K964653 · Respironics, Inc. · BTT · Feb 14, 1997 · Anesthesiology
Device Facts
Record ID
K964653
Device Name
OASIS HUMIDIFIER
Applicant
Respironics, Inc.
Product Code
BTT · Anesthesiology
Decision Date
Feb 14, 1997
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 868.5450
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Oasis Humidifier is an accessory for Respironics CPAP and BiPAP systems to provide moisture to the patient circuit. The Oasis is the only humidifier with the acoustical design that enables effective operation of Respironics' Auto-CPAP mode. As such, it is the only humidifier that can be used with Respironics' Auto-CPAP products (i.e., Virtuoso Smart CPAP System (K953930) and Quartet Clinical System (K963761, currently under review)). The Oasis can also be used with standard CPAP devices from other manufacturers which have maximum operating pressures of 20 cm H₂O, and do not have bi-level or automatic pressure titration capabilities. Performance of non-Respironics CPAP systems with the Oasis humidifier should be verified before use.
Device Story
Passive respiratory gas humidifier; accessory for CPAP/BiPAP systems. Input: dry inspiratory air from pressure-generating unit. Operation: air passes through water reservoir; evaporation increases humidity; no heating element. Output: humidified air delivered to patient circuit. Features acoustically-tuned J-shaped tube to maintain impedance for Auto-CPAP mode functionality. Used in home or clinical settings; operated by patient or clinician. Device consists of dishwasher-safe plastic lid, reservoir, gasket, and flexible tubing. Benefits patient by providing moisture to prevent airway dryness during positive pressure therapy.
Clinical Evidence
Bench testing only. Performance verified through environmental testing and compliance with FDA standards. Pressure drop measured at 0.17 cm H2O (30 L/min) and 0.5 cm H2O (60 L/min). Humidity output validated to >25% RH.
Technological Characteristics
Passive evaporation humidifier. Materials: dishwasher-safe plastic (reservoir/lid), gasket, flexible tubing. Dimensions: 440 ml capacity. Connectivity: pneumatic connection to CPAP/BiPAP via 24" flexible tubing. Sterilization: EtO compatible; disinfection via pasteurization. Acoustically-tuned internal passage for Auto-CPAP impedance matching.
Indications for Use
Indicated for patients requiring CPAP or BiPAP therapy who need supplemental moisture in the patient circuit. Compatible with Respironics Auto-CPAP, CPAP, and BiPAP systems, and standard CPAP devices from other manufacturers with max operating pressures of 20 cm H2O without automatic titration.
Regulatory Classification
Identification
A respiratory gas humidifier is a device that is intended to add moisture to, and sometimes to warm, the breathing gases for administration to a patient. Cascade, gas, heated, and prefilled humidifiers are included in this generic type of device.