K123614 · Trudell Medical Intl. · CAF · Mar 11, 2013 · Anesthesiology
Device Facts
Record ID
K123614
Device Name
SOLARYS VENTILATOR AEROSOL DELIVERY SYSTEM (AS)
Applicant
Trudell Medical Intl.
Product Code
CAF · Anesthesiology
Decision Date
Mar 11, 2013
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 868.5630
Device Class
Class 2
Attributes
Therapeutic, Pediatric
Indications for Use
The Solarys* Ventilator Aerosol Delivery System (AS) is a nebulizer intended to deliver prescribed aerosolized medication to mechanically ventilated pediatric and adults patients able to use an Endotracheal Tube equal to or greater than 4.5 mm in diameter. The device is designed to operate in-line with ventilator circuits and mechanical ventilators in acute and subacute care environments.
Device Story
Solarys Ventilator Aerosol Delivery System (AS) is a single-use, in-line nebulizer for mechanically ventilated patients. Device atomizes bulk liquid medication into aerosol for inhalation via ventilator circuit. Unlike vibrating mesh predicates, Solarys is pneumatically driven; requires no cleaning during circuit life. Used in acute/subacute care settings by clinicians. Input is liquid medication; output is aerosolized drug delivered through endotracheal tube. Performance verified via cascade impactor and simulated adult patient lung models to measure total delivered dose and particle size distribution. Benefits include continuous delivery without breaking patient circuit for refills.
Clinical Evidence
No clinical data. Substantial equivalence based on bench testing, including aerosol characterization using cascade impactor methodology and simulated adult mechanical ventilation patient models. Testing measured total delivered dose and particle size distribution (MMAD, GSD) for multiple medications, confirming performance comparable to predicate.
Technological Characteristics
Pneumatically driven nebulizer; single-use/disposable; in-line ventilator circuit integration. Materials evaluated per ISO 10993-1. No electrical components or software. Designed for endotracheal tubes ≥ 4.5 mm.
Indications for Use
Indicated for pediatric and adult patients on mechanical ventilation requiring aerosolized medication, using an endotracheal tube ≥ 4.5 mm diameter. Contraindicated for patients requiring smaller endotracheal tubes.
Regulatory Classification
Identification
A nebulizer is a device intended to spray liquids in aerosol form into gases that are delivered directly to the patient for breathing. Heated, ultrasonic, gas, venturi, and refillable nebulizers are included in this generic type of device.
Predicate Devices
Aeroneb Professional Nebulizer System Aerogen Limited -Aeroneb Solo (K070642)
Submission Summary (Full Text)
{0}------------------------------------------------
K123614
Trudell Medical International
# MAR 1 1 2013
### Premarket Notification 510(k) Solarys* Ventilator Aerosol Delivery System
# Section 5 – 510(k) Summary
Prepared: 20 November 2012
#### Trudell Medical International 510(k) Owner 725 Third Street London, Ontario N5V 5G4 CANADA Contact Person Darryl Fischer Associate Director, Global Regulatory Affairs Phone 1-519-455-7060 ext 2140 Fax 1-519-455-6329 e-mail dfischer@trudelimed.com
#### Device Name
Solarys* Ventilator Aerosol Delivery System Proprietary Common/Classification Nebulizer, Direct Patient Interface
| Product Code | CAF |
|--------------|-----|
|--------------|-----|
Classification Regulation 868.5630
Predicate Device 510(k) Number Trade/Model Name(s) Manufacturer K070642 Aeroneb Professional Nebulizer System Aerogen Limited -Aeroneb Solo
#### Device Description
The Solarys* Ventilator Aerosol Delivery System (AS)
The Solarys* Ventilator Aerosol Delivery System (AS) is a single-use (disposable) continuous nebulizer system designed for use with mechanically ventilated patients to aerosolize physician-prescribed medications for inhalation.
Amended 11 March 2013
{1}------------------------------------------------
### Section 5 -- 510(k) Summary
#### Intended Use
The Solarys* Ventilator Aerosol Delivery System (AS) is a nebulizer intended to deliver prescribed aerosolized medication to mechanically ventilated pediatric and adults patients able to use an Endotracheal Tube equal to or greater than 4.5 mm in diameter. The device is designed to operate in-line with ventilator circuits and mechanical ventilators in acute and subacute care environments.
#### Technological Characteristic Comparison to Predicate Device(s)
Common characteristics to both the Solarys* AS and the predicate device ;
- operate primarily by atomizing bulk liquid containing medication into an aerosol form that can be inhaled
- both have a reservoir that is part of the medication delivery system for extended treatments
- both devices allow for refilling during extended treatments, without having to break the patient circuit
Relevant differences in operating principles of the Solarys* AS and the predicate device; - the Solarys* AS is a pneumatically driven nebulizer, whereas the Aerogen Solo system is a vibrating mesh nebulizer that is electrically driven.
- No cleaning of the Solarys* AS system is required to maintain nebulizer performance (remains in the patient circuit for the duration of circuit life). The predicate device requires cleaning to prevent clogging of the vibrating mesh to maintain its performance with many formulations.
#### Non-Clinical Test Summary
Evaluation of the Solarys* AS and the predicate device was performed in accordance with the relevant sections of the CDRH Guidance Document "Reviewers Guidance for Nebulizers, Metered Dose Inhalers, Spacers and Actuators" (FDA/CDRH/ODE/DCRD/ADDB -1993).
