U-RIGHT Compressor Nebulizer, model TD-7013/ TD-7012, is designed to provide a compressed air source to aerosolize physician-prescribed liquid medication when used in combination with the packaged Nebulizer kit, except for Pentamidine. The packaged Nebulizer kit is intended for single use by single patient. U-RIGHT Compressor Nebulizer, model TD-7013/ TD-7012, is intended for use with children, adolescents ( 2 years to 18 years old), and adult patients in the homecare settings.
Device Story
U-RIGHT Compressor Nebulizer (models TD-7013/TD-7012) is a portable aerosol delivery device for home use. It utilizes a pneumatic piston compressor to generate pressurized air, which is directed into a nebulizer kit containing liquid medication. The airflow disperses the medication into aerosol particles for patient inhalation. Operated by patients or caregivers, the device provides a compressed air source for respiratory therapy. Clinical benefit is derived from the effective delivery of prescribed liquid medications to the respiratory tract. The device is a mechanical compressor system; it does not involve complex software or AI-driven decision support.
Clinical Evidence
Bench testing only. Aerosol characterization via cascade impactor compared the device to the predicate using three drugs (Ipratropium bromide, Ventolin, Cromolyn sodium), showing no significant difference (p>0.05) in particle distribution. Gas path analysis confirmed no toxic emissions (CO, CO2, ozone, VOCs) and compliance with PM2.5 standards. Biocompatibility testing (ISO 10993-3, -5, -6, -10) and electrical safety/EMC testing (EN 60601-1, CISPR 11) were performed.
Technological Characteristics
Pneumatic piston compressor; portable form factor. Gas path components tested for biocompatibility per ISO 10993. Electrical safety/EMC compliance with EN 60601-1 and CISPR 11. No software-based algorithm or connectivity features.
Indications for Use
Indicated for children (2-18 years) and adults requiring aerosolized physician-prescribed liquid medication (excluding Pentamidine) via a compressor nebulizer in homecare settings.
Regulatory Classification
Identification
A nebulizer is a device intended to spray liquids in aerosol form into gases that are delivered directly to the patient for breathing. Heated, ultrasonic, gas, venturi, and refillable nebulizers are included in this generic type of device.
{1}------------------------------------------------
## B2. 510(k) Summary
#### 510(K) SUMMARY
This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of 21 CFR 807.92.
K121969 The Assigned 510(k) number is:
- 1. Submitter's Identification: TaiDoc Technology Corporation 3F,5F, No.127, Wugong 2nd Rd., Wugu District, New Taipei City, 24888, Taiwan
Correspondence: Pinjung Chen Regulatory Affairs Specialist Tel: +886-2-6625-8188 #1176 Fax: +886-2-6625-0288 Email: pinjung.chen(@taidoc.com.tw
Date of submission: June 30th, 2012
- 2. Device name:
Proprietary name: U-RIGHT Compressor Nebulizer, model TD-7013/ TD-7012
Regulatory information:
- A. Regulation section: 21 CFR § 868.5630
- B. Classification: Class II
- C. Product Code: CAF, nebulizer (direct patient interface)
- D. Panel: Anesthesiology
- 3. Intended Use:
- 4. U-RIGHT Compressor Nebulizer, model TD-7013/ TD-7012, is designed to provide a compressed air source to aerosolize physician-prescribed liquid medication when used in combination with the packaged Nebulizer kit, except for Pentamidine. The packaged Nebulizer kit is intended for single use by single patient. U-RIGHT
{2}------------------------------------------------
Compressor Nebulizer, model TD-7013/ TD-7012, is intended for use with children, adolescents ( 2 years to 18 years old), and adult patients in the homecare settings.
## 5. Device Description:
U-RIGHT Compressor Nebulizer is a light weight portable aerosol nebulizer, which uses a pneumatic piston that compresses air, forcing it to flow into the nebulizer container. And the force of the air flowing into the nebulizer container disperses the liquid medicine into aerosol particles for inhalation treatment of a physician's prescription medicine.
The only difference between model TD-7012 and TD-7013 is the outer casing design. The operating principle, the compressor type, working mechanism, and the accessories Nebulizer kit are all the same in these two models.
## 6. Substantial Equivalence Information:
A. Predicate device name: Bestneb Portable Aerosol Therapy Unit.
Model AP-10010
- B. Predicate K number: K990435
- C. Comparison with predicate:
The modified U-RIGHT Compressor Nebulizer has the following similarities to the predicate device:
. Same intended use
■ Same patient population and same environment of use
1 Same operating principle
- Same working mechanism 1
- Same fundamental scientific technology 트
The differences encompass:
- 트 Physical appearance of unit and components
- . Slight difference in storage condition
- Device weight 1
- Power requirement and power consumption
{3}------------------------------------------------
- 7. Summary of Performance Testing:
## 1) Aerosol Characterization Testing
The particle size distribution test via Cascade Impactor of U-RIGHT Compressor Nebulizer was performed in comparison to the predicate device K990435 with three drugs (Ipratropium bromide, Ventolin, and Cromolyn sodium). The test has shown U-RIGHT Compressor Nebulizer consistent in repeatability tests for each three classes of drug, and demonstrated equivalent performance to the predicate device K990435 that no significant difference (p>0.05) in the particle distributions.
