The FSC Anti-Static Valved Collapsible Holding Chamber is intended to be used by patients who are under the care or treatment of a licensed health care professional. The device is intended to be used by these patients to administer aerosolized medication from most pressurized Metered Dose Inhalers, (pMDIs). Environment of use - Home, hospitals and clinics. This product is intended for patients who can follow verbal instructions.
Device Story
The FSC Anti-Static Valved Collapsible Holding Chamber is a non-sterile, single-patient, multi-use spacer device used in homes, hospitals, and clinics. It acts as an interface between a pressurized Metered Dose Inhaler (pMDI) and the patient's respiratory system. The device features a collapsible housing, an integral mouthpiece, and a one-way valve that prevents the patient from exhaling into the chamber. To operate, the patient attaches the pMDI to the chamber, actuates the inhaler, and inhales the aerosolized medication through the mouthpiece. The collapsible bag serves as a visual indicator of inhalation progress. By reducing the velocity of the aerosol and providing a holding space, the device facilitates the delivery of corticosteroids, anti-cholinergics, beta-2-sympathomimetics, and non-steroidal chromones to the upper and lower respiratory tracts. The anti-static properties of the chamber help ensure consistent medication delivery by minimizing drug adherence to the device walls. The device is intended for patients who can follow verbal instructions, aiding in effective medication administration and potentially improving therapeutic outcomes.
Clinical Evidence
Bench testing only. Performance was evaluated using an 8-stage cascade impactor at 28 lpm and 12 lpm flow rates with three different MDI drugs (Ventolin HFA, Atrovent HFA, QVAR 40). Metrics included Total Delivered Dose, Total Respirable Dose (0.5-5.0 microns), MMAD, and GSD. Results showed equivalent performance to the predicate. Additional testing included simulated lifetime cycle, cleaning validation, environmental/mechanical stress (high/low temp, drop test), differential pressure, and anti-static surface resistivity. Biocompatibility testing (cytotoxicity, sensitization, genotoxicity, extractables/leachables) was performed per ISO 10993-1.
Technological Characteristics
Collapsible housing with integral mouthpiece; one-way inhalation valve; anti-static material construction. Designed for prolonged (>30 days) mucosal surface contact. Tested per ISO 10993-1 for biocompatibility. Operates via passive inhalation; no external energy source. Dimensions/form factor collapsible for portability.
Indications for Use
Indicated for patients under the care of a licensed healthcare professional capable of following verbal instructions for the administration of aerosolized medication from pressurized Metered Dose Inhalers (pMDIs).
Regulatory Classification
Identification
A nebulizer is a device intended to spray liquids in aerosol form into gases that are delivered directly to the patient for breathing. Heated, ultrasonic, gas, venturi, and refillable nebulizers are included in this generic type of device.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized eagle with three heads in profile, facing right. The eagle is enclosed in a circle with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" around the perimeter.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
October 3, 2014
FSC Laboratories, Inc. C/O Mr. Paul Dryden Promedic, Inc., President Regulatory Consultant for FSC Laboratories 6000 Fairview Rd., Suite 600 Charlotte, NC 28210
Re: K140062
Trade/Device Name: FSC Anti-Static Valved Collapsible Holding Chamber Regulation Number: 21 CFR 868.5630 Regulation Name: Nebulizer Regulatory Class: II Product Code: NVP Dated: September 4, 2014 Received: September 4, 2014
Dear Mr. Dryden:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Page 2 - Mr. Dryden
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
*Tejashri Purohit-Sheth, M.D.*
Tejashri Purohit-Sheth, M.D. Clinical Deputy Director DAGRID/ODE/CDRH FOR
Erin I. Keith, M.S Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
#### Indications for Use Statement 4
We have prepared the Indications for Use statement utilizing Form 3881 which follows.
{3}------------------------------------------------
#### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration Indications for Use
Form Approved: OMB No. 0910-0120 Expiration Date: December 31, 2013 See PRA Statement on last page.
510(k) Number (if known)
Device Name
FSC Anti-Static Valved Collapsible Holding Chamber
#### Indications for Use (Describe)
The FSC Anti-Static Valved Collapsible Holding Chamber is intended to be used by patients who are under the care or treatinent of a licensed health care professional. The device is intended to be used by these patients to administer acrosolized medication from nost pressurized Metered Dose Inhalers, (pMDIs).
