SPROTTE NRFit, Quincke NRFit Lumbar Puncture needles

K160294 · PAJUNK GmbH Medizintechnologie · BSP · Oct 21, 2016 · Anesthesiology

Device Facts

Record IDK160294
Device NameSPROTTE NRFit, Quincke NRFit Lumbar Puncture needles
ApplicantPAJUNK GmbH Medizintechnologie
Product CodeBSP · Anesthesiology
Decision DateOct 21, 2016
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 868.5150
Device ClassClass 2
AttributesTherapeutic, Pediatric

Indications for Use

The SPROTTE® NRFit™, Quincke NRFit™ lumbar puncture needles are intended to gain entry into or puncture the spinal cavity permitting injection (including anesthesia) / withdrawal of fluids for purposes of diagnostic lumbar puncture, myelography/ discography procedures. The device is intended for adult and pediatric patients.

Device Story

Lumbar puncture needles (SPROTTE® and Quincke styles) designed for spinal cavity access; used for diagnostic fluid withdrawal or anesthetic injection. Device consists of stainless steel cannula, polycarbonate hub, and stylet. Operates via manual insertion by clinician; NRFit™ hub design ensures compatibility with neuraxial connectors per ISO 80369-6. Output is physical access to spinal space for fluid collection or drug delivery. Benefits include standardized, secure connection to neuraxial equipment, reducing risk of misconnection and contamination. Used in clinical settings by trained healthcare professionals.

Clinical Evidence

No clinical data. Substantial equivalence is supported by bench testing, including bending rigidity (ISO 9626), bonding-to-hub pull-off force (≥22N, ISO 7864), and penetration force comparisons. Biocompatibility (ISO 10993) and sterilization (ISO 11135-1) were validated.

Technological Characteristics

Materials: Stainless steel (cannula/tubing), Polycarbonate (hub/knob), Epoxy resin (glue). Standards: ISO 9626 (tubing), ISO 7864 (needles), ISO 80369-6 (neuraxial connectors). Energy: Manual. Sterilization: Ethylene Oxide (SAL 10^-6). Form factor: 18G-29G (SPROTTE), 20G-27G (Quincke); lengths 25mm-285mm.

Indications for Use

Indicated for adult and pediatric patients requiring spinal cavity entry for diagnostic lumbar puncture, myelography, discography, or injection of anesthesia.

