← Product Code [NHK](/submissions/AN/subpart-f%E2%80%94therapeutic-devices/NHK) · K092687

# VENTI.PLUS HYPERINFLATION BAG SYSTEM WITH PRESSURE MANOMETER (K092687)

_A Plus Medical · NHK · Nov 23, 2009 · Anesthesiology · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/AN/subpart-f%E2%80%94therapeutic-devices/NHK/K092687

## Device Facts

- **Applicant:** A Plus Medical
- **Product Code:** [NHK](/submissions/AN/subpart-f%E2%80%94therapeutic-devices/NHK.md)
- **Decision Date:** Nov 23, 2009
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 868.5905
- **Device Class:** Class 2
- **Review Panel:** Anesthesiology
- **Attributes:** Therapeutic, Pediatric

## Indications for Use

Single patient use non self-inflating manual resuscitator for use in hospital and transport to temporary ventilate neonate, infant and pediatric patients. The pressure manometer in this product provides a visual indication of airway pressure during ventilation.

## Device Story

Single patient use hyperinflation bag system; includes integrated 0-60 cm H2O pressure manometer. Clinician manually manipulates gas flow, respiratory rate, and exhalation port to control gas volume and inspiratory pressure delivered to patient. Used in hospital and transport settings for neonate, infant, and pediatric patients. Manometer provides visual airway pressure feedback to clinician, assisting in manual ventilation control and monitoring.

## Clinical Evidence

Bench testing only. No clinical data provided. Performance characteristics and manometer accuracy compared to predicate devices via bench testing.

## Technological Characteristics

Manual hyperinflation bag system; 0.25, 0.50, and 1.0 liter sizes. Integrated 0-60 cm H2O mechanical pressure manometer. Conical connectors compliant with ISO 5356-1:2004. Single patient use.

## Regulatory Identification

A noncontinuous ventilator (intermittent positive pressure breathing-IPPB) is a device intended to deliver intermittently an aerosol to a patient's lungs or to assist a patient's breathing.

## Predicate Devices

- SIMS Hyperinflation Bag System ([K970785](/device/K970785.md))
- 1st Response Disposable Manometer ([K961318](/device/K961318.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

(a) A Plus Medical
Solutions for Respiratory Care

510(k) Summary

K092687

NOV 2 8 2009

### Owner:

A Plus Medical 5431 Avenida Encinas, STE G Carlsbad, CA 92008-4411 + 760-930-4025 Tel: + 760-930-0040 Fax:

# Owner/Operator Number:

10023166

# Official Contact:

Thomas C. Loescher + 760-930-4025 Tel:

+ 760-930-0040 Fax:

# Trade Names:

VESSA.Plus" Hyperinflation Bag System with Pressure Manometer

# Common/Usual Name:

Hyperinflation Bag with Pressure Manometer

# Classification Name:

| Device Name:  | Resuscitator, Manual, Non Self-Inflating |
|---------------|------------------------------------------|
| Product Code: | NHK                                      |
| Regulation:   | CFR 868.5905                             |
| Device Class: | II                                       |
| Device Name:  | Monitor, Airway Pressure                 |
| Product Code: | CAP                                      |
| Regulation:   | CFR 868.2600                             |
| Device Class: | II                                       |

#### Device:

4.2

Venti*Plus Hyperinflation Bag System with Pressure Manometer

### Predicate Devices:

| Number:        | K970785                                                                                                                                              |
|----------------|------------------------------------------------------------------------------------------------------------------------------------------------------|
| Product Name:  | SIMS Hyperinflation Bag System                                                                                                                       |
| Manufacturer:  | Smiths Medical (Intertech Resources)                                                                                                                 |
| Product Codes: | 008330DM, 008330VM, 008332DM, 008335DM,<br>008430, 008430DM, 00843M, 008430T,,<br>008430VM, 008431T, 008432, 008432DM,<br>008432T, 008435 & 008435DM |
| Number:        | K961318                                                                                                                                              |
| Product Name:  | 1st Response Disposable Manometer                                                                                                                    |
| Manufacturer:  | Smiths Medical (Intertech Resources)                                                                                                                 |
| Product Codes: | 008201                                                                                                                                               |

5411 Avenida Encinas, STE G Carlsbad, CA 92008-4411 盆 +740-930-4025 Fax: +760-930-0040

{1}------------------------------------------------

Image /page/1/Picture/0 description: The image shows the logo for A Plus Medical. The logo features the letter A inside of a circle on the left. To the right of the circle is the text "A Plus Medical" in a bold font. Below the company name is the text "Solutions for Respiratory Care" in a smaller font.

