K200063 · Trudell Medical International · CAF · Jul 17, 2020 · Anesthesiology
Device Facts
Record ID
K200063
Device Name
AeroEclipse* ONE BAN
Applicant
Trudell Medical International
Product Code
CAF · Anesthesiology
Decision Date
Jul 17, 2020
Decision
SESE
Submission Type
Abbreviated
Regulation
21 CFR 868.5630
Device Class
Class 2
Attributes
Therapeutic, Pediatric
Indications for Use
The AeroEclipse* ONE BAN is intended to be used with adults and pediatric patients over 5 years old, who are under the care or treatment of a licensed health care provider or physician. The device is intended to be used by these patients to administer aerosolized medication prescribed by a physician or health care professional. The intended environments for use include the home, hospitals and clinics.
Device Story
AeroEclipse* ONE BAN is a pneumatic jet nebulizer for aerosolized medication delivery. Device connects to external compressed air/oxygen source. Principle of operation: high-velocity gas jet creates negative pressure, drawing liquid medication to nozzle; baffle shears liquid into aerosol. Breath-actuation mechanism: patient inhalation moves actuator/diaphragm, initiating aerosol production; cessation of inhalation stops aerosol, reducing fugitive emissions. Used in home, hospital, or clinic by patients under physician supervision. Output is inhaled aerosol via mouthpiece. Benefits include targeted medication delivery and reduced environmental drug exposure.
Clinical Evidence
No clinical data. Substantial equivalence based on bench testing, including aerosol characterization per FDA 1993 guidance, design verification (flow, drop, environmental, life cycle), and ISO 18562 gas pathway testing (VOCs, particulate matter, inorganic gases).
Technological Characteristics
Pneumatic jet nebulizer; plastic construction identical to predicate. Operates via compressed air/oxygen (2.75–8 lpm). Features breath-actuated mechanism with diaphragm/actuator. Max fill volume 6ml. Single-patient use. Non-sterile. No software or electronic components.
Indications for Use
Indicated for adults and pediatric patients >5 years old under physician care for administration of prescribed aerosolized medication in home, hospital, or clinic settings.
Regulatory Classification
Identification
A nebulizer is a device intended to spray liquids in aerosol form into gases that are delivered directly to the patient for breathing. Heated, ultrasonic, gas, venturi, and refillable nebulizers are included in this generic type of device.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: a symbol on the left and the FDA name on the right. The symbol on the left is a stylized image of a human figure, while the FDA name on the right is written in blue letters. The words "U.S. FOOD & DRUG ADMINISTRATION" are written in a clear, sans-serif font.
July 17, 2020
Trudell Medical International Marianne Tanton Director, Ouality and Regulatory Affairs 725 Baransway Drive London, N5V 5G4 Ca
# Re: K200063
Trade/Device Name: AeroEclipse* ONE BAN Regulation Number: 21 CFR 868.5630 Regulation Name: Nebulizer Regulatory Class: Class II Product Code: CAF Dated: June 16, 2020 Received: June 18, 2020
# Dear Marianne Tanton:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part
{1}------------------------------------------------
801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
James J. Lee Ph.D. Assistant Director DHT1C: Division of ENT, Sleep Disordered Breathing, Respiratory and Anesthesia Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K200063
Device Name AeroEclipse* ONE BAN
Indications for Use (Describe)
The AeroEclipse* ONE BAN is intended to be used with adults and pediatric patients over 5 years old, who are under the care or treatment of a licensed health care provider or physician. The device is intended to be used by these patients to administer aerosolized medication prescribed by a physician or health care professional. The intended environments for use include the home, hospitals and clinics.
Type of Use (Select one or both, as applicable)
| <span style="text-decoration: underline;">Prescription Use (Part 21 CFR 801 Subpart D)</span> | <div> <input checked="true" type="checkbox"/> </div> |
|-----------------------------------------------------------------------------------------------|------------------------------------------------------|
| Over-The-Counter Use (21 CFR 801 Subpart C) | <div> <input type="checkbox"/> </div> |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Prepared: 17 July 2020
#### 1. Submitter
Trudell Medical International 725 Baransway Drive London, Ontario N5V 5G4, Canada
Contact: Marianne Tanton Director, Quality and Requlatory Affairs Phone: 1-519-455-7060 Email: mtanton@trudellmed.com
# 2. Device Name
| Trade Name: | AeroEclipse* ONE BAN |
|----------------------|---------------------------|
| Common Name: | Nebulizer |
| Classification Name: | Nebulizer 21 CFR 868.5630 |
| Regulatory Class: | II |
| Product Code: | CAF |
#### 3. Predicate Device
AeroEclipse* II BAN - K053605 Manufacturer: Trudell Medical International (TMI)
The predicate device has not been subject to a recall.
# 4. Device Description
The subject device. AeroEclipse* ONE BAN (Breath Actuated Nebulizer) is a jet nebulizer intended to administer medication (prescribed by a physician or health care provider) in the form of an aerosol for patient inhalation. It operates in conjunction with a compressed air source and a liquid medication. The subject device is designed to produce and deliver aerosolized medication only during the patient inspiratory cycle, thus significantly reducing fugitive emissions to the environment. It can be used in hospitals, clinics and in the home.
Liquid medication is placed in the medication cup of the nebulizer bottom, into which the nebulizer top sub-assembly is attached. The air supply tubing provides the interface for the pressurized air. Aerosol is inhaled by the patient via the oral interface of the mouthpiece. As the patient inhales, ambient air travels down through the nebulizer, creating a pressure drop which causes the actuator, connected to the diaphragm to move down into the nebulizing position. As inhalation continues, aerosol is created and inhaled through the mouthpiece. When inhalation ceases, the actuator returns to the up position thereby eliminating the pressure drop which leads to the device discontinuing the creation of an aerosol. As the patient inhales through the mouthpiece, the actuator drops, and the creation of aerosol begins aqain.
