BIRD TBIRD VELA VENTILATOR

K032451 · Bird Products Corp. · CBK · Mar 17, 2004 · Anesthesiology

Device Facts

Record IDK032451
Device NameBIRD TBIRD VELA VENTILATOR
ApplicantBird Products Corp.
Product CodeCBK · Anesthesiology
Decision DateMar 17, 2004
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 868.5895
Device ClassClass 2
AttributesTherapeutic, Pediatric

Indications for Use

The TBird VELA Ventilator is intended to provide continuous or intermittent mechanical ventilatory support for individuals who require mechanical ventilation. The ventilator is a restricted medical device intended for use by qualified, trained personnel under the direction of a physician. Specifically, the ventilator is applicable for adult and pediatric patients weighing at least 10 kg (22 lbs.), who require the following general types of ventilatory support, as prescribed by an attending physician: - Positive pressure ventilation . - Assist/Control, SIMV, CPAP modes of ventilation . The ventilator is suitable for use in institutional and transport settings. It is not intended for use as an emergency medical transport ventilator.

Device Story

TBird VELA Ventilator is a turbine-based electromechanical system providing mechanical ventilation to pediatric and adult patients (≥10 kg). System comprises flow delivery, exhalation, safety, and inspiratory hold subsystems. Microprocessor-controlled electronics manage graphical user interface, power, and main control functions via pressure transducers and solenoid valves. Device operates on AC power or internal/external batteries. Clinicians use touchscreen/SVGA interface to set ventilation modes (e.g., PRVC/Vsync, APRV/BiPhasic, NPPV, Machine Volume, leak compensation). Output provides respiratory support; clinicians monitor patient status via interface to adjust therapy. Benefits include advanced ventilation modes and portability for non-emergency transport.

Clinical Evidence

Bench testing only. Performance testing verified that the TBird VELA Ventilator meets its performance requirements and is substantially equivalent to legally marketed devices.

Technological Characteristics

Turbine gas delivery system; microprocessor-controlled electromechanical system. Subsystems: flow delivery, exhalation, safety, inspiratory hold. Electronics: SVGA/touchscreen interface, pressure transducers, A/D and D/A converters, solenoid valves, watchdog/hardware fault monitors. Power: AC, internal battery, external battery pack. Connectivity: Standalone. Software: Microprocessor-based control.

Indications for Use

Indicated for adult and pediatric patients weighing at least 10 kg (22 lbs.) requiring continuous or intermittent mechanical ventilatory support (positive pressure, Assist/Control, SIMV, CPAP) in institutional or transport settings. Not for emergency medical transport.