- . Aerosol characterization testing for the Solarys* AS and the predicate device has been conducted with 3 commonly prescribed drugs using the cascade impactor method. The total mass of each drug emitted was determined by simulating normal device usage in a ventilator circuit and demonstrates that the performance of the Solarys* AS raises no new issues of safety or effectiveness from the legally marketed predicate device. Constant sampling through an
Amended 11 March 2013
{2}------------------------------------------------
Premarket Notification 510(k) Solarys* Ventilator Aerosol Delivery System
# Section 5 – 510(k) Summary
impactor at a fixed flow rate does not provide an accurate indication of the amount of medication potentially available in a clinically relevant situation, so in order to obtain a representative measurement of total emitted mass (TEM), a simulated adult mechanically ventilated patient was used to assess the mass of medication available at the exit of a 7.0 mm I.D. ETT. The aerosolized medication exiting the ETT was collected directly onto a polypropylene microfiber filter, connected to an adult mechanical test lung that was used as the patient model.
- . Biocompatibility testing was conducted on the Solarys* AS device according to ISO 10993-1
- Mechanical safety testing consisted of the following; .
- 0 Simulated use
- 0 Maximum pressure testing
- Environmental limit testing (operating limits and environmental storage and o transportation testing)
- Drop testing O
- o Connection leakage
- Strain relief fatigue and tensile testing o
The aerosol characterizations below have been defined based on statistical analysis (95% confidence interval) of the observed data collected using cascade impactor methodology.
| Particle Size Characterization | | | | | |
|-------------------------------------------------|-----------|-----------------------|-----------|---------------------|--|
| | | | Combivent | | |
| | Albuterol | Ratio-<br>Ipratropium | Albuterol | Ipratropium Bromide | |
| Filter / Breathing Simulator Data | | | | | |
| Total Delivered Dose ex ETT (µg) | 724 | 296 | 1593 | 390 | |
| NGI Data | | | | | |
| Particle Fraction Greater Than<br>8.64µm (%) | 24.3 | 21.6 | 26.3 | 26.5 | |
| Particle Fraction Between 0.98 -<br>8.64 µm (%) | 70.2 | 73.4 | 68.2 | 68.0 | |
| Particle Fraction Less Than 0.98<br>µm (%) | 5.5 | 5.0 | 5.5 | 5.5 | |
| MMAD (µm) | 4.3 | 4.0 | 4.6 | 4.7 | |
| GSD | 2.8 | 2.8 | 2.7 | 2.8 | |
{3}------------------------------------------------
Premarket Notification 510(k) Solarys * Ventilator Aerosol Delivery System
#### Trudell Medical International
# Section 5 - 510(k) Summary
#### Clinical Performance Summary
Not applicable, the determination of substantial equivalence is not based on Clinical Performance Data.
#### Conclusions from Testing
The Solarys* AS has been evaluated against a currently marketed (predicate) device for the determination of substantial equivalency. The Solarys* AS and the predicate device share common indications for use and usage environments. The devices are both single patient use, non-sterile, disposable and are available by prescription.
Performance data, gathered in accordance with "Reviewers Guidance for Nebulizers, Metered Dose Inhalers, Spacers and Actuators" (FDA/CDRH/ODE/DCRD/ADDB -1993), demonstrate that the Solarys* AS raises no new issues of safety or effectiveness from the legally marketed predicate device.
Amended 11 March 2013
{4}------------------------------------------------
Image /page/4/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized eagle with three lines forming its body and wings. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular fashion around the eagle.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
March 11, 2013
Mr. Darryl Fischer, COM Associate Director, Global Regulatory Affairs Trudell Medical International 725 Third Street London Canada N5V 5G4
Re: K123614
Trade/Device Name: Solarys * Ventilator Aberosol Delivery System Regulation Number: 21 CFR 868.5630 Regulation Name: Nebulizer Regulatory Class: II Product Code: CAF Dated: December 21, 2012 Received: December 26, 2012
#### Dear Fischer:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act
{5}------------------------------------------------
Page 2 - Mr. Fischer
or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH0ffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
Kwame O. Ulmer for
Anthony D. Watson, B.S., M.S., M.B.A. Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
### Indications for Use
510(k) Number:
Device Name: Solarys* Ventilator Aerosol Delivery System
K123614
Indications for Use:
Solarys* Ventilator Aerosol Delivery System (AS) is a nebulizer intended to deliver prescribed aerosolized medication to mechanically ventilated pediatric and adults patients able to use an Endotracheal Tube equal to or greater than 4.5 mm in diameter. The device is designed to operate in-line with ventilator circuits and mechanical ventilators in acute and sub-acute care environments.
Prescription Use: (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
CONCURRENCE OF CDRH, OFFICE OF DEVICE EVALUATION (ODE) Albert E. Moyal: 5 Jan 1992 - 1992 - 1994 - 1994 - 1994 - 1994 - 1994 - 1994 - 1 for LS
'Division Sign-Off) rision of Anestheslology, General Hospital ection Control, Dental Devices
10(k) Number:
.
Page 1 of 1
* Trade marks and registered trademarks of Trudell Medical International 11:00:00 14
: : : : , , , , , , , , , , , , , , , , , , , , , , , , , , , , , , , , , , , , , , , , , , , , , , , , , , , , , , , , , , , , , , , , , , , , , , , , , , , , , , , , , , ,
:::
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.