A7 Anesthesia System

K142552 · Shenzhen Mindray Bio-Medical Electronics Co., Ltd. · BSZ · Jan 23, 2015 · Anesthesiology

Device Facts

Record IDK142552
Device NameA7 Anesthesia System
ApplicantShenzhen Mindray Bio-Medical Electronics Co., Ltd.
Product CodeBSZ · Anesthesiology
Decision DateJan 23, 2015
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 868.5160
Device ClassClass 2
AttributesTherapeutic, Pediatric

Indications for Use

The A7 Anesthesia System is a device used to administer to a patient, continuously or intermittently, a general inhalation anesthetic and to maintain a patient's ventilation. The A7 is intended for use by licensed clinicians, for patients requiring anesthesia within a health care facility, and can be used in adult and pediatric populations.

Device Story

A7 Anesthesia System is a continuous flow inhalation gas anesthesia system; delivers anesthetic vapor; provides automatic and manual ventilation modes. Inputs: O2, N2O, air, and anesthetic agents (Desflurane, Isoflurane, Enflurane, Sevoflurane, Halothane). System monitors gas concentrations (O2, CO2, N2O, agents) and airway pressure. Used in clinical settings by licensed clinicians. Output: controlled anesthetic delivery and ventilation support. Features include fresh gas control, negative pressure suction, agent consumption calculation, auxiliary common gas outlet, sample gas return, quick-release APL valve, and auxiliary work surface. Clinicians use real-time monitoring and spirometry loops (pressure-volume, flow-volume) to adjust ventilation and anesthesia delivery, ensuring patient stability during surgical procedures.

Clinical Evidence

Bench testing only. Compliance with IEC 60601-1, IEC 60601-1-2, IEC 60601-1-8, IEC 60601-2-13, ISO 10993 (biocompatibility), and ISO 14971 (risk management) verified. Software verified and validated per FDA guidance. No clinical data presented.

Technological Characteristics

Continuous flow gas anesthesia system; variable bypass vaporizers; automatic/manual ventilator; bellows volume 1500mL. Materials tested per ISO 10993-1 (cytotoxicity, irritation, sensitization, VOC). Connectivity: integrated gas module. Standards: IEC 60601-1, IEC 62304, ISO 5356-1, ISO 21647, ASTM F1101-90. Paramagnetic O2 sensor. Software life cycle per IEC 62304.

Indications for Use

Indicated for adult and pediatric patients requiring general inhalation anesthesia and ventilation maintenance within a health care facility.

Regulatory Classification

Identification

A gas machine for anesthesia is a device used to administer to a patient, continuously or intermittently, a general inhalation anesthetic and to maintain a patient's ventilation. The device may include a gas flowmeter, vaporizer, ventilator, breathing circuit with bag, and emergency air supply. A gas machine for analgesia is a device used to administer to a patient an analgesic agent, such as a nitrous oxide-oxygen mixture (maximum concentration of 70 percent nitrous oxide).

