NELLCOR PURITAN BENNETT 700 SERIES VENTILATOR SYSTEM, AND EASYNEB NEBULIZER
Applicant
Puritan-Bennett Ireland, Ltd.
Product Code
CAF · Anesthesiology
Decision Date
Mar 15, 2000
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 868.5630
Device Class
Class 2
Attributes
Therapeutic, Pediatric
Indications for Use
The 700 Series Ventilator System is used to provide continuous ventilation to patients requiring respiratory support. This device is used for a wide range of patients from pediatric to adult and for a wide variety of clinical conditions. It also has the capability to provide nebulization through use of an optional accessory, the EasyNeb Nebulizer.
Device Story
700 Series Ventilator provides continuous respiratory support to pediatric and adult patients in hospital settings or during transport. Optional EasyNeb Nebulizer accessory delivers aerosolized medication. Nebulizer uses piezoelectric crystal to generate ultrasonic waves, converting liquid medication (up to 10 mL) into fine particles. Ventilator software controls nebulizer power, duration, and mode via a dedicated port on the communications interface. Operated by trained clinicians; intended for prescription use only. Device supports patient care by providing both ventilation and targeted medication delivery without compromising ventilator performance or patient data.
Clinical Evidence
Bench testing only. Verification and validation testing conducted per FDA software and environmental guidance. Compliance with international standards (EN60601-1, EN60601-1-2, CAN/CSA C22.2 No. 601-1M90, UL 2601-1) demonstrated device meets specifications.
Technological Characteristics
Critical care ventilator with optional ultrasonic nebulizer. Nebulizer utilizes piezoelectric crystal for aerosol generation. Connectivity via communications interface port. Software-controlled power and operation. Compliant with EN60601-1, EN60601-1-2, CAN/CSA C22.2 No. 601-1M90, and UL 2601-1 standards.
Indications for Use
Indicated for pediatric and adult patients (tidal volume 0.04 - 2 L) requiring continuous invasive or non-invasive respiratory support. Contraindicated for use in the presence of flammable anesthetics.
Regulatory Classification
Identification
A nebulizer is a device intended to spray liquids in aerosol form into gases that are delivered directly to the patient for breathing. Heated, ultrasonic, gas, venturi, and refillable nebulizers are included in this generic type of device.
{0}------------------------------------------------
# SUMMARY OF SAFETY AND EFFECTIVENESS
K990897
| SUBMITTER: | Nellcor Puritan Bennett Ireland,<br>a subsidiary of Mallinckrodt Inc., |
|-------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------|
| on behalf of: | Puritan-Bennett Corp.,<br>a subsidiary of Mallinckrodt Inc. |
| DATE: | 16 March 1999 |
| COMMON NAME: | Continuous Ventilator |
| PROPRIETARY NAME: | 700 Series Ventilator |
| CONTACT: | Robbie Walsh, Regulatory Affairs Manager<br>Nellcor Puritan Bennett Ireland,<br>Mervue,<br>Galway.<br>Ireland.<br>Tel: + 353 91 753771<br>Fax: + 353 91 770156 |
| CLASSIFICATION: | Class II per 21 CFR 868.5895<br>Continuous Ventilator |
### PREDICATE DEVICES:
Puritan-Bennett Corp. is claiming substantial equivalence to the following two predicate medical devices:
| Predicate Device | 510(k) Number | Classification |
|--------------------------------------|---------------|-----------------------------------------------------------|
| Siemens Servo<br>Ventilator 300 | K902859 | Class II,<br>Continuous Ventilator<br>per 21 CFR 868.5895 |
| Siemens Servo Ultra<br>Nebulizer 345 | K960854 | Class II,<br>Nebulizer<br>per 21 CFR 868.5630 |
{1}------------------------------------------------
#### A Device Description:
The device is a low-cost, critical care ventilator intended to provide continuous ventilation for pediatric and adult patients, with the provision of nebulization functionality, through use of an optional accessory, the EasyNeb Nebulizer.
The nebulization functionality is implemented by:
a) modifying the 700 Series software to activate a nebulizer interface which directly controls the power and operation of the EasyNeb Nebulizer and b) the provision of an external nebulizer, the EasyNeb Nebulizer, as an optional accessory. which is intended to be used solely with the 740 and 760 ventilators.
The modification to the ventilator is the activation through software of the nebulizer port on an existing Communications interface option. This nebulizer port facilitates electrical connection of the EasyNeb Nebulizer to the ventilator, the operation of which is controlled completely by ventilator software. The ventilator modification in essence then is a software modification which allows control of the nebulizer power on/off, the nebulization period and mode of operation.
