K171549 · Qingdao Future Medical Technology Co., Ltd. · CAF · Aug 31, 2018 · Anesthesiology
Device Facts
Record ID
K171549
Device Name
Intelligent Mesh Nebulizer
Applicant
Qingdao Future Medical Technology Co., Ltd.
Product Code
CAF · Anesthesiology
Decision Date
Aug 31, 2018
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 868.5630
Device Class
Class 2
Attributes
Therapeutic, Pediatric
Indications for Use
The Intelligent Mesh Nebulizer designed to aerosolized liquid medications for inhalation by patient, the device may be used with pediatric (> 4 years of age) and adult patients in the home, hospital and sub-acute care settings. It is not intended for use with Pentamidine.
Device Story
Vibrating mesh nebulizer; uses low-frequency piezoelectric ceramic vibration to oscillate metal mesh plate; liquid medication passes through micro-mesh holes to create low-velocity aerosol particles for inhalation via mask or mouthpiece. Device is portable, battery-powered (3.7V DC), and holds up to 8ml medication. Model NEB001 includes Bluetooth connectivity to a mobile app; NEB002 is standalone. App provides interactive games, aerosol flow rate adjustment, and treatment information display. Flow sensors in the atomizing cup detect inhalation/exhalation to trigger app-based interactive games. Healthcare providers use device to deliver respiratory medication; patients use it for home treatment. App enhances adherence through gamification and dosage tracking. Benefits include portable, efficient medication delivery to lower respiratory tract.
Clinical Evidence
No clinical data. Bench testing only, including particle size distribution (cascade impactor and laser light scattering), electrical safety (IEC 60601-1), EMC (IEC 60601-1-2), biocompatibility (ISO 10993 series), and software validation.
Technological Characteristics
Vibrating mesh nebulizer; piezoelectric ceramic actuator; 3.7V DC lithium battery; 8ml capacity. Materials: PVC patient contact parts. Connectivity: Bluetooth (NEB001). Standards: IEC 60601-1, IEC 60601-1-2, IEC 60601-1-11, ISO 10993-5/10/3/6/11. Software: App-based control for flow rate and interactive features.
Indications for Use
Indicated for aerosolizing liquid medications for inhalation by pediatric (>4 years) and adult patients in home, hospital, and sub-acute care settings. Not for use with Pentamidine.
Regulatory Classification
Identification
A nebulizer is a device intended to spray liquids in aerosol form into gases that are delivered directly to the patient for breathing. Heated, ultrasonic, gas, venturi, and refillable nebulizers are included in this generic type of device.
Predicate Devices
Micro Air Vibrating Mesh Nebulizer (NE-U22) (K062263)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, with the letters "FDA" in a blue square. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
August 31, 2018
Qingdao Future Medical Technology Co., Ltd. % Ray Wang Official Correspondent Beijing Believe Technology Service Co., Ltd. 5-402, Bldg #27, YangGuangYiShang, No.56, LiangXiang East Rd., FangShan Beijing, 102401 CHINA
Re: K171549
Trade/Device Name: Intelligent Mesh Nebulizer Regulation Number: 21 CFR 868.5630 Regulation Name: Nebulizer Regulatory Class: Class II Product Code: CAF Dated: July 23, 2018 Received: July 30, 2018
Dear Ray Wang:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal
{1}------------------------------------------------
statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.html; good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
# James J. Lee -S
for Tina Kiang, Ph.D. Acting Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
# Indications for Use
510(k) Number (if known) K171549
Device Name Intelligent Mesh Nebulizer
Indications for Use (Describe)
The Intelligent Mesh Nebulizer designed to aerosolized liquid medications for inhalation by patient, the device may be used with pediatric (>4 years of age) and adult patients in the home, hospital and sub-acute care settings. It is not intended for use with Pentamidine.
| Type of Use (Select one or both, as applicable) | |
|------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------|
| <span style="font-size: 12px;"> ☑ </span> Prescription Use (Part 21 CFR 801 Subpart D) | <span style="font-size: 12px;"> ☐ </span> Over-The-Counter Use (21 CFR 801 Subpart C) |
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
***DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.***
The burden time for this collection of information is estimated to average 79 hours per response, including the
time to review instructions, search existing data sources, gather and maintain the data needed and complete
and review the collection of information. Send comments regarding this burden estimate or any other aspect
of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2020 See PRA Statement below.
