K223840 · Pari Respiratory Equipment, Inc. · CAF · Aug 11, 2023 · Anesthesiology
Device Facts
Record ID
K223840
Device Name
eRapid Nebulizer System
Applicant
Pari Respiratory Equipment, Inc.
Product Code
CAF · Anesthesiology
Decision Date
Aug 11, 2023
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 868.5630
Device Class
Class 2
Attributes
Therapeutic, Pediatric
Indications for Use
The eRapid NCP Nebulizer System is to be used with patients for whom doctors have prescribed medicine for nebulization. It is intended for adult and pediatric patients age 4 and older who can coordinate breathing through a mouthpiece, and may be used in hospitals, hospital-type facilities, nursing homes, sub-acute institutions and home environments.
Device Story
The eRapid NCP Nebulizer System is a portable, handheld electronic nebulizer for aerosolizing liquid medications. It consists of a controller and a nebulizer handset connected by a cord. The device uses micro-perforated vibrating membrane technology: a piezoelectric actuator, driven by an electronic circuit in the controller, vibrates a stainless steel membrane with laser-drilled holes against a fluid reservoir. This vibration creates liquid jets that disperse into a stream of droplets. The device is intended for use in hospitals, nursing homes, and home environments by patients or caregivers. The user initiates treatment via an On/Off button; the device operates continuously until the reservoir is empty, an error is detected, or a maximum time limit is reached. The controller provides audio-visual feedback regarding status and failures. By delivering medication as an aerosol, the device enables inhalation therapy for respiratory diseases, benefiting patients by facilitating targeted drug delivery to the lungs.
Clinical Evidence
No clinical data provided. Substantial equivalence is supported by bench testing, including aerosol performance (Total Output Rate, Mass Median Diameter, Fine Particle Fraction), biocompatibility, EMC/electrical safety (ANSI/AAMI ES60601-1, IEC 60601-1-2), software verification/validation, simulated use testing, and validation of cleaning/disinfection methods.
Technological Characteristics
Materials: Polypropylene, silicone rubber, stainless steel, ceramics, ABS, PC, TPE. Principle: Piezo-driven micro-perforated vibrating membrane (117 kHz). Power: 5V DC (rechargeable NiMH battery pack or AC adapter). Form factor: Handheld controller with tethered handset. Connectivity: Standalone. Software: Moderate level of concern, Class B per IEC 62304. Sterilization: Manual cleaning/disinfection (chemical or thermal).
Indications for Use
Indicated for adult and pediatric patients age 4 and older who can coordinate breathing through a mouthpiece and have been prescribed medication for nebulization. No specific contraindications listed.
Regulatory Classification
Identification
A nebulizer is a device intended to spray liquids in aerosol form into gases that are delivered directly to the patient for breathing. Heated, ultrasonic, gas, venturi, and refillable nebulizers are included in this generic type of device.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: a symbol on the left and the FDA name on the right. The symbol on the left is a stylized image of a human figure, while the FDA name on the right is written in blue letters. The words "U.S. FOOD & DRUG ADMINISTRATION" are written in a clear, sans-serif font.
August 11, 2023
PARI Respiratory Equipment, Inc. Michael Judge VP, Operations and Regulatory Affairs 2412 Pari Way Midlothian, Virginia 23112
Re: K223840
Trade/Device Name: eRapid Nebulizer System Regulation Number: 21 CFR 868.5630 Regulation Name: Nebulizer Regulatory Class: Class II Product Code: CAF Dated: July 5, 2023 Received: July 10, 2023
Dear Michael Judge:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal
{1}------------------------------------------------
statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
# Ethan L. Nyberg -S
Ethan Nyberg, Ph.D. Assistant Director DHT1C: Division of Sleep Disordered Breathing, Respiratory and Anesthesia Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K223840
Device Name eRapid NCP Nebulizer System
### Indications for Use (Describe)
The eRapid NCP Nebulizer System is to be used with patients for whom doctors have prescribed medicine for nebulization. It is intended for adult and pediatric patients age 4 and older who can coordinate breathing through a mouthpiece, and may be used in hospitals, hospital-type facilities, nursing homes, sub-acute institutions and home environments.
