CODE 150 BLOOD COMPONENT COLLECTION SET AND WITH DIRECT DRAW LINE
K963758 · Sorin Biomedical, Inc. · CAC · Feb 20, 1998 · Anesthesiology
Device Facts
Record ID
K963758
Device Name
CODE 150 BLOOD COMPONENT COLLECTION SET AND WITH DIRECT DRAW LINE
Applicant
Sorin Biomedical, Inc.
Product Code
CAC · Anesthesiology
Decision Date
Feb 20, 1998
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 868.5830
Device Class
Class 2
Attributes
Therapeutic
Intended Use
Code 150 Blood Component Collection Set is intended for separation of Platelet-Poor Plasma (PPP) and Platelet-Rich Plasma (PRP) from whole blood collected into transfer bags. Code 151 Blood Component Collection Set with Direct Draw Line is intended for separation of Platelet-Poor Plasma (PPP) and Platelet-Rich Plasma (PRP) from whole blood collected directly from the patient. The process of PRP/PPP is to be performed in the immediate preoperative period, prior to the surgical procedure and before the establishment of cardiopulmonary bypass. The collected PRP/PPP shall be administered during or after the operation to aid in the normalization of the patient’s hemocoagulative state.
Device Story
Blood Component Collection Set (Code 150/151) comprises four-way adapter, bag spike assembly, and collection bags. Device integrates with preassembled surgical wash systems to sequester PRP/PPP from whole blood. Input: whole blood collected via transfer bag (Code 150) or direct patient draw via filtered burette (Code 151). Operation: adapter inserted between wash bowl and waste bag; blood processed to separate plasma components. Used in clinical setting during immediate preoperative period by surgical staff. Output: sequestered PRP/PPP for re-administration to patient during or after surgery to aid hemocoagulation. Benefits: provides autologous blood components to normalize patient coagulation status.
Clinical Evidence
No clinical data. Substantial equivalence supported by bench testing including pull testing for connection security, volume accuracy testing for burette assemblies, leak testing at 15 psi, and biocompatibility testing for externally communicating, blood-path direct devices with short-term contact duration.
Technological Characteristics
System includes four-way adapter, bag spike, and collection bags. Materials tested for biocompatibility per Tripartite Guidelines for short-term (5 min to 29 days) blood contact. Operates via integration with surgical wash systems. No electronic components, software, or energy sources.
Indications for Use
Indicated for patients undergoing surgical procedures requiring normalization of hemocoagulative state via administration of autologous Platelet-Rich Plasma (PRP) or Platelet-Poor Plasma (PPP) collected in the immediate preoperative period.
Regulatory Classification
Identification
An autotransfusion apparatus is a device used to collect and reinfuse the blood lost by a patient due to surgery or trauma.
Predicate Devices
Electromedics Plasma Sequestration Set
Electromedics AT-1000 Sequestration Set
Related Devices
K241236 — XTRA Collection sets; XTRA Sequestration set X · Sorin Group Italia S.R.L. · Aug 16, 2024
K012178 — FRESENIUS C.A.T.S; ATI AUTOTRANSFUSION SET; PSQ PLASMA SEQUESTRATION SET; PSQ-DD PLASMA SEQUESTRATION DIRECT DRAW · Fresenius Hemotechnology, Inc. · Jul 23, 2001
K022348 — ACCESS CONCENTRATION SYSTEM, MODEL ACC-100, ACCESS DISPOSABLE SET, MODEL ADS-2000 · Interpore Cross Intl. · Aug 15, 2002
K240584 — XTRA Autotransfusion System (with XTRA Bowl sets) · Sorin Group Italia S.R.L. · Apr 29, 2024
Submission Summary (Full Text)
{0}
K963758
510(k) Notification
Sorin Biomedical Inc.
Dideco Blood Component Collection Sets - Code 150 and Code 151
## XIV. 510(k) SUMMARY
### A. Name and Address of Submitter
Sorin Biomedical Inc.
17600 Gillette Avenue
P.O. Box 19503
Irvine, California 92713-9503
FEB 20 1998
### B. Telephone and Fax Numbers of Submitter
Telephone: (714) 250-8322
Fax: (714) 757-8644
### C. Name of Contact Person
Susan Reimers
Manager, Clinical and Regulatory Affairs
### D. Date Summary was Prepared:
September 16, 1996
### E. Device Name
*Trade or Proprietary Name:*
Dideco Compact-A Autotransfusion System
*Common Name:*
Autotransfusion Apparatus
*Classification Name:*
Autotransfusion Devices
### F. Summary of Substantial Equivalence
The Blood Component Collection set is substantially equivalent in intended use, materials, design, and performance characteristics to the Electromedics Plasma Sequestration Set.
SORIN BIOMEDICAL INC. CONFIDENTIAL
27
{1}
Sorin Biomedical Inc.
Dideco Blood Component Collection Sets - Code 150 and Code 151
510(k) Notification
## G. Device Description
The Blood Components Collection Set consists of a four-way adapter, a bag spike assembly, and two (2) collection bags. The four way adapter is inserted between the wash bowl and waste bag components of a preassembled surgical wash. The collection bags are attached to the free arms of the adapter. Whole blood to be processed may be collected in the immediate preoperative period into a transfer bag, which is then attached to the inlet side of the wash set using the bag spike assembly. In addition, blood may be collected directly from the patient by utilization of a filtered burette assembly intended for the administration of anticoagulation solution during preoperative collection of patient blood for sequestration of PRP/PPP. Alternately, whole blood from the extracorporeal circuit (e.g. cardiotomy reservoir or oxygenator) may be processed during the intraoperative period in the same manner.