{3}------------------------------------------------
### 510(k) Summary
This 510(k) Summary of 510(k) safety and effectiveness information is being submitted in accordance with requirements of SMDA 1990 and Title 21, CFR Section 807.92.
The assigned 510(k) Number: K171549
- 1. Date of Preparation
08/30/2018
- Sponsor 2.
# Qingdao Future Medical Technology Co., Ltd.
Room 401 Building F, No. 169 Zhuzhou Road, Laoshan District, QingDao City Shangdong Province, China Contact Person: Wang WeiLai Position: General Manager Tel: 86-18669802350 Fax:86-532-55677725 Email: wang.weilai@smartdean.com
- Submission Correspondent 3.
### Beijing Believe Technology Service Co., Ltd.
5-402, Building #27, YangGuangYiShang, No. 56, LiangXiang East Rd., FangShan District, BeiJing, China 102401 Contact Person: Ray Wang Position: General Manager Tel: 86-18910677558 Fax:86-10-59120817 Email: Ray.wang@believe-med.com
- 4. Identification of Proposed Device
Trade Name: Intelligent Mesh Nebulizer Common Name: Nebulizer Model(s): NEB001/NEB002
Regulatory Information: Classification Name: Nebulizer (Direct Patient Interface)
{4}------------------------------------------------
Classification: II; Product Code: CAF; Regulation Number: 21 CFR 868.5630; Review Panel: Anesthesiology;
Intended Use:
The Intelligent Mesh Nebulizer designed to aerosolized liquid medications for inhalation by patient, the device may be used with pediatric (> 4 years of age) and adult patients in the home, hospital and sub-acute care settings.
It is not intended for use with Pentamidine.
- Device Description 5.
The proposed device is a vibrating mesh nebulizer that uses low frequency vibration to create aerosol and delivers aerosolized medication to the lower respiratory tract by using a vibrating mesh to create aerosol and provide fine particles to the patient's lungs. The mesh plate has holes to create low velocity aerosol.
The proposed device is a portable size, curvaceous body design that is convenient to hold, and ability to detect the amount of medications available. Which is battery powered, 3.7V d.c. internally lithium battery. The medication container capacity is 8ml maximum.
The proposed device consists of two parts: physical device and APP software.
For physical device, it is pushed by certain circuit frequency vibration to make piezoelectric ceramic vibrate harmony that caused high speed vibration of metal mesh. And the medicine liquid will be quickly popped through micro mesh hole of metal mesh plate to be countless micro atomizing particles. This will be further transferred by inhalation treatment using masks or mouthpieces to patients' respiratory system.
For APP software, it consists of interactive games, Aerosol Flow rate adjustment, power off and information display which could enhance entertainment and users' adherence to the nebulizer during treatment. It could be connected to the physical device by bluetooth.
Firstly, when users inhale and exhale, the flow pressure could trigger two single-way switches(inhalation switch and exhalation switch) seprately on atomizing cup, then it could transmit signal to CPU of the device by sensor and the signal will be transmitted to the APP further by bluetooth to control playing interactive games.
Secondly, the APP could calculate atomizing time and dosage according to the information transmitted from physical device, and display them in the games.
There are 2 models included, NEB001/NEB002, the two models have same intended use, mechanism of action, principle and specification.
