MERITS MODEL N SERIES NEBULIZER COMPRESSORS

K092859 · Merits Health Products Co., Ltd. · CAF · Jul 2, 2010 · Anesthesiology

Device Facts

Record IDK092859
Device NameMERITS MODEL N SERIES NEBULIZER COMPRESSORS
ApplicantMerits Health Products Co., Ltd.
Product CodeCAF · Anesthesiology
Decision DateJul 2, 2010
DecisionSESE
Submission TypeAbbreviated
Regulation21 CFR 868.5630
Device ClassClass 2
AttributesTherapeutic, Pediatric

Indications for Use

The Merits Model N Series Nebulizer Compressors are AC-powered air compressors intended to provide a source of compressed air for medical purposes for use in home health care. These devices are used in conjunction with pneumatic nebulizers to produce fine aerosol mists of medication for respiratory therapy, for both children and adults suffering from respiratory disorders such as asthma, allergies, etc.

Device Story

Merits Model N Series Nebulizer Compressors are AC-powered, piston-type air compressors designed for home health care. The device consists of a motor-driven piston housed in a plastic cabinet with rubber bumpers. Operation involves connecting the compressor to a power source, attaching air delivery tubing and a nebulizer kit to the outlet hose barb, and activating the switch. The compressor generates compressed air, which passes through a replaceable filter and is delivered to the nebulizer kit to convert liquid medication into a fine aerosol mist for patient inhalation. The device is operated by the patient or caregiver. It provides respiratory therapy for children and adults with respiratory conditions like asthma or allergies. The device contains no microprocessors or electronic components.

Clinical Evidence

Bench testing only. Evidence includes comparative performance testing against the predicate (SanUp Model 3050), electrical safety testing (IEC/UL 60601-1), electromagnetic compatibility testing (IEC 60601-1-2), emission testing for VOCs, CO, O3, and PM2.5, and cascade impaction testing for particle size analysis.

Technological Characteristics

AC-powered (115VAC/60Hz) motor-driven piston compressor. Housed in plastic cabinet with rubber bumpers. Includes tubing, nebulizer kit, and replaceable filters. No microprocessors or electronic components. Operates via mechanical compression of air.

Indications for Use

Indicated for children and adults suffering from respiratory disorders, including asthma and allergies, requiring aerosolized medication therapy via pneumatic nebulizer in home health care settings.

