ResScan is intended to augment the standard follow-up care of patients by providing transfer of machine and therapeutic information, including the ability to remotely change settings. It is intended to be used by Clinicians in conjunction with ResMed compatible flow generators, using ResMed's proprietary communications protocol.
Device Story
ResScan software facilitates clinical management of patients using ResMed flow generators (S8, VPAP, Stellar). Input: patient/machine therapeutic data via removable media (SD cards, USB) or direct PC connection. Operation: clinician downloads/views data, stores patient details, adjusts device parameters, and generates compliance reports (per US CMS guidelines). Output: clinical reports, updated device settings. Context: clinical setting; operated by clinicians. Benefit: enables remote monitoring and adjustment of therapy settings, supporting patient compliance management.
Clinical Evidence
No clinical testing performed. Evidence consists of bench testing (end-to-end verification of data transfer and setting adjustments) and risk analysis. Clinical trials were deemed unnecessary as the device only processes data from therapeutic devices with established clinical history.
Technological Characteristics
Software-based data management system. Interfaces with ResMed flow generators via removable media (SD/USB) or direct PC connection. Supports Windows Vista (64-bit) and Windows 7 (32/64-bit). Implements proprietary ResMed communications protocol. Features include data downloading, parameter setting, and report generation.
Indications for Use
Indicated for patients requiring follow-up care for therapy managed by ResMed flow generators (S8, VPAP, Stellar series). Used by clinicians to monitor therapeutic data and adjust device settings.
Regulatory Classification
Identification
A continuous ventilator (respirator) is a device intended to mechanically control or assist patient breathing by delivering a predetermined percentage of oxygen in the breathing gas. Adult, pediatric, and neonatal ventilators are included in this generic type of device.
{0}------------------------------------------------
·
**ResMed**
## 510(k) Summary - ResScan
[As required by 21 CFR 807.92 (c)]
u113815
:
| Date Prepared | 22nd February 2012 |
|--------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Submitter | Nicole Gaddi<br>Regulatory Affairs Manager<br>ResMed Ltd, Australia |
| Official Contact | Mr. David D'Cruz<br>V.P., Clinical & Regulatory Affairs<br>ResMed Corp<br>9001 Spectrum Center Boulevard<br>San Diego, CA 92123<br>USA<br>Tel: (858) 836 5984<br>Fax: (858) 836 5522 |
| Classification Reference | 21 CFR 868.5895 |
| Product Code | 73 BZD |
| Common/Usual Name | Non continuous ventilator (IPPB) |
| Proprietary Name | ResScan |
| Predicate Device(s) | ResScan (K050775) |
New Device
Reason for submission
:
:
{1}------------------------------------------------
Image /page/1/Picture/0 description: The image shows the word "RESMED" in all capital letters. The font is bold and sans-serif. The letters are black against a white background.
#### Indication for Use
ResScan is intended to augment the standard follow-up care of patients by providing transfer of machine and therapeutic information, including the ability to remotely change settings.
It is intended to be used by Clinicians in conjunction with ResMed compatible flow generators, using ResMed's proprietary communications protocol.
#### Substantial Equivalence
The new device has the following similarities to the previously cleared predicate device.
- A Similar intended use
- A Same operating principle
- Same technologies A
- A Same manufacturing process
Design and Verification activities were performed on ResScan as a result of the risk analysis and design requirements. Verification testing included end-to-end testing to confirm that settings were successfully transferred between the flow generator and ResScan, and data captured by the flow generator was sent to ResScan. All tests confirmed the product met the predetermined acceptance criteria. ResMed has determined that the new device is substantially equivalent to the predicate device. The modified ResScan has not altered the safety and effectiveness when used for patient compliance management as an adjunct with ResMed flow generators. The new device complies with the applicable requirements referenced in the FDA quidance documents:
- > FDA Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices (May 11, 2005)
- > FDA Off-the-Shelf Software Use in Medical Devices (September 9,1999)
#### Non-Clinical Testing:
Performance testing of ResScan has been conducted using End bench testing methodology to demonstrate that the modified ResScan performs to design input specifications.
ResScan device met the predetermined pass/fail criteria as defined in the ResScan System Verification Report.
#### Clinical Testing:
Clinical testing was not deemed necessary as indentified in the Risk Analysis, as ResScan only obtains patient and machine information from therapeutic devices for which clinical trials have already been conducted, or compared with previous predicate comparison test results. Accordingly no clinical testing is required.
#### Device Description
The performance and functional characteristics of ResScan includes all the user friendly features of the predicate device.
ResScan allows the clinician to:
- Download and view patient and machine data from ResMed flow generators >
- Store patient details >
- Set machine parameters (Using Removal Media or PC direct connection) 入
- Create and print reports >
{2}------------------------------------------------
# RESMED
- Uses Removal Media or PC direct connection as the interface between the flow generator ➤ and ResScan
- > Support for Data Card Reader
Summary of additional features from the ResScan (K050775):
- > ResMed compatible flow generators include, S8 series flow generators (73 BZD), VPAP Bilevel devices (73 MNS) and Stellar (73 MNT).
- Display/reporting of additional modes such as S, ST, ASV, iVAPS & PAC. 入
- Additional Windows Vista (64 bit) and Windows 7 (32 & 64 bit), and added support for remote A settings via removable media - SD cards and USB sticks.
- > Added support for display of alarm events.
- > Added the capability to generate patient compliance reports based on US CMS guidelines both on a per patient basis and across all patient data.
The inclusion of these features has been assessed within the risk analysis and no additional safety risks have been found as a result of the inclusion of the features.
#### Conclusion
The modified ResScan is substantially equivalent to the predicate device, ResScan (K050775).
{3}------------------------------------------------
## DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized eagle with three lines forming its body and wings. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular pattern around the eagle.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
ResMed, Limited c/o Mr. David D'Cruz Vice President, Clinical & Regulatory Affairs ResMed Corporation 9001 Spectrum Center Boulevard San Diego, California 92123
Re: K113815
Trade/Device Name: ResScan Regulation Number: 21 CFR 868.5895 Regulation Name: Continuous Ventilator Regulatory Class: II Product Code: BZD, MNS, MNT Dated: February 23, 2012 Received: February 27, 2012
Dear Mr. D'Cruz:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
MAR 2 7 2012
{4}------------------------------------------------
### Page 2- Mr. D'Cruz
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers. International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
Nh for
Anthony D. Watson, B.S., M.S., M.B.A. Director
Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
## Enclosure
{5}------------------------------------------------
## RESMED
#### Indication for Use
510(k) Number (if known):
Device Name: ResScan
Indication for Use
ResScan is intended to augment the standard follow-up care of patients by providing transfer of machine and therapeutic information, including the ability to remotely change settings.
It is intended to be used by Clinicians in conjunction with ResMed compatible flow generators, using ResMed's proprietary communications protocol.
L. Shuler
(Division Sign-Off) " ivision of Anesthesiology, General Hospital nfection Control, Dental Devices
510(k) Number: K113815
AND/OR Over-The-Counter Use_ Prescription Use __ x (Part 21 CFR 807 Subpart C) (Part 21 CFR 801 Subpart D) (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH; Office of Device Evaluation (ODE)
Page 1 of__1
Panel 1
/
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.