The only difference between two models is that the model NEB001 has two parts, Physical Device
{5}------------------------------------------------
and App software, the NEB001 has the function of Bluetooth for connection between the physical device and App; the model NEB002 only has physical devcie. The detailed difference shown as following:
| Model | NEB001 | NEB002 |
|----------------------------------------|----------------------------------------------------------------------------------------------------------|-------------------|
| Main Parts | Physical Device &<br>APP software | Physical Device |
| Nebulization<br>Rate | Adjustable as 0.21 ml/min., 0.27ml/min.,/<br>0.30ml/min., 0.32 ml/min., 0.35ml/min.<br>Controlled by APP | 0.2ml/min minimum |
| Data<br>Transmission | Bluetooth | N/A |
| The OS of<br>connected smart<br>device | Andriod 5.0 or higher | N/A |
The proposed device has the components shown as following illustration:
Image /page/5/Figure/5 description: This image shows a medical nebulizer and its parts. The nebulizer is composed of a mesh plate, atomizing cup, main unit, and a bottom piece. The main unit has an indicator light, charge indicator light, USB interface, and an active/stand-by button.
Figure 1 Physical Device Components illustration
{6}------------------------------------------------
The contents provided in the Table 2 are description about the components.
| Main Unit | Provide electricity. |
|-----------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Atomizing Cup | Filled with medicine liquid. |
| Mesh plate | Create low velocity aerosol |
| PUSH button | Disassemble the nebulizer. |
| (ACTIVE/STAND<br>-BY)button | Active/stand-by the nebulizer. |
| Indicator light | flash blue light when the battery is running out<br>electricity.<br>Flash green light for 3s when there's no medicine<br>liquid in the cup.<br>Normally on green when the nebulizer working<br>regularly. |
| Charge indicator<br>light | Charge indicator .normally on orange when<br>charging and tuns out to green while the charging<br>finished. |
| USB interface | Connect wire and charging adaptor. |
| Mask | Transport the aerosol to patient's respiratory<br>system. |
| Mouthpiece | Transport the aerosol to patient's respiratory<br>system. |
| Adapter | Battery charging |
Table 2 Main Components of Physical Device introduction
#### Identification of Predicate Device 6.
Predicate Device #1: 510(k) Number K062263 Predicate Device Name Micro Air Vibrating Mesh Nebulizer (NE-U22) Manufacturer Omron Healthcare, Inc.
Predicate Device #2: 510(k) Number K132247 Predicate Device Name Ultrasonic Mesh Nebulizer NE105
{7}------------------------------------------------
## Manufacturer
Foshan Gaunying Electronics Co., Ltd.
- 7. Non-Clinical Test Conclusion
Non clinical tests were conducted to verify that the proposed device met all design specifications as was Substantially Equivalent (SE) to the predicate device. The test results demonstrated that the proposed device complies with the following standards:
- A IEC 60601-1:2012 Medical Electrical Equipment - Part 1: General Requirements For Basic Safety And Essential Performance:
- > IEC 60601-1-2:2014, Medical electrical equipment- Part 1-2: General requirements for basic safety and essential performance- Collateral standard: Electromagnetic compatibility-Requirements and tests; which includes the tests as following:
- IEC CISPR 11:2016 Industrial, scientific and medical equipment - Radio-frequency disturbance characteristics - Limits and methods of measurement;
- . IEC 61000-4-2:2008 Electromagnetic compatibility (EMC) - Part 4-2: Testing and measurement techniques – Electrostatic discharge immunity test;
- IEC 61000-4-3:2010 Electromagnetic compatibility (EMC) - Part 4-3: Testing and measurement techniques - Radiated, radio-frequency, electromagnetic field immunity test;
- IEC 61000-4-8:2009 Electromagnetic compatibility (EMC) - Part 4-8: Testing and measurement techniques - Power frequency magnetic field immunity test;
- > IEC 60601-1-11:2015, Medical Electrical Equipment -- Part 1-11: General Requirements For The Basic Safety And Essential Performance - collateral standard: Requirements for medical electrical equipment and medical electrical system used in the home healthcare environment;
- ISO 10993-5:2009 Biological Evaluation Of Medical Devices Part 5: Tests For In Vitro A Cytotoxicity;
- > ISO 10993-10:2010 Biological Evaluation Of Medical Devices - Part 10: Tests For Irritation And Skin Sensitization;
- > ISO 10993-3:2014 Biological Evaluation Of Medical Devices - Part 3: Tests For Genotoxicity, Carcinogenicity And Reproductive Toxicity;
- > ISO 10993-6:2007 Biological Evaluation Of Medical Devices - Part 6: Tests For Local Effects After Implantation;
- > ISO 10993-11:2006 Biological Evaluation Of Medical Devices - Part 11: Tests For Systemic Toxicity;
- > Software Validation Test;
- > Particle Size Distribution testing with Cascade Impactor testing method AND Laser Light Scattering testing method, and three kind of drugs used testing, Pulmicort-contains Budesonide, BRICANYL-contains Terbutaline and Tobramycin in Physiological saline;
- Wireless Data Integrity Testing (only for NEB001); A
- A Wireless Coexistence Testing (only for NEB001);
- Breath Detection Testing (only for NEB001); >
{8}------------------------------------------------
- > Aerosol Flow rate testing (only for NEB001);
- > Excessive Temperatures Test for battery, evaluate the device surface temperature when the inter battery under the "short" condition.