Environment of use - Home, hospitals and clinics.
This product is intended for patients who can follow verbal instructions.
Type of Use (Select one or both, as applicable) [x] Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.
FOR FDA USE ONLY CONLY CONSTITUTION OF
Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
FORM FDA 3881 (9/13)
Page 1 of 2
PSC Publishing Services (301) 443-6740
{4}------------------------------------------------
#### 510(k) Summary Page 1 of 8
| Date Prepared: | 03-Sep-2014 | |
|--------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------|--|
| FSC Laboratories, Inc.<br>6000 Fairview Rd., Suite 6000<br>Charlotte, NC 28210 | Tel - 704-941-2500<br>Fax - 866-289-0893 | |
| Official Contact: | Karla Worley Ham<br>Quality and Regulatory Affairs Officer | |
| Proprietary or Trade Name: | FSC Anti-Static Valved Collapsible Holding Chamber | |
| Common/Usual Name: | Spacer / Holding Chamber | |
| Classification Name: | Holding Chambers, Direct Patient Interface<br>Product Classification - NVP<br>21 CFR 868.5630<br>Class II | |
| Predicate Devices: | K052332 - Trudell - AeroChamber Plus Z-stat<br>K933090 - FSC Laboratories - E-Z Spacer | |
#### Device Description:
The FSC Anti-Static Valved Collapsible Holding Chamber is intended for use in the inhalation of medications delivered via an MDI and for which the medication is to be delivered to the upper and lower respiratory system. The device consists of a collapsible housing and mouth piece and a one-way valve to prevent exhaling into the chamber.
The FSC Anti-Static Valved Holding Chamber is intended to inhale aerosolized drugs of approved MDIs from the following groups of active substances:
- · Corticosteroids (anti-inflammatory medications)
- · Anti-cholinergics and ß2-sympathomimetics (bronchodilator medications)
- · Non-steroidal chromones (DNCG)
It is a single patient, multi-use, non-sterile device.
#### Indications for Use:
The FSC Anti-Static Valved Collapsible Holding Chamber is intended to be used by patients who are under the care or treatment of a licensed health care professional. The device is intended to be used by these patients to administer aerosolized medication from most pressurized Metered Dose Inhalers, (pMDIs).
Environment of use - Home, hospitals and clinics.
This product is intended for patients who can follow verbal instructions.
#### 510(k) Summary
{5}------------------------------------------------
#### Page 2 of 8 03-Sep-2014
## Comparison to Predicates
We have chosen two (2) predicates for our substantial equivalence claim. The following is a rationale for this selection.
AeroChamber Plus Z-stat (K052332) and E-Z Spacer (K933090)
Table 1 is a table which highlights the reason for selecting each predicate.
| SE Element | AeroChamber<br>Plus Z-stat (K052332) | E-Z Spacer<br>(K933090) |
|-----------------------------------|--------------------------------------|--------------------------------------|
| Indications for Use | X | X |
| Environment of Use | X | X |
| Patient Population | X | X |
| Technology of collapsible chamber | | X |
| Feature inhalation progress | X | X |
| | Has a whistle | Bag collapses as visual<br>indicator |
| Anti-static feature | X | |
| Design and usability | | X |
## Table 2 - Comparison to Predicate - AeroChamber Plus Z-Stat (K052332)
| Attribute | AeroChamber Plus Z-stat<br>K052332 | Proposed<br>FSC Chamber |
|---------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Intended Use | For use with MDIs | For use with MDIs |
| Indications for Use | Intended to be used by patients who<br>are under the care or treatment of a<br>licensed health care provider or<br>physician. The device is intended to<br>be used by these patients to<br>administer aerosolized medication<br>from most pressurized Metered<br>Dose Inhalers, prescribed by a<br>physician or health care<br>professional. | The FSC Anti-Static Valved<br>Collapsible Holding Chamber is<br>intended to be used by patients who are<br>under the care or treatment of a<br>licensed health care professional. The<br>device is intended to be used by these<br>patients to administer aerosolized<br>medication from most pressurized<br>Metered Dose Inhalers, (pMDIs).<br>Environment of use - Home, hospitals<br>and clinics.<br>This product is intended for patients<br>who can follow verbal instructions. |
