K981371 · Pulmonetic Systems, Inc. · CBK · Oct 30, 1998 · Anesthesiology
Device Facts
Record ID
K981371
Device Name
LTV 1000 VENTILATOR
Applicant
Pulmonetic Systems, Inc.
Product Code
CBK · Anesthesiology
Decision Date
Oct 30, 1998
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 868.5895
Device Class
Class 2
Attributes
Therapeutic, Pediatric
Indications for Use
The LTV1000 ventilator is intended to provide continuous or intermittent ventilatory support for the care of individuals who require mechanical ventilation. The ventilator is a restricted medical device intended for use by qualified, trained personnel under the direction of a physician. Specifically, the ventilator is applicable for adult and pediatric patients weighing at least 10 kg (22 lbs), who require the following types of ventilatory support: . Positive Pressure Ventilation, delivered invasively (via ET tube) or non-invasively (via mask). . . Assist/Control, SIMV, or CPAP modes of ventilation. The ventilator is suitable for use in institutional and transport settings.
Device Story
LTV 1000 is a self-contained, portable mechanical ventilator for continuous life support. It uses an internal flow generator (brushless DC motor) to provide pressurized gas; an electromechanical inspiratory flow valve delivers breath types (Volume, Pressure Control, Pressure Support, Spontaneous). Modes include Assist/Control, SIMV, CPAP, and Apnea Backup. The device monitors parameters like tidal volume, airway pressure, and I:E ratio, displaying them via 7-segment/alpha-numeric LEDs. Operated by qualified personnel in institutional or transport settings; powered by AC or internal battery. Healthcare providers use real-time monitoring and alarm data (e.g., apnea, pressure limits) to adjust ventilation settings, ensuring patient respiratory stability.
Clinical Evidence
Bench testing only. Validation included system specification testing, software requirements verification, system hazards analysis, and adherence to ASTM F 1100-90 standards. Performance testing covered the full range of operational parameters (e.g., tidal volume 50-2000 ml) under varying conditions of patient compliance, resistance, temperature, and humidity. No clinical data provided.
Technological Characteristics
Microprocessor-based control system; internal flow generator (brushless DC motor); electromechanical inspiratory flow valve; external piloted exhalation valve; external spring-loaded PEEP valve. Materials: sheetmetal chassis with thermoplastic elastomer side panels. Connectivity: standalone. Power: AC or internal rechargeable battery. Dimensions: portable, slightly larger than a laptop.
Indications for Use
Indicated for adult and pediatric patients weighing at least 10 kg (22 lbs) requiring continuous or intermittent mechanical ventilatory support (invasive or non-invasive) in institutional or transport settings.
Regulatory Classification
Identification
A continuous ventilator (respirator) is a device intended to mechanically control or assist patient breathing by delivering a predetermined percentage of oxygen in the breathing gas. Adult, pediatric, and neonatal ventilators are included in this generic type of device.
K101643 — MODEL LTV 1100 VENTILATOR · Care Fusion · Jan 26, 2011
K032226 — LTV-1000 VENTILATOR · Pulmonetic Systems, Inc. · Oct 16, 2003
K040540 — MODIFICATION TO LTV 1000 VENTILATOR · Pulmonetic Systems, Inc. · Apr 29, 2004
K101803 — NEWPORT E360 VENTILATOR · Newport Medical Instruments, Inc. · Sep 8, 2010
Submission Summary (Full Text)
{0}------------------------------------------------
510(4) K 18,371
OCT 30 1998
## Submitter
| Pulmonetic Systems, Inc | |
|-------------------------------|-----------------|
| 1016 E. Cooley Drive, Suite B | |
| Colton, CA 92324 | |
| Ph. | (909) 783-2280 |
| Fax | (909) 783-2975 |
| Contact Person: | Douglas DeVries |
| Date Prepared: | April 15, 1998 |
## DEVICE NAME
| Trade Name: | LTV 1000 |
|----------------------|-------------------------------------|
| Common Name: | Ventilator |
| Classification Name: | Ventilator, Continuous (Respirator) |
## PREDICATE DEVICE
The primary predicate device is the TBird Ventilator (K950484), manufactured by Bird Products Corp. of Palm Springs, CA.
## DESCRIPTION
The LTV 1000 ventilator is intended to provide continuous or intermittent ventilatory support for the care of individuals who require mechanical ventilator is suitable for use in institutional and attended transport settings, and is applicable for adult and pediatric patients weighing at least 10 kg (22 lbs), who require the following types of ventilatory support:
- . Positive Pressure Ventilation, delivered invasively (via ET tube) or non-invasively (via mask).
- . Assist/Control, SIMV, or CPAP modes of ventilation.
- . Breath types including Volume, Pressure Control, and Pressure Support
{1}------------------------------------------------
The LTV 1000 is a self-contained mechanical ventilator suitable for continuous life support in institutional and attended transport settings. The ventilator is slightly larger than a "laptop computer" and is selfcontained in that it can be operated without the need for externally supplied compressed air. The unit may be operated from external AC power through the use of an external AC/DC converter, or may be operated for approximately one hour using an internal rechargeable battery pack.
