The Pajunk Epidural MiniFilter NRFit (ISO80369-6) and Epidural MiniFilter LUER (ISO80369-7) is an anesthesia conduction filter. An anesthesia conduction filter is a microporous filter used while administering to a patient injection of local anaesthetics to minimize particulate (foreign material) contamination of the injected fluid. The anaesthesia conduction filter is intended for use for patients that weigh 10kg and above.
Device Story
The Epidural MiniFilter is a sterile, disposable, microporous anesthesia conduction filter. It is placed in-line during the administration of local anesthetics to remove particulate matter from the injected fluid. The device features a modified acrylic housing and a hydrophilic HI-FLO PES membrane (0.2µm). It is available with either an NRFit (ISO 80369-6) or LUER (ISO 80369-7) connector to ensure secure attachment to syringes and needles. The device is used in clinical settings by healthcare professionals. By filtering the anesthetic solution, it reduces the risk of foreign material entering the patient's epidural space. The device is supplied sterile (EtO) or non-sterile for further processing by manufacturers.
Clinical Evidence
Bench testing only. No clinical data provided. Performance testing included connectivity, leak tightness, flow rate, burst pressure, and bubble point testing, all of which met internal protocols and relevant ISO standards (ISO 80369-6/7). Biocompatibility was evaluated per ISO 10993-1, including cytotoxicity and chemical analysis.
Technological Characteristics
Materials: Modified Acrylic housing, Hydrophilic HI-FLO PES membrane (0.2µm). Connectivity: NRFit (ISO 80369-6) or LUER (ISO 80369-7). Sterilization: Ethylene Oxide (ISO 11135). Form factor: Round, flat housing with rotating locking hub. Filtration area: 1.45cm². Flow rate: ≥ 200ml/Min. Bubble point: ≥ 46psi.
Indications for Use
Indicated for patients weighing 10kg and above requiring administration of local anesthetics via an anesthesia conduction filter to minimize particulate contamination of the injected fluid.
Regulatory Classification
Identification
An anesthesia conduction filter is a microporous filter used while administering to a patient injections of local anesthetics to minimize particulate (foreign material) contamination of the injected fluid.
K083451 — PORTEX EPIDURAL FILTER, MODEL 100/386/010, VARIOUS STANDARD AND CUSTOM TRAYS · Smiths Medical Asd, Inc. · Mar 4, 2009
K143581 — Arrow Epidural Catheter Kit · Teleflexmedical, Inc. · Jun 26, 2015
K161075 — Arrow Epidural Catheter Kit · Teleflexmedical, Inc. · Oct 4, 2016
K172410 — Portex NRFit Epidural Needles, Portex NRFit Spinal Needles, Portex NRFit Loss of Resistance (L.O.R.) Syringes., Portex EpiFuse Catheter Connector, Portex Regional Anaesthesia Portfolio Filters · Smiths Medical Asd, Inc. · Jun 1, 2018
K974661 — 0.2 U BACTERIAL FILTER · Sims Deltec, Inc. · Feb 11, 1998
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logos of the Department of Health & Human Services and the Food and Drug Administration (FDA). The Department of Health & Human Services logo is on the left, and the FDA logo is on the right. The FDA logo is a blue square with the letters "FDA" in white, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue.
June 17, 2019
PAJUNK GmbH Medizintechnologie Christian Quass Director Regulatory Affairs, Safety Officer Karl-Hall-Str. 1 78187 Geisingen. Baden-Wuerttemberg, Germany
Re: K190663
Trade/Device Name: Epidural MiniFilter LUER (ISO80369-7), Epidural MiniFilter NRFit (ISO80369-6) Regulation Number: 21 CFR 868.5130 Regulation Name: Anesthesia Conduction Filter Regulatory Class: Class II Product Code: BSN Dated: May 17, 2019 Received: May 20, 2019
Dear Christian Quass:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
Todd Courtney Assistant Director DHT1C: Division of ENT, Sleep Disordered Breathing, Respiratory and Anesthesia Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K190663
Device Name Epidural MiniFilter NRFit (ISO80369-6) and LUER (ISO80369-7)
#### Indications for Use (Describe)
The Pajunk Epidural MiniFilter NRFit (ISO80369-6) and Epidural MiniFilter LUER (ISO80369-7) is an anesthesia conduction filter. An anesthesia conduction filter is a microporous filter used while administering to a patient injection of local anaesthetics to minimize particulate (foreign material) contamination of the injected fluid. The anaesthesia conduction filter is intended for use for patients that weigh 10kg and above.
