Servo-u Ventilator System 4.1, Servo-n Ventilator System 4.1, Servo-u MR Ventilator System 4.1
K201874 · Maquet Critical Care AB · CBK · Apr 20, 2021 · Anesthesiology
Device Facts
Record ID
K201874
Device Name
Servo-u Ventilator System 4.1, Servo-n Ventilator System 4.1, Servo-u MR Ventilator System 4.1
Applicant
Maquet Critical Care AB
Product Code
CBK · Anesthesiology
Decision Date
Apr 20, 2021
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 868.5895
Device Class
Class 2
Attributes
Therapeutic, Pediatric
Indications for Use
The Servo-u Ventilator System is: - · intended for respiratory support, monitoring and treatment of neonatal, pediatric and adult patients - to be used only by healthcare providers - · to be used only in professional healthcare facilities and for transport within these facilities The Servo-n Ventilator System is: - · intended for respiratory support, monitoring and treatment of neonatal and pediatric patients - · to be used only by healthcare providers - · to be used only in professional healthcare facilities and for transport within these facilities The Servo-u MR Ventilator System is: - · intended for respiratory support, monitoring and treatment of neonatal, pediatric and adult patients - · to be used only by healthcare providers - · to be used only in professional healthcare facilities and for transport within these facilities - · to be used in MR environment according to specified conditions - with 1.5 T or 3 T MR scanners - outside magnetic fields >20 mT/200 Gauss
Device Story
Ventilator systems (Servo-u, Servo-n, Servo-u MR) provide respiratory support via controlled/supported breaths (constant flow/pressure) or High Flow therapy. Systems consist of a patient unit (gas mixing/delivery) and user interface. Input includes patient respiratory effort measured via Electrical activity of the diaphragm (Edi) catheters. NAVA (Neurally Adjusted Ventilatory Assist) mode delivers assist synchronized with patient respiratory drive. Used in professional healthcare facilities/transport; Servo-u MR includes magnetic field indicator and shielded components for 1.5T/3T MR environments. Output includes ventilation delivery, monitoring data, and visual/audible alarms. Healthcare providers use output to adjust therapy, monitor patient status, and manage respiratory failure. Benefits include synchronized ventilation and continuous monitoring.
Clinical Evidence
Bench testing only. Verification and validation activities included software code review, static analysis, unit/integration testing, waveform testing, biocompatibility (ISO 10993-1, ISO 18562-1), and human factors validation. Compliance with standards including ANSI/AAMI ES 60601-1, IEC 60601-1-2, IEC 60601-1-8, ISO 80601-2-12, and ISO 80601-2-55.
Technological Characteristics
Critical care ventilators with patient unit and user interface. Features: Edi monitoring, NAVA mode, High Flow therapy. Servo-u MR includes magnetic field indicator, shielded panel cables, and auto-lock mobile cart. Software-controlled gas delivery. Connectivity: battery-powered for transport. Sterilization: per ISO 17664. Software: Major Level of Concern, developed per IEC 62304.
Indications for Use
Indicated for respiratory support, monitoring, and treatment of neonatal, pediatric, and adult patients (adults excluded for Servo-n). Used by healthcare providers in professional facilities and during transport. Servo-u MR version indicated for use in 1.5T or 3T MR environments outside magnetic fields >20 mT/200 Gauss.