| Type of Use (Select one or both, as applicable) | |
|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------|
| <div> <span> <svg fill="none" height="16" viewbox="0 0 16 16" width="16" xmlns="http://www.w3.org/2000/svg"> <path d="M13.25 4.75L6.75 11.25L2.75 7.25" stroke="#000" stroke-linecap="round" stroke-linejoin="round" stroke-width="1.5"></path> </svg> Prescription Use (Part 21 CFR 801 Subpart D) </span> </div> | <div> <span> <input type="checkbox"/> Over-The-Counter Use (21 CFR 801 Subpart C) </span> </div> |
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 1. Submitter Information
| Name: | PARI Respiratory Equipment, Inc. |
|----------------------|---------------------------------------|
| Address: | 2412 Pari Way<br>Midlothian, VA 23112 |
| Phone: | (804) 253-7274, ext. 269 |
| Fax: | (804) 253-0260 |
| Contact Name: | Michael Judge |
| Email: | mike.judge@pari.com |
| Device Name | |
| Name of Device: | eRapid NCP Nebulizer System |
| Common Name: | Nebulizer |
| Classification Name: | Nebulizer (21 CFR 868.5630) |
| Regulatory Class: | II |
| Product Code: | CAF |
# 3. Legally Marketed Predicate Device
eRapid Nebulizer System, K112859
The following predicate device has not been subject to a design-related recall.
# 4. Device Description
2.
Both the eRapid NCP and the predicate eRapid device are identical in purpose, function, core technology and method of operation. They are single-patient use, reusable electronic nebulizers, using micro-perforated vibrating membrane technology to aerosolize liquid medications. They are for adult and pediatric inhalation therapy in a home care, nursing home, sub-acute institution, or hospital environment. The devices are hand-held and portable, consisting of a controller and a nebulizer handset, connected with a connection cord. Power input for both devices is provided by either AA batteries, or a DC or AC adapter. Alternate power cords, plugs and adapters allow their use in any country.
# 5. Mechanism of Action
The medication is filled into the medication reservoir of the nebulizer handset for presenting its liquid content to the vibrating membrane of the nebulizer to generate aerosol for inhalation. There is only the On/Off button on the controller to turn on the device, and start the treatment. When the treatment is done the device automatically shuts off.
This eFlow technology uses a wafer-thin plate or membrane of stainless steel, which is perforated with numerous laser-drilled holes (Figure 3). The size of the membrane holes determines the aerosol particle size. There is no other mechanism in place that changes the aerosol particle size, whether through software and/or user interface. This micro-perforated membrane vibrates at a high frequency against a body of fluid. The vibration source is the piezoelectric actuator that is activated by an electronic drive circuit of the Controller. The actuator and the perforated membrane are the main components of the aerosol head that is in contact
{4}------------------------------------------------
with the liquid medication to be aerosolized. Liquid jets are created as an inertial response to the vibration of the membrane. Surface tension and hydrodynamic effects then cause these jets to disperse to produce a stream of precisely controlled droplets.
Image /page/4/Figure/1 description: This image shows a diagram of a drug delivery system. The system consists of a reservoir containing the drug, electronics, a membrane with laser-drilled holes, a piezo actuator, and batteries. The drug is released as an aerosol through the laser-drilled holes.
# Figure 3: Principle of function of the vibrating membrane technology for aerosol generation
# 6. Indication for Use
The eRapid NCP Nebulizer System is to be used with patients for whom doctors have prescribed medicine for nebulization. It is intended for adult and pediatric patients age 4 and older who can coordinate breathing through a mouthpiece, and may be used in hospitals, hospital-type facilities, nursing homes, sub-acute institutions and home environments.