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized caduceus symbol, which is a staff with two snakes coiled around it, representing medicine and healing. The symbol is placed to the right of the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA", which is arranged in a circular fashion around the symbol. The logo is simple and professional, reflecting the department's role in public health and human services. Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002 January 23, 2015 Shenzhen Mindray Bio-Medical Electronics Co., Ltd Tang Hao Product Approval Engineer, Technical Regulation Department Mindrav Building. Keii 12th Road South High-tech Industrial Park, Nanshan Shenzhen, Guandong 518057 CHINA Re: K142552 > Trade/Device Name: A7 Anesthesia System Regulation Number: 21 CFR 868.5160 Regulation Name: Gas Machine For Anesthesia Or Analgesia Regulatory Class: Class II Product Code: BSZ Dated: December 19, 2014 Received: December 22, 2014 Dear Mr. Hao, We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {1}------------------------------------------------ Page 2 - Tang Hao Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Sincerely yours, Tejashri Purohit-Sheth, M.D. Tejashri Purohit-Sheth, M.D. Clinical Deputy Director DAGRID/ODE/CDRH FOR Erin Keith Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ ## Indications for Use 510(k) Number (if known) K142552 Device Name A7 Anesthesia System #### Indications for Use (Describe) The A7 Anesthesia System is a device used to administer to a patient, continuously or intermittently, a general inhalation anesthetic and to maintain a patient's ventilation. The A7 is intended for use by licensed clinicians, for patients requiring anesthesia within a health care facility, and can be used in adult and pediatric populations. Type of Use (Select one or both, as applicable) 2 Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C) #### PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED. #### FOR FDA USE ONLY Concurrence of Center for Devices and Radiological Health (CDRH) (Signature) This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ K142552 # 510(K) SUMMARY In accordance with 21 CFR 807.87(h) and (21 CFR 807.92) the 510(k) Summary for the A7 Anesthesia System is provided below. | Device Common Name: | Gas-Machine, Anesthesia | |------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Device Proprietary Name: | A7 Anesthesia System | | Submitter: | SHENZHEN MINDRAY BIO-MEDICAL<br>ELECTRONICS CO., LTD.<br>Mindray Building, Keji 12th Road South<br>High-tech Industrial Park, Nanshan<br>Shenzhen 518057, P.R. China<br>Tel: +86 755 81885635<br>Fax: +86 755 26582680 | | Contact: | Yanhong Bai<br>Manager Regulatory Affairs<br>SHENZHEN MINDRAY BIO-MEDICAL<br>ELECTRONICS CO., LTD.<br>Mindray Building, Keji 12th Road South<br>High-tech Industrial Park, Nanshan<br>518057, P.R. China<br>Tel: +86 755 81885635<br>Fax: +86 755 26582680<br>E-mail: baiyanhong@mindray.com | | Date Prepared: | January 23, 2015 | | Panel: | Anesthesiology | | Classification Regulation<br>and Product Code: | Primary:<br>868.5160- BSZ Anesthesia Gas Machine Class II<br>Secondary:<br>868.1400 - CCK - Carbon Dioxide Gas Analyzer<br>868.1500 – NHO/CBQ/NHQ/NHP - Enflurane gas analyzer<br>868.1620 - CBS - Halothane Gas Analyzer<br>868.1700 - CBR - Nitrous Oxide Gas Analyzer<br>868.1720 - CCL- Oxygen Gas Analyzer<br>880.6740 - KDP- Vacuum Regulator | | Predicate Devices: | | K123211 - A5 Anesthesia Delivery System, Mindray DS USA, Inc. K042607 - Primus US, Draeger Medical Inc. K110213 - GE Datex-Ohmeda Aisys, Datex-Ohmeda Inc. {4}------------------------------------------------ K123125 - Avance CS2, GE ## Indication for Use: The A7 Anesthesia System is a device used to administer to a patient, continuously or intermittently, a general