ESPRIT VENTILATOR WITH NICO-ESPRIT INTERFACE OPTION, MODEL V1000
K041412 · Respironics California, Inc. · CBK · Jun 18, 2004 · Anesthesiology
Device Facts
Record ID
K041412
Device Name
ESPRIT VENTILATOR WITH NICO-ESPRIT INTERFACE OPTION, MODEL V1000
Applicant
Respironics California, Inc.
Product Code
CBK · Anesthesiology
Decision Date
Jun 18, 2004
Decision
SESE
Submission Type
Special
Regulation
21 CFR 868.5895
Device Class
Class 2
Attributes
Therapeutic, Pediatric
Indications for Use
The Esprit ventilator is a microprocessor controlled, electrically powered, mechanical ventilator. It is intended for use by qualified medical personnel in providing continuous or intermittent ventilatory support to adult and pediatric patients as prescribed by a physician. The Esprit Ventilator is intended for use in either invasive or non-invasive applications. Esprit is not intended for use in the presence of flammable anæsthetics. Esprit is a prescription use device that is intended for sale by or on the order of a physician.
Device Story
The Esprit Ventilator with NICO-Esprit Interface is a software-upgraded mechanical ventilator. It facilitates bi-directional communication between the Esprit Ventilator and the NICO Monitor. The interface enables the transfer of NICO parameters to the ventilator and the transfer of breath type information and FiO2 from the ventilator to the NICO monitor. The device is operated by qualified medical personnel in clinical settings to provide ventilatory support. The integration allows for synchronized data exchange between the two systems, assisting clinicians in monitoring patient status and adjusting ventilatory support, potentially improving patient management.
Clinical Evidence
Bench testing only. Safety testing conducted per IEC 60601-1-2 and UL 2601. Software validation performed in accordance with FDA's 1998 Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices.
Technological Characteristics
Microprocessor-controlled, electrically powered mechanical ventilator. Software-based interface for bi-directional data communication. No new hardware materials or sensing principles introduced. Complies with IEC 60601-1-2 and UL 2601 standards.
Indications for Use
Indicated for adult and pediatric patients requiring continuous or intermittent ventilatory support in invasive or non-invasive applications. Contraindicated for use in the presence of flammable anesthetics.
Regulatory Classification
Identification
A continuous ventilator (respirator) is a device intended to mechanically control or assist patient breathing by delivering a predetermined percentage of oxygen in the breathing gas. Adult, pediatric, and neonatal ventilators are included in this generic type of device.
{0}------------------------------------------------
JUN 1 8 2004
Special 510(k): Device Modification Esprit Ventilator with NICO-Esprit Interface
### 510(k) SUMMARY 6
| Company<br>Information: | Respironics, California Inc.<br>2271 Cosmos Court<br>Carlsbad, CA. 92009 | |
|-------------------------|--------------------------------------------------------------------------|-------------------------------------|
| Contact: | Mary Funk, Regulatory Affairs Project Manager | |
| | Phone Number: (760) 918-7328<br>Fax Number: (760) 918-0169 | |
| Date Prepared: | May 11, 2004 | |
| Product Name: | Esprit Ventilator with the NICO-Esprit Interface | |
| Common Name: | Ventilator | |
| Classification: | Class II<br>Continuous Ventilator (per 21 CFR 868.5895) | |
| Predicate<br>Devices: | Respironics Esprit Ventilator | K981072 K001208<br>K023350, K034040 |
| | Infrasonics Adult Star Ventilator | K964543 |
| | NICO Monitor | K030886 |
### Device Description: 6.1
The NICO-Esprit Interface is a software upgrade to the Esprit Ventilator, which allows the bi-directional communication of the devices to facilitate the transfer of NICO parameters to the Esprit Ventilator as well as the transfer of a patient's breath type information and FiO2 from the Esprit to the NICO Monitor.
### 6.2 Intended Use:
The addition of the NICO-Esprit Interface has not changed the original intended use as described in the labeling. The Esprit ventilator is a microprocessor controlled, electrically powered, mechanical ventilator. It is intended for use by qualified medical personnel in providing continuous or intermittent ventilatory support to adult and pediatric patients as prescribed by a physician. The Esprit Ventilator is intended for use in either invasive or non-invasive applications.
Esprit is not intended for use in the presence of flammable anæsthetics. Esprit is a prescription use device that is intended for sale by or on the order of a physician.
### Technological Characteristics: 6.3
The Esprit ventilator does not incorporate any new technological characteristics with the addition of the NICO-Esprit Interface Option.