- > Usability Study, performed for NEB001 and NEB002 respectively;
- > Particle Size Distribution testing for pediatric patients (mouthpiece and face mask), simulation flow rate at 7L/min for pediatric population (>4 years of old). Cascade Impactor testing method AND Laser Light Scattering testing method, and three kind of drugs used for each testing, Pulmicort-contains Budesonide, BRICANYL-contains Terbutaline and Tobramycin in Physiological saline;
- Clinical Test Conclusion 8.
No clinical study is included in this submission.
| Table 3 General Comparison | | | | |
|----------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------|
| ITEM | Proposed Device | Predicate Device #1<br>K062263 | Predicate Device #2<br>K032849 | Remark |
| Product Code | CAF | CAF | CAF | Same |
| Regulation No. | 21 CFR 868.5630 | 21 CFR 868.5630 | 21 CFR 868.5630 | Same |
| Class | 2 | 2 | 2 | Same |
| Prescription/<br>OTC | Prescription | Prescription | Prescription | Same |
| Intended Use | The Intelligent Mesh Nebulizer<br>designed to aerosolized liquid<br>medications for inhalation by<br>patient, the device may be used<br>with pediatric and adult patients in<br>the home, hospital and sub-acute<br>care settings.<br>It is not intended for use with<br>Pentamidine. | The Omron NE-U22 is an ultrasonic<br>(vibrating mesh) nebulizer system<br>designed to aerosolize liquid<br>medications for inhalation by<br>patient.<br>The device may be used with<br>pediatric and adult patients in the<br>home, hospital, and sub-acute care<br>settings.<br>It is not intended for use with<br>Pentamidine. | The Aeroneb Go<br>Nebulizer, for use by<br>pediatric and adult patients,<br>is intended to aerosolize<br>physician-prescribed<br>solutions for inhalation that<br>are approved for use with a<br>general-purpose nebulizer | Same |
- Substantially Equivalent (SE) Comparison 9.