Regulatory Classification
Identification
A continuous ventilator (respirator) is a device intended to mechanically control or assist patient breathing by delivering a predetermined percentage of oxygen in the breathing gas. Adult, pediatric, and neonatal ventilators are included in this generic type of device.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
Maquet Critical Care AB % Mark Smith Sr. Regulatory Affairs Specialist Getinge 45 Barbour Pond Drive Wayne, New Jersey 07470
Re: K201874
Trade/Device Name: Servo-u Ventilator System 4.1, Servo-n Ventilator System 4.1, Servo-u MR Ventilator System 4.1 Regulation Number: 21 CFR 868.5895 Regulation Name: Continuous ventilator Regulatory Class: Class II Product Code: CBK Dated: July 6, 2020 Received: July 7, 2020
Dear Mark Smith:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal
{1}------------------------------------------------
statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Brandon Blakely, PhD Acting Assistant Director DHT1C: Division of ENT, Sleep Disordered Breathing, Respiratory and Anesthesia Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) TBD
Device Name Servo-u Ventilator System 4.1 Servo-n Ventilator System 4.1 Servo-u MR Ventilator System 4.1
Indications for Use (Describe)
The Servo-u Ventilator System is:
- · intended for respiratory support, monitoring and treatment of neonatal, pediatric and adult patients
- to be used only by healthcare providers
- · to be used only in professional healthcare facilities and for transport within these facilities
The Servo-n Ventilator System is:
- · intended for respiratory support, monitoring and treatment of neonatal and pediatric patients
- · to be used only by healthcare providers
- · to be used only in professional healthcare facilities and for transport within these facilities
The Servo-u MR Ventilator System is:
- · intended for respiratory support, monitoring and treatment of neonatal, pediatric and adult patients
- · to be used only by healthcare providers
- · to be used only in professional healthcare facilities and for transport within these facilities
- · to be used in MR environment according to specified conditions
- with 1.5 T or 3 T MR scanners
- outside magnetic fields >20 mT/200 Gauss
| Type of Use (Select one or both, as applicable) | |
|---------------------------------------------------------------|--------------------------------------------------------------|
| <div> Prescription Use (Part 21 CFR 801 Subpart D) </div> | <div> Over-The-Counter Use (21 CFR 801 Subpart C) </div> |
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for Getinge, a company that provides equipment and systems for healthcare and life sciences. The logo consists of the word "GETINGE" in a bold, sans-serif font, followed by a stylized symbol. The symbol is composed of four curved shapes arranged in a circular pattern, creating a sense of movement and connection.
## 510(k) SUMMARY
as required by section 21 CFR 807.92
| Device owner | Maquet Critical Care AB<br>Röntgenvägen 2<br>SE-171 54 Solna, Sweden<br>Tel: (011) 46 10 335 7300 |
|--------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Contact Persons for this submission: | Mr. David Ardanius<br>Regulatory Affairs Manager<br>Phone: direct: (011) 46 10 335 7300<br>Email: david.ardanius@getinge.com |
| Application Correspondent: | Mr. Mark N. Smith<br>Manager, Regulatory Affairs<br>Getinge<br>45 Barbour Pond Drive<br>Wayne, NJ 07470<br>Email: mark.n.smith@getinge.com<br>Phone: 585-272-5274 |
Date prepared: August 31, 2020
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the logo for GETINGE. The word "GETINGE" is written in a bold, sans-serif font. To the right of the word is a stylized symbol that looks like four curved arrows pointing inward towards a central point. The logo is simple and modern.
| Trade Name : | Model: | Model no: |
|----------------------------------|------------|-----------|
| Servo-u 4.1 Ventilator System | Servo-u | 66 94 800 |
| Servo-n 4.1 Ventilator System | Servo-n | 66 88 600 |
| Servo-u MR 4.1 Ventilator System | Servo-u MR | 68 88 800 |
## Device Classification
Common Name Classification Number Class Regulation Number
Ventilator, continuous, facility use CBK II 21 CFR 868.5895
## Predicate Device Identification
| | Maquet Servo-u/n Ventilator System | K180098 |
|--|------------------------------------|---------|
|--|------------------------------------|---------|
## Reference Device Identification
| Maquet Servo-i Ventilator System | K123149 |
|----------------------------------|---------|
| Nihon Kohden NKV-550 | K181695 |
| Hamilton-G5 | K193228 |
## Indications for Use
The Servo-u/n/u MR Ventilator Systems are:
- . intended for respiratory support, monitoring and treatment of neonatal, pediatric and adult patients (adult not applicable for Servo-n)
- to be used only by healthcare providers .
- to be used only in professional healthcare facilities and for transport within these . facilities
The Servo-u MR Ventilator System to be used in MR environment according to specified conditions:
- . with 1.5 T or 3 T MR scanners
- outside magnetic fields >20 mT/200 Gauss .
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the logo for Getinge, a company that provides equipment and systems to healthcare and life science industries. The logo consists of the word "GETINGE" in a bold, sans-serif font, followed by a stylized symbol resembling four arrows pointing inward towards a central point. The color scheme is primarily a dark blue, giving the logo a professional and modern appearance.
