The CapMedic is intended to be used in adults and children 5 years of age and above, who are prescribed CapMedic to be used with their MDIs by their physician. The CapMedic device is an accessory intended for single-patient use to assist physicians and patients in recording and monitoring the MDI (Metered Dose Inhaler) actuations and conducting PEF (Peak Expiratory Flow) and FEV1 (Forced Expiratory Volume for 1 second) spirometry tests. CapMedic measures of MDI use and records it for review by the physician and/or the patient. Furthermore, CapMedic reminds the patient on important steps of MDI use through audio-visual-haptic cues during their MDI use. With the provided PEF/FEV 1 adapter, CapMedic can also perform PEF/FEV1 spirometry test and provides LED feedback based on the measured FEV1. CapMedic can securely transmit MDI usage data and PEF/FEV1 test results to a Bluetooth enabled device running data collection software meeting the CapMedic secure interface protocol over a Bluetooth connection. CapMedic can be used in the home, work, healthcare, and clinical use environments/settings. The CapMedic device may also be used in clinical trials where researchers need to know information about the use of MDI medication(s) by a participant. CapMedic may be used with any of the following MDIs: Ventolin HFA, Advair HFA, Flovent HFA, Xopenex HFA, Symbicort HFA, Atrovent HFA, Proventil HFA, Alvesco HFA, and Asmanex HFA and is attached with a specific adapter for the MDIs.
Device Story
CapMedic is a reusable sensing module attaching to MDIs to monitor usage and provide spirometry testing. It tracks MDI shaking, orientation, actuation-inspiration coordination, and inhalation duration. The device provides real-time audio-visual-haptic cues to guide patients through correct MDI technique and dosing. For spirometry, a separate mouthpiece adapter enables PEF and FEV1 measurement with LED feedback. Data is transmitted via Bluetooth to a secure software application for physician/patient review. Used in home, clinical, or research settings, it helps clinicians assess medication adherence and lung function, potentially improving patient outcomes through better inhaler technique and monitoring.
Clinical Evidence
No clinical data. Substantial equivalence supported by bench testing, including MDI use measurement, UI performance, spirometry accuracy per 2005 ATS guidance, Aerosol Particle Size Distribution (APSD), Delivered Drug Uniformity (DDU), biocompatibility (ISO 10993-1), electrical safety (IEC 60601 series), and cybersecurity/wireless coexistence testing.
Technological Characteristics
Battery-powered (rechargeable Li-Polymer) sensing module; Bluetooth Low Energy connectivity; includes PEF/FEV1 adapter. Complies with IEC 60601-1, 60601-1-2, 60601-1-11, 60601-1-8, and 60601-1-6. Biocompatible per ISO 10993-1. Features integrated audio-visual-haptic UI for user guidance.
Indications for Use
Indicated for adults and children 5 years and older prescribed MDI medication by a physician to assist in monitoring MDI usage and performing PEF/FEV1 spirometry tests.
Regulatory Classification
Identification
A nebulizer is a device intended to spray liquids in aerosol form into gases that are delivered directly to the patient for breathing. Heated, ultrasonic, gas, venturi, and refillable nebulizers are included in this generic type of device.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food & Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
September 13, 2019
Cognita Labs, LLC Rajoshi Biswas Architect and Clinical Lead 700 N Main St. Ste C1 Santa Ana, California 92701
Re: K183586
Trade/Device Name: CapMedic Regulation Number: 21 CFR 868.5630 Regulation Name: Nebulizer Regulatory Class: Class II Product Code: CAF, BZG Dated: August 14, 2019 Received: August 15, 2019
# Dear Rajoshi Biswas:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part
{1}------------------------------------------------
801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
James Lee Assistant Director DHT1C: Division of ENT, Sleep Disordered Breathing, Respiratory and Anesthesia Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K183586
Device Name CapMedic
#### Indications for Use (Describe)
The CapMedic is intended to be used in adults and children 5 years of age and above, who are prescribed CapMedic to be used with their MDIs by their physician.
The CapMedic device is an accessory intended for single-patient use to assist physicians and patients in recording and monitoring the MDI (Metered Dose Inhaler) actuations and conducting PEF (Peak Expiratory Flow) and FEV1 (Forced Expiratory Volume for 1 second) spirometry tests. CapMedic measures of MDI use and records it for review by the physician and/or the patient. Furthermore, CapMedic reminds the patient on important steps of MDI use through audio-visual-haptic cues during their MDI use. With the provided PEF/FEV 1 adapter, CapMedic can also perform PEF/FEV1 spirometry test and provides LED feedback based on the measured FEV1. CapMedic can securely transmit MDI usage data and PEF/FEV1 test results to a Bluetooth enabled device running data collection software meeting the CapMedic secure interface protocol over a Bluetooth connection.
CapMedic can be used in the home, work, healthcare, and clinical use environments/settings.