inhalation anesthetic and to maintain a patient's ventilation. The A7 is intended for use by licensed clinicians, for patients requiring anesthesia within a health care facility, and can be used in adult and pediatric populations. ## De vice De scription: The A7 Anesthesia System is a continuous flow inhalation gas anesthesia system that delivers anesthetic vapor and provides for automatic and manual modes of ventilation. The A7 incorporates 02, CO2, N2O and Agent concentration monitoring (Desflurane, Isoflurane, Enflurane, Sevoflurane and Halothane). The A7 consists of a main unit (includes an anesthetic ventilator and flow meter monitor assembly) and a detachable breathing system. ## Performance Data: Biocompatibility - Each patient contacting material and gas path material has been tested to comply with the following applicable requirements per ISO 10993-1: - Cytotoxicity: ISO10993-5:2009 Biological evaluation of medical devices -. Part 5: Biological evaluation of medical devices - Part 5: Tests for in vitro cytotoxicity - Irritation: ISO10993-10:2010 Biological evaluation of medical devices Part 10: Tests . for irritation and skin sensitization - Sensitization: ISO10993-10:2010 Biological evaluation of medical devices Part 10: ● Tests for irritation and skin sensitization - Volatile Organic Compound Testing: ISO10993-18:2005 Biologica1 evaluation of . medical devices -- Part 18: Chemical characterization of materials Software - The A7 Anesthesia System software has been fully verified and validated and documentation in accordance with FDA Guidance "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices: May 11, 2005" has been provided in this submission. Performance Testing - The A7 Anesthesia system has been tested to comply with its product specifications and intended use. The following standards were adhered to for performance testing: - IEC 60601-1:1988+A1:1991+A2:1995: Medical electrical equipment - Part 1: General Requirements for Basic Safety and Essential Performance - IEC 60601-1-1:2000 General requirements for safety - Collateral standard: Safety requirements for medical electrical systems - IEC 60601-1-2:2007 Medical electrical equipment Part 1-2: General requirements ● for basic safety and essential performance - Collateral standard: Electromagnetic compatibility - Requirements and tests - IEC 60601-1-4:2000 Medical electrical equipment Part 1-4: General requirements ● for safety - Collateral standard: Programmable electrical medical systems {5}------------------------------------------------ - . IEC 60601-1-8:2003 Medical electrical equipment - Part 1-8: General requirements for safety - Collateral Standard: General requirements, tests and guidance for alarm systems in medical electrical equipment and medical electrical systems - IEC 60601-2-13:2009 Medical electrical equipment - Part 2-13: Particular requirements for the safety and essential performance of anesthetic systems - IEC 62304: 2006 Medical device software Software life cycle processes ● - IEC 62366:2007 Medical devices - Application of usability engineering to medical devices - ISO 14971:2007 Medical devices Application of risk management to medical ● devices - ISO 15223-1:2012 Medical devices Symbols to be used with medical devices labels, ● labeling, and information to be supplied - Part 1: General requirements - ISO 5356-1:2004 Anesthetic and respiratory equipment Conical connectors Part 1: ● Cones and sockets - ISO 21647:2004 Medical electrical equipment - Particular requirements for the basic safety and essential performance of respiratory gas monitors - ASTM F1101-90:2003 Standard Specification for Ventilators Intended for Use ● During Anesthesia - CGA V-1:2005 Standard for Compressed Gas Cylinder Valves Outlet and Inlet . Connections - CGA V-5:2008 Diameter-Index