Type of Use (Select one or both, as applicable)
| <div> <span></span>Prescription Use (Part 21 CFR 801 Subpart D) </div> |
|---------------------------------------------------------------------------|
| <div> <span></span>Over-The-Counter Use (21 CFR 801 Subpart C) </div> |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the text "Premarket Notification Submission 510(k)". The text is in bold and black font. The text is centered on the image and takes up most of the space.
Image /page/3/Picture/1 description: The image shows the word "PAJUNK" in white letters against a teal background. The letters are slightly blurred, giving them a soft, glowing effect. A small registered trademark symbol is visible to the right of the letter "K".
Pioneering medical technology
### 510(k) Summary as required by 21 CFR 807.92(c).
Date of Preparation: 2019-06-17
# Document Control Number: K190663
510(k) owner:
PAJUNK® GmbH Medizintechnologie Karl-Hall-Str. 1 78187 Geisingen, Baden-Wuerttemberg, Germany
Establishment Registration Number: 9611612
#### Submitter Information/ production site:
PAJUNK® GmbH Medizintechnologie Karl-Hall-Str. 1 78187 Geisingen Baden-Wuerttemberg, Germany Fon: +49(0)7704-9291-586 Fax: +49(0)7704-9291-605 Establishment Registration Number: 9611612
# USA Contact:
PAJUNK MEDICAL SYSTEMS 6611 Bay Circle, Suite 140 Norcross, GA 30071 Phone +1 (888) 9PAJUNK (72-5865) Fax +1 (678) 514-3388
Establishment Registration Number: 3004076349
#### Contact:
Christian G. H. Quass Director Regulatory Affairs, Safety Officer Patricia Weisbrod Regulatory Affairs Manager Fon: +49(0)7704-9291-586 Fax: +49(0)7704-9291-605 E-Mail: christian.quass@pajunk.com E-Mail: patricia.weisbrod@pajunk.com
# Contact
Marco Wohniq President PAJUNK® Medical Systems L.P. Phone: (770) 493-6832 Direct: (470) 448-4070 Cell: (770) 757-2449 (678) 514-3388 Fax: E-Mail: marco.wohnig@pajunk-usa.com
{4}------------------------------------------------
Image /page/4/Picture/1 description: The image features the word "PAJUNK" in a bold, sans-serif typeface. The letters are white and stand out against a teal background. A small circle with an "R" inside, indicating a registered trademark, is located to the upper right of the letter "K".
| Device Information: | |
|------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Device Name: | Epidural MiniFilter NRFit™ (ISO80369-6)<br>Epidural MiniFilter LUER (ISO80369-7) |
| Sterilization method: | Ethylene Oxide<br>disposable device, supplied sterile to the end<br>user and non-sterile intended to be sterilized<br>prior to use to repackagers/ medical device<br>manufacturers |
| Contract Sterilizer: | Sterigenics Germany GmbH<br>Kasteler straße 45<br>65203 Wiesbaden<br>Germany, Hessen<br>Establishment Registration Number:<br>3002807090 |
| Document Control Number | K190663 |
| Classification Name: | filter, conduction, anesthetic |
| Classification Reference: | 21 CFR 868.5130 |
| Product Code: | BSN |
| Establishment Registration Number: | 9611612 |
| Regulatory Class: | II |
| Panel: | Anesthesiology |
| Predicate Device | K083451, Portex® Epidural Filter |
PAJUNK® GmbH Medizintechnologie is submitting this 510(k) for 2 types of Anaesthesia Conduction Filter called Epidural MiniFilter with either NRFit™ Connector according to ISO 80369-6 or LUER-Connector according to ISO 80369-7. The types only differ in shape of the hub.