| Environments of use | Home, hospitals and clinics. | Home, hospitals and clinics |
| Prescriptive | Yes | Yes |
| Patient population | All - not specified | Intended for patients who can follow<br>verbal instructions |
{6}------------------------------------------------
# 510(k) Summary Page 3 of 8
03-Sep-2014
| Attribute | AeroChamber Plus Z-stat<br>K052332 | Proposed<br>FSC Chamber |
|-------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Single patient, multi-use | Yes | Yes |
| Patient interface | Mouthpiece<br>Face Mask | Integral Mouthpiece |
| Basic components | Housing<br>One-way valve to prevent<br>exhalation into chamber<br>End caps - removable | Housing<br>One-way valve to prevent<br>exhalation into chamber<br>End caps - removable |
| Anti-static properties | Yes | Yes |
| Housing | Rigid | Collapsible |
| Indicator of inhalation | Whistle to alert user | Visual collapse of the bag |
| Single patient, multi-use | Yes | Yes |
| Materials Patient<br>Contact | Materials in the gas pathway<br>External Communicating<br>Tissue<br>Prolonged<br><br>Materials in direct patient contact<br>Surface Contact<br>Mucosal<br>Prolonged | Materials in the gas pathway<br>External Communicating<br>Tissue<br>Prolonged Duration of Use<br><br>Materials in direct patient contact<br>Surface Contact<br>Mucosal<br>Prolonged Duration of Use |
| Performance testing | Particle characterization<br>Comparison results found to be<br>equivalent | Particle characterization<br>Mechanical<br>Environmental<br>Simulated lifetime cycle (cleaning)<br>Differential Pressure<br>ISO 10993 testing |
| Attribute | E-Z Spacer<br>K933090 | Proposed<br>FSC Chamber |
| Intended Use | For use with MDIs | For use with MDIs |
| Indications for Use | Intended to be used by patients who<br>are under the care or treatment of a<br>licensed health care provider or<br>physician. The device is intended to<br>be used by these patients to<br>administer aerosolized medication<br>from most pressurized Metered<br>Dose Inhalers, prescribed by a<br>physician or health care<br>professional. (Not specified in FDA<br>database) | The FSC Anti-Static Valved<br>Collapsible Holding Chamber is<br>intended to be used by patients who are<br>under the care or treatment of a<br>licensed health care professional. The<br>device is intended to be used by these<br>patients to administer aerosolized<br>medication from most pressurized<br>Metered Dose Inhalers, (pMDIs).<br>Environment of use - Home, hospitals<br>and clinics.<br>This product is intended for patients<br>who can follow verbal instructions. |
| Environments of use | Home, hospitals and clinics. | Home, hospitals and clinics |
| Prescriptive | Yes | Yes |
| Patient population | All - not specified | This product is intended for patients<br>who can follow verbal instructions |
| Single patient, multi-<br>use | Yes | Yes |
| Patient interface | Integral Mouthpiece | Integral Mouthpiece |
| Basic components | Housing<br>End caps - removable | Housing<br>One-way valve to prevent exhalation<br>into chamber<br>End caps - removable |
| Anti-static properties | No | Yes |
| Housing | Collapsible | Collapsible |
| Indicator of inhalation | Visual collapse of the bag | Visual collapse of the bag |
| Single patient, multi-<br>use | Yes | Yes |
| Materials Patient<br>Contact | Materials in the gas pathway<br>External Communicating<br>Tissue<br>Prolonged<br>Materials in direct patient contact<br>Surface Contact<br>Mucosal | Materials in the gas pathway<br>External Communicating<br>Tissue<br>Prolonged<br>Materials in direct patient contact<br>Surface Contact<br>Mucosal |
| | Prolonged | Prolonged |
{7}------------------------------------------------
# 510(k) Summary Page 4 of 8
03-Sep-2014
## Table 3 - Comparison to Predicate - E-Z Spacer (K933090)
{8}------------------------------------------------
510(k) Summary Page 5 of 8 03-Sep-2014
#### Substantial Equivalence Discussion
Tables 2 to 3 above compare the key features of the proposed FSC Anti-static Valved Collapsible Holding Chamber with the identified predicates and demonstrates that the device can be found to be substantially equivalent.