Safety System (Noninterchangeable Low Pressure ● Connections for Medical Gas Applications) #### Substantial Equivalence : The subject device has the exact same indication statement as the primary predicate device. the A5 as cleared in K123211. The subject A7 Anesthesia System and the primary predicate are both gas anesthesia machines used to administer to a patient, continuously or intermittently, a general inhalation anesthetic and to maintain a patient's ventilation. These systems may be used in adult and pediatric populations. Both of the devices are prescription use, and are not suitable for use in an MRI environment. A technical comparison of the subject device to the primary predicate device, the A5 as cleared in K123211 is provided in the table below. | Technical<br>Characteristics | A7 Anesthesia Delivery System<br>Shenzhen Mindray Bio-Medical<br>Electronics Co., Ltd | A5 Anesthesia Delivery System<br>Mindray DS USA, Inc.<br>K123211 | | |--------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------| | Vaporizers | Two, variable bypass | Two, variable bypass | | | Agent - Sevoflurane | Yes | Yes | | | Agent - Isoflurane | Yes | Yes | | | | Technical<br>Characteristics | A7 Anesthesia Delivery System<br>Shenzhen Mindray Bio-Medical<br>Electronics Co., Ltd | A5 Anesthesia Delivery System<br>Mindray DS USA, Inc.<br>K123211 | | | Agent - Desflurane | Yes | Yes | | | Agent - Halothane | Yes | Yes | | | Agent - Enflurane | Yes | Yes | | | Automatic Ventilator | Yes | Yes | | | Bellows | Yes | Yes | | | Bellows Volume | 1500mL | 1500mL | | Ventilation Modes | | | | | | VCV | Yes | Yes | | | PCV | Yes | Yes | | | PCV – VG | Yes | Yes | | | SIMV - VC | Yes | Yes | | | SIMV - PC | Yes | Yes | | | PS | Yes | Yes | | | Tidal Volume | Yes | Yes | | Specifications | | | | | | Range, ml | 20 - 1500 | 20 - 1500 | | | Minute Volume | Yes | Yes | | | Rate, bpm | 4-100 bpm | 4-100 bpm | | | Inspiratory Flow, L/min | 110 L/min + fresh gas flow | 110 L/min + fresh gas flow | | | I:E Ratio | 4:1 to 1:8 with 0.5 increment | 4:1 to 1:8 with 0.5 increment | | | Inspiratory Pause | Off, 5 - 60% of insp. Period | Off, 5 - 60% of insp. Period | | Fresh<br>Gas | Air Flow Range | 0~15 L/min | 0~15 L/min | | | N₂O Flow Range | 0~12 L/min | 0~12 L/min | | | O₂ Flow Range | 0~15 L/min | 0~15 L/min | | | Individual Gas<br>Flow Accuracy | ±50 ml/min or ±5% of setting value,<br>whichever is greater | ±120 ml/min or ±10% of setting<br>value, whichever is greater | | | Pressure Limit, cm H2O | 0 - 100 | 0 - 100 | | | PEEP, cm H20 | Off, 3-30, 1 cmH2O increment | Off, 3-30, 1 cmH2O increment | | | System Checks | Auto at start | Auto at start | | | Airway Pressure<br>Measured at | Inspiratory | Inspiratory | | | High/Low Airway | Yes | Yes | | Technical<br>Characteristics | A7 Anesthesia Delivery System<br>Shenzhen Mindray Bio-Medical<br>Electronics Co., Ltd | A5 Anesthesia Delivery System<br>Mindray DS USA, Inc.<br>K123211 | | | Pressure Alarm | Yes | Yes | | | Pressure Limiting Alarm | Yes | Yes | | | Sub Atmospheric Pressure<br>Alarm | Yes | Yes | | | Continuous Press Alarm | Yes | Yes | | | Apnea >2 Minute Alarm | Yes | Yes | | | Apnea Alarm | Yes | Yes | | | High/Low Minute Volume<br>Alarm | Yes | Yes | | | High/Low O2<br>Concentration Alarm | Yes | Yes | | | Type of O2 Sensor | Paramagnetic | Paramagnetic or Galvanic (The<br>Paramagnetic sensor is utilized<br>when using the integrated gas<br>module. The Galvanic sensor is<br>utilized when the gas module is<br>not connected.) | | | Heated Breathing Circuit | Yes | Yes | | | Spirometry: Pressure-<br>Volume and Flow-Volume<br>loops | Yes | Yes | | | Anesthetic Gas Module<br>Sampling Rate | P/N 9200-10-10530 water trap: 120,<br>150, 200mL/min<br>P/N 9200-10-10574 water trap: 70,<br>90, 120mL/min | P/N 9200-10-10530 water trap:<br>120, 