← Product Code [NFB](/submissions/AN/subpart-f%E2%80%94therapeutic-devices/NFB) · K022865

# 6203 (K022865)

_Contemporary Products, Inc. · NFB · Feb 13, 2003 · Anesthesiology · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/AN/subpart-f%E2%80%94therapeutic-devices/NFB/K022865

## Device Facts

- **Applicant:** Contemporary Products, Inc.
- **Product Code:** [NFB](/submissions/AN/subpart-f%E2%80%94therapeutic-devices/NFB.md)
- **Decision Date:** Feb 13, 2003
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 868.5905
- **Device Class:** Class 2
- **Review Panel:** Anesthesiology
- **Attributes:** Therapeutic

## Indications for Use

The 6203 is indicated for use to conserve oxygen for patients prescribed 1 to 6 liters per minute of supplemental oxygen and use nasal cannulas and USP bottled oxygen.

## Device Story

6203 is a battery-operated, microprocessor-controlled oxygen conserving device. Installed between USP bottled oxygen supply and patient nasal cannula; device uses internal breath sensor to detect patient inhalation via 0.5cm negative pressure. Upon detection, microprocessor opens normally closed valve to deliver preset bolus of oxygen based on flow rate setting (1-6 lpm). Valve closes during end of inhalation and exhalation to conserve oxygen. Device features audible alarm for cannula disconnection or failure to detect inhalation within 45 seconds; LED indicators for battery status and flow fault/apnea. Used in clinical or home settings by patients prescribed supplemental oxygen. Output allows for reduced oxygen consumption compared to continuous flow, potentially extending oxygen supply duration.

## Clinical Evidence

Bench testing only. Testing included functional performance, environmental conditions (impact/drop, storage temperature), electromagnetic interference, electrostatic discharge, surface temperature, and power supply (battery life, low power indicator).

## Technological Characteristics

Battery-operated electronic device; microprocessor-controlled. Features breath sensor (diaphragm-based) and normally closed solenoid valve. Dimensions/form factor not specified. Connectivity: standalone. Sterilization: not specified. Software: microprocessor-based control logic.

## Regulatory Identification

A noncontinuous ventilator (intermittent positive pressure breathing-IPPB) is a device intended to deliver intermittently an aerosol to a patient's lungs or to assist a patient's breathing.

## Predicate Devices

- Contemporary Products, Inc., OCD2001 ([K992935](/device/K992935.md))
- Invacare IPD Oxygen Conserving Device ([K953852](/device/K953852.md))
- Chad Therapeutics Oxymatic Electronic Oxygen Conserver ([K852650](/device/K852650.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

K022865

# Demand Oxygen Conserving Device 510(k) Summary

FEB 1 3 2003

# Submitter's Name, Address, Telephone Number and Contact Person

| Submitter                      |              |  |
|--------------------------------|--------------|--|
| Barry A. Schwartz              |              |  |
| Contemporary Products, Inc.    |              |  |
| 470 Riverside Street           |              |  |
| Portland, ME 04103             |              |  |
| Telephone:                     | 207-797-4647 |  |
| Facsimile:                     | 207-797-9411 |  |
| FDA Establishment No.: 1223412 |              |  |

| <b>Contact Person</b>       |
|-----------------------------|
| Barry A. Schwartz           |
| Contemporary Products, Inc. |
| 470 Riverside Street        |
| Portland, ME 04103          |
| Telephone: 207-797-4647     |
| Facsimile: 207-797-9411     |

### Date Prepared

August 27, 2002

## Name of Device

| Trade Name:          | 6203                                                      |
|----------------------|-----------------------------------------------------------|
| Common Name:         | Oxygen Conserver                                          |
| Classification Name: | Ventilator, Non-Continuous (Respirator)<br>21CFR 868.5905 |
| Product Code:        | 73 NFB                                                    |

### Predicate Device

Contemporary Products, Inc., OCD2001 (K992935)

## Intended Use

The 6203 is indicated for use to conserve oxygen for patients prescribed 1 to 6 liters per minute of supplemental oxygen and use nasal cannulas and USP bottled oxygen.

{1}------------------------------------------------

## Technological Characteristics and Substantial Equivalence

The 6203 with integrated pressure regulator is intended to be used as an accessory to an oxygen supply system to reduce or conserve the amount of oxygen used by the patient. The 6203 is a battery operated electronic device that is microprocessor controlled and contains a breath sensor and a normally closed valve. When installed between the oxygen supply and patient's nasal cannula, the device detects the patient's inhalation, opens the valve according to the flow rate set on the device and delivers a preset bolus of oxygen to the patient as determined by the device flow rate algorithm. The valve closes and conserves the oxygen that would have been wasted during the end of inhalation and during exhalation.