The CapMedic device may also be used in clinical trials where researchers need to know information about the use of MDI medication(s) by a participant.
CapMedic may be used with any of the following MDIs: Ventolin HFA, Advair HFA, Flovent HFA, Xopenex HFA, Symbicort HFA, Atrovent HFA, Proventil HFA, Alvesco HFA, and Asmanex HFA and is attached with a specific adapter for the MDIs.
Type of Use (Select one or both, as applicable)
| <span style="font-family: sans-serif;"> <svg class="bi bi-check-square" fill="currentColor" height="1em" viewbox="0 0 16 16" width="1em" xmlns="http://www.w3.org/2000/svg"> <path d="M14 1H2a1 1 0 0 0-1 1v12a1 1 0 0 0 1 1h12a1 1 0 0 0 1-1V2a1 1 0 0 0-1-1zM2 0a2 2 0 0 0-2 2v12a2 2 0 0 0 2 2h12a2 2 0 0 0 2-2V2a2 2 0 0 0-2-2H2z" fill-rule="evenodd"></path> <path d="M10.97 4.97a.75.75 0 0 1 1.071 1.05l-3.992 4.99a.75.75 0 0 1-1.08.02L4.324 8.384a.75.75 0 1 1 1.06-1.06l2.094 2.093 3.473-4.425a.236.236 0 0 1 .02-.022z" fill-rule="evenodd"></path> </svg> </span> Prescription Use (Part 21 CFR 801 Subpart D) |
|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| <span style="font-family: sans-serif;"> <svg class="bi bi-square" fill="currentColor" height="1em" viewbox="0 0 16 16" width="1em" xmlns="http://www.w3.org/2000/svg"> <path d="M14 1H2a1 1 0 0 0-1 1v12a1 1 0 0 0 1 1h12a1 1 0 0 0 1-1V2a1 1 0 0 0-1-1zM2 0a2 2 0 0 0-2 2v12a2 2 0 0 0 2 2h12a2 2 0 0 0 2-2V2a2 2 0 0 0-2-2H2z" fill-rule="evenodd"></path> </svg> </span> Over-The-Counter Use (21 CFR 801 Subpart C) |
# CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
# *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# K183586 510(k) Summary
Image /page/3/Picture/2 description: The image shows the logo for Cognita Labs. The word "Cognita" is written in blue, except for the "o", which is a green drawing of a brain. Below the word "Cognita" is the word "Labs" in gray.
September 13th, 2019 Cognita Labs, LLC 700 N Main St. Ste C1 Santa Ana CA 92701 info@cognitalabs.com
This summary of 510(k) information is submitted in accordance with the requirements of 21 CFR 807.92:
#### (a)(1) Submitter Information
| Company | Cognita Labs LLC<br>700 N Main St Ste C1<br>Santa Ana<br>California 92701 |
|---------------------|---------------------------------------------------------------------------|
| Contact Person | Rajoshi Biswas, Architect and Clinical Lead |
| Telephone # | +1-832-538-3042 |
| Date of preparation | September 13, 2019 |
#### (a)(2) Name of the Device
| Device Name | CapMedic |
|-----------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------|
| Common Name | Inhaler Guidance Tool and Home Spirometer (PEF/FEV1) |
| Product Code and<br>Classification Name | Primary product code: CAF, 21 CFR 868.5630, Class II, Nebulizer<br>Secondary product code: BZG, 21 CFR 868.1840, Class II, Diagnostic spirometer |
#### (a)(3) An identification of the legally marketed devices
The predicate devices to which substantial equivalence is claimed are: K990185, MDLog System MDC-512 marketed by Medtrac Technologies, Inc., K140638, Propeller Model 2 Sensor marketed by Reciprocal Labs Corporation, and K133722 Asthma Monitor AM3 GSM (PEF and FEV1 meter) marketed by eResearchTechnology.
#### (a)(4) A description of the device
The CapMedic device includes a reusable sensing module that attaches to the Metered Dose Inhaler (MDI) to measure when and how patients use their inhalers. CapMedic has on-board visual, audio and haptic user interfaces (Uls) to remind patients to use their inhaler correctly through Ul prompts. The recorded parameters of MDI use are MDI shaking, orientation, coordination between MDI actuation and the start of inspiration, and inhalation duration. During MDI use, CapMedic can provide just-in-time reminders, in the form of Ul prompts, for various steps of MDI use, namely, MDI shaking, MDI upright orientation, correct coordination, breathe long and hold breath at the end. CapMedic also optionally provides dosing reminders through UI prompts at times preset by the user.