## H. Device Intended Use
Code 150 Blood Component Collection Set is intended for separation of Platelet-Poor Plasma (PPP) and Platelet-Rich Plasma (PRP) from whole blood collected into transfer bags.
Code 151 Blood Component Collection Set with Direct Draw Line is intended for separation of Platelet-Poor Plasma (PPP) and Platelet-Rich Plasma (PRP) from whole blood collected directly from the patient.
The process of PRP/PPP is to be performed in the immediate preoperative period, prior to the surgical procedure and before the establishment of cardiopulmonary bypass. The collected PRP/PPP shall be administered during or after the operation to aid in the normalization of the patient’s hemocoagulative state.
## I. Summary of Comparison of Technological Characteristics
The technological characteristics of the Blood Component Collection sets are similar to the Electromedics Plasma Sequestration Set. Both sets are identical in respect to their process steps, method of operation, materials, and suggested flow rates. They are both indicated for separation of Platelet-Poor Plasma (PPP) and Platelet-Rich Plasma (PRP) from whole blood collected from the patient.
SORIN BIOMEDICAL INC. CONFIDENTIAL
28
{2}
Sorin Biomedical Inc.
510(k) Notification
Dideco Blood Component Collection Sets - Code 150 and Code 151
## J. Summary of Nonclinical Tests
Substantial equivalence was based on a comparison of test results from the following *in vitro* physical and functional tests:
### Physical Tests
#### Pull Testing
Pull testing was performed to ensure the blood components collection set can be securely attached to the preassembled surgical wash set to prevent being inadvertently disconnected. The results of the physical testing demonstrate that a secure connection can be established between the blood components collection set and the reassembled surgical wash set.
#### Volume Accuracy
The purpose of this test was to verify the accuracy of the volume label on the burette assembly by recording the volume indication at specific volume levels. All burettes recorded accurately at the specified volumes.
#### Leak Testing of the Filtered Burette
Leak testing was conducted on the blood component burette assembly under pressure. Each unit was slowly pressurized to $15 \pm 0.5$ psi and submerged under water.
### Biocompatibility
Biocompatibility testing was performed on finished, sterilized devices in accordance with Tripartite Guidelines for the category of “externally communicating devices, blood path direct: with a short term contact duration (5 minutes to 29 days)”. Results demonstrate that the blood contacting components of the Blood Component Collection Set are safe for their intended use.
## K. Conclusions
Based upon the above information, Sorin Biomedical Inc. concludes that the blood component collection set is substantially equivalent to the Electromedics AT-1000 Sequestration Set for the plasma sequestration indication for use.
SORIN BIOMEDICAL INC. CONFIDENTIAL
{3}
DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
FEB 20 1998
Food and Drug Administration
9200 Corporate Boulevard
Rockville MD 20850
Ms. Sharon Thompson
Director, Regulatory Affairs/Quality
Sorin Biomedical, Inc.
17600 Gillette Avenue
P.O. Box 19503
Irvine, CA 92713-9503
Re: K963758
Code 150 Blood Component Collection Set
Code 151 Blood Component Collection Set with Direct Draw Line
Regulatory Class: II (two)
Product Code: 74 CAC
Dated: January 30, 1998
Received: February 2, 1998
Dear Ms. Thompson:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.
{4}
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for *in vitro* diagnostic devices), please contact the Office of Compliance at (301) 594-4648. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html".
Sincerely yours,
Thomas J. Callahan, Ph.D.
Director
Division of Cardiovascular, Respiratory, and Neurological Devices
Office of Device Evaluation
Center for Devices and Radiological Health
Enclosure
{5}
Page 1 of 1
NIBICATE
510(k) Number (if known): K963758
Device Name: CODE 150 BLOOD COMPONENT COLLECTION SET AND WITH DIRECT DRAW LINE
Indications For Use:
Code 150 Blood Component Collection Set is intended for separation of Platelet-Poor Plasma (PPP) and Platelet-Rich Plasma (PRP) from whole blood collected into transfer bags.
Code 151 Blood Component Set with Direct Draw Line is intended for separation of Platelet-Poor Plasma (PPP) and Platelet-Rich Plasma (PRP) from whole blood collected from directly the patient.
The process of PRP/PPP is to be performed in the immediate preoperative period, prior to the surgical procedure and before the establishment of cardiopulmonary bypass. The collected PRP/PPP shall be administered during or after the operation to aid in the normalization of the patient's hemocoagulative state.
The Electronmedics PRP/PPP Sequestration Kits are indicated for the same use as the Code 150 and Code 151 Blood Component Collection Sets.
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Prescription Use ☑ (Per 21 CFR 801.109)
OR
Over-The-Counter Use ☐ (Optional Format 1-2-96)
(Division Sign-Off)
Division of Cardiovascular, Respiratory, and Neurological Devices
510(k) Number K963758
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.