The Filter is a Class II medical device substantially equivalent to the predicate device (K083451, Portex Epidural Filter) as defined in 21 CFR §868.5130, product code BSN
The device is a sterile finished disposable device, supplied sterile to the end user and non-sterile intended to be sterilized prior to use to re-packagers/ medical device manufacturers.
The predicate device chosen for demonstrating substantial equivalence is the Portex Epidural Filter manufactured by SMITHS MEDICAL ASD, INC., 10 bowman dr., keene, NH 03431 and cleared by the Food and Drug Administration for market under K083451, product code BSN, review panel Anaesthesiology.
{5}------------------------------------------------
Image /page/5/Picture/1 description: The image shows the word "PAJUNK" in white letters against a teal background. The letters are bold and slightly blurred, giving them a soft appearance. A small circled R symbol is located to the upper right of the K, indicating a registered trademark.
#### Indications for use
The Pajunk Epidural MiniFilter NRFit™ (ISO80369-6) and Epidural MiniFilter LUER (ISO80369-7) is an anesthesia conduction filter. An anesthesia conduction filter is a microporous filter used while administering to a patient injection of local anaesthetics to minimize particulate (foreign material) contamination of the injected fluid. The anaesthesia conduction filter is intended for use for patients that weigh 10kg and above.
#### Determination methods and results of Substantial Equivalence Determination:
| Intended Use | Result: Substantially Equivalent |
|--------------|----------------------------------|
| Design | Result: Substantially Equivalent |
| Performance | Result: Substantially Equivalent |
| | Predicate device | Subject Device | Result of comparison,<br>if necessary with<br>rationale |
|------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------|
| Characteristics | K083451 | Anaesthesia Conduction Filter | |
| Biocompatibility | Both devices are classified as externally communicating. Therefore and<br>based upon ISO10993-1 biocompatibility applies. | | Substantially equivalent |
| Packaging | Individually sterile in soft blister<br>pouch or as part of a kit | Individually sterile in soft blister<br>pouch or as part of a kit | Same packaging<br>The PAJUNK®<br>individual packaging<br>was selected according<br>to a validated<br>sterilization and<br>transportation process |
| Indications for<br>Use | An anesthesia conduction filter is<br>a microporous filter used while<br>administering to a patient<br>injections of local anesthetics to<br>minimize particulate (foreign<br>material) contamination of the<br>injected fluid. | The Pajunk Epidural MiniFilter<br>NRFitTM (ISO80369-6) and<br>Epidural MiniFilter LUER<br>(ISO80369-7) is an anesthesia<br>conduction filter.<br>An anesthesia conduction filter is<br>a microporous filter used while<br>administering to a patient<br>injections of local anesthetics to<br>minimize particulate (foreign<br>material) contamination of the<br>injected fluid. | Substantially equivalent |
| Overall design: | Round, flat housing<br>Female Lock and male connector<br>with rotating locking hub<br>Filter membrane | flat housing<br>Female Lock and male connector<br>with rotating locking hub<br>Filter membrane | identical<br>Both devices do not<br>have direct patient<br>contact and are made<br>from plastic materials |
#### Equivalence in materials used
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image shows the word "PAJUNK" in white letters against a teal background. The letters are bold and sans-serif. A small circle with an "R" inside is located to the right of the "K", indicating that it is a registered trademark.