Regulatory Classification

Identification

A nebulizer is a device intended to spray liquids in aerosol form into gases that are delivered directly to the patient for breathing. Heated, ultrasonic, gas, venturi, and refillable nebulizers are included in this generic type of device.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ 092859 page 1 of 2 #### 9.0 SUMMARY OF SAFETY AND EFFECTIVENESS #### 9.1 Submitter: ## "510(k) SUMMARY" JUL -- 2 2010 Steve Chao Merits Health Products Co., LTD. 9. Road 36. Taichung Industrial Park Taichung, Taiwan R.O.C. Tel: 886-4-2359-4985 ext. 240 Fax: 886-4-2359-4992 E-mail: stevechao@merits.com.tw | 9.2 | Date of Preparation: | April 13, 2010 | |-----|-------------------------|--------------------------------------------| | 9.3 | Trade/Proprietary Name: | Merits Model N Series Nebulizer Compressor | | 9.4 | Common/Usual Name: | Nebulizer Compressor | Classification Name: Portable Air Compressor તે. ક #### Comparison to Currently Marketed Devices તે છે. કિ Both the Merits Model N Series Nebulizer Compressors and the predicate devices are AC-powered, contain the same filter material, meet Environmental, Safety and EMC requirements, and are in the similar compressor operating pressure and flow range. Performance characteristics are basically the same. That is, the Merits Model N Series Nebulizer Compressors are substantially equivalent to the currently marketed SAN UP S.A. Nebulizer Compressor Model: 3050 (K002468). #### 9.7 Device Description The Merits Model N Series Nebulizer Compressors consist of AC powered piston-type compressors that generates compressed air. They are small, lightweight and designed for convenience. The piston-type compressor is housed in a plastic cabinet with rubber bumpers. The motor driven piston compressor is connected with a power cord and a switch. The devices contain no microprocessors or other electronic components. They operate from 115VAC/60Hz. Each of them is supplied with tubing, a nebulizer kit and several replaceable filters. Model N281 and N282 are identical in specifications and components. The only difference is they use different cabinets which are different in size and shape. When using, the compressor is placed on a flat surface. Air delivery tubing and a nebulizer kit are connected to the outlet hose barb. The air passes the replaceable filter is compressed and delivered to the nebulizer kit through the tubing. {1}------------------------------------------------ #### 9.8 intended Use The Merits Model N Series Nebulizer Compressors are AC-powered air compressors intended to provide a source of compressed air for medical purposes for use in home health care. These devices are used in conjunction with pneumatic nebulizers to produce fine aerosol mists of medication for respiratory therapy, for both children and adults suffering from respiratory disorders such as asthma, allergies, etc. #### ರಿ ರಿ Technological Characteristics The Merits Model N Series Nebulizer Compressors operates by using AC-powered motor-driven piston compressor to drive liquid medication via Nebulizer. Compressor type nebulizer has been in use on portable model for many years. Technologies utilized by the Model N Series Nebulizer Compressors bring forth no new questions of safety and effectiveness. These technologies are well established and have been used in other legally marketed products. There are no major technologies differences. ### 9.10 Performance Data The Performance & Safety tests performed are summarized in the following. - a. Comparative Performance Test of Merits N281/N282 Nebulizer Compressor & SanUp Model 3050 Compressor - b. IEC/UL 60601-1, Medical Electrical Equipment Part1: General Requirements for Safety, - c. IEC 60601-1-2, Medical Electrical Equipment Part 1-2: General Requirements for Safety - Collateral Standard: Electromagnetic Compatibility, - d. Emission of VOCs, CO, 03 & PM2.5 - e. Cascade Impaction Testing for Particle Size Analysis The results of the testing confirm that the device meets specifications and is substantially equivalent to the predicate device. #### 9.11 Conclusion Based on the design, performance specifications and intended use, the Merits Model N Series Nebulizer Compressors are substantially equivalent to the currently marketed device. {2}------------------------------------------------ Image /page/2/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle or bird-like figure with three curved lines forming its body and wings. The logo is surrounded by text that reads "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" in a circular arrangement. Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002 JUL - 2 2010 Mr. Steve Chao Manager Merits Health Products Company, Limited 9, Road 36, Taichung Industrial Park Taichung China TAIWAN 407 Re: K092859 Trade/Device Name: Merits Model N Series Nebulizer Compressor Regulation Number: 21 CFR 868.5630 Regulation Name: Nebulizer Regulatory Class: II Product Code: CAF Dated: June 28, 2010 Received: July 2, 2010 Dear Mr. Chao: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {3}------------------------------------------------ Page 2- Mr. Chao Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Sincerely yours, Wh for Anthony D. Watson, B.S., M.S., M.B.A. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {4}------------------------------------------------ # Indications for Use 510(k) File Number: Device Name: Merits Model N Series Nebulizer Compressor Indications for Use: The Merits Model N Series Nebulizer Compressors are AC-powered air compressors intended to provide a source of compressed air for medical purposes for use in home health care. These devices are used in conjunction with a pneumatic nebulizers to produce fine aerosol mists of medication for respiratory therapy, for both children and adults suffering from respiratory disorders such as asthma, allergies, etc. Prescription Use V (Part 21 CFR 801 Subpart D) AND/OR Over-The-Counter Use (Part 21 CFR 801 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) (Division Sign-Off) Division of Anesthesiology, General Hospital Infection Control, Dental Devices 510(k) Number: K092859
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...