150, 200mL/min<br>P/N 9200-10-10574 water trap:<br>70, 90, 120mL/min | | | Anesthetic Gas Module<br>Sampling Delay Time: | <4 seconds | <4 seconds | | | Anesthetic Gas Module<br>Refresh Rate: | 1 second | 1 second | | | Anesthetic Gas Module<br>Warm-up Time: | 45 seconds to warm-up status 10<br>minutes to ready-to-measure status | 45 seconds to warm-up status<br>10 minutes to ready-to-measure<br>status | | | Anesthetic Gas Module<br>Accuracy CO2: | 0-1%: +/-.1%<br>1-5%: +/-.2%<br>5-7%: +/-.3%<br>7-10%: +/-.5%<br>>10%: unspecified | 0-1%: +/-.1%<br>1-5%: +/-.2%<br>5-7%: +/-.3%<br>7-10%: +/-.5%<br>>10%: unspecified | | | Technical<br>Characteristics | A7 Anesthesia Delivery System<br>Shenzhen Mindray Bio-Medical<br>Electronics Co., Ltd | A5 Anesthesia Delivery System<br>Mindray DS USA, Inc.<br>K123211 | | | Anesthetic Gas Module<br>Accuracy N2O: | 0-20%: +/-2%<br>20-100%: +/-3% | 0-20%: +/-2%<br>20-100%: +/-3% | | | Anesthetic Gas Module<br>Accuracy Desflurane: | 0-1%: +/-.15%<br>1-5%: +/-.2%<br>5-10%: +/-.4%<br>10-15%: +/-.6%<br>15-18%: +/-1%<br>>18%: unspecified | 0-1%: +/-.15%<br>1-5%: +/-.2%<br>5-10%: +/-.4%<br>10-15%: +/-.6%<br>15-18%: +/-1%<br>>18%: unspecified | | | Anesthetic Gas Module<br>Accuracy Sevflurane: | 0-1%: +/-.15%<br>1-5%: +/-.2%<br>5-8%: +/-.4%<br>>8%: unspecified | 0-1%: +/-.15%<br>1-5%: +/-.2%<br>5-8%: +/-.4%<br>>8%: unspecified | | | Anesthetic Gas Module<br>Accuracy<br>Enflurane/Isoflurane/<br>Halothane: | 0-1%: +/-.15%<br>1-5%: +/-.2%<br>>5%: unspecified | 0-1%: +/-.15%<br>1-5%: +/-.2%<br>>5%: unspecified | | | Anesthetic Gas Module<br>Accuracy O2: | 0-25%: +/-1%<br>25-80%: +/-2%<br>80-100%: +/-3% | 0-25%: +/-1%<br>25-80%: +/-2%<br>80-100%: +/-3% | | | Anesthetic Gas Module<br>Accuracy awRR: | 2-60rpm: +/-1rpm<br>>60rpm: unspecified | 2-60rpm: +/-1rpm<br>>60rpm: unspecified | | | Anesthetic Gas Module<br>Measurement Rise Time: | CO2: ≤250ms<br>N2O: ≤250ms<br>O2: ≤500ms<br>Hal/Iso/Sev/Des: ≤300ms<br>Enf: ≤350ms | CO2: ≤250ms<br>N2O: ≤250ms<br>O2: ≤500ms<br>Hal/Iso/Sev/Des: ≤300ms<br>Enf: ≤350ms | | | Measurement Range CO2 | 0-30% | 0-30% | | | Measurement Range N2O | 0-100% | 0-100% | | | Measurement Range Des | 0-30% | 0-30% | | | Measurement Range Sev | 0-30% | 0-30% | | | Measurement Range<br>Enf/Iso/Hal | 0-30% | 0-30% | | | Measurement Range O2 | 0-100% | 0-100% | | {6}------------------------------------------------ K142552 {7}------------------------------------------------ K142552 {8}------------------------------------------------ A7 Anesthesia System {9}------------------------------------------------ A7 Anesthesia System ## Substantial Equivalence Conclusion There are seven new features that have been added to the A7 when compared to the A5 as follows: - 1. Fresh Gas Control System - 2. Negative Pressure Suction Device - 3. Agent Consumption Calculation - Auxiliary Common Gas Outlet 4. - Sample Gas Return 5. - Quick Release APL Valve 6. - 7. Auxiliary Work Surface A detailed comparison of these 7 new features is provided in the submission and performance testing was completed to establish that the new features do not raise new issues of safety and effectiveness. The A7 Anesthesia System has the same intended use as the predicate device, to administer to a patient, continuously or intermittently, a general inhalation anesthetic and to maintain a patient's ventilation. The A7 is a line extension of the FDA-cleared A5 Anesthesia System (the primary predicate device), and has the identical indications for use as the A5. The differences in the technological characteristics of these features when compared to the predicate device do not raise new types of safety and effectiveness questions. Therefore the A7 can be found substantially equivalent to the predicate device.
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...