Regulatory Classification

Identification

A continuous ventilator (respirator) is a device intended to mechanically control or assist patient breathing by delivering a predetermined percentage of oxygen in the breathing gas. Adult, pediatric, and neonatal ventilators are included in this generic type of device.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ K032451 ## MAR 1 7 2004 | Manufacturing Site: | Bird Products Corporation<br>1100 Bird Center Drive<br>Palm Springs, CA 92262 | |---------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Contact: | Tom Gutierrez (760) 778-7255 (phone); (760) 778-7274 (fax) | | Summary Date | August 11, 2000 | | Device Trade Name: | TBird VELA Ventilator | | Device<br>Common/Classification Name: | Classification name: 868.5895 Continuous Ventilator, 73 CBK | | Establishment<br>Registration Number | 2021710 | | Device Class: | Class II | | Classification Panel: | Anesthesiology | | Predicate Device: | The predicate devices are:<br>1 K020746 TBIRD VELA Ventilator Bird Products Corporation<br>2 K013642 AVEA Ventilator Bird Products Corporation<br>3 K981072 Esprit Respironics | | Device Description: | The TBird VELA Ventilator employs a turbine gas delivery system along with microprocessor<br>control to provide support for pediatric to adult patients. Capable of delivering clinically<br>advanced modes of ventilation like Pressure Support with an internal battery or AC power th<br>TBird VELA Ventilator has an extensive patient range.<br>The TBird VELA Ventilator ventilator pneumatic system is an electromechanical system<br>comprised of four major subsystems, each containing several components. These systems<br>include the flow delivery system, the exhalation system, the safety system and the inspiratory<br>hold valve.<br>This electromechanical system controls all inspiratory flow to the patient. The exhalation<br>system controls the flow of gas from the patient's lungs during the exhalation phase. The<br>mechanical safety system ensures that the patient can breath spontaneously from room air<br>and that the patient pressure is limited to a maximum value in the event of a ventilator<br>malfunction. When activated, the inspiratory hold valve blocks flow between the flow delivery<br>system and the patient.<br>The TBird VELA Ventilator ventilator electronic system is comprised of several subsystems,<br>each containing numerous components. These subsystems include the Graphical User<br>Interface System, the Power System, the Main Controller System, Exhalation, and Flow<br>Delivery systems. The Display System is comprised of SVGA and Touchscreen. The Main<br>Controller System is comprised of three Pressure Transducers, an Analog-to-Digital<br>Converter, two Digital-to-Analog Converters, the Input-Output Processor, Solenoid Valves, | Converter, wo Digital to Androg Onlyonors. The Power System conditions and controls. and the Watchdog and Hardware Fault Monitors. The Power System conditions and controls and the Watchoog and Piardware Padin Monitorer - The Porton of external battery pack. errergy tron the RO into input, the interrial battery, and the operates from this source, as well as recharging the internal battery. {1}------------------------------------------------ The TBird VELA Ventilator is intended to provide continuous or intermittent mechanical Intended Use: I he TBIRG VEEN Ventilate is individuals who require mechanical ventilation. The ventilatory support for the baro of internation for use by qualified, trained personnel under ventilator is a restholou modioally, the ventilator is applicable for adult and pediatric patients weighing at least 10 kg (22 lbs.), who require the following general types of ventilatory support, as prescribed by an attending physician: - Positive pressure ventilation . - Assist/Control, SIMV, CPAP modes of ventilation . The ventilator is suitable for use in institutional and transport settings. It is not intended for use as an emergency medical transport ventilator Substantial Equivalence The TBird VELA Ventilator is the same device as the TBird VELA Ventilator, which was cleared for market under 510(k) K020746. The name of the device was changed to the TBird VELA. Modifications to the TBird VELA Ventilator are associated with this submittal - Addition of the mode PRVC / Vsync . - Addition the mode APRV / BiPhasic . - Addition of dual limb non-invasive ventilation (NPPV) as a mode . - Remove the mode VAPS and re-implement it as Machine Volume � - Add leak Compensation . - Change the nebulizer on time from a fixed 60 minutes to user settable time from 1 to . 60 minutes. The modified TBird VELA Ventilator have the following similarities to those which previously received 510(k) concurrence: - have the same indicated use, . - use the same ventilation operating principle, ● - incorporate the same basic ventilator design with the exception of modifications � described in this submittal. - incorporate the same basic electronic control system . - are manufactured and packaged utilizing the same basic processes. . - In summary, the TBird VELA Ventilator described in this submission is, in our opinion, ● substantially equivalent to the predicate device(s). Summary of Testing and Validation: Performance testing verified that the TBird VELA Ventilator meets it's performance requirements and that this device is substantially equivalent to medical devices currently legally marketed in the United States. D3 {2}------------------------------------------------ Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an emblem featuring a stylized eagle with three wavy lines emanating from its body. Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 MAR 1 7 2004 Mr. Tom Gutierrez Regulatory Affairs Manager Bird Products Corp. 1100 Bird Center Drive Palm Springs, CA 92262 Re: K032451 TBird Vela Ventilator Regulation Number: 868.5895 Regulation Name: Continuous Ventilator Regulatory Class: II (two) Product Code: CBK Dated: December 22, 2003 Received: December 24, 2003 Dear Mr. Gutierrez: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {3}------------------------------------------------ Page 2 - Mr. Tom Gutierrez Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807): labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (301) 594-4646. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html. Sincerely yours! fthwh for, Chiu Lin, Ph.D. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {4}------------------------------------------------ ## Indication For Use 510 (k) Number (if known):__K03245| __________________________________________________________________________________________________________________________________________ Page ________ 1 ___ of _______________________________________________________________________________________________________________________________________________________ Device Name: ___ TBird VELA Ventilator Indication For Use: The TBird VELA Ventilator is intended to provide continuous or intermittent mechanical The TBIRG VED . VORMARE . Internatividuals who require mechanical ventilation. The ventilator is a restricted medical device intended for use by qualified, trained personnel ventilator to a roothere a ntvsician. Specifically, the ventilator is applicable for adult and ander the unionton of a least 10 kg (22 lbs.), who require the following general types of ventilatory support, as prescribed by an attending physician: - Positive pressure ventilation . - Assist/Control, SIMV, CPAP modes of ventilation . The ventilator is suitable for use in institutional and transport settings. It is not intended for use as an emergency medical transport ventilator. (PLEASE DO NOT WRITE BELOW THIS LINE – CONTINUE ON ANOTHER PAGE IF NEEDED) ------------------------------------------------------------------------------------------------------------------------------------------------------------------------------ Concurrence of CDRH, Office of Device Evaluation (ODE) (Division Sign-Off) Division of Anesthesiology, General Hospital, Infection Control, Dental Devices | 510(k) Number: | K032451 | |----------------|---------| |----------------|---------| | Prescription Use | <div> <span style="text-decoration: overline;"></span> ✓ </div> | OR | Over-The-Counter Use | <span style="text-decoration: overline;"></span> | |----------------------|------------------------------------------------------------------------|----|--------------------------|--------------------------------------------------| | (Per 21 cfr 801.109) | | | (Optional Format 1-2-96) | |
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...