### Pioneering medical technology
| | Predicate device | Subject Device | Result of comparison,<br>if necessary with<br>rationale |
|-----------------|------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Characteristics | K083451 | Anaesthesia Conduction Filter | |
| Picture | Image: Predicate device | Image: Subject Device | Both systems allow<br>easy handling and<br>connecting |
| Materials used | Filter housing: Modified Acrylic<br>Filter Media: Supor® 200- 0.2µm<br>(PES) | Filter housing: Modified Acrylic<br>Filter Media: Hydrophilic HI-FLO<br>PES membrane 0.2µm | identical<br>Each of the materials<br>used either in the<br>Predicate Devices or<br>the Subject Device are<br>established materials<br>used for manufacturing<br>medical devices. |
| Connectivity | ISO 80369-7 | ISO 80369-6 NRFit™<br>ISO 80369-7 LUER | Both systems do have<br>identical connectivity<br>towards syringe and<br>needle; 2017's state of<br>the art connection type<br>is NRFit™ which slightly<br>differs in dimensions of<br>the male and female<br>connectors in order to<br>avoid misconnections. |
| Filtration Area | 5,25cm2 | 1,45cm2 | Due to smaller size the<br>subject devices filtration<br>area is smaller. The<br>filtration properties i. e.<br>the intended use are<br>not affected by this<br>difference. |
| Flow Rate | ≥ 200ml/Min | ≥ 200ml/Min | identical |
| Bubble Point | ≥ 46psi. | ≥ 46psi. | identical |
# Equivalence in the Indications for use
### Subject Device:
The Pajunk Epidural MiniFilter NRFit™ (ISO80369-6) and Epidural MiniFilter LUER (ISO80369-7) is an anesthesia conduction filter. An anesthesia conduction filter is a microporous filter used while administering to a patient injection of local anaesthetics to minimize particulate (foreign material) contamination of the injected fluid.
{7}------------------------------------------------
Image /page/7/Picture/1 description: The image features the word "PAJUNK" in bold, white letters against a teal background. The font is sans-serif and appears to be a modern, clean design. A registered trademark symbol is present to the right of the letter "K".
#### Predicate Devices:
An anesthesia conduction filter is a microporous filter used while administering to a patient injection of local anaesthetics to minimize particulate (foreign material) contamination of the injected fluid.
#### Discussion
The indications for use as well as the intended use of the predicate devices and of the subject device are identical. Pajunk's IFU statement is amended by an introducing statement of identify identifying the tradenames for the common term "anaesthesia conduction filter".
Conclusion: Substantially Equivalent
#### Sterilization
#### Subject Device:
Sterilized with Ethylene Oxide. Shelf Life: 5 years.
#### Predicate Devices:
Sterilized with Ethylene Oxide. Shelf Life: 5 years.
#### Discussion
Both devices are sterilized using Ethylene Oxide and have a shelf life of 5 years.
Conclusion: Substantially Equivalent
#### Performance Testing
#### Subject Device:
Connectivity: LUER and NRFit™ (ISO 80369-6 and ISO 80369-7). Leak Tightness: Complies with internal protocol Flow Rate: Complies with internal protocol Burst Pressure: Complies with internal protocol Bubble Point: Complies with internal protocol
{8}------------------------------------------------
Image /page/8/Picture/1 description: The image features the word "PAJUNK" in bold, white letters against a teal background. The font is sans-serif and appears to be slightly rounded. A small registration mark is visible to the upper right of the letter "K".
Pioneering medical technology
### Predicate Devices:
Connectivity: LUER (ISO594/ ISO80369-7).
Leak Tightness: Complies with internal protocol of the sponsor
Flow Rate: Complies with internal protocol of the sponsor
Burst Pressure: Complies with internal protocol of the sponsor
Bubble Point: Complies with internal protocol of the sponsor
# Technology Characteristics/ Performance Testing
#### Sterilization
The contract sterilizer and the sterilizing process are identical to the process and sterilizer used for all PAJUNK® - manufactured devices which are already cleared for market or exempt.
Sterilization parameters are
| SAL | 10-6 |
|-----------------|----------------------------------------------------------------------------------|
| Type of gas | Ethylene Oxide 99,99% |
| Exposure time | 300 min. |
| Aeration method | evacuation<br>2 airwashes |
| Aeration period | residual EtO-gas is removed in circulating<br>air at 40° C (±5) for at least 48h |
Sterilization has been validated according to ISO 11135-1 Overkill Approach (1 sublethal cycle, 2 half cycle, 1 full cycle)
Residuals of EO and ECH are in compliance with ISO 10993-7.
Cleaning and Sterilization method, which ensures an SAL of 10° as well as compliance with limits for chemical burden, bioburden (i.e. LAL) and EtO-residuals as well as shelf life have been validated and are safe and effective.
The limits listed below are met by each device:
Limits for Residuals of Ehyleneoxide and Ethylene chlorhydrine are met.
Limit for Pyroburden/ endotoxine is met.
