The Hailie™ sensor is intended for single-patient use in the home environment as a medication reminder and electronic data capture accessory for recording usage of prescribed inhaler medication. This may be used in the following applications: · In clinical trials, where researchers need to know when a patient has used their trial inhaler medication. · In clinical practice, where specialists, general practitioners, nurses, and educators need to know if a patient has used their prescribed inhaler medication. · In self-management, where patients need to track their medication use as part of their management plan. The Hailie™ sensor is compatible only with the HandiHaler® inhaler and is for use by adults.
Device Story
Hailie Sensor is a clip-on accessory for HandiHaler inhalers; monitors medication adherence. Uses mechanical, optical, and acoustic sensors to detect inhaler presence and actuation. Contains internal clock/calendar to log date/time of use. Features status button, multi-color LEDs, and buzzer for user notifications/reminders. Data transmitted via Bluetooth to paired mobile applications. Used by patients in home environments; also used by clinicians/researchers to track usage history. Benefits include improved medication adherence tracking and management. Device is battery-powered and rechargeable.
Clinical Evidence
No clinical data. Substantial equivalence established via non-clinical testing, including biocompatibility (ISO 10993), electrical safety (IEC 60601-1), EMC (IEC 60601-1-2), software verification/validation, and summative usability testing.
Technological Characteristics
Microprocessor-controlled clip-on sensor; mechanical, optical, and acoustic sensing; Bluetooth connectivity; rechargeable battery; non-volatile memory for event logging. Materials meet ISO 10993-1, -5, -10, -12. Complies with ANSI/AAMI ES60601-1, IEC 60601-1-11, IEC 60601-1-2, and FCC Part 15.
Indications for Use
Indicated for adult patients using the HandiHaler® inhaler for tracking medication adherence in home, clinical practice, or clinical trial settings.
Regulatory Classification
Identification
A nebulizer is a device intended to spray liquids in aerosol form into gases that are delivered directly to the patient for breathing. Heated, ultrasonic, gas, venturi, and refillable nebulizers are included in this generic type of device.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
June 19, 2019
Adherium (NZ) Ltd Chris Mander Head of Regulatory & Quality Level 2, 204 Quay Street Auckland, 1010 NZ
## Re: K182638
Trade/Device Name: Hailie Sensor Regulation Number: 21 CFR 868.5630 Regulation Name: Nebulizer Regulatory Class: Class II Product Code: CAF Dated: May 16, 2019 Received: May 20, 2019
## Dear Chris Mander:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part
{1}------------------------------------------------
801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
James Lee Assistant Director DHT1C: Division of ENT, Sleep Disordered Breathing, Respiratory and Anesthesia Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
#### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
# Indications for Use
510(k) Number (if known) K182638
Device Name Hailie™ Sensor
Indications for Use (Describe)
The Hailie™ sensor is intended for single-patient use in the home environment as a medication reminder and electronic data capture accessory for recording usage of prescribed inhaler medication. This may be used in the following applications:
· In clinical trials, where researchers need to know when a patient has used their trial inhaler medication.
· In clinical practice, where specialists, general practitioners, nurses, and educators need to know if a patient has used their prescribed inhaler medication.
· In self-management, where patients need to track their medication use as part of their management plan.
The Hailie™ sensor is compatible only with the HandiHaler® inhaler and is for use by adults.
Type of Use (Select one or both, as applicable)
| Prescription Use (Part 21 CFR 801 Subpart D)
X Over-The-Counter Use (21 CFR 801 Subpart C)
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the word "adherium" in a lowercase, sans-serif font. The word is colored in a light orange hue. To the right of the word, there is a stylized graphic that resembles two overlapping shapes, one in light gray and the other in the same light orange as the word. The shapes are positioned in a way that suggests movement or a dynamic interaction.
Adherium (NZ) Limited Level 2, 204 Quay Street, Auckland 1010, New Zealand PO Box 106-612, Auckland 1143, New Zealand Phone +64 9 307 2771 contact@hailie.com www.hailie.com
20 June 2019
## 510(k) SUMMARY
#### I. Submitter
| Company Details: | Adherium (NZ) Ltd<br>Level 2, 204 Quay Street, Auckland 1010, New Zealand<br>PO Box 106-612, Auckland 1143, New Zealand<br>Phone +64 9 307 2771 |
|------------------|-------------------------------------------------------------------------------------------------------------------------------------------------|
| Contact Person: | Chris Mander, Head of Regulatory & Quality |
#### II. Device
| Device Name: | Hailie™ Sensor |
|----------------------|------------------------------------------------------------------------------------------------|
| Model Number: | NF0101 |
| Classification Name: | Nebulizer (Direct Patient Interface)<br>Anesthesiology Devices, 21 CFR 868.5630, Class II, CAF |
#### III. Predicate Device
The predicate device to which substantial equivalence is: K180407, Smartinhaler™, manufactured by Adherium (NZ) Limited.