Regulatory Classification

Identification

An anesthesia conduction needle is a device used to inject local anesthetics into a patient to provide regional anesthesia.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/1 description: The image shows the seal of the U.S. Department of Health & Human Services. The seal features the department's name encircling a symbol. The symbol consists of three stylized human profiles facing right, stacked on top of each other. The text reads "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA". Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002 October 21, 2016 PAJUNK® GmbH Medizintechnologie Christian Quass Director Regulatory Affairs, Safety Officer Karl-Hall-Str. 1 Geisingen, 78187 GERMANY Re: K160294 Trade/Device Name: SPROTTE® NRFit™ and Quincke NRFit™ lumbar puncture needles Regulation Number: 21 CFR 868.5150 Regulation Name: Anesthesia Conduction Needle Regulatory Class: Class II Product Code: BSP Dated: September 19, 2016 Received: September 21, 2016 Dear Christian Quass: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in {1}------------------------------------------------ the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Sincerely yours, Tejashri Purohit-Sheth, M.D. Tejashri Purohit-Sheth, M.D. Clinical Deputy Director DAGRID/ODE/CDRH FOR Tina Kiang, Ph.D. Acting Director Division of Anesthesiology, General Hospital. Respiratory. Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ ## Indications for Use 510(k) Number (if known) K160294 Device Name SPROTTE® NRFit™, Quincke NRFit™ lumbar puncture needles Indications for Use (Describe) The SPROTTE® NRFit™, Quincke NRFit™ lumbar puncture needles are intended to gain entry into or puncture the spinal cavity permitting injection (including anesthesia) / withdrawal of fluids for purposes of diagnostic lumbar puncture, myelography/ discography procedures. The device is intended for adult and pediatric patients. Type of Use (Select one or both, as applicable) | <input checked="" type="checkbox"/> Prescription Use (Part 21 CFR 801 Subpart D) | |----------------------------------------------------------------------------------| | <input type="checkbox"/> Over-The-Counter Use (21 CFR 801 Subpart C) | #### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ Image /page/3/Picture/1 description: The image shows the word "PAJUNK" in white letters against a teal background. The letters are in a bold, sans-serif font. A registered trademark symbol is located to the right of the letter K. Pioneering medical technology 510(k) Summary as required by 21 CFR 807.92(c). Date of Preparation: October 20th 2016 Document Control Number: K160294 ## 510(k) owner: PAJUNK® GmbH Medizintechnologie Karl-Hall-Str. 1 78187 Geisingen Baden-Wuerttemberg, Germany Fon: +49(0)7704-9291-586 Fax: +49(0)7704-9291-605 Establishment Registration Number: 9611612 ## Submitter Information/ production site: PAJUNK® GmbH Medizintechnologie Karl-Hall-Str. 1 78187 Geisingen Baden-Wuerttemberg, Germany Fon: +49(0)7704-9291-586 Fax: +49(0)7704-9291-605 Establishment Registration Number: 9611612 ## Contact: Christian G. H. Quass Director Regulatory Affairs, Safety Officer Patricia Weisbrod Regulatory Affairs Manager Fon: +49(0)7704-9291-586 Fax: +49(0)7704-9291-605 E-Mail: christian.quass@pajunk.com E-Mail: patricia.weisbrod@pajunk.com ## USA Contact: PAJUNK MEDICAL SYSTEMS 6611 Bay Circle, Suite 140 Norcross, GA 30071 Phone +1 (888) 9PAJUNK (72-5865) Fax +1 (678) 514-3388 Establishment Registration Number: 3004076349 ## Contact Marco Wohnig President PAJUNK® Medical Systems L.P. Phone: (770) 493-6832 Direct: (470) 448-4070 Cell: (770) 757-2449 Fax: (678) 514-3388 E-Mail: marco.wohnig@pajunk-usa.com {4}------------------------------------------------ Image /page/4/Picture/1 description: The image features the word "PAJUNK" in a bold, sans-serif font. The letters are white, and they stand out against a teal background. A registered trademark symbol is located to the right of the letter "K". Pioneering medical technology | Device Information: | | |------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Device Name: | SPROTTE® NRFit™ and Quincke NRFit™<br>lumbar puncture needles | | Sterilization method: | Ethylene Oxide<br>disposable device, supplied sterile to the end<br>user and non-sterile intended to be sterilized<br>prior to use to repackagers/ medical device<br>manufacturers | | Contract Sterilizer: | Sterigenics Germany GmbH<br>Kasteler straße 45<br>65203 Wiesbaden<br>Germany, Hessen<br>Establishment Registration Number:<br>3002807090 | | Document Control Number | K160294 | | Classification Name: | Anesthesia Conduction Needle | | Classification Reference: | 21 CFR § 868.5150 | | Product Code: | BSP | | Establishment Registration Number: | 9611612 | | Regulatory Class: | II | | Panel: | Anesthesiology | | Predicate Devices: | K911260<br>(Owner: PAJUNK® GmbH Medizintechnologie)<br>ATRAUMATIC STANDARD SPROTTE<br>NEEDLE<br>K040965<br>(Owner: PAJUNK® GmbH Medizintechnologie)<br>PAJUNK TUOHY NEEDLES, QUINCKE<br>NEEDLES, CHIBA NEEDLES & CRAWFORD<br>NEEDLES | PAJUNK® GmbH Medizintechnologie is submitting this 510(k) for SPROTTE® NRFit™ and Quincke NRFit™ lumbar puncture needles. The intended use as well as the basic technical description of the needles which is relevant to clinical use is identical to the predicate devices and has been cleared in 510(k)s sent in earlier by the sponsor. The clinical technique, the indications for use, the technical specification, the materials used, the sterility status (validation and sterility assurance level) as the biocompatibility status is identical. None of these is affected by the altered hub as it is subject to this submission. {5}------------------------------------------------ Image /page/5/Picture/1 description: The image shows the word "PAJUNK" in all caps. The letters are white and the background is teal. There is a registered trademark symbol to the right of the K. Pioneering medical technology ## 1 Identification of Predicate devices In this section SPROTTE® NRFit™ and Quincke NRFit™ lumbar puncture needles are compared to the predicate device. In order to avoid confusion and to make the information more readable, the predicate devices shall be identified as Predicate Device I (which is for the SPROTTE® NRFit™) and Predicate Device II (which is for the Quincke NRFit™) since the Subject devices are identical in Indications for use while the predicates show differences in wording. ## Predicate Device I The predicate device for the SPROTTE® NRFit™ lumbar puncture needles is: - K911260 ATRAUMATIC STANDARD SPROTTE NEEDLE (Owner: PAJUNK® GmbH Medizintechnologie) ## Predicate Device II The predicate device for the Quincke NRFit™ lumbar puncture needles is: - K040965 PAJUNK TUOHY NEEDLES, QUINCKE NEEDLES, CHIBA NEEDLES & -CRAWFORD NEEDLES (Owner: PAJUNK® GmbH Medizintechnologie) ## Determination methods and results of Substantial Equivalence Determination: ## 2 Determination of Substantial Equivalence ## 2.1 Intended Use ## Intended Use Subject Device The SPROTTE® NRFit™, Quincke NRFit™ lumbar puncture needles are intended to gain entry into or puncture the spinal cavity permitting injection (including anesthesia) / withdrawal of fluids for purposes of diagnostic lumbar puncture, myelography/ discography procedures. The device is intended for adult and pediatric patients. ## Intended Use K911260 (Predicate Device I) To gain entry into or puncture the spinal cavity permitting injection/ withdrawal of fluids for purposes of diagnostic lumbar puncture, myelography/discography and chemonucleolysis procedures. Intended Use K040965 (Predicate Device II) Pajunk's anesthesia conduction needles - Tuohy, Quincke, Chiba, and Crawford - are intended for the transient delivery of anesthetics to provide regional anesthesia or to facilitate placement of an epidural catheter. {6}------------------------------------------------ Image /page/6/Picture/1 description: The image shows the word "PAJUNK" in white letters against a green background. The letters are bold and sans-serif. A registered trademark symbol is located to the right of the letter "K". ## Discussion of differences For the SPROTTE® NRFit™ lumbar puncture needles the indications for use is exactly the same as in K911260. For the Quincke NRFit™ lumbar puncture needles the indication for use is covered by the indications for use as given in the predicate K040965. Rationale: K040965 covers needles for regional anaesthesia (spinal and epidural) and placement of epidural catheters. The Quincke Needle subject to this Premarket Notification is intended for diagnostic lumbar puncture. Instead of injecting agents it is intended for collecting samples of cerebrospinal fluid. The anatomical region of the same. Furthermore