The following major features are included:
- Modes: Assist/Control, SIMV, CPAP, Apnea Backup .
- Breath Types: Volume Control, Pressure Control, Pressure Support, Spontaneous .
- Flow Triggering .
- . Oxygen Blending
- . PEEP
- Monitors: Calculated Peak Flow, Exhaled Tidal Volume, I:E Ratio, Mean Airway Pressure, Real Time . Airway Pressure, Peak inspiratory Pressure, PEEP, Total Breath Rate, Total Minute Volume
- Alarms: Apnea, High Pressure Limit, Low Peak Pressure, Low Minute Volume, Disconnect, Low & . Lost External Power, Low & Empty Internal Battery, Oxygen Inlet Pressure
The ventilator uses an internal flow generator to provide the pressurized gas source. All breath types are delivered by an electromechanical inspiratory flow valve. An oxygen blender meters oxygen as required to meet the current setting of the O2% control. Mechanical valves are provided internally for overpressure relief and sub-ambient relief functions.
The patient circuit is comprised of a single leg inspiratory tube connected to an exhalation system located proximal to the patient connection. The exhalation valve system will consist of a piloted exhalation valve, a PEEP valve, and a flow transducer combined in a compact package. Additional small bore tubing is included to transmit the flow, pressure and exhalation drive signals.
#### INTENDED USE
The LTV1000 ventilator is intended to provide continuous or intermittent ventilatory support for the care of individuals who require mechanical ventilation. The ventilator is a restricted medical device intended for use by qualified, trained personnel under the direction of a physician. Specifically, the ventilator is applicable for adult and pediatric patients weighing at least 10 kg (22 lbs), who require the following types of ventilatory support:
- . Positive Pressure Ventilation, delivered invasively (via ET tube) or non-invasively (via mask).
- . Assist/Control, SIMV, or CPAP modes of ventilation.
The ventilator is suitable for use in institutional and and transport settings.
{2}------------------------------------------------
# Comparison of Technological Characteristics to the Predicate Device
The following table sets forth a comparison of the major technological characteristics of the TBird Ventilator (K950484), manufactured by Bird Products Corp. of Palm Springs, CA.
| Characteristic | LTV 1000 Vent | Predicate Device<br>Bird TBird Ventilator<br>(K950484) | Discussion of Differences and Similarities |
|--------------------------|-------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Compressed Air Source | Internal, Flow Generator driven by a<br>brushless DC motor. | Internal, Flow Generator driven by a<br>brushless DC motor. | The flow generator is similar in that both devices<br>a rotating element driven by a brushless DC motor.<br>The LTV differs in that the flow generator operates<br>at a constant speed and is not used to control the<br>beginning and end of inspiration as does the<br>predicate device. This constant speed operation<br>equivalent to the operation of other ventilators,<br>as the Quantum (K962517) mfg by Healthdyne<br>Technologies. |
| Inspiratory Flow Control | Variable poppet valve positioned by a<br>rotary actuator. | No separate valve, breath control is<br>effected through variable speed control<br>of the flow generator. | The rotary actuator driven flow control valve uses<br>the LTV 1000 is similar to flow valves used on other<br>currently marketed ventilators the Servo Ventilator<br>900C ( K811102) mfg by Siemens. |
| Exhalation Valve | External Piloted balloon type. | Internal, diaphragm driven by a linear<br>actuator. | The external piloted exhalation valve used on the<br>LTV 1000 is similar to types used on other currently<br>marketed ventilators, for example the Nellcor<br>Bennet LP 10 ventilator, (K905244). |
| PEEP Valve | External, spring loaded diaphragm<br>type. | Internal, controlled by closed loop<br>control of the linear actuator. | The external spring loaded PEEP valve used on the<br>LTV 1000 is similar to types used on other currently<br>marketed ventilators, for example the Nellcor<br>Bennet LP 10 ventilator, (K905244). |
Pulmonetic Systems, Inc.
{3}------------------------------------------------
| Characteristic | LTV 1000 Vent | Predicate Device<br>Bird TBird Ventilator<br>(K950484) | Discussion of Differences and Similarities |
|------------------------|---------------------------------------------------------|------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------|
| Exhalation Flow Xducer | Differential Pressure Type, fixed<br>orifice. | Differential Pressure Type, variable<br>orifice. | Transducer is similar. |
| Control System | Microprocessor based analog and<br>digital electronics. | Microprocessor based analog and<br>digital electronics. | Control system is similar. |
| Package | Sheetmetal and thermoplastic<br>elastomer. | Sheetmetal, thermoplastic, and<br>thermoplastic elastomer. | The LTV 1000 employs a sheetmetal chassis,<br>protected by elastomeric side panels to provide<br>enhanced durability. |
| Displays | 7-segment, alpha-numeric, and single<br>lamp LEDs. | 7-segment, alpha-numeric, and single<br>lamp LEDs. | Displays are similar |
:
{4}------------------------------------------------
## NON-CLINICAL PERFORMANCE TESTING
Pursuant to establishing functional and technical equivalence to the predicate device, a non clinical test program was completed to insure that the device met its stated specifications. Validation activities included running test procedures written to validate the requirements as specified in the following documents:
- System Specification1 .