{4}------------------------------------------------
Further, CapMedic has a built-in peak-flow meter functionality with a separate mouthpiece attachment to the bottom of the sensing module to measure peak expiratory flow (PEF) and forced expiratory volume in 1 second (FEV1). CapMedic can securely transmit MDI usage data and PEF/FEV1 test results to a Bluetooth enabled device running data collection software meeting the CapMedic secure interface protocol over a Bluetooth connection. The CapMedic is powered by a rechargeable Lithium Polymer battery and can be charged via a micro-USB cable.
# (a)(5) Intended use of the device
The CapMedic is intended to be used in adults and children 5 years of age and above, who are prescribed CapMedic to be used with their MDIs by their physician.
The CapMedic device is an accessory intended for single-patient use to assist physicians and patients in recording and monitoring the MDI (Metered Dose Inhaler) actuations and conducting PEF (Peak Expiratory Flow) and FEV1 (Forced Expiratory Volume for 1 second) spirometry tests. CapMedic measures the parameters of MDI use and records it for review by the physician and/or the patient. Furthermore, CapMedic reminds the patient on important steps of MDI use through audio-visual-haptic cues during their MDI use. With the provided PEF/FEV1 adapter, CapMedic can also perform PEF/FEV1 spirometry test and provides LED feedback based on the measured FEV1. CapMedic can securely transmit MDI usage data and PEF/FEV1 test results to a Bluetooth enabled device running data collection software meeting the CapMedic secure interface protocol over a Bluetooth connection.
CapMedic can be used in the home, work, healthcare, and clinical use environments/settings.
The CapMedic device may also be used in clinical trials where researchers need to know information about the use of MDI medication(s) by a participant.
CapMedic may be used with any of the following MDIs: Ventolin HFA, Advair HFA, Flovent HFA, Xopenex HFA, Symbicort HFA, Atrovent HFA, Proventil HFA, Alvesco HFA, Dulera HFA, and Asmanex HFA and is attached with a specific adapter for the MDIs.
# (a)(6) Technological Characteristics
The CapMedic device is technologically equivalent to the three predicates identified above for the three main functions, MDI use measurement, lung function measurement, and data transfer. For the MDI use measurement functions, the CapMedic is similar to both MDLog system and Propeller sensors for monitoring MDI medication use. Both predicates and CapMedic measure the timestamp of MDI use. MDILog also measures MDI use parameters such as MDI shaking, canister actuation and inhalation, all of which are also measured by CapMedic. All devices are battery-powered. Further, both the Propeller device and CapMedic transfer data using Bluetooth Low Energy. The Asthma Monitor measures spirometry parameters of PEF and FEV1, is battery powered and transfers data through Bluetooth Low Energy, same as CapMedic.
The CapMedic device measures the MDI orientation as an additional parameter of MDI use as compared with MDILog. Further, CapMedic also provides UI prompts to remind the user of the correct steps for MDI use and contains a rechargeable battery. All of these parameters were extensively tested, the performance bench-marked and thorough risk analysis conducted to demonstrate that the difference does not affect the substantial equivalence.
# (b) 510(k) summaries for those premarket submissions in which a determination of substantial equivalence is also based on an assessment of performance data shall contain the following information:
#### (1) A brief discussion of the nonclinical tests submitted:
- The MDI use measurement parameters and UI are tested extensively to demonstrate the performance and substantial equivalence.
- The lung function parameters measured by CapMedic have been tested according to the ATS guidance on standardization of spirometry, published in 2005.
- The Aerosol Particle Size Distribution (APSD) and Delivered Drug Uniformity (DDU) tests have been conducted to demonstrate the compatibility of CapMedic's use with the supported MDIs.
- · Biocompatibility tests have been conducted and passed according to ISO 10993-1.
{5}------------------------------------------------
- CapMedic has been tested for the electrical requirements: IEC 60601-1:2005 + A1:2012, IEC 60601-1-2:2014, IEC 60601-1-11:2015, IEC 60601-1-8:2006+AMD1:2012, and IEC 60601-1-6:2010+AMD1:2013.
- Bluetooth data transfer and interfaces are tested to comply with "Radio Frequency Wireless Technology in Medical Devices guidance for industry and FDA, 2013" and FDA's guidance on "Management of Cybersecurity in Medical Devices, 2014".
- · Additionally, usability testing, software, and system verification activities have been conducted.
The tests demonstrate the performance and compatibility of CapMedic with the MDIs Ventolin HFA, ProAir HFA, Advair HFA, Flovent HFA, Xopenex HFA, Symbicort HFA, Atrovent HFA, Proventil HFA, Alvesco HFA, Dulera HFA and Asmanex HFA.
# (2) A brief discussion of the clinical tests submitted
Clinical testing was not required to determine substantial equivalence of CapMedic.
# (3) The conclusions drawn from the nonclinical and clinical tests
Based upon the foregoing performance testing and comparison to the legally marketed predicate devices for indications for use, technology, and performance it is demonstrated that the CapMedic is substantially equivalent to the predicates.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.