### IV. Device Description
The Hailie™ sensor is used to provide a medication reminder and actuation recording function for use as an accessory to the inhaler specified on the device label. Under the current 510(k), the Hailie™ sensor is indicated for use only with the HandiHaler® inhaler.
The Hailie™ sensor is a clip-on device that attaches externally around the housing of the inhaler. Mechanical, optical and acoustic sensors are used to detect the presence and actuation of the inhaler. The Hailie™ sensor contains an electronic clock and calendar that are used to log the date and time of inhaler actuation.
The user interface consists of a single Status Button and dual multi-color LED indicators to check device status, initiate communications functions, and provide reminder features. The Hailie™ sensor has a Bluetooth interface to wirelessly exchange medication and reminder setting data with a paired communications device and compatible mobile software applications.
{4}------------------------------------------------
### V. Indications for Use
The Hailie™ sensor is intended for single-patient use in the home environment as a medication reminder and electronic data capture accessory for recording usage of prescribed inhaler medication. This may be used in the following applications:
- In clinical trials, where researchers need to know when a patient has used their trial inhaler . medication.
- . In clinical practice, where specialists, general practitioners, nurses, and educators need to know if a patient has used their prescribed inhaler medication.
- . In self-management, where patients need to track their medication use as part of their management plan.
The Hailie™ sensor is compatible only with the HandiHaler® inhaler and is for use by adults.
### VI. Comparison of Technological Characteristics with the Predicate Device
Technological characteristics of the Hailie™ sensor are equivalent to the predicate device listed above. They are both microprocessor-controlled electronic devices that clip on to an inhaler, using a combination of sensors to detect inhaler use which is logged to compile a usage history.
| Device Feature | Proposed Device | Comparison to Predicate Device |
|--------------------------|-------------------------------------------------------------|---------------------------------------------------------------------------------------------------------|
| <b>Intended Use...</b> | | |
| Purpose and function: | Remind and monitor medication usage | Same |
| Type of use: | Over-the-counter use | Same |
| User population: | HandiHaler® users | Substantially equivalent. User population appropriate for intended inhaler. |
| Device usage: | Single-patient use | Same |
| Environment of use: | Home use | Same |
| Dose counting: | No dose counting function | Same |
| <b>Features...</b> | | |
| Data collected: | Date and time for inhaler usage | Same |
| User notifications: | Audio/visual | Same |
| Medication reminders: | Configurable times | Same |
| Data transfer interface: | Automatic upload via Bluetooth | Same |
| <b>User Interface...</b> | | |
| Controls: | Status button | Same |
| Display: | LED lights | Substantially equivalent. Proposed device LEDs located for better visibility. |
| Audio: | Buzzer | Same |
| <b>Hardware...</b> | | |
| Size and weight: | Small and lightweight, suitable for use with inhaler | Substantially equivalent. Proposed device shape and attachment method modified to fit intended inhaler. |
| Configuration: | Attaches around outside of inhaler housing with clip-on fit | Same |
{5}------------------------------------------------
| Device Feature | Proposed Device | Comparison to Predicate Device |
|--------------------------|-------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Sensor technology: | • Optical sensor for inhaler presence<br>• Mechanical and acoustic sensors for<br>inhaler usage | Same<br>Substantially equivalent. Inhaler usage<br>sensors appropriate to type of inhaler.<br>Testing verifies functional performance. |
| Wireless technology: | Bluetooth | Same |
| Inhaler Interaction... | | |
| Inhaler compatibility | Attaches to inhaler without affecting<br>mechanism or operation | Same |
| Inhaler access: | Inhaler readily removed from device<br>for cleaning and label access | Substantially equivalent. Usability testing<br>verifies users understand instructions for<br>review and removal of inhaler. |
| Electronics/Software | | |
| Internal clock: | Yes | Same |
| Event record resolution: | 1 second | Same |
| Event storage: | Non-volatile memory | Same |
| Limited event overwrite: | Yes | Same |
| Reject spurious events: | Yes | Same |
| Power source: | Rechargeable battery | Substantially equivalent. Proposed device<br>only has rechargeable model. |
| Battery life: | 2-3 months on full charge | Substantially equivalent. Proposed device<br>only has rechargeable model. Both<br>devices meet intended battery life. |
| Battery level indicator: | LED light color | Same |
| Power save before use: | Deep-sleep mode | Same |
| Materials... | | |
| Device components: | Components tested to meet ISO 10993<br>cytotoxicity, sensitization, irritation | Substantially equivalent. Proposed device<br>color matches intended inhaler. All<br>materials and colorants are safe for<br>intended use and pass biocompatibility<br>testing. |
The design changes described in the above table were verified by non-clinical testing to establish equivalent performance to the predicate device.
