K202970 · Inovytech Medical Solutions , Ltd. · CBK · Jan 29, 2021 · Anesthesiology
Device Facts
Record ID
K202970
Device Name
Ventway Sparrow
Applicant
Inovytech Medical Solutions , Ltd.
Product Code
CBK · Anesthesiology
Decision Date
Jan 29, 2021
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 868.5895
Device Class
Class 2
Attributes
Therapeutic, Pediatric
Indications for Use
The Ventway Sparrow ventilator is intended to provide continuous or intermittent ventilatory support for the care of individuals who require mechanical ventilation. Specifically, the ventilator is applicable for adult and pediatric patients weighing at least 5 kg (11 lb.), who require the following types of ventilatory support: SIMV VC PS, SIMV PC PS or CPAP. The Ventway Sparrow lung ventilator is intended for emergency use and during transportation. It may be used for invasive (via an endotracheal tube and tracheostomy) or noninvasive (full non-vented ventilation face mask) ventilation presets. It may be used in hospital (transport) and field environments. Models VWSP100MR and VWSP900MR may be used in a Magnetic Resonance (MR) environment up to 3 Tesla. The ventilator is a restricted medical device intended for use by qualified, trained personnel under the direction of a physician.
Device Story
Portable ventilator for invasive/non-invasive support; used in hospital, transport, and field environments by qualified personnel. Inputs: atmospheric air; user-set parameters (Tidal Volume, Rate, PEEP, FiO2). Operation: blower-based positive pressure delivery; controlled via rotary encoder/push-button interface with 1.8" TFT LCD display. Outputs: mechanical ventilation; real-time monitoring of PEEP, PIP, Tidal Volume, I:E ratio, Breath rate, Minute Volume. Alarms: extensive set including disconnect, apnea, pressure, and battery alerts. Benefits: provides life-sustaining respiratory support in mobile/emergency settings; MR-conditional models (VWSP100MR/VWSP900MR) allow use in 3 Tesla MRI environments. Healthcare providers use output data to adjust ventilation settings and monitor patient status.
Clinical Evidence
No clinical data. Bench testing only. Performance validated via cleaning/disinfection, durability, biocompatibility (ISO 10993-5/10, ISO 18562-2/3), electrical safety (IEC 60601-1), EMC, RFID immunity, environmental testing, and adherence to ISO 80601-2-12 and ISO 10651-3. Comparative waveform testing and human factors/usability studies performed.
Technological Characteristics
Blower-based positive pressure ventilator. Materials: biocompatible (ISO 10993/18562). Power: AC/DC, rechargeable Li-Ion or non-rechargeable Li-Mn batteries. Interface: 1.8" TFT LCD (160x128, ST7735 driver) with rotary encoder. Connectivity: standalone. MR-conditional (ASTM F2119) up to 3 Tesla. Standards: IEC 60601-1, IEC 60601-1-2, IEC 60601-1-12, ISO 80601-2-12, IEC 62304, EN 1789, RTCA DO-160G.
Indications for Use
Indicated for adult and pediatric patients weighing ≥5 kg requiring mechanical ventilation (SIMV-VC, SIMV-PC, or CPAP) in emergency, transport, hospital, or field environments. Contraindications not specified.
Regulatory Classification
Identification
A continuous ventilator (respirator) is a device intended to mechanically control or assist patient breathing by delivering a predetermined percentage of oxygen in the breathing gas. Adult, pediatric, and neonatal ventilators are included in this generic type of device.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
Inovytech Medical Solutions Ltd. % Paul Dryden Consultant c/o ProMedic, LLC 131 Bay Point Dr. NE St. Petersburg, Florida 33704
# Re: K202970
Trade/Device Name: Ventway Sparrow Regulation Number: 21 CFR 868.5895 Regulation Name: Continuous Ventilator Regulatory Class: Class II Product Code: CBK Dated: September 29, 2020 Received: September 30, 2020
# Dear Paul Dryden:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part
{1}------------------------------------------------
801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4. Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Brandon Blakely, PhD Acting Assistant Director DHT1C: Division of ENT, Sleep Disordered Breathing, Respiratory and Anesthesia Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known)
### K202970
Device Name
### Ventway Sparrow
Indications for Use (Describe)
The Ventway Sparrow ventilator is intended to provide continuous or intermittent ventilatory support for the care of individuals who require mechanical ventilation. Specifically, the ventilator is applicable for adult and pediatric patients weighing at least 5 kg ( 1 1b.), who require the following types of ventilatory support: SIMV VC PS, SIMV PC PS or CPAP.