### Device Description:

Single patient use hyperinflation bag with pressure manometer offered in 0.25, 0.50 and 1.0 liter ventilation bag sizes. End user manipulates gas flow and respiratory rate and exhalation port to control amount of gas, inspiratory pressure supplied to the patient. Each device has an integrated 0 - 60 cm H2O pressure manometer.

#### Indications for Use:

Single patient use non self-inflating manual resuscitator for use in hospital and transport to temporary ventilate neonate, infant and pediatic patients. The pressure manometer in this provides a visual indication of arway pressure during ventilation.

#### Contraindications:

None identified.

### Patient Population:

Patient populations of neonate, newborn and pediatric.

#### Environment of Use:

Hospital and patient transport

### Comparative of Technological Characteristics:

The A Plus Medical Ventit Plus Hyperinflation System with Pressure Manometer is substantially equivalent in indications for use, environment of use, patient population, material and function to the identified predicate. The A Plus Medical Ventif Plus Hyperinflation System with Pressure Manometer and identified predicate device meet the requirements set forth ISO 5356- 1:2004 entitled "Anaesthetic and respiratory equipment - Conical connectors: Part 1: Cones and sockets". Bench testing confirmed that the A Plus Medical Venti*Plus Hyperinflation System with Pressure Manometer and predicate device have similar performance characteristics and the accuracy of the integrated pressure manometer in the Venti*Plus Hyperinflation System with Pressure Manometer is equivalent to the predicate device when used with its associated pressure manometer.

#### Conclusion:

The Venti*Plus Hyperinflation System with Pressure Manometer is substantially equivalent to the identified predicates. The Ventif Plus Hyperinflation System with Pressure Manometer and the identified predicates have substantially equivalent performance.

The Ventif Plus Hyperinflation System with Pressure Manometer and the identified predicates are made from substantially equivalent material, intended use, patient populations and environment of use

> 5411 Avenida Encinas, STE G Carlsbad, CA 92008-4411 2. +740-930-4025 Fax: +760-930-0040 ~ 14~

{2}------------------------------------------------

Image /page/2/Picture/1 description: The image shows the seal of the U.S. Department of Health & Human Services. The seal features a stylized eagle-like symbol on the right. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" is arranged in a circular pattern around the symbol.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Room W-O66-0609 Silver Spring, MD 20993-0002

Thomas C. Loescher, R.R.T. President A Plus Medical 5431 Avenida Encinas, Suite G Carlsbad, California 92008-4411

NOV 2 3 2009

Re: K092687

Trade/Device Name: Venti.Plus™ Hyperinflation Bag System with Pressure Manometer Regulation Number: 21 CFR 868.5905 Regulation Name: Noncontinuous Ventilator (IPPB) Regulatory Class: II Product Code: NHK Dated: November 4, 2009 Received: November 10, 2009

Dear Mr. Loescher:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{3}------------------------------------------------

# Page 2- Mr. Loescher

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to

http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/cdrh/mdr/ for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers. International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely yours,

for

Susan Runner, D.D.S., M.A. Acting Division Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

### Enclosure

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image shows the logo for A Plus Medical. The logo features the letter 'a' inside of a parenthesis. The text 'A Plus Medical' is to the right of the parenthesis, and the text 'Solutions for Respiratory Care' is below 'A Plus Medical'.

# Indications for Use Statement

510(k) Number:

."

KD92689 (To be assigned)

Device Name: Indications for Use: VENEA. Plus" Hyperinflation Bag System with Pressure Manometer

Single patient use non self-inflating manual resuscitator for use in hospital and transport to temporary ventilate neonate, infant and pediatric patients. The pressure manometer in this product provides a visual indication of airway pressure during ventilation.

X Prescription Use (Part 21 CFR 801 Subpart D)

Over-the-counter use (21 CFR 807 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

or

Concurrence of CDRH, Office of Device Evaluation (ODE)

Y. Schutte

(Division Sign-Off) Division of Anesthesiology, General Hospital Infection Control, Dental Devices

510(k) Number: K092687

---

**Source:** [https://fda.innolitics.com/submissions/AN/subpart-f%E2%80%94therapeutic-devices/NHK/K092687](https://fda.innolitics.com/submissions/AN/subpart-f%E2%80%94therapeutic-devices/NHK/K092687)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