| Table 4 Performance Comparison |
|--------------------------------|
|--------------------------------|
| ITEM | Proposed Device | Predicate Device #1<br>K062263 | Predicate Device #2<br>K032849 | Remark |
|--------------|----------------------------------------------------------------------|-----------------------------------------------------|----------------------------------------------------------------------|---------|
| Power Source | ZN-103450 Lithium battery:<br>3.7Vd.c.<br>AC Adapter AC 100-240Va.c. | Battery - 2 “AA”<br>AC Adapter AC 120V (60Hz/DC 3V) | Battery—3 “AA”<br>disposable or rechargeable batteries, 4.5 VDC, 350 | Similar |
{9}------------------------------------------------
| | 47-63Hz 0.4-0.2A | | mA | |
|--------------------------------------|------------------------------|------------------------------|--------------------------------|------------|
| Nebulizing<br>Method | Vibrating Mesh | Vibrating Mesh | Vibrating mesh | Same |
| Vibration<br>Frequency | Approx. 110KHz | Approx. 180KHz | Approx. 100 KHz | Similar |
| Aerosol Flow rate | 0.2ml/min minimum | 0.25 ml/min to 0.9 ml/min | / | Similar |
| Particle Size | 2.2µm±25% | Approx. 3 um | 1.96-2.26 um | Similar |
| Medicine Capacity | 8ml maximum,0.5ml minimum | 7 ml | 6 mL | Similar |
| Nebulizer<br>Components<br>Cleanable | Yes | Yes | Yes | Same |
| Use | Single Patient | Single Patient | Single Patient | Same |
| Patient Connector | Mouthpiece or mask | Mouthpiece | Mouthpiece or optional<br>mask | Same |
| Wireless<br>Connection | Bluetooth<br>(Model NEB001) | NA | NA | Difference |
| Mobile<br>Application | Yes<br>(Model NEB001) | NA | NA | Difference |
| Dimensions (mm) | 50mm(L)×74mm(W)×111mm(<br>H) | 38(W) x 51(D) x 104(H) | 40 x 105 x 95 mm | Similar |
| Weight (kg) | Approx.106g | 97 g | 325 g | Similar |
| Operating<br>Conditions | 5°C to 40°C, 15% to 90% RH | 0°C to 45°C, 30% to 85% RH | 5°C to 45°C to 95% RH | Similar |
| Storage<br>Conditions | -25°C to 70°C, ≤90% RH | -25°C to 70°C, 10% to 90% RH | -20°C to 60°C to 95% RH | Similar |
# Table 5 Safety Comparison
| Item | Proposed Device | Predicate Device #1 | Predicate Device #2 | Remark |
|------------------------------|---------------------------|---------------------------|---------------------------|---------|
| | | K062263 | K032849 | |
| Patient Contact<br>Materials | PVC | Polypropylene | PVC | Similar |
| Cytotoxicity | Comply with ISO 10993-1 | Comply with ISO 10993-1 | Comply with ISO 10993-1 | |
| Sensitization<br>Irritation | | | | Same |
| Electrical Safety | Comply with IEC 60601-1 | Comply with IEC 60601-1 | Comply with IEC 60601-1 | Same |
| EMC | Comply with IEC 60601-1-2 | Comply with IEC 60601-1-2 | Comply with IEC 60601-1-2 | Same |
{10}------------------------------------------------
| Item | NEB001 | | | | | | NEB002 | | | | | | Aeroneb Go (K032849) | | | | | |
|-------------------------------------------------|---------------|---------------|-----------------|---------------|---------------|-----------------|---------------|---------------|-----------------|---------------|---------------|-----------------|----------------------|---------------|-----------------|---------------|---------------|-----------------|
| | mouthpiece | | | mask | | | mouthpiece | | | mask | | | mouthpiece | | | mask | | |
| | Budesonide | Terbutaline | Tobramycin | Budesonide | Terbutaline | Tobramycin | Budesonide | Terbutaline | Tobramycin | Budesonide | Terbutaline | Tobramycin | Budesonide | Terbutaline | Tobramycin | Budesonide | Terbutaline | Tobramycin |