The Ventway Sparrow lung ventilator is intended for emergency use and during transportation. It may be used for invasive (via an endotracheal tube and tracheostomy) or noninvasive (full non-vented ventilation face mask) ventilation presets. It may be used in hospital (transport) and field environments.
Models VWSP100MR and VWSP900MR may be used in a Magnetic Resonance (MR) environment up to 3 Tesla.
The ventilator is a restricted medical device intended for use by qualified, trained personnel under the direction of a physician.
Type of Use (Select one or both, as applicable)
XX Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
| Date Prepared: | 29-Jan-21 |
|---------------------------------|-------------------------------------------------------------------------|
| Inovytec Medical Solutions Ltd. | |
| | 5 HaTidhar St., |
| | Raanana 4366507, Israel |
| | Tel: +972 9 779 41 35 |
| Official Contact: | Dana Hofeller – QA&RA Manager |
| Submission Correspondent: | Paul Dryden |
| | ProMedic, LLC |
| Proprietary or Trade Name: | Ventway Sparrow |
| Common/Usual Name: | Ventilator, Continuous, Facility Use |
| Regulation Number: | 21CFR 868.5895 |
| Regulation Code: | Ventilator, Continuous, Facility Use |
| Product Code: | CBK |
| Regulatory Class: | II |
| Predicate Device: | K083688 - LTV® 1200 MR Conditional Ventilator - Pulmonetic Systems Inc. |
### Device Description:
The Ventway Sparrow ventilator (a.k.a. Ventway Ventilator / Ventway System) is a portable ventilator, used for transport, in prehospital, field hospital, and hospital settings.
The ventilator is suitable for non-invasive ventilation for a full non-vented ventilation face mask or for invasive ventilation via an endotracheal tube or tracheostomy.
The Ventway is available in 2 models plus MR conditional use:
- . VWSP-100 Civil Model
- VWSP-100MR Civil Model ●
- VWSP-900 Robust Model ●
- VWSP-900MR Robust Model
### Indications for Use:
The Ventway Sparrow ventilator is intended to provide continuous or intermittent ventilatory support for the care of individuals who require mechanical ventilation. Specifically, the ventilator is applicable for adult and pediatric patients weighing at least 5 kg (11 lb.), who require the following types of ventilatory support: SIMV - VC (PS), SIMV-PC, or CPAP.
The Ventway Sparrow lung ventilator is intended for emergency use and during transportation. It may be used for invasive (via an endotracheal tube and tracheostomy) or noninvasive (full nonvented ventilation face mask) ventilation presets. It may be used in hospitals, pre-hospital (transport) and field environments.
{4}------------------------------------------------
### 510(k) Summary Page 2 of 6
Models VWSP100MR and VWSP900MR may be used in a Magnetic Resonance (MR) environment up to 3 Tesla.
The ventilator is a restricted medical device intended for use by qualified, trained personnel under the direction of a physician.
# Patient Population:
Adult and pediatric patients weighing at least 5 kg (11 lb.)
### Environments of use:
Hospitals, pre-hospital (transport) and field environments
Table 1 is a comparison - Subject Device vs. the Predicate, K083688 - LTV 1200.
| Table | 1 |
|-------|---|
| | |
| Technological | Ventway Sparrow<br>Inovytec Ltd. | Predicate - LTV 1200 MR Conditional<br>Ventilator Pulmonetic Systems Inc. -<br>K083688 |
|--------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Characteristic | | |
| Product Code, Class | CBK, Class II | CBK, Class II |