| PRODUCT | FDA-Cleared eRapid<br>Nebulizer System | eRapid NCP Nebulizer System |
|---------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Device Classification | Nebulizer 21 CFR 868.5630 | Nebulizer 21 CFR 868.5630 |
| 510k No. | K112859 | K223840 |
| Manufacturer (Reg.<br>No.) | PARI Respiratory Equipment, Inc.<br>(2954963) | PARI Respiratory Equipment, Inc.<br>(2954963) |
| Regulatory Class | II | II |
| Product Code | CAF | CAF |
| Intended Use / | The eRapid Nebulizer is a handheld | The eRapid NCP Nebulizer System |
| Indications for Use | nebulizer designed to aerosolize any<br>medications prescribed by a physi-<br>cian. The eRapid Nebulizer System<br>is intended for adult and pediatric<br>patients, and may be used in hospi-<br>tals, hospital-type facilities, nursing<br>homes, sub-acute institutions, and<br>home environments. | is to be used with patients for whom<br>doctors have prescribed medicine<br>for nebulization. It is intended for<br>adult and pediatric patients age 4<br>and older who can coordinate<br>breathing through a mouthpiece,<br>and may be used in hospitals, hospi-<br>tal-type facilities, nursing homes,<br>sub-acute institutions and home en-<br>vironments. |
| Prescription Use | Rx Only | Rx Only |
| | | |
| METHOD OF OPERATION | | |
| Technology Used | Micro-perforated vibrating mem-<br>brane technology to aerosolize liq-<br>uid medications | Micro-perforated vibrating mem-<br>brane technology to aerosolize liq-<br>uid medications |
| Aerosol Direction/<br>Output | Sprayed into aerosol chamber,<br>cleared by inspiratory flow | Sprayed into aerosol chamber,<br>cleared by inspiratory flow |
| Breath Enhanced/<br>Controlled/ Triggered | Breath enhanced two-way valve<br>system. | Breath enhanced two-way valve<br>system. |
| Fluid Delivery –<br>Generator | Direct contact between aerosol head<br>and fluid in sealed chamber (medi-<br>cation reservoir). | Direct contact between aerosol head<br>and fluid in sealed chamber (medi-<br>cation reservoir). |
| Configuration | Remote (tethered) only. | Remote (tethered) only. |
| Automatic Shut-off | Yes. Automatic shut-off when:<br>(1) medication reservoir is empty;<br>(2) out-of-range parameters are de-<br>tected; or (3) programmed maxi-<br>mum operating time) is reached. | Yes. Automatic shut-off when:<br>(1) medication reservoir is empty;<br>(2) out-of-range parameters are de-<br>tected; or (3) programmed maxi-<br>mum operating time) is reached. |
| DESIGN CAPACITIES | | |
| Medication Reservoir | | |
| Min. Fill | 2.0 mL | 2.0 mL |
| Max. Fill | 6.0 mL | 6.0 mL |
| Residue | Approx. 1.0 ml residue (depending<br>on filled volume) | Approx. 1.0 ml residue (depending<br>on filled volume) |
| AEROSOL PERFORMANCE | | |
| Total Output Rate<br>(mg/min) | 619 (AVG) SD=72 | 663 (AVG) SD=60 |
| Mass Median Diameter<br>( $\mu$ m) | 4.35 (AVG) SD=0.09 | 4.37 (AVG SD=0.10 |
| Geometrical<br>Standard Deviation | 1.53 | 1.53 |
| Fine Particle Fraction<br><5 (%) | 63.56 | 63.15 |
| ELECTRICAL COMPARISON (eBase Controller) | | |
| | eBase (Round) 1st Generation | eBase (NCP) 2nd Generation |
| Mean Frequency (Hz) | 117040 | 117650 |
| Mean Power (Watts) | 1.255 | 1.253 |
| | | |
| SOFTWARE | | |
| Level of Concern | Moderate | Moderate |
| Device-Specific | Yes | Yes |
| Dependent on External<br>Devices | No | No |
| Function | By continuous loop: (1) conducts<br>power pre-processing; (2) supervis-<br>es the nebulizer's aerosol genera-<br>tion; and (3) communicates to the<br>user on the operational status of the<br>device and out-of-range parameters.<br>These functions are performed con-<br>tinuously until the medication res-<br>ervoir is empty, out-of-range pa-<br>rameters are detected, or pro-<br>grammed maximum operating time<br>(Tmax) is reached. | By continuous loop: (1) conducts<br>power pre-processing; (2) supervis-<br>es the nebulizer's aerosol genera-<br>tion; and (3) communicates to the<br>user on the operational status of the<br>device and out-of-range parameters.<br>These functions are performed con-<br>tinuously until the medication res-<br>ervoir is empty, out-of-range pa-<br>rameters are detected, or pro-<br>grammed maximum operating time<br>(Tmax) is reached. |
| Audio-Visual Signals | LED, tone sounds, and illuminated<br>display showing graphical symbols<br>concerning battery level, operation- | Tone sounds, and illuminated dis-<br>play showing graphical symbols<br>concerning battery level, operation- |
| | al status and failure mode. | al status and failure mode. |
| ELECTRICAL CONECTION | | |
| CISPR 11 Group | II | II |
| CISPR 11 Class | B | B |
| Protection Class | B-with power supply. Internally<br>powered - with batteries | B-with power supply. Internally<br>powered - with batteries |