The EasyNeb Nebulizer is an ultrasonic nebulizer which is designed to deliver aerosol medications to the patient without affecting ventilator performance or patient data. It is powered and controlled by a 700 Series Ventilator, with the Communications Option installed. It contains a piezoelectric crystal that generates ultrasonic waves. The crystal's vibrations are transmitted through buffer water to a medication cup and convert the liquid medication into fine particles. The nebulizer medication cup is designed for single-patient use only, and holds up to 10 mL of medication.
#### Intended Use: B
The intended use is provided below;
Purpose and function of device:
- · The 700 Series Ventilator is intended to provide continuous ventilation to patients requiring respiratory support.
- · It offers nebulization functionality through use of an optional accessory, the EasyNeb Nebulizer
- This product is intended for a wide range of patients from pediatric to adult and for a wide variety of clinical conditions.
Intended patient population:
- · The intended patient population includes pediatric and adult patients (tidal volume 0.04 - 2 L) who require continuous respiratory support.
- Intended for patients who require either invasive or non-invasive . ventilation.
{2}------------------------------------------------
Intended environment of use:
- · The 700 Series Ventilator is intended for use in hospitals and hospital type facilities which provide respiratory care for patients requiring respiratory support.
- · The 700 Series Ventilator may be used during hospital and hospital type facility transport
- · The 700 Series Ventilator is not to be used in the presence of flammable anesthetics
- · The 700 Series Ventilator is intended for sale by or on the order of a physician only.
- · This product is intended for operation by trained and qualified clinicians only.
#### C Substantial Equivalence
The intended use of the 700 Series Ventilator, incorporating the EasyNeb Nebulizer as an optional accessory, is the same as that for a standard, currently marketed critical care ventilator. The materials and design of these device are similar to those of the predicate devices. The technical characteristics of the 700 Series Ventilator do not introduce new questions regarding safety or effectiveness of critical care ventilators. Furthermore, the labeling associated with the 700 Series Ventilator and EasyNeb Nebulizer provides similar information as the predicate devices.
Information provided in the 510(k) submission supports the determination of substantial equivalence. Software design and development, (including verification and validation testing. test and software quality procedures) was conducted using FDA's Guidance for the Content of Premarket Submissions for Software contained in medical devices, dated May 29 1998, as a guidance and per internal company requirements. Environmental testing was conducted using FDA's Reviewers Guidance for Premarket Notification Submissions, Nov. 1993 draft as a guideline and per internal company requirements. The 700 Series Ventilator and EasyNeb Nebulizer device design and testing are also compliant with various voluntary, international standards including: EN60601-1:1990, EN 60601-1-2:1993, CAN/CSA C22.2 No. 601-1M90:1994, UL 2601-1:1994 and 93/42/EEC Medical Device Directive.
The combined testing and analysis of results provides assurance that the device meets its specifications and is safe and effective for its intended use.
In summary Puritan-Bennett Corp. has demonstrated the 700 Series Ventilator, incorporating the EasyNeb Nebulizer as an optional accessory, to be safe and effective. This device is considered to be substantially equivalent to currently marketed devices which have been previously cleared by FDA.
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is circular and contains the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an emblem that resembles a stylized bird or eagle with three wing-like shapes.
Public Health Service
MAR 1 5 2000
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Mr. Robbie Walsh Nellcor Puritan Bennett Ireland Mervue Galway Ireland
Re: K990897 700 Series Ventilator incorporating EasyNeb Nebulizer Regulatory Class: II (two) Product Code: CAF, CBK December 10, 1999 Dated: December 16, 1999 Received:
Dear Mr. Walsh:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in requlatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.
{4}------------------------------------------------
Page 2 - Mr. Robbie Walsh
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling requlation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4648. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification"(21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597, or at its internet address "http://www.fda.gov/cdrh/dsma/dsmamain.html".
Sincerely yours,
James E. Dillard III Director Division of Cardiovascular, Respiratory, and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
## INDICATION FOR USE
| 510(k) Number: | K990897 |
|---------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Device Name: | 700 Series Ventilator System (incorporating the EasyNeb<br>Nebulizer as an optional accessory) |
| Indication for Use: | The 700 Series Ventilator System is used to provide<br>continuous ventilation to patients requiring respiratory support.<br>This device is used for a wide range of patients from pediatric<br>to adult and for a wide variety of clinical conditions. It also has<br>the capability to provide nebulization through use of an optional<br>accessory, the EasyNeb Nebulizer. |
Prescription Use: Yes (Per 21 CFR 801.109)
L.A.Ketron
(Division Sign-off)
Division of Cardiovascular, Respiratory,
and Neurological Devices
510(k) Number
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.