| Indications for Use | The Ventway Sparrow ventilator is intended to<br>provide continuous or intermittent ventilatory<br>support for the care of individuals who require<br>mechanical ventilation. Specifically, the<br>ventilator is applicable for adult and pediatric<br>patients weighing at least 5 kg (11 lb.), who<br>require the following types of ventilatory<br>support: SIMV - VC (PS), SIMV-PC, or CPAP.<br><br>The Ventway Sparrow lung ventilator is<br>intended for emergency use and during<br>transportation. It may be used for invasive (via<br>an endotracheal tube and tracheostomy) or<br>noninvasive (full non-vented ventilation face<br>mask) ventilation presets. It may be used in<br>hospitals, pre-hospital (transport) and field<br>environments.<br><br>Models VWSP100MR and VWSP900MR may<br>be used in a Magnetic Resonance (MR)<br>environment up to 3 Tesla.<br><br>The ventilator is a restricted medical device<br>intended for use by qualified, trained personnel<br>under the direction of a physician. | The LTV® 1200 ventilator is intended to<br>provide continuous or intermittent ventilatory<br>support for the care of the individuals who<br>require mechanical ventilation. Specifically,<br>the ventilator is applicable for adult and<br>pediatric patients weighing at least 5kg (11<br>lbs.), who require the following types of<br>ventilatory support: PPV, Assist/Control,<br>SIMV, CPAP, or NPPV. |
| User Population | Qualified, trained personnel under the direction<br>of a physician. | Qualified, trained personnel under the direction<br>of a physician. |
| Target Population | Adults and pediatric patients weighing ≥5 kg | Adults and pediatric patients weighing ≥5 kg |
| Environment Used | Hospitals, field, transport | Institutional, home or transport settings |
| Energy Used /<br>Delivered | No energy delivered | No energy delivered |
| Design: | The Ventway Sparrow consists of a ventilator<br>unit with an LCD screen and a single use<br>patient circuit | The LTV® 1200 consists of a ventilator unit<br>with LCD displays and a reusable patient<br>circuit |
| Technological<br>Characteristic | Ventway Sparrow<br>Inovytec Ltd. | Predicate - LTV 1200 MR Conditional<br>Ventilator Pulmonetic Systems Inc. -<br>K083688 |
| -<br>Mechanism of<br>Action | The Ventway Sparrow provides continuous or<br>intermittent ventilatory support. | The LTV® 1200 provides continuous or<br>intermittent ventilatory support. |
| -<br>Components | The Ventway Sparrow consists of the following<br>components:<br>- Ventilator<br>- Patient Circuit<br>- AC/DC Power Supply<br>- DC/DC Power Adapter<br>- Rechargeable Battery Pack<br>- Non-rechargeable Battery Pack<br>- Oxygen Mixer (external)<br>- Oxygen enrichment kit<br>- Inlet Filter | The LTV® 1200 consists of the following<br>components:<br>- Ventilator<br>- Patient Circuit<br>- AC/DC Power Adapter<br>- DC/DC Power Adapter |
| -<br>Dimensions<br>(HxWxD) | 2.36" x 6.5" x 6.57"<br>(6.0 cm x 16.5 cm x 16.7 cm) | 3.25" x 10.5" x 13.5"<br>(8.4 cm x 27 cm x 38 cm) |
| -<br>Weight | Standard version - 2.2 lbs. (1 kg)<br>Robust version - 2.64 lbs. (1.2 kg) | 14.5 lbs. (6.5 kg) |
| -<br>Battery | Rechargeable: Lithium Ion<br>Non-rechargeable: Li-Mn | Rechargeable: Sealed Lead Acid |
| -<br>Duration per charge | 4 hr. @ VT=500ml, 12bmp, PEEP=5cmH2O,<br>FIO2=21% | 1hr @ VT=800ml, 15bmp, PEEP=5cmH2O,<br>FIO2=21% |
| -<br>External power<br>supply | Input: 100 - 264 VAC/ 50-60 Hz, max 1.6 A<br>Output: 16-28 VDC, 120 W | Input: 100 to 250 VAC/ 50-60 Hz<br>Output: 13 VDC |
| -<br>Display/User<br>Interface | LCD: SPI 1.8" TFT 160x128 driver st7735,<br>Rotary encoder selector integrated with push<br>button | 7- segment control display, LEDs, push buttons<br>and a control knob |
| Performance | • SIMV-VC (PS) (a.k.a SIMV)<br>• CPAP<br>• Apnea Backup Ventilation<br>• Control (Trigger sensitivity "off")<br>• Assist/Control (AC)<br>• SIMV-PC | • SIMV<br>• CPAP<br>• Apnea Backup Ventilation<br>• Control (Trigger sensitivity "--")<br>• Assist/Control (AC)<br>• NPPV |
| -<br>Basic and Advanced<br>User Modes | Yes | Yes |
| -<br>Manual Triggered<br>Breath | Yes | Yes |
| -<br>Air Source | Atmospheric | Atmospheric |
| -<br>Controls | "Multi-function" control knob | Dedicated control buttons and knob |
| -<br>Monitors | PEEP, PIP ,Tidal Volume (Vt), I:E ratio,<br>Breath rate (bpm), Minute Volume | PEEP, PIP, Tidal Volume (Vt), I:E Ratio,<br>Breath rate (bpm), MAP, Minute Volume |
| Settings: | | |
| -<br>Tidal Volume<br>[mL/breath] | 50-2000 | 50-2000 |
| -<br>Respiratory Rate<br>[bpm] | 1-60 | 5-80 |
| -<br>Inspiratory Time<br>(sec) | 0.3-4 ±10% | 0.3-9.9 ±0.05 s |
| -<br>PIP Limit [cmH2O] | 11-60 ±5 | 5-100 |