## Device Description
The Servo-u/n/u MR Ventilator Systems 4.1 consist of a Patient Unit where gases are mixed and administered, and a User Interface where the settings are made and ventilation is monitored.
The Servo-u/n/u MR Ventilator Systems 4.1 are based on the cleared predicate device Servo-u/n Ventilator Systems 2.1 (K180098) with some improvements. The ventilation modes in the Servo-u/n/u MR 4.1 are the same as the predicate device. Standard configurations of available modes and optional modes do differ between the devices, i.e. Servo-u/n/u MR 4.1.
The ventilators deliver controlled or supported breaths to the patient, with constant flow, constant pressure, using a set oxygen concentration. The ventilators can also deliver High Flow therapy with a constant flow.
The Electrical activity of the diaphragm (Edi) is a measurement of the patients own breathing efforts. The Edi functionality makes it possible to monitor Edi activity in all ventilation modes, High Flow therapy as well as in Standby.
NAVA stands for Neurally Adjusted Ventilatory Assist and is a supported mode of ventilation based on the Edi, delivering assist in proportion to and synchronized with the patient's respiratory drive. NAVA is available as an invasive and a non-invasive mode. The included parts related to this mode, such as Edi module and Edi catheters are identical to the cleared predicate devices Servo-u/n 2.1 (K180098).
Servo-u/n contain a dedicated controller circuit for the Aerogen Solo nebulizer (included as standard). It is identical to the cleared predicate devices Servo-u/n 2.1 (K180098). Not available on Servo-u MR.
Accessories for CO2 monitoring and flow and pressure measurements at the Y piece (Y sensor) are integrated as options. It is identical to the cleared predicate devices Servo-u/n 2.1 (K180098).
The Servo-u/n/u MR Ventilator Systems will produce visual and audible alarms if any parameter varies beyond pre-set or default limits and log alarm recordings. The alarm handling is similar to the one used in the cleared predicate devices Servo-u/n 2.1 (K180098).
The Servo-u/n/u MR Ventilator Systems contain provisions for battery modules to supply the system in the case of mains power failure or during intra-hospital transport. The batteries are identical to the one used for the cleared predicate devices Servo-u/n 2.1 (K180098).
{6}------------------------------------------------
# GETI
## COMPARISSON TO PREDICATE
## Comparison of Intended Use/Indications Use
The Indications for Use for the proposed Servo-u/n/u MR Ventilator Systems 4.1 are identical to the predicate device Servo-u/n 2.1 Ventilator System (K180098 with the addition of MR environment to Servo-u MR.
MAQUET bases this 510(k) notification on that the Servo-u/n/u MR 4.1 fulfil the appropriate product standards and is substantially equivalent with the already cleared predicate device Servo-u/n Ventilator Systems 2.1 (K180098).
### Comparison of Technology Characteristics
The Servo-u/n/u MR Ventilator Systems 4.1 are modifications of the previously cleared Servo-u/n Ventilator Systems 2.1 (K180098).
The following changes have been made to Servo-u/n/u MR Ventilator Systems 4.1 compared to the cleared predicate device Servo-u/n Ventilator Systems 2.1 (K180098):
#### Servo-u MR, Mechanical adaptions and Magnetic Field Indicator -
Compared to the predicate device, a magnetic field indicator and mobile cart with autolock are added. Also the panel is modified.
The patient unit is permanently attached to the mobile cart and can therefore not be detached once assembled.
Cable Shielding ferrites have been added to the panel cable and panel connector to protect the components from the induced magnetic field. The panel cable is fixated in the panel to increase the robustness.
The panel arm on the user interface is not adjustable. However, it is possible to rotate and tilt the user interface.
A Magnetic Field Indicator (MFI) is introduced to the system. The intent of the MFI is to assist the user when placing the ventilator in the MR environment by continuously indicating the magnetic field.
An auto-lock handle is added. When the handle is released, the brakes automatically lock all four wheels. When the handle is depressed, the wheels are unlocked, enabling the user to move the ventilator.
{7}------------------------------------------------
## GETINGE
Additionally, the nebulizer port is plugged and it is not possible to attach drawers or place compressor mini on the mobile cart.