In summary one can conclude that substantial equivalence is met based upon the following:
#### Indications for Use -
The indications for use are identical for the proposed device when compared to the predicate -K052332 - AeroChamber Plus Z-Stat. The predicate K933090 - E-Z spacer does not have published indications but is known to be similar.
Discussion - Each device is indicated for use with MDIs of the same category of medications.
#### Technology and construction -
The design, fabrication, shape, size, etc. are equivalent to the predicate - K933090 - E-Z Spacer which is a collapsible chamber with integral mouthpiece.
Discussion - This design incorporates a collapsible housing, end caps, one way valve for inhalation, and patient interface of a mouthpiece which are the same as the predicate.
We are offering Anti-static properties for the whole device, which is identical to the predicate K052332 - AeroChamber Plus Z-Stat.
Discussion - We have performed testing to support the anti-static claim.
#### Environment of Use -
The environments of use are identical to both predicates - K052332 - AeroChamber Plus Z-Stat and K933090 - E-Z spacer.
Discussion - The environments of use are identical to both predicates - K052332 - AeroChamber Plus Z-Stat and K933090 - E-Z spacer.
#### Patient Population -
The patient population are those who are able to follow verbal instructions. This is similar to the predicates which do not specify the patient population, however they offer face mask for pediatric patients.
Discussion - The patient populations are equivalent to both predicates - K052332 -AeroChamber Plus Z-Stat and K933090 - E-Z spacer.
#### Non-clinical Testing Summary -
#### Particle Characterization -
We performed comparative particle characterization testing via Cascade Impactor and the results demonstrated equivalent performance concluding that the proposed device is equivalent to the predicate K052332 - AeroChamber Plus Z-Stat.
{9}------------------------------------------------
#### 510(k) Summary Page 6 of 8 03-Sep-2014
In addition, we performed testing related to simulated life testing, cleaning validation, environmental and mechanical, differential pressure, and anti-static property testing. The results demonstrated that the proposed device either passed or met its performance specifications after each test and where appropriate was equivalent or better than the predicate, K052332 -AeroChamber Plus Z-Stat.
#### Materials -
We have performed ISO 10993-1 testing on the component materials of the FSC Anti-Static Valved Collapsible Holding Chamber which is considered as having two(2) patient contact classifications, External Communicating and Surface Contact. Per ISO 10993-1 and G95-1 the required tests for the duration of use are the same. The following are details.
- External Communication (Indirect contact) for all materials not in direct contact .
- Tissue communicating .
- · Permanent duration (>30 days)
For those materials in direct contact, the mouthpiece, it is considered
- · Surface Contact
- · Mucosal membrane
- · Permanent duration (> 30 days)
We performed: cytotoxicity, sensitization, genotoxicity, and exhaustive leachable and extractable testing and included a risk assessment.
Discussion - We have tested the materials which are common to chambers.
#### Performance Testing including Comparative Testing:
We performed comparative particle characterization testing via Cascade Impactor and the results demonstrated equivalent performance between the proposed device and the predicate. This testing included:
- Particle Characterization testing via Cascade Impactor .
- 0 Adult 28 lpm
- o Pediatric 12 lpm
- o Intra- and Inter-sample variance
- Simulated lifetime testing .
- Pre and post- exposure o
- Cleaning o
- Environmental and mechanical testing (part of Simulated Life Cycle testing) ●
- o High and Low temperature
- o Drop test
- Anti-static surface resistivity ●
- Differential Pressure comparative ●
{10}------------------------------------------------
#### 510(k) Summary Page 7 of 8 03-Sep-2014
A series of aerosol performance tests were performed using an 8 stage cascade impactor at sampling flow rates of 28 lpm and 12 lpm, equipped with a USP <601> induction port throat. Aerosol was sampled directly from the outlet. A summary of the results is listed below with 95% confidence intervals.