| MMAD( $\mu$ m) | 2.09-2.29 | 2.09-2.25 | 2.11-2.27 | 2.10-2.30 | 2.11-2.27 | 2.17-2.25 | 2.08-2.28 | 2.09-2.25 | 2.09-2.25 | 2.11-2.27 | 2.15-2.25 | 2.10-2.28 | 2.09-2.19 | 2.00-2.12 | 1.96-2.06 | 2.04-2.12 | 1.87-2.01 | 1.95-2.09 |
| GSD ( $\mu$ m) | 2.05-2.13 | 2.02-2.10 | 2.04-2.12 | 2.12-2.20 | 1.92-2.00 | 2.11-2.19 | 2.03-2.11 | 2.01-2.09 | 2.02-2.10 | 1.98-2.10 | 1.89-2.07 | 1.94-2.02 | 2.01-2.13 | 1.91-2.05 | 1.81-1.93 | 2.14-2.24 | 1.78-1.86 | 2.05-2.15 |
| Particle mass collected by the cascade impactor | 443.53-536.77 | 668.64-801.88 | 1293.08-1466.32 | 409.32-522.00 | 667.78-777.42 | 1198.98-1284.48 | 441.53-534.77 | 672.93-806.17 | 1286.83-1460.07 | 439.90-516.88 | 671.39-778.11 | 1209.44-1322.10 | 503.01-571.57 | 757.27-851.79 | 1414.18-1510.96 | 476.37-562.91 | 788.88-844.16 | 1394.47-1462.83 |
| respirable particle fraction(%) (0.5-5 $\mu$ m) | 83.24-88.84 | 84.11-91.23 | 80.99-88..11 | 83.91-90.03 | 84.09-91.21 | 84.12-91..24 | 83.39-88.99 | 83.56-90.68 | 79.89-87..01 | 81.70-88.54 | 82.63-88.33 | 82.42-87..90 | 85.21-88.89 | 84.26-88.64 | 82.56-88..26 | 83.80-88.94 | 81.07-85.87 | 81.92-86..60 |
| coarse particle fraction(%) (>4.7 $\mu$ m) | 12.89-18.97 | 14.14-14.44 | 14.14-14..42 | 15.58-15.86 | 15.14-15.42 | 13.07-13..35 | 13.88-19.96 | 13.05-13.33 | 15.2-15.48 | 13.74-17.82 | 14.01-14.23 | 16.21-16..35 | 13.69-14.03 | 14.69-14.86 | 17.21-17..47 | 16.50-16.84 | 18.17-18.29 | 16.91-17..19 |
| ultra-fine particle fraction(%) | 11.37-15.47 | 9.99-14..89 | 7.67-12.57 | 10.67-15.57 | 7.78-12..68 | 7.80-11.90 | 11.09-15.19 | 9.99-14..89 | 8.67-13.57 | 9.27-11..43 | 8.20-12..36 | 9.33-10.45 | 13.57-18.31 | 14.60-17.08 | 14.38-16..86 | 10.12-14.22 | 11.32-15.50 | 9.89-15.83 |
| <1um) | | | | | | | | | | | | | | | | | | |
| | | | | | | | | | | | | | | | | | | |
# Table 6 Comparative particle test Comparison
{11}------------------------------------------------
510(K) Summary
{12}------------------------------------------------
Analysis for difference
The proposed device has the same Intended Use, Classification, Nethod, Operation and use with the predicate device.
The proposed device has the similar specifications with the predicate device, such as Power Source, Vibration Frequency, Aerosol Flow rate, Particle Size, Medicine Capacity, based on the nonclinical tests performed, those minor differences for similar specifications do not effects the safety and effective of the device.
The main differences are the functions of wireless connection and mobile application between the proposed device and predicate, the model NEB 001. For this difference, we have designed the proposed device as the FCC Part 15 rules and addressed the requirements of "Radio Frequency Wireless Technology in Medical Devices – Guidance for Industry and Food and Drug Administration" and "Content of Premarket Submissions for Management of Cybersecurity in Medical Devices". We have conducted the non-clinical test for such requirements or standards, such as software V&V testing, Testing of FCC Part 15 rules, IEC 60601-1-2 testing, Data Integrate Testing, Wireless Coexistence Testing, and we also conducted the risk management activities about the risks caused by wireless and mobile application as the guidance mentioned above.
The proposed device is substantially equivalent to the predicate device. Based on the nonclinical tests performed, the subject device is as safe, as effective, and performs as well as the legally marketed predicate device.
## 10. Substantially Equivalent (SE) Conclusion
Based on the comparison and analysis above, the proposed device is determined to be Substantially Equivalent (SE) to the predicate device.
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.