QFB · Isocapnic Ventilation Device

General Hospital · 21 CFR 868.5480 · Class 2

Overview

Product CodeQFB
Device NameIsocapnic Ventilation Device
Regulation21 CFR 868.5480
Device ClassClass 2
Review PanelGeneral Hospital

Identification

An isocapnic ventilation device is a prescription device used to administer a blend of carbon dioxide and oxygen gases to a patient to induce hyperventilation. This device may be labeled for use with breathing circuits made of reservoir bags (§ 868.5320), oxygen cannulas (§ 868.5340), masks (§ 868.5550), valves (§ 868.5870), resuscitation bags (§ 868.5915), and/or tubing (§ 868.5925).

Classification Rationale

Class II (special controls). The special controls for this device are:

Special Controls

In combination with the general controls of the FD&C Act. the isocapnic ventilation device is subject to the following special controls:

*Classification.* Class II (special controls). The special controls for this device are:(1) Nonclinical performance testing data must demonstrate that the device performs as intended under anticipated conditions of use, including the following performance characteristics: (i) Gas concentration accuracy testing for the range of intended concentrations; (ii) Airway pressure delivery accuracy testing; (iii) Supplemental O 2 flowrate accuracy testing;(iv) Alarm testing; and (v) Use life testing. (2) The patient-contacting components of the device must be demonstrated to be biocompatible. (3) Labeling must include the following: (i) Instructions for use; (ii) A precaution that monitoring of capnography is necessary during treatment with non-spontaneously breathing patients; and (iii) Use life specification.

Recent Cleared Devices (1 of 1)

RecordDevice NameApplicantDecision DateDecision
DEN170044ClearMateThornhill Research, Inc.Mar 14, 2019DENG

Top Applicants

Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...