U-RIGHT Compressor Nebulizer has the same performance characteristics as the predicate device and meets its product specification as well.
## 2) Airpath Testing
Gas sample analysis of U-RIGHT Compressor Nebulizer has performed and shown the device does not emit potential toxic gases that may cause harmful influences to human, including carbon monoxide, carbon dioxide, ozone, or volatile organic compounds (VOCs). The output of particulate matter conformed to EPA requirements of the PM2.5 standard.
#### 3) Materials
FDA authority considers the gas path contact components the external communicating components with tissue contact. Biocompatibility evaluations of cytotoxicity (ISO 10993-5), irritation (ISO 10993-10), sensitization (ISO 10993-10), implantation (ISO 10993-6) and genotoxicity (ISO 10993-3) tests have been conducted for the gas path contact materials.
#### 4) Safety and EMC
The electromagnetic compatibility and electric safety of the proposed device are tested to meet the following standards:
- 트 EN 60601-1:2006
- EN 60601-1-1-2:2007, CISPR 11: 2009+A1:2010
### 8. Conclusion:
Based on the information provided in this submission, U-RIGHT Compressor
{4}------------------------------------------------
Nebulizer is substantially equivalent to the predicate device in safety and effectiveness. :
.
:
.
1. 1. 1. 1.
.
{5}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized eagle or bird-like symbol with three overlapping, curved lines forming its body and wings. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" is arranged in a circular fashion around the symbol.
#### DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
February 28, 2013
Ms. Pinjung Chen Regulatory Affairs Specialist TaiDoc Technology Corporation 3F,5F, No.127, Wugong 2nd Road Wugu District New Taipei City, Taiwan 24888
Re: K121969
Trade/Device Name: U-RIGHT Compressor Nebulizer, Model TD-7013/ TD-7012 Regulation Number: 21 CFR 868.5630 Regulation Name: Nebulizer Regulatory Class: II Product Code: CAF Dated: January 15, 2013 Received: January 22, 2013
Dear Mr. Chen:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{6}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely vours.
Kwame O. Ulmer for
Anthony D. Watson, B.S., M.S., M.B.A. Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{7}------------------------------------------------
#### B1. Indications for Use Statement
## Indications for Use
# 510(k) Number:
Device Name: U-RIGHT Compressor Nebulizer, Model TD-7013/ TD-7012
Indications for Use:
U-RIGHT Compressor Nebulizer, model TD-7013/ TD-7012, is designed to provide a compressed air source to aerosolize physician-prescribed liquid medication when used in combination with the packaged Nebulizer kit, except for Pentamidine. The packaged Nebulizer kit is intended for single use by single patient. U-RIGHT Compressor Nebulizer, model TD-7013/ TD-7012, is intended for use with children, adolescents ( 2 years to 18 years old), and adult patients in the homecare settings.
Prescription Use _ X
(21 CFR Part 801 Subpart D)
And/Or Over the Counter Use (21 CFR Part 801 Subpart C)
# (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Albert E
Moyal-S
c=US, o=U.S. Government,
ou=HHS, ou=FDA, ou=People,
0.9.2342.19200300.100.1.1=13
00059331, cn=Albert E. Moyal
2013.02.28 11:43:43 -05'00'
heslotogy, General Hospital
**510(k) Number:** k121969
BI- Page 1 of 1
Page 1 of ____________________________________________________________________________________________________________________________________________________________________
{8}------------------------------------------------
## B1. Indications for Use Statement
## Indications for Use
# 510(k) Number: x | 21969
Device Name: U-RIGHT Compressor Nebulizer, Model TD-7013/ TD-7012
Indications for Use:
U-RIGHT Compressor Nebulizer, model TD-7013/ TD-7012, is designed to provide a compressed air source to aerosolize physician-prescribed liquid medication when used in combination with the packaged Nebulizer kit, except for Pentamidine. The packaged Nebulizer kit is intended for single use by single patient. U-RIGHT Compressor Nebulizer, model TD-7013/ TD-7012, is intended for use with children, adolescents ( 2 years to 18 years old), and adult patients in the homecare settings.
Prescription Use __ X (21 CFR Part 801 Subpart D)
Over the Counter Use (21 CFR Part 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
And/Or
Concurrence of CDRH, Office of Device Evaluation (ODE)
c=US, o=U.S. Government,
ou=HHS, ou=FDA, ou=People,
0.9.2342.19200300.100.1.1=13
00059331,cn=Albert E. Moyal
Albert EFDA
Moyal S
112
2013.02.28 11:43:43 -05'00'
for LS
estology, General Hospital ol. Dental Device
**510(k) Number:** k121969
B1- Page 1 of 1
Page 1 of l
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.