K961126 · Devilbiss Health Care, Inc. · NFB · Oct 21, 1996 · Anesthesiology
Device Facts
Record ID
K961126
Device Name
PULSE DOSE SERIES
Applicant
Devilbiss Health Care, Inc.
Product Code
NFB · Anesthesiology
Decision Date
Oct 21, 1996
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 868.5905
Device Class
Class 2
Attributes
Therapeutic
Intended Use
The devices assist patients, who require prescribed doses of supplemental gases such as oxygen, to conserve gas from the various gaseous cylinders or liquid oxygen reservoirs. By allowing oxygen to flow only from the cylinder or liquid oxygen reservoirs during the inspiratory segment of a patient's breathing cycle, conservation occurs.
Device Story
Pulse Dose Series oxygen management system; provides bolus of oxygen at start of inhalation; conserves gas by restricting flow to inspiratory phase of breathing cycle; battery-powered; includes battery level indicator; microprocessor-controlled; repackaged version of OMS 20 and 50; intended for patient use to improve oxygen delivery efficiency and device usability.
Clinical Evidence
Bench testing only. Comparison of dosage delivery between Pulse Dose Series and predicate devices (OMS 20/50) confirms identical performance.
Technological Characteristics
Battery-powered; microprocessor-controlled; pneumatic oxygen delivery system; includes battery level indicator; repackaged for improved cosmetics and assembly.
Indications for Use
Indicated for patients requiring prescribed supplemental oxygen therapy to conserve gas from cylinders or liquid oxygen reservoirs.
Regulatory Classification
Identification
A noncontinuous ventilator (intermittent positive pressure breathing-IPPB) is a device intended to deliver intermittently an aerosol to a patient's lungs or to assist a patient's breathing.
Predicate Devices
OMS 20
OMS 50
Related Devices
K071928 — THE INVACARE FLYER, MODEL IPC 100 · Invacare Corp. · Dec 12, 2007
K023420 — PULSED OXYGEN CONSERVING DEVICE - EVERY BREATH, POCDEB, REACTEB · Medical Electronic Devices, Inc. · Oct 25, 2002
K962944 — TRANSTRACHEAL DEMAND OXYGEN CONTROLLER (DOC-2000) · Transtracheal Systems · Oct 23, 1996
K081468 — DEVILBISS PORTABLE OXYGEN CONCENTRATOR, 306 SERIES · Sunrise Medical · Aug 25, 2008
K051540 — INOVO, INC. INDEPENDENCE OXYGEN CONSERVING REGULATOR · Inovo, Inc. · Sep 21, 2005
Submission Summary (Full Text)
{0}
10/09/96 WED 14:25 FAX 814 443 7571
DeVILBISS ENGINEERING
003
DeVILBISS
a division of
SUNRISE MEDICAL
DeVilbiss Health Care, Inc.
1200 East Main Street
P.O. Box 635
Somerset, PA 15501-0635
U.S.A.
(814) 443-4881
Fax: (814) 443-7572
K961126
OCT 21 1996
# 16.0 510 (k) Summary
## Summary of Safety and Effectiveness
**Firms name:** DeVilbiss Health Care, Inc.
Pulsair Division
P.O. Box 2609
4106 Avenue D
Fort Pierce, Fla. 34954
**Submitter's name:** Gregory W. Good
**Phone:** (814) 443-7466
**Fax:** (814) 443-7571
**Date of Preparation:** 9 October 1996
**Trade name:** Oxygen Management Systems
**Common name:** Pulse Dose Series
**Classification name:** Non-Continuous Ventilators 73BZD (21 CFR 868.5905)
The Pulse Dose Series devices are fundamentally repackaged versions of the OMS 20 and 50, DeVilbiss' current Oxygen Management Systems. The devices provide a bolus of oxygen to a patient at the start of inhalation. The devices assist patients, who require prescribed doses of supplemental gases such as oxygen, to conserve gas from the various gaseous cylinders or liquid oxygen reservoirs. By allowing oxygen to flow only from the cylinder or liquid oxygen reservoirs during the inspiratory segment of a patient's breathing cycle, conservation occurs.
The gas dosage methodology, oxygen delivery specifications, and performance of the devices in the Pulse Dose Series are identical to those of the OMS 20 and 50. This is evidenced in the various comparison matrices and testing reports contained in this 510(k). This equivalency is also evidenced in the non-clinical testing which reveals the same dosage delivery when comparing the predicate devices with the devices in the Pulse Dose Series. The device's performance to the end user are identical as compared to the OMS 20 and 50.
Exhibit 1A
{1}
10/09/96 WED 14:26 FAX 814 443 7571
DeVILBISS ENGINEERING
004
When compared to the predicate devices, the OMS 20 and 50, the devices in the Pulse Dose Series utilize the same type of energy source, and there are no significant changes in the materials or features.
All the devices in the Pulse Dose series utilize batteries as the energy source, as do the OMS 20 and 50. The fundamental operational features are identical; however, the devices in the Pulse Dose Series include some additional features, e.g., battery level indicator, which make the devices more user friendly.
The differences between the devices in the Pulse Dose Series and the OMS 20 and 50 can be categorized as additional patient features (battery level indicator), technological improvements (microprocessor control), or re-packaging (improved cosmetics and manufacturing / servicing assembly). Tests and analysis indicate that there is no greater risk to the end user than there is with the predicate devices. These device improvements were intended to make the devices in the Pulse Dose Series more user friendly while increasing their reliability, without altering their performance for the end user.
Therefore, based on the above mentioned similarities, especially the dosage methodology, the Pulse Dose Series devices and the OMS 20 and 50 are substantially equivalent devices.
Exhibit 1A
(cont.)
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.