Quincke needles (and SPROTTE needles) are state of the art in lumbar puncture for application of anesthetics as well as harvesting spinal fluid. So only the nature of fluid passing through the needle is different, but the anatomical regions are identical. By indicating the needle for collecting of specimens instead of injecting anesthetic agents the risk for the patient especially regarding contamination through fluid path is lowered. Conclusion: Substantially Equivalent The predicate device as well as the subject device only differs in design of the hub. This difference does not have any impact on the performance tests listed below due to the fact, that the materials used, the manufacturing processes employed and the techniques described are identical. ## 2.2 Technical Description ## 2.2.1 Technical Description SPROTTE | Materials used in SPROTTE® NRFit™ lumbar puncture needles: | | | | Predicate Device | |------------------------------------------------------------|--------------------------|-----------------|-------------------|------------------| | | NAME OF COMPONENT | MATERIAL | BODY CONTACT | MATERIAL K911260 | | 01 | Needle/cannula<br>Tubing | Stainless Steel | Direct, limited | Stainless Steel | | | Hub | Polycarbonate | Indirect, limited | Polycarbonate | | | Optional: Glue | Epoxy resin | No contact at all | Epoxy resin | | 02 | Introducer<br>Tubing | Stainless Steel | Direct, limited | Stainless Steel | | | Hub | Polycarbonate | No contact at all | Polycarbonate | | | Optional: Glue | Epoxy resin | No contact at all | Epoxy resin | | 03 | Stylet<br>Tubing | Stainless Steel | Direct, limited | Stainless Steel | | | Knob | Polycarbonate | No contact at all | Polyamide | | | Optional: Glue | Epoxy resin | No contact at all | Epoxy resin | | 04 | Retaining plate<br>Plate | Polycarbonate | No contact at all | Polycarbonate | {7}------------------------------------------------ Image /page/7/Picture/1 description: The image shows the word "PAJUNK" in white letters against a teal background. The letters are bold and sans-serif. A circled "R" symbol is present to the right of the "K", indicating that the name is a registered trademark. Pioneering medical technology Based on substantial equivalence testing listed below the following devices are subject to this premarket notification and shall be cleared: SPROTTE® NRFit™ in a range from 18G to 29G at a rage of length from 50mm to 150mm. Dimensions cleared in Predicate Device's Premarket Notification 510(k): K911260 Length: 70mm -150mm Diameter: 19,5G - 24G Dimensions intended in the subject devices Premarket Notification K160194: Length: 50mm to 150mm Diameter: 18G - 29G ## 12.2.2.1.1 Discussion of differences 1. The material of the hub has been altered in 1994. Before that date Polycarbonate yyy has been used which then was switched to Polycarbonate xxx. The biocompatibility test reports provided for the SPROTTE contain needles equipped with hubs made from xxx. However, the base substance for both is Polycarbonate Medical Grade. Note that yyy has a significant higher flow pattern index. 2. The material of the stylet's knob has been switched from Poliamide to Polycarbonate. Since the knob does not have any patient contact this change is non-significant to form, fit and function. | Materials used in Quincke NRFit™ lumbar puncture needles: | | | | Predicate Device | | |-----------------------------------------------------------|--------------------|-------------------|-----------------|-------------------|------------------| | | MATERIAL K911260 | | MATERIAL | BODY CONTACT | MATERIAL K040965 | | 01 | Needle/<br>cannula | Tubing | Stainless Steel | Direct, limited | Stainless steel | | | | Hub | Polycarbonate | Indirect, limited | Polycarbonate | | | | Optional:<br>Glue | Epoxy resin | No contact at all | Epoxy Resin | | 02 | Introducer | Tubing | Stainless Steel | Direct, limited | Stainless steel | | | | Hub | Polycarbonate | No contact at all | Polycarbonate | | | | Optional:<br>Glue | Epoxy resin | No contact at all | Epoxy Resin | | 03 | Stylet | Tubing | Stainless Steel | Direct, limited | Stainless steel | | | | Knob | Polycarbonate | No contact at all | Polycarbonate | | | | Optional:<br>Glue | Epoxy resin | No contact at all | Epoxy Resin | ## 2.2.2 Technical Description Quincke {8}------------------------------------------------ Image /page/8/Picture/1 description: The image