| Technological | Ventway Sparrow<br>Inovytec Ltd. | Predicate - LTV 1200 MR Conditional<br>Ventilator Pulmonetic Systems Inc. -<br>K083688 |
| - PEEP [cmH2O] | 0-20 | 5-20 |
| - Supplemental<br>Oxygen (FiO2) | 21% to 95% ±5 | 21% to 50% ±3<br>51% to 95% ±5 |
| - Alarms | Blower malfunction<br>Tube disconnect<br>Patient disconnect<br>Apnea<br>System recovered from a crash<br>Battery empty<br>Sensor disconnect<br>Low respiratory rate<br>High minute volume<br>Low minute volume<br>High inspiratory pressure<br>Low inspiratory pressure<br>Leak<br>Inverse I:E ratio<br>High temperature<br>Expiratory valve blocked<br>High PEEP<br>Low PEEP<br>High respiratory rate<br>Tidal Volume Limit Reached<br>Low tidal volume<br>Low pressure<br>Battery low<br>High Voltage<br>Replace filter<br>Service needed<br>Altitude out of range<br>Unexpected Restart | Start-up Self-Test<br>Circuit Disconnect<br>Over Pressure/Blockage<br>Apnea<br>External power indicator<br>Low and Empty Battery<br>Detect Spontaneous Breath<br>High PEEP |
| Patient Breathing<br>Circuit<br>Disposable | Single limb<br>Exhalation valve and PEEP valve<br>User to add legally marketed non-rebreathing<br>valve | Exhalation valve and PEEP valve |
| Accessories – User<br>supplied | Bacterial / Viral Filter<br>HME<br>Non-rebreathing Valve (22mm O.D. inlet x<br>22mm I.D. outlet, resistance < 0.5 cmH2O) | Bacterial / Viral Filter<br>HME |
| Standards Met | IEC 60601-1<br>IEC 60601-1-2<br>IEC 60601-1-12<br>ISO 80601-2-12<br>IEC 62304<br>EN 1789<br>RTCA DO-160G<br>RFID AIM7351731<br>ASTM F2119 - 3 Tesla | IEC 601-1<br>IEC 68-2-27<br>IEC 68-2-6<br>IEC 68-2-34<br>MIL-STD-810E |
| Technological<br>Characteristic | Ventway Sparrow<br>Inovytec Ltd. | Predicate - LTV 1200 MR Conditional<br>Ventilator Pulmonetic Systems Inc. -<br>K083688 |
| Biocompatibility | Externally communicating, tissue, prolonged<br>duration<br>ISO 10993-5<br>ISO 10993-10<br>ISO 18562-2<br>ISO 18562-3 | Externally communicating, tissue, prolonged<br>duration |
| Human Factors /<br>Usability | Preformed with the identified user group | |
{5}------------------------------------------------
# 510(k) Summary
{6}------------------------------------------------
### 510(k) Summary Page 4 of 6
{7}------------------------------------------------
### 510(k) Summary Page 5 of 6
### Substantial Equivalence Discussion
The Ventway Sparrow has the same intended use and similar technological characteristics and principles of operation as the predicate LTV 1200, K083688.
### Indications for Use
The Ventway Sparrow ventilator is intended to provide continuous or intermittent ventilatory support for the care of individuals who require mechanical ventilation. It may be used for invasive or noninvasive ventilation presets. It may be used in hospital (transport) and field environments.
This is similar to the predicate.
### Technological Characteristics
Both devices have blower-based technology to provide positive pressure and have applicable alarms and sensors to maintain performance within their specifications.
### Non-clinical Testing
Performance testing included:
- Cleaning and Disinfection ●
- Durability ●
- Biocompatibility ●
- Software Verification and Validation ●
- Electrical Safety ●
- Electromagnetic Compatibility
- RFID Immunity
- Environmental
- ISO 80601-2-12 Medical electrical equipment Part 2-12: Particular requirements for the . safety of lung ventilators - Critical care ventilators [Including: Technical Corrigendum 1 (2011)]
- . ISO 10651-3: 1997 Lung ventilators for medical use - Part 3: Particular requirements for emergency and transport ventilators
- Comparative waveform testing
- Conical connectors ●
- MRI
- Altitude ●
- Human Factors / Usability
{8}------------------------------------------------
### 510(k) Summary Page 6 of 6
The results demonstrate similar performance.
# Substantial Equivalence Conclusion
The sponsor has demonstrated through performance testing, design and features, and non-clinical testing that the proposed device and predicate have been found to substantially equivalent.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.