### VII. Performance Data
Non-clinical testing of the Hailie™ sensor has been carried out to cover biocompatibility testing, electrical safety and electromagnetic compatibility, software verification and validation testing, performance testing, and usability evaluation.
### Biocompatibility Testing
The biocompatibility evaluation for the Hailie™ sensor was conducted in accordance with the FDA guidance Use of International Standard ISO 10993-1, "Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process" (16 Jun 2016), to meet requirements from the following standards: ANSI/AAMI/ISO 10993-1:2009 (biocompatibility), ANSI/AAMI/ISO 10993-5:2009 (cytotoxicity), ANSI/AAMI/ISO 10993-10:2010 (sensitization and intracutaneous irritation), and ANSI/AAMI/ISO 10993-12:2012 (sample preparation for biocompatibility testing). The
{6}------------------------------------------------
Hailie™ sensor was tested according to requirements for a surface device contacting intact skin for limited duration ≤ 24 hours.
### Electrical Safety and Electromagnetic Compatibility (EMC)
Electrical safety and EMC testing were conducted by external laboratories on the Hailie™ sensor. The device complies with the following standards and regulations: ANSI/AAMI ES60601-1:2005 +A1:2012, C1:2009, A2:2010 (general safety), IEC 60601-1-11:2015 (home-use safety), IEC 60601-1-2:2014 (electromagnetic compatibility), and ANSI C63.10:2013 / 47 CFR Part 15 (FCC regulations for radiofrequency (RF) devices). General safety testing was conducted according to applicable requirements for a home use, battery-powered device, that may be recharged via a USB connector. EMC testing was conducted according to applicable requirements for an internally powered, nonpatient coupled, Bluetooth radio device that may be recharged from an AC adaptor. Information was provided according to FDA guidance Information to Support a Claim of Electromagnetic Compatibility (EMC) of Electrically-Powered Medical Devices (11 Jul 2016), and Radio-Frequency Wireless Technology in Medical Devices (13 Aug 2013).
### Software Verification and Validation Testing
Software verification and validation testing were conducted to ensure correct functionality for the Hailie™ sensor software release, for all software modules (real-time clock, user interface, power management, event recording, Bluetooth communications, USB power, non-volatile parameters, medication reminders, medication, and user functionality). Documentation was provided as recommended by FDA guidance Content of Premarket Submissions for Software Contained in Medical Devices (11 May 2005), Off-The-Shelf Software Use in Medical Devices (9 Sep 1999), and Content of Premarket Submissions for Management of Cybersecurity in Medical Devices (2 Oct 2014).
### Performance Testing
Performance testing was conducted to establish correct functionality and compatibility of the Hailie™ sensor with the HandiHaler® according to requirements, covering:
- Optical inhaler presence detection determined optical calibration limits and confirmed . accurate detection of an installed inhaler.
- Mechanical inhaler usage detection sensor confirmed reliability of sensor over device lifetime. .
- Acoustic inhaler usage detection sensor confirmed accurate detection of inhaler usage and . prevention of detection from other inputs.
- Spurious log prevention testing determined circuit parameters and confirmed prevention of . erroneous medication usage logs.
- General performance testing confirmed acceptable performance over the specified shelf life . and specified Bluetooth communications range.
- . User interface testing - confirmed visibility of device display and audibility of device buzzer.
### Usability Evaluation
Summative usability validation testing was carried out to evaluate critical tasks indicated by the usability risk analysis process. A description of the usability engineering process and the results obtained from the testing were provided in accordance with the FDA guidance Applying Human Factors and Usability Engineering to Medical Devices (3 Feb 2016).
### Clinical Testing
Clinical testing was not required for a determination of substantial equivalence of the Hailie™ sensor. The product functionality has been adequately assessed by non-clinical testing as above.
### VIII. Conclusions
Finished device testing carried out for the Hailie™ sensor indicates it meets design and performance functional requirements. Software verification demonstrates that device functions are substantially equivalent to the predicate device meets standard requirements for biocompatibility,
{7}------------------------------------------------
#### 510(k) Summary continued - Hailie™ Sensor
electrical safety, electromagnetic compatibility, and wireless communications. The usability evaluation demonstrated that the Hailie™ sensor can be used successfully with the compatible inhaler.
This information indicates that the Hailie™ sensor is substantially equivalent to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.