- . Software Requirements Specification
- . System Hazards Analysis
- . ASTM F 1100-90 Standard Specification for Ventilators for Use in Critical Care
The test procedures were divided into 3 primary categories:
| ASTM tests | Tests that verify the ventilator meets ASTM standards |
|-------------------|---------------------------------------------------------------------------------------------------------------|
| Performance Tests | Tests that verify the ventilator meets performance specifications including<br>tolerance and accuracy issues. |
| Functional Tests | Tests that verify the ventilator meets behavioral specifications (Primarily<br>Software) |
Test procedures were written for individual tests specifying: the requirement being tested, the number of units to be tested, the equipment to be used, test set-up instructions, the data to be recorded. In general, test procedures were written to test each parameter over the entire range of operation while varying other relevant parameters such as patient lung condition and environment. For example, the Set Tidal Volume function was tested over its entire range (50 - 2000 ml) and over worst case combinations of the following parameters:
- . Patient Compliance
- Patient Resistance .
- Temperature/Humidity of Patient breathing gas. .
- Ventilator Operating Temperature. .
A validation report was written that summarizes the results of these tests. The ventilator was found to perform to specifications with exceptions as noted in the validation report. If a parameter was found to be non-compliant, a procedure was described in the report to ensure the anomaly does not present a hazard to the patient or user.
## CONCLUSION
The comparison analysis and validation test results demonstrate the device is substantially equivalent to the predicate device and appropriate for the intended use.
See Section 15: Software.
{5}------------------------------------------------
Image /page/5/Picture/1 description: The image is a seal for the Department of Health & Human Services USA. The seal is circular and contains the words "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" around the perimeter. In the center of the seal is a stylized image of three human profiles facing to the right. Below the profiles are three wavy lines.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
OCT 30 1998
Mr. Douglas DeVries Chief Executive Officer Pulmonetic Systems 930 S. Mt. Vernon Avenue Suite 100 Colton, CA 92324
Re: K981371 LTV 1000 Ventilator Regulatory Class: II (two) Product Code: 73 CBK August 26, 1998 Dated: Received: August 27, 1998
Dear Mr. DeVries:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments. You may, therefore, market the device, subject to the general controls provisions of the Federal Food, Drug, and Cosmetic Act (Act). The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, and labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval) it may be subject to such additional controls. Existing major requlations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General requlation (21 CFR Part 820) and that, through periodic QS inspections, FDA will verify such assumptions. Failure to comply with the GMP requlation may result in regulatory action. In addition, the Food and Drug Administration (FDA) may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under section 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal Laws or Regulations.
{6}------------------------------------------------
Page 2 - Mr. Douglas DeVries
On August 16, 1993 the Final Rule for Device Tracking was published in the Federal Register, pages 43442-43455 (copy enclosed). Be advised that under Section 519(e) of the Act as amended by the Safe Medical Devices Act of 1990, FDA has identified the above device as a device which requires tracking. Because the device is subject to tracking, you are required to adopt a method of tracking that follows the devices through the distribution chain and then identifies and follows the patients who receive them. The specific requirements of the regulation are found in 21 CFR 821 as described in the August 16, 1993 Federal Register beginning on page 43447.
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a leqally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4648. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please not the regulation entitled, "Misbranding by reference to premarket notification"(21 CFR 807.97). Other general information on your responsibilities under the Act, may be obtained from the Division of Small Manufacturers Assistance at its toll free number (800) 638-2041 or (301) 443-6597, or at its internet address "http://www.fda.gov/cdrh/dsma/dsmamain.html".
Sincerely yours,
, Ath. A. Giall.
Yo Thomas J. Callahan, Ph.D. Director Division of Cardiovascular, Respiratory, and Neurological Deivces Office of Device Evaluation Center for Devices and Radiological Health
Enclosures
{7}------------------------------------------------
# STATEMENT OF INTENDED USE
The LTV1000 ventilator is intended to provide continuous or intermittent ventilatory support for the care of individuals who require mechanical ventilation. The ventilator is a restricted medical device intended for use by qualified, trained personnel under the direction of a physician. Specifically, the ventilator is applicable for adult and pediatric patients weighing at least 10 kg (22 lbs), who require the following types of ventilatory support:
- Positive Pressure Ventilation, delivered invasively (via ET tube) or non-invasively (via mask). .
- Assist/Control, SIMV, or CPAP modes of ventilation.
The ventilator is suitable for use in institutional and transport settings.
revised 8/5/98.
Mark Kram
(Division Sign-Off) Division of Cardiovascular, Respiratory, and Neurological Devices
510(k) Number K98i371
Prescription Use ✓
(Per 21 CFR 801.109)
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.