#### Software and user interface -
Software modifications have been made to add new functionality such as:
- -OLT trends (Open Lung Tool)
- OLT RM (Recruitment maneuver) -
- OLT SRM (Stepwise RM) -
- TP (Transpulmonary pressure) -
- SI (Stress index) -
- -Heliox
- -The possibility of setting PEEP to 0 cmH2O
- Minor improvements of the Total PEEP and Dynamic compliance calculations and the Circuit compliance compensation.
- -Minor corrections
The Servo-u/n/u MR software is considered a Major Level of Concern. Software verification was performed according to FDA's Guidance for Industry and FDA Staff, "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices" and according to the standard IEC 62304 "Medical device software - Software life-cycle processes".
#### -Components
The label of the Y sensor module pressure port is changed from PAW to PAUX. Compared to the predicate device an adapter for Heliox connection is added. CO2 module updated due to last-time-buy.
Y-piece, angled label change to disposable
Increased stability of Support arm by added clamp holder
#### -Alarms
The following adjustments and changes have been implemented for the alarms.
The algorithm linked to triggering of the alarm "Patient circuit disconnected" has been improved and split into two alarm types of the same priority; "Patient circuit disconnected" and "Check Tubing".
Minor improvements of the alarm handling for the Y sensor
## Non-clinical Testing and Performance
Maquet Critical Care has conducted risk analysis and performed necessary verification and validation activities to demonstrate that the design output of the modified devices meet the design input requirements:
{8}------------------------------------------------
# GETINGE
### Software
- Code review
- Static code analysis ●
- Unit tests ●
- Integration tests ●
## Performance
- Specification and system-level verification testing
- Waveform testing
## Biocompatibility
- . Volatile Organic Compounds
- Particulate testing
- Leachable testing
Human Factors Validation Testing
The following product standards are included in the verification:
- ANSI/AAMI ES 60601-1:2005 + A1:2012, Medical electrical equipment ● Part 1: General requirements for basic safety and essential performance, Recognition Number 19-4
- IEC 60601-1-2:2014, Medical electrical equipment Part 1-2: General . requirements for basic safety and essential performance - Collateral Standard: Electromagnetic disturbances - Requirements and Test, Recognition Number 19-8
- . IEC 60601-1-8:2006 + A1:2012, Medical electrical equipment - Part 1-8: General requirements for basic safety and essential performance - Collateral Standard: General requirements, tests and guidance for alarm systems in medical electrical equipment and medical electrical systems, Recognition Number 5-76
- ISO 80601-2-12:2011, Medical electrical equipment -- Part 2-1: Particular ● Requirements For The Safety Of Lung Ventilators - Critical Care Ventilators, Recognition Number 1-98
- ISO 80601-2-55:2018, Medical electrical equipment -- Part 2-55: Particular . requirements for the basic safety and essential performance of respiratory gas monitors, Recognition Number 1-140
- . IEC 62133-1:2017, Secondary Cells And Batteries Containing Alkaline Or Other Non-Acid Electrolytes - Safety Requirements For Portable Sealed Secondary Cells, And For Batteries Made From Them, For Use In Portable Applications -Part 1: Nickel Systems
{9}------------------------------------------------
## GETING
The connector to High-Pressure gas is in accordance with CGA V-5:2008, Recognition Number 1-81.
Biocompatibility evaluation of the Servo-u/n/u MR 4.1 is in accordance with ISO 10993-1:2018, recognition number 2-258 and ISO 18562-1:2017, recognition number 1-134.
Processing of Health Care Products evaluation of the Servo-u/n/u MR 4.1 is in accordance with ANSI AAMI ISO 17664:2017, recognition number 14-515 included the extent of recognition.
## Conclusion for Substantial Equivalence
The subject devices and the predicate device have the same intended use. There are no new type questions of safety and effectiveness for the subject devices that pose significant safety or effectiveness concerns as compared to the cleared predicate device. MAQUET has conducted risk analysis and performed the necessary verification and validation activities to demonstrate that the design outputs meet the design input requirements and the appropriate product standards. MAQUET concludes that the performance data for the subject devices shows that they are substantially equivalent to the cleared predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.