The 6203 is intended to be used with USP bottled oxygen and nasal cannulas and is installed as shown in the following figure:

Image /page/1/Figure/3 description: This figure shows a diagram of an oxygen conservation device. The diagram includes a nasal cannula, an oxygen conserving device, and an oxygen supply. The figure is labeled as "Figure 1" and includes the number "1" and "23456" in the upper right corner.

The side panel of the 6203 has a selector switch, battery status light and flow fault/apnea light. The battery LED winks every two (2) seconds when about 3 hours remain in the batteries and once every second when about 30 minutes

{2}------------------------------------------------

remain. It then lights steady if batteries can no longer properly power the unit informing the patient that the unit has shut down.

When the selector switch is set to "ON" the device operates as follows:

The sensing diaphragm closes its circuit in response to .5cm negative pressure produced by the user's inhalation effort. This circuit closing is input into the microprocessor. The microprocessor then opens the valve to allow the appropriate bolus of oxygen to flow (based upon the chart that follows) and then waits for the next negative pressure.

Chart 1

| Liter Flow Setting | Bolus Size |
|--------------------|------------|
| 1 Ipm              | 15 ml      |
| 2 Ipm              | 28 ml      |
| 3 Ipm              | 41 ml      |
| 4 Ipm              | 55 ml      |
| 5 Ipm              | 68 ml      |
| 6 lpm              | 81 ml      |

The 6203 contains an alarm package that is designed to alert the user in the event of disconnection of the cannula or unit malfunction. The 6203 will produce an audible alarm tone to alert the user if it has not detected sufficient negative pressure to close the circuit on the sensing diaphragm within 45 seconds. The 6203 will cause the flow/apnea light to light a steady red if the microprocessor should fail.

The 6203 is substantially equivalent in intended use and principal of operation to other oxygen conserving devices including our own device, the OCD2001 referenced above (K992935 now marketed as the 6200); the Invacare IPD Oxygen Conserving Device (K953852) and the Chad Therapeutics Oxymatic Electronic Oxygen Conserver (K852650). These predicate devices, like the 6203, are electronic products that use a breath sensor and normally closed valve. Additionally, the Invacare IPD Oxygen Conserving Device, like the 6203 opens the valve and delivers the oxygen to the patient on every detected inhalation.

{3}------------------------------------------------

### Performance Data

Extensive functional testing of the 6203 has been performed. In addition, testing of the device has been performed under various environmental conditions, including impact/drop testing, storage temperature testing, electromagnetic interference testing, electrostatic discharge testing and surface temperature testing. Power supply testing was also performed; these tests included battery life testing and low power indicator testing. The functional, environmental and power supply testing performed on the device demonstrated that it meets its performance objectives and complies with applicable FDA guidelines.

The electronic components and software design are identical to the Contemporary Products, Inc. predicate device, OCD2001 (6200).

{4}------------------------------------------------

Image /page/4/Picture/1 description: The image is a black and white logo for the U.S. Department of Health & Human Services. The logo features a stylized eagle with its wings spread, facing left. The eagle is enclosed within a circular border that contains the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" in a sans-serif font. The text is arranged around the upper half of the circle.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

FEB 1 3 2003

·Mr. Barry A. Schwartz President Contemporary Products, Incorporated 470 Riverside Street Portland, Maine 04103

Re: K022865

Trade/Device Name: 6203 Demand Oxygen Conservation Device Regulation Number: 21 CFR 868.5905 Regulation Name: Oxygen Conserver Regulatory Class: II Product Code: 73 NFB Dated: December 3, 2002 Received: December 4, 2002

Dear Mr. Schwartz:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{5}------------------------------------------------

Page 2 - Mr. Schwartz

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510/k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (301) 594-4646. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrl/dsma/dsmamain.html

Sincerely yours,

Susan Runoe

Susan Runner, DDS, MA Interim Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

{6}------------------------------------------------

Contemporary Products, Inc.

Wave Conserver Model 6203

# Indications for Use Statement

# 510(k) Reference Number:

This is an initial submission; no number has yet been assigned

#### Statement of Indications for Use:

The Wave Conserver Model 6203 is indicated for use to conserve oxygen for patients prescribed 1 to 6 liters per minute of supplemental oxygen and use of a nasal cannula and USP bottled oxygen.

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED.)

# CONCURRENCE OF CDRH, OFFICE OF DEVICE EVALUATION (ODE)

(Division Sign-Off) Division of Cardiovascular, Respiratory and Neurological Devices

510(k) Number:

| Prescription Use     | OR |
|----------------------|----|
| (Per 21 CFR 801.109) |    |

Over-the Counter Use

Hua Ruto 2/12/03

(Optional Format 1-2-96)

. General Hospital

510(k) Number ..

---

**Source:** [https://fda.innolitics.com/submissions/AN/subpart-f%E2%80%94therapeutic-devices/NFB/K022865](https://fda.innolitics.com/submissions/AN/subpart-f%E2%80%94therapeutic-devices/NFB/K022865)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