{1}------------------------------------------------
# Special 510(k): Device Modification Esprit Ventilator with NICO-Esprit Interface
### Determination of Substantial Equivalence: 6.4
Ventilator with NICO-Esprit Interface has similar performance The Esprit characteristics, the same intended use and the same environment of use and patient populations as the currently marketed predicate devices. The labeling and instructional information, including warning and caution statements, is similar to that informaticate devices. The addition of this new feature does not raise new questions of safety or effectiveness for the Esprit.
### Summary of Performance Testing: 6.5
Safety testing was conducted per the applicable sections of IEC 60601-1-2 and UL 2601. Software validation testing was performed in accordance with FDA's Guidance for the Content of Premarket Submissions for Software contained in Medical Devices (1998). The results of all verification and validation testing demonstrate that all design and system requirements for the Esprit ventilator with NICO-Esprit Interface have been met.
### Conclusion: 6.6
The technological characteristics of the Esprit ventilator with NICO-Esprit Interface and the results of the performance testing do not raise new questions of safety and effectiveness and demonstrate that the device is substantially equivalent to the legally marketed predicate devices.
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the seal of the Department of Health & Human Services (HHS) of the United States. The seal features a stylized caduceus, a symbol often associated with medicine and healthcare, with three overlapping lines forming the wings. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the caduceus.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
JUN 1 8 2004
Ms. Mary Funk Regulatory Affairs Project Manager Respironics California, Incorporated 2271 Cosmos Court Carlsbad, California 92009
Re: K041412
Ko41412
Trade/Device Name: Esprit Ventilator with NICO-Esprit Interface Option, Model V1000 Regulation Number: 868.5895 Regulation Name: Continuous Ventilator Regulatory Class: II Product Code: CBK Dated: May 26, 2004 Received: May 27, 2004
Dear Ms. Funk:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate for the sured in the May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). and Cosmette For (110) that the device, subject to the general controls provisions of the Act. The 1 ou may, aters, provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it n your de vice to such additional controls. Existing major regulations affecting your device can may be subject to back as a sateral Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that I'DA's issuance of a substantial equivalence determination does not mean r rease be act nou mail termination that your device complies with other requirements of the Act that + Drimas intatutes and regulations administered by other Federal agencies. You must
{3}------------------------------------------------
Page 2 – Ms. Mary Funk
comply with all the Act's requirements, including, but not limited to: registration and listing (21 comply with all the Act 3 requirements, niems manufacturing practice requirements as set
CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice the electronic CFR Part 807); labeling (21 CFR Part 017), good and if applicable, the electronic
forth in the quality systems (QS) regulation (21 CFR Part 820); and if app 1000, 1050 - This forth in the quality systems (QS) regulation (21-542 of the Act); 21 CFR 1000-1050. This product radiation confor provisions (Sections For Sections of Section 510(k) letter will anow you to begin marketing your substantial equivalence of your device to a legally premarket notification. The FDA miding of dassification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please if you desire specific advice for your ac not on one on one one the regulation entitled,
contact the Office of Compliance at (301) 594-4646. Also, please note Contact the Office of Comphance at (So Frenarket notification" (21CFR Part 807.97). You may obtain "Misbranding by relevelec to premanter notifications in the Act from the Division of Small other general Information on your responsion.com at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html.
Sincerely yours,
Chiu Lin, Ph.D.
Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
Special 510(k): Device Modification Esprit Ventilator with NICO-Esprit Interface
## STATEMENT OF INDICATIONS FOR USE 8
Koy1412-510(k) Number (if known):
Esprit Ventilator with NICO Esprit Interface Device Name:
The Esprit ventilator is a microprocessor controlled, electrically Indications for use: The Eoph vechanical ventilator. It is intended for use by powered, moonanisal personnel in providing continuous or qualificu - modioa. Support for adult and pediatric patients intermittent vontilatory sician. The Esprit Ventilator is intended for use in either invasive or non-invasive applications.
Prescription Use XX (Per 21 CFR 801 subpart D) AND/OR
Over-the-counter Use (21 CFR 807 Subpart C)
PLEASE DO NOT WRITE BELOW THIS LINE – CONTINUE ON ANOTHER PAGE IF NEEDED
Concurrence of CDRH, Office of Device Evaluation (ODE)
Hna U. Pinto
Division Sign-Off) Division of Anesthesiology, General Hospital, Infection Control, Dental Devices
510(k) Number _
Page 1 of