The filters are also available in bulk non-sterile. If appropriately packed and sterilized with Ethyleneoxide according to the parameters above the technological parameters remain unchanged. However, final responsibility for sterilization remains with the customer of filters purchased bulk non-sterile.
{9}------------------------------------------------
Image /page/9/Picture/1 description: The image features the word "PAJUNK" in bold, white letters against a teal background. The font is sans-serif and appears to be a modern, clean typeface. A registered trademark symbol is present to the right of the letter "K".
#### Shelf Life
Efficacy of sterile product's lifecycle has been validated.
Sterility tests have been performed. Performance of the essential performance of the device (NRFit and LUER connection, stability of connections) has been tested with real time aged devices (1 year) and devices subject to accelerated aging (1 year, 3 years). There is no decrease in performance after 5 years.
Shelf-life is set to 5 years.
#### Biocompatibility:
All products comply with ISO 10993-1, 2nd and 3rd edition.
The Subject Device is an external communicating device with no direct patient contact. Therefore based upon ISO10993-1 biocompatibility is evaluated and tested.
Biocompatibility has been evaluated and tested according to ISO 10993-1. The tests conducted are testing for Cytotoxicity according to ISO10993-5 and Chemical Analysis and Evaluation according to ISO10993-12. The evaluation and qualification method complies with ISO10993-1 as well as with FDA-2013-D-0350 "Use of International Standard ISO 10993-1, "Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process".
Therefore and based upon sterilization and residuals validation the devices are considered to be biocompatible.
#### Standards/ Requirements
The Subject Device has been tested to comply with the state-of-the-art standards listed below. For connector standards both, the male and female connectors have been tested:
| Test Detail | Standard | FDA-Rec.-No. | Result |
|-------------------|------------------|--------------|--------|
| Sterilization | ISO 11135 | 14-452 | Pass |
| Residuals | ISO 10993-7 | 14-408 | Pass |
| Biocompatibility | ISO 10993-1 | 2-220 | Pass |
| Liquid Leakage | ISO 80369-7, 6.1 | 5-115 | Pass |
| Air Leakage | ISO 80369-7, 6.2 | 5-115 | Pass |
| Stress Cracking | ISO 80369-7, 6.3 | 5-115 | Pass |
| Axial Load | ISO 80369-7, 6.4 | 5-115 | Pass |
| Unscrewing torque | ISO 80369-7, 6.5 | 5-115 | Pass |
| Overriding | ISO 80369-7, 6.6 | 5-115 | Pass |
{10}------------------------------------------------
Image /page/10/Picture/1 description: The image shows the word "PAJUNK" in white letters against a teal background. The letters are bold and sans-serif. There is a registered trademark symbol to the right of the letter K.
Pioneering medical technology
| Test Detail | Standard | FDA-Rec. -No. | Result |
|----------------------------------------|-------------------|---------------|--------|
| Liquid Leakage | ISO 80369-6, 6.1 | 5-108 | Pass |
| Air Leakage | ISO 80369-6, 6.2 | 5-108 | Pass |
| Stress Cracking | ISO 80369-6, 6.3 | 5-108 | Pass |
| Axial Load | ISO 80369-6, 6.4 | 5-108 | Pass |
| Unscrewing torque | ISO 80369-6, 6.5 | 5-108 | Pass |
| Overriding | ISO 80369-6, 6.6 | 5-108 | Pass |
| Burst Pressure | Internal protocol | n.a. | Pass |
| Bubble Point | Internal Protocol | n.a. | Pass |
| Leak Tightness | Internal Protocol | n.a. | Pass |
| Flow Rate | Internal Protocol | n.a. | Pass |
| ISTA Procedure 2A (ASTM D 6653 et al.) | | | |
| ASTM F1886 - 09 | | | |
| ASTM F 1929 – 98 | | | |
| ASTM E 515 – 05 (4a)) | | | |
| DIN EN 868-5:2009-09 | | | |
| DIN EN 868-10:2009-09 | | | |
# Conclusion:
The comparison between the predicate device and the subject device, as well as the validated sterilization process and the results of the standard testing and performance testing demonstrates that the subject device is substantially equivalent to the predicate device. The subject device is safe and effective as the legally marketed predicate device.
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.