The overall device is comprised of a nebulizer bottom with nozzle feature, a nebulizer top sub-assembly, a mouthpiece and air supply tubing.
{4}------------------------------------------------
# 5. Principle of Operation
The subject device. AeroEclipse* ONE BAN, is considered a jet nebulizer, with the driving gas and liquid traveling through separate orifice is located in the center of the nebulizer. The driving gas (air or oxygen), passes through the small orifice and is deflected by a baffle on the actuator. The high velocity jet of gas produces a negative pressure behind it, which draws liquid up to the nozzle area. The deflected gas flow shears off the liquid being drawn up which generates the aerosol.
# 6. Indications for Use
The AeroEclipse* ONE BAN is intended to be used with adults and pediatric patients over 5 years old, who are under the care or treatment of a licensed health care provider or physician. The device is intended to be used by these patients to administer aerosolized medication prescribed by a physician or health care professional. The intended environments for use include the home, hospitals and clinics.
# 7. Comparison to Predicate Device
The subject device, AeroEclipse* ONE BAN, is equivalent in terms of indications for use and technological characteristics to that of the predicate device, AeroEclipse* II BAN (K053605).
The difference between the subject and predicate devices involves the aerosolization mode which does not affect the safety or effectiveness of the subject device. The subject device operates in breath actuated mode only while the predicate device operates in both breath actuated and continuous mode. Table 1 provides a comparison of the subject and predicate devices.
| Element of Comparison | AeroEclipse* ONE BAN<br>(Subject Device) | AeroEclipse* II BAN<br>(Predicate Device - K053605) |
|-----------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Indications for Use | The AeroEclipse* ONE Breath<br>Actuated Nebulizer is intended to<br>be used with adults and pediatric<br>patients over 5 years old, who are<br>under the care or treatment of a<br>licensed healthcare provider or<br>physician. The device is intended<br>to be used by these patients to<br>administer aerosolized medication<br>prescribed by a physician or health<br>care professional. The intended<br>environments for use include the<br>home, hospitals and clinics. | The AeroEclipse* II Breath Actuated<br>Nebulizer is intended to be used by<br>patients who are under the care or<br>treatment of a licensed healthcare<br>provider or physician. The device is<br>intended to be used by these patients<br>to administer aerosolized medication<br>prescribed by a physician or health<br>care professional. The intended<br>environments for use include the<br>home, hospitals and clinics. |
# Table 1: Comparison to Predicate Device
{5}------------------------------------------------
| Principle of Operation | Pneumatic Jet Nebulizer |
|------------------------|--------------------------------------------------------------------------------------------|
| Environment of use | Hospital, Clinic or Home |
| Single Patient Use | Yes |
| Aerosolization Mode | Breath Actuated Mode only<br>Built-in mode selector for breath actuated or continuous mode |
| Type of Device | Single patient use, multiple use, prescription only, non-sterile |
| Manufacturing process | Plastic molding |
| Type of gas source | Compressed air or oxygen |
| Flow Rate | 2.75 – 8 lpm (liters per minute) |
| Maximum Fill Volume | 6ml |
# 8. Performance Data
#### 8.1. Aerosol Characterization
Aerosol characterization testing was performed in accordance with relevant sections of the CDRH Guidance Document "Reviewer Guidance for Nebulizers, Metered Dose Inhalers, Spacers and Actuators" (FDA/CDRH - 1993). The results of the aerosol characterization testing demonstrated substantially equivalent in-vitro performance subject device, AeroEclipse* ONE BAN, and the predicate device, AeroEclipse* II BAN.
# 8.2 . Design Verification Testing
The following design verification tests were performed on the subject device, AeroEclipse* ONE BAN:
- Flow Performance ●
- . Drop Testing
- Environmental Testing ●
- Life Cycle Testing ●
Results demonstrate the subject device, AeroEclipse* ONE BAN meets the requirements outlined in the design verification plan. No new issues of safety and efficacy were identified as a result of the testing performed.
{6}------------------------------------------------
#### 9. Biocompatibility and Dry Gas Pathway
#### 9.1. Biocompatibility Testing
No additional biocompatibility testing was performed to demonstrate compliance with ISO 10993-1 since the materials of construction of the subject device, AeroEclipse* ONE BAN, in its final finished form are identical to the materials of construction of the cleared commercial device, AeroEclipse* II BAN K053605, in formulation, processing and geometry and no other chemicals have been added (e.g., plasticizers, fillers, additives, cleaning agents, mold release agents).
# 9.2. Dry Gas Pathway Testing
To support the safe use of the subject device, AeroEclipse* ONE BAN, testing pertaining to the dry gas pathway and associated risk assessments/conclusions per ISO 18562 were conducted. Testing included the following assessments:
- · Emissions of volatile organic compounds (VOCs), and
- · Fine particles (particulate matter PM2.5)
- · Inorganic gases (ozone, CO2, and CO)
Testing results and risk assessment demonstrated that exposure during use is unlikely to result in toxicological effects.
#### 10. Clinical Performance Summary
Not applicable. The determination of substantial equivalence is not based on Clinical Performance data.
#### 11. Conclusion
The AeroEclipse* ONE BAN is substantially equivalent to the predicate device, AeroEclipse* II BAN, cleared under K053605.
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.