@ 28 lpm - 3 samples of the device were tested with 3 drugs, 3 times for a total of 9 sample points.
@ 12 1pm - 3 samples of the device were tested with 3 drugs, 3 times for a total of 9 sample points.
MDI only - 3 samples were tested with 3 drugs.
. '
Ventolin HFA Atrovent HFA QVAR 40
#### Table 4 -- Total Respirable Dose Delivered @ 28 lpm flow rate MDI vs. MDI-Spacer - 95% Confidence Intervals
| Total Respirable Dose Delivered (0.5-5.0 microns) ug/burst | | | |
|------------------------------------------------------------|--------------|--------------|-----------|
| | Ventolin HFA | Atrovent HFA | QVAR 40 |
| MDI only | 37.5-46.8 | 7.5-8.1 | 12.1-14.8 |
| MDI - Spacer | 50.6-55.0 | 6.3-8.3 | 10.7-15.4 |
### Table 5 - Total Dose Delivered @ 28 lpm flow rate MDI vs. MDI-Spacer - 95% Confidence Intervals
| Total Dose Delivered - ug/burst | | | |
|---------------------------------|---------------|--------------|-----------|
| | Ventolin HFA | Atrovent HFA | QVAR 40 |
| MDI only | 102.7 - 104.8 | 19.0-21.0 | 33.6-38.4 |
| MDI - Spacer | 61.3 - 66.1 | 13.2-15.4 | 26.4-34.1 |
#### Table 6 - MDI - Spacer with 3 Drugs @ 28 lpm
| | Ventolin<br>HFA | Atrovent HFA | Qvar 40 |
|-----------------------------------------------------|-----------------|--------------|-----------|
| Particle Size (MMAD)<br>(um) | 1.68-1.8 | 0.76-0.92 | 0.4-0.49 |
| Geometric Standard<br>Deviation (GSD) | 1.87-2.15 | 3.03-3.51 | 2.84-3.33 |
| Total Delivered Dose by<br>Device - ug / burst | 61.3-66.1 | 13.2-15.4 | 26.4-34.1 |
| Total Respirable Dose<br>(0.5 - 5 um) - ug/burst | 50.6-55.0 | 6.3-8.3 | 10.7-15.4 |
| Coarse Particle Dose<br>>4.7 microns - ug/burst | 9.0-10.1 | 3.1-4 | 1.8-2.3 |
| Fine Particle Dose<br><4.7 microns - ug/burst | 51.6-56.7 | 9.5-12.0 | 24.4-32.0 |
| Ultra-Fine Particle Dose<br><1.0 microns - ug/burst | 10.3-14.7 | 5.4-6.8 | 19.2-25.1 |
{11}------------------------------------------------
### 510(k) Summary Page 8 of 8 03-Sep-2014
| | Ventolin HFA | Atrovent HFA | Qvar 40 |
|-----------------------------------------------------|--------------|--------------|-----------|
| Particle Size (MMAD) (um) | 1.97-2.1 | 0.99-1.21 | 0.6-0.73 |
| Geometric Standard<br>Deviation (GSD) | 1.85-2.02 | 2.71-3.0 | 2.14-3.0 |
| Total Delivered Dose by<br>Device - ug / burst | 56.0-68.9 | 12.7-13.8 | 21.7-28.5 |
| Total Respirable Dose<br>(0.5 - 5 um) - ug/burst | 49.0-57.2 | 6.5-7.4 | 13.1-17.2 |
| Coarse Particle Dose<br>>4.7 microns - ug/burst | 8.2-13.2 | 3.3-3.5 | 0.8-1.9 |
| Fine Particle Dose<br><4.7 microns - ug/burst | 47.8-55.7 | 9.4-10.3 | 20.9-26.5 |
| Ultra-Fine Particle Dose <1.0<br>microns - ug/burst | 6.3-8.3 | 4.7-5.6 | 13.9-17.8 |
## Table 5.7 – MDI – Spacer with 3 Drugs @ 12 lpm
## Substantial Equivalence Conclusion -
The sponsor has demonstrated through performance testing, design and features, and non-clinical testing that the proposed device and predicate have been found to be substantially equivalent.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.