| Liquid Protection Rating<br>(controller) | IPX0 (No Protection) | IP22 |
| Rated Supply Voltage<br>(controller) | 12V | 5V |
| Rated Input Current<br>(controller) | 300 mA | 750 mA |
| Rated Input Power<br>(Controller) | 3.6 W | 3.75 W |
| Electrical Requirement<br>(power supply) | 100V - 240V, 50 Hz/60 Hz | 100V - 240V, 50 Hz/60 Hz |
| Battery Operation | Four AA disposable or rechargeable<br>batteries | Three NiMH AA rechargeable bat-<br>tery pack |
| Frequency power for<br>vibrating mesh mem-<br>brane | 1.8w, 117 kHz | 1.8w, 117 kHz |
| NEBULIZER MATERIALS | | |
| Mouthpiece | Polypropylene | Polypropylene |
| Valve(s) | Silicone rubber | Silicone rubber |
| Aerosol chamber | Polypropylene | Polypropylene |
| Fluid feed | Polypropylene, Thermoplastic elas-<br>tomer | Polypropylene, Thermoplastic elas-<br>tomer |
| Aerosol Head | Stainless steel, ceramics, adhesive | Stainless steel, ceramics, adhesive |
| Medication Reservoir | Polypropylene, Thermoplastic elastomer | Polypropylene, Thermoplastic elastomer |
| Medication Cover | Polypropylene, POM, Thermo-plastic elastomer | Polypropylene, POM, Thermo-plastic elastomer |
| <b>MECHANICAL</b> | | |
| Weight<br>Controller, Nebulizer<br>Handset, w/Batteries | Approx. 9.5 oz. (269g) | Approx. 9.3 oz. (264g) |
| Dimensions<br>Controller | H 43.5 mm Ø 117 mm | L 157mm x W 66mm x H 22mm |
| Handset | L 145 x W 50 x H 63 mm | L 145 x W 50 x H 63 mm |
| Temperature Range | 59° to 95°F (15° – 35°C) | 59° to 95°F (15° – 35°C) |
| Humidity | 30% to 95% | 30% to 95% |
| Ambient Temperature | - 4° to 140°F (-25° – 70°C) | - 4° to 140°F (-25° – 70°C) |
| Ambient Humidity | 10% to 95% | 10% to 95% |
| <b>CLEANING / DISINFECTION</b> | | |
| Cleaning | Cleaning with detergent and water. | Cleaning with detergent and water. |
| Disinfection | Disinfection with chemical (Control III) or thermal means (boiling) | Disinfection with chemical (Control III) or thermal means (boiling) |
| Cleaning Aid | PP/TPE accessory to mechanically<br>rinse the aerosol head membrane<br>pores by means of backwashing<br>(back flushing). | PP/TPE accessory to mechanically<br>rinse the aerosol head membrane<br>pores by means of backwashing<br>(back flushing). |
7. Comparison of eRapid NCP Nebulizer system to the Predicate Device
{5}------------------------------------------------
{6}------------------------------------------------
{7}------------------------------------------------
### eRapid NCP Nebulizer System and Predicate eRapid Nebulizer system a.
Similarities are that both the eRapid NCP and the predicate eRapid device are identical in purpose, function, core technology and energy sources. That is, they are portable, reusable, single-patient use, handheld electronic nebulizers that use a piezo-driven, microperforated vibrating membrane technology to aerosolize liquid medications for the treatment or prophylaxis of respiratory diseases. They are to be used with adult and pediatric patients for whom doctors have prescribed medication, i.e. they are for prescription use only. They may be used in hospitals, hospital-type facilities, nursing homes, sub-acute institutions and home environments. Energy sources for the devices are provided by: (1) connection to a mains power source, via an AC Power Supply; or, batteries.
Both the eRapid NCP and the predicate eRapid employ breath enhanced, two-way valve systems. Further, both devices have permanent modes of action, i.e., once the on/off switch on the controller is pressed, continuous nebulization occurs until the medication is nebulized. Neither has any breath trigger or interrupter. Fluid delivery / generation for both devices is by direct contact between the aerosol head and the fluid in sealed medication reservoir.
The difference between the two devices are a 2nd generation eBase controller. The redesigned eBase Controller features: (1) Different housing design with a more modern user interface, (2) Integrated rechargeable battery pack, (3) Higher resolution LCD display, (4)
{8}------------------------------------------------
Adjustable settings for audio and visual features. In addition, The eBase NCP Controller will require a new Connection Cord and Power Supply.
Only the part of the Connection Cord that plugs into the eBase NCP Controller was modified. The other end of the Connection Cord, which is already part of the FDA-cleared device, has not been changed.
Differences from FDA-cleared device are thus confined to the new control unit, connection cord, and power supply.
# 8. Performance Data
The following performance data were provided in support of a substantial equivalence determination.
### Biocompatibility and Airpath Testing a.
The eRapid NCP Nebulizer System u…
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.