shows the word "PAJUNK" in white letters against a green background. The letters are in a bold, sans-serif font. A registered trademark symbol is located to the upper right of the letter K. The logo is simple and clean, with a focus on readability. Pioneering medical technology | Materials used in Quincke NRFit™ lumbar puncture needles: | | | | | Predicate Device | |-----------------------------------------------------------|--------------------|---------------|-------------------|------------------|------------------| | | MATERIAL K911260 | MATERIAL | BODY CONTACT | MATERIAL K040965 | | | 04 | Retaining<br>plate | Polycarbonate | No contact at all | Polycarbonate | | | 05 | Fixation<br>Clip | Polycarbonate | No contact at all | Polycarbonate | | Based on substantial equivalence testing listed below the following devices are subject to this premarket notification and shall be cleared: Quincke NRFit™ in a range from 20G to 27G at a rage of length from 50mm to 120mm. Dimensions cleared in Predicate Device's Premarket Notification 510(k): K040965 Length: 35mm – 200mm Diameter: 14G - 23G Dimensions intended in the subject devices Premarket Notification K160194: Length: 25mm – 285mm Diameter: 20G - 27G ## 2.2.3 Devices under test For demonstrating substantial equivalence the items listed below have been compared: ## SPROTTE® NRFfit™ | # | Devices/ Materials | Item-number | Length<br>[mm] | Gauge | |---|--------------------|-------------|----------------|-------| | 1 | SPROTTE® NRFit™ | 321163-30F | 120 | 18 | | 2 | SPROTTE® NRFit™ | 321163-31C | 90 | 19 | | 3 | SPROTTE® NRFit™ | 321163-31B | 90 | 20 | | 4 | SPROTTE® NRFit™ | 321163-31A | 90 | 21 | | 5 | SPROTTE® NRFit™ | 141163-30C | 150 | 22 | | 6 | SPROTTE® NRFit™ | 131163-30A | 150 | 24 | | 7 | SPROTTE® NRFit™ | 261163-29A | 150 | 25 | | 8 | SPROTTE® NRFit™ | 231163-27A | 120 | 27 | | 9 | SPROTTE® NRFit™ | 501163-28A | 90 | 29 | ## SPROTTE® | # | Devices/ Materials | Item-number | Length<br>[mm] | Gauge | |---|--------------------|-------------|----------------|-------| | 1 | SPROTTE® | 321151-30F | 120 | 18 | | 2 | SPROTTE® | 321151-31C | 90 | 19 | | 3 | SPROTTE® | 321151-31B | 90 | 20 | | 4 | SPROTTE® | 321151-31A | 90 | 21 | | 5 | SPROTTE® | 141151-30C | 150 | 22 | | 6 | SPROTTE® | 141151-30A | 150 | 24 | | 7 | SPROTTE® | 061151-29A | 150 | 25 | | 8 | SPROTTE® | 231151-27A | 120 | 27 | | 9 | SPROTTE® | 501151-28A | 90 | 29 | ## Quincke NRFit™ {9}------------------------------------------------ Image /page/9/Picture/1 description: The image features the word "PAJUNK" in a bold, sans-serif font. The word is white and stands out against a teal background. A registered trademark symbol is located to the right of the letter "K". ## Pioneering medical technology | # | Devices/ Materials | Item-number | Length<br>[mm] | Gauge | |---|--------------------|-------------|----------------|-------| | 1 | Quincke NRFit™ | 1163-3G080 | 80 | 20 | | 2 | Quincke NRFit™ | 1163-3E090 | 90 | 22 | | 3 | Quincke NRFit™ | 1163-7Y090 | 90 | 24 | | 4 | Quincke NRFit™ | 1163-7C090 | 90 | 25 | | 5 | Quincke NRFit™ | 1163-7B090 | 90 | 27 | #### Quincke | # | Devices/ Materials | Item-number | Length<br>[mm] | Gauge | |---|--------------------|-------------|----------------|-------| | 1 | Quincke | 1149-3G080 | 80 | 20 | | 2 | Quincke | 1149-3E090 | 90 | 22 | | 3 | Quincke | 1149-7Y090 | 90 | 24 | | 4 | Quincke | 1149-7C090 | 90 | 25 | | 5 | Quincke | 1149-7B090 | 90 | 27 | Both, the subject devices as well as the predicate devices have been tested ready to use right after sterilization and have been subject to testing after accelerated aging (1year, 3years, 5years) in order to demonstrate substantial equivalence and Shelf Life. ## 2.3 Technology/ Performance ## 2.3.1 Performance Testing: Summary The predicate device as well as the subject device only differs in design of the hub. This difference does not have any impact on the performance tests listed below. Therefore performance testing is still valid and applies to both, subject device and predicate device. Additional benchmark tests of the predicate device and the subject device is not necessary and has not been performed due to the fact, that the materials used, the manufacturing processes employed and the techniques described are identical. Therefore no further benchmark testing is required in order to verify substantial equivalence. The performance tests