.5750 | 868.5750 | | Class | II | II | | User | Health professional (doctor or nurse) specifically trained in this tool | Trained and qualified professionals under the supervision of a doctor and within the limits of the technical specifications indicated | | Patient population | Adult and pediatric | Adults, pediatrics, and neonates, who are intubated with ETT or TT | | Contraindication | None known. Note that any contraindication to the use of the endotracheal tube will result in a contraindication to the use of the CuffGuard | None identified in the IFU | | **Principle of operation and performances** | | | | Principe of operation | Software-controlled pressure supervision and automatic pressure adjustment Pressure sensor Use of ambient / room air | Software-controlled pressure supervision and automatic pressure adjustment Pressure sensor Use of ambient / room air | | Type of control | Automatic | Automatic | | Pressure range | 5 – 40 cmH2O | 5 – 50 cmH2O | | Accuracy of the control | +/- 2cmH2O | +/- 2cmH2O | | Adjustment of accuracy | +/- 1cmH2O | +/- 1cmH2O | | Display of accuracy | +/- 1cmH2O | +/- 1cmH2O | | **Technology characteristics** | | | | Leak alarm | Yes | Yes | | Pressure alarm | Yes | Yes | | Stop sound alarm | Yes | Yes | | Audible alarm | Yes | Yes | | Visual alarm | Yes | Yes | | Deflate Cuff | Yes | Yes | | Inflate Cuff | Yes | Yes | | Time-limited hold | Yes | Yes | | Choice of unit | Yes | Yes | | Energy | Battery and AC power supply | Battery and AC power supply | | Portable device | Yes | Yes | 510(k) – CuffGuard 5-3 {7} idmed an eye on your patient | Features | CuffGuard | Intellicuff | | --- | --- | --- | | Product code | BSK | BSK | | Regulation number | 868.5750 | 868.5750 | | Class | II | II | | General safety | Conpliant to IEC 60601-1, IEC 60601-1-2, IEC 60601-1-8 | Conpliant to IEC 60601-1, IEC 60601-1-2, IEC 60601-1-8, IEC 60601-1-12 | | Biological characteristics | | | | Patient contact material | NA, no patient contact | NA, no patient contact | ## 1.2. Indications for use The CuffGuard is an endotracheal tube cuff controller designed to continuously monitor and maintain user-defined endotracheal tube cuff pressure for patients (adults and pediatrics) who require mechanical ventilation and who are intubated with endotracheal tube including cuff in the ICU, OR and for intra-hospital transport. The following differences exist between subject and predicate device: - The subject device is not indicated to be used on tracheostomy tube (TT) - The subject device is not indicated to be used on neonate population of patient - The subject device is not indicated to be used during transfer by rescue vehicles, ship, jet or helicopter However, these differences do not raise questions of safety and effectiveness, because: - the indication not applicable for the subject device (Tracheostomy tube), - the population not applicable for the subject device (neonate), - the environment of use not applicable for the subject device (transfer by rescue vehicles, ship, jet, or helicopter), are not considered for the comparison. The subject device and its predicate device are both used in the same clinical conditions or purposes (adult or pediatric patients requiring mechanical ventilation for whom the cuff pressure of an endotracheal tube (ETT) is monitored) and are used by the same kind of users (trained medical professional). They have the same principle of operation, using a pressure sensor and air pump). ## 4. Comparison of technological characteristics with the predicate device The technological principle of the subject device is to maintain and automatically adjust thanks to embedded software, the cuff pressure of intubating tubes using a pressure sensor and a pump. The following technological difference exists between subject device and its predicate device: 510(k) – CuffGuard 5-4 {8} idmed an eye on your patient - The pressure range of the subject device is not as wide as its predicate device However, this difference is supported by performance data (Software verification or safety test reports). Furthermore, this difference does not raise different questions of safety and effectiveness and does not negatively affect the clinical performances for the following reason: - Difference of pressure range has been evaluated on the SW verification and validation. Furthermore, this difference is not clinically significant as the use of pressure range until 40 cm H2O appears as a realistic physiological limit and as a common clinical practice used as shown on various publications. # 5. Performance data The following performance data were provided in support of the substantial equivalence determination. ## Electrical safety and electromagnetic compatibility (EMC) Electrical safety and EMC testing were conducted on the CuffGuard device. The system complies with the IEC 60601-1, IEC 60601-1-6 and IEC 60601-1-8 standards for safety and the IEC 60601-1-2 standard for EMC. ## Software Verification and Validation Testing Software verification and validation testing were conducted and documentation was provided as recommended by FDA’s Guidance for Industry and FDA Staff, “Content of Premarket Submissions for Device Software Functions.” The documentation level for CuffGuard software is enhanced. ## Animal and Clinical Studies No animal or clinical testing was required to demonstrate the substantial equivalence of this device to its predicate, nor its safety and effectiveness. # 6. Conclusions Based on its intended use, design principles, and technological characteristics, the CuffGuard device was found to be as safe, as effective, and performs comparably to the predicate device. The technological differences identified do not raise new questions of safety and effectiveness as the non-clinical and clinical literature data support the safety of the device and the hardware and software verification and validation demonstrate that the CuffGuard device should perform as intended in the specified use conditions. 510(k) – CuffGuard
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...