have been accomplished in order to comply with the standards listed below (even though 6-366 and 6-362 are not directly linked with Product Code BSP): | Recognition-# | Standard Number | Title | |---------------|-----------------|-------------------------------------------------------------------------------------------------------------------------------| | 6-366 | ISO 9626 | Stainless steel needle tubing for the manufacture of<br>medical devices - Requirements and test methods | | 6-362 | ISO 7864 | Sterile hypodermic needles for single use -<br>Requirements and test methods | | 5-108 | ISO 80369-6 | Small bore connectors for liquids and gases in<br>healthcare applications - Part 6: Connectors for<br>neuraxial applications. | Stability test Bonding-to-Hub has to be conducted in order to compare the stability of the LUER Hub with the stability of the NRFit™ hub. Furthermore compliance of the needle hub with ISO 80369-6 has to be proven. {10}------------------------------------------------ Image /page/10/Picture/1 description: The image features the word "PAJUNK" in a bold, sans-serif typeface. The word is white and is set against a teal background. A small registered trademark symbol is located to the right of the letter "K". Pioneering medical technology Standard testing regarding the remaining sections of the recognized performance standards can be conducted using either the predicate device or the subject device because the needle tubing is identical. {11}------------------------------------------------ PAJUNK® #### Needle: stability test bending rigidity Reason for test: The needle has to demonstrate bending stability and resistance against breakage in order to resist forces reasonably assumed to the needle in situ under the defined intended use Procedure of test: The test procedure is defined by international standard ISO/FDIS 9626:2016 Stainless steel needle tubing for manufacture of medical devices. Pass/ Fail criteria: The acceptance criterion of bending rigidity for the cannula is defined in above mentioned standard. Results: The bending rigidity of the predicate device's needles and the subject device's needles is compliant with the standard. Conclusion: Substantially Equivalent ## Needle: stability test bonding to hub Reason for test: The needle has to demonstrate stability at the bonding of the hub in order to resist forces reasonably assumed to be applied to the needle in situ under the defined intended use. Procedure of test: The test procedure is defined by international standard ISO 7864:2016 Sterile hypodermic needles for single use Pass/ Fail criteria: The acceptance criterion for the bond between hub and needle tube (pull-off force) is ≥22N. Results: For the needles of the subject device and the needles of the predicate device a force significantly higher than the target value has to be applied. Therefore the predicate device 's needles as well as the subject device 's needles are substantially equivalent. Conclusion: Substantially Equivalent ## Needle: Penetration force Reason for test: The needles have to demonstrate less trauma when applied with the patient under the intended use. ISO 7864:2016 Sterile hypodermic needles for single use recommends penetration force testing only without giving a normative test method. Procedure of test: According to international European standard EN 13097. Pass/ Fail criteria: -none- objective comparison only. Results: The subject device's needles as well as the predicate device's needles show identical penetration/ insertion forces. Conclusion: Substantially Equivalent {12}------------------------------------------------ Image /page/12/Picture/1 description: The image features the word "PAJUNK" in a bold, sans-serif font. The letters are white, and they stand out against a teal background. A small, encircled "R" symbol is located to the upper right of the letter "K", indicating that the name is a registered trademark. ## 2.4 Sterilization The predicate device as well as the subject device only differs in design of the hub. This difference does not have any impact on Sterilization. Therefore the Sterilization is still valid. The contract sterilizer and the sterilizing process are identical to the contract sterilizer and the sterilizing process used for all PAJUNK® - manufactured devices which are already cleared for market or exempt. Sterilization parameters are | SAL | 10-6 | |-----------------|----------------------------------------------------------------------------------| | Type of gas | Ethylene Oxide 99,99% | | Exposure time | 300 min. | | Aeration method | evacuation<br>2 airwashes | | Aeration period | residual EtO-gas is removed in circulating<br>air at 40° C (±5) for at least 48h | Sterilization has been validated according to ISO 11135-1 Overkill Approach (1 sublethal cycle, 2 half cycle, 1 full cycle) Residuals of EO and ECH are in compliance with ISO 10993-7. Cleaning and Sterilization method, which ensures an SAL of 10° as well as compliance with limits for chemical burden, bioburden (i.e. LAL) and EtO-residuals as well as shelf life have been validated and are safe and effective. The limits listed below are met by each device: The needles are also available in bulk non sterile. If appropriately packed and sterilized with Ethyleneoxide according to the parameters above the technological parameters remain unchanged. However, final responsibility for sterilization remains with the customer of needles purchased bulk non-sterile. ## 2.5 Shelf Life The predicate device as well as the subject device only differs in design of the hub. This difference does not have any impact on Shelf Life. Therefore the Shelf Life is still valid. Efficacy of sterile product´s lifecycle has been validated using similar products and worst case devices. Sterility tests have been performed using worst case devices with similar characteristics made from identical material after 5 years. The devices were found to be sterile after 5 years, the sterile barrier system is efficient. Performance of the essential performance of the device (LUER/ NRFit™ connection, stability of bonding connections, catheter's tensile strength, needle 's bending rigidity) has been tested with real time aged needles and catheters made from identical material employing identical processes and those are found to work properly. There is no decrease in performance after 5 years. Shelf-life is set to 5 years. {13}------------------------------------------------ Image /page/13/Picture/1 description: The image features the word "PAJUNK" in a bold, sans-serif typeface. The letters are white, set against a solid teal background. A registered trademark symbol is present to the right of the letter "K". ## 2.6 Biocompatibility: The predicate device as well as the subject device only differs in design of the hub. The manufacturing processes of the subject device and the predicate device are identical. The difference in design does not have any impact on Biocompatibility of the device. Therefore the biocompatibility testing conducted with the predicate device is still valid. All devices comply with ISO 10993-1, 2nd and 3rd edition. The components of the SPROTTE® NRFit™ and Quincke NRFit™ are identical to the components of the SPROTTE and Quincke as they were approved/cleared in K911260 and K040965 in formulation, processing, and sterilization, and no other chemicals have been added (e.g., plasticizers, fillers, color additives, cleaning agents, mold release agents, etc.). The components of the subject devices SPROTTE® NRFit™ and Quincke NRFit™ and the components of the predicate devices SPROTTE and Quincke are identical to the components of the worst case device which has been tested in formulation, processing, and sterilization, and no other chemicals have been added (e.g., plasticizers, fillers, cleaning agents, mold release agents, etc.). The tests listed below have been conducted and accomplished successfully by the worst case device: - In vitro Cytotoxicity ISO 10993-5 - . Irritation ISO 10993-10 - Sensitization ISO 10993-10 - . Acute systemic toxicity - . Hemocompatibility Therefore and based upon sterilization validation and residuals validation the devices are considered to be biocompatible. ## 3. Conclusion All tests are passed successfully. There are no significant differences between the baseline and the aged needles Subject device and Predicate device nor are there significant differences between the predicate device and the subject device in performance. Compliance to international standards and Shelf life of 5 years is proven. The conclusions drawn from the nonclinical and clinical tests that the device is as safe, as effective, and performs as well as the legally marketed device identified.
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...