← Product Code [MNT](/submissions/AN/subpart-f%E2%80%94therapeutic-devices/MNT) · K113640

# STELLAR 150 (K113640)

_Resmed Germany, Inc. · MNT · Apr 5, 2012 · Anesthesiology · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/AN/subpart-f%E2%80%94therapeutic-devices/MNT/K113640

## Device Facts

- **Applicant:** Resmed Germany, Inc.
- **Product Code:** [MNT](/submissions/AN/subpart-f%E2%80%94therapeutic-devices/MNT.md)
- **Decision Date:** Apr 5, 2012
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 868.5895
- **Device Class:** Class 2
- **Review Panel:** Anesthesiology
- **Attributes:** Therapeutic, Pediatric

## Indications for Use

The device is intended to provide ventilation for non-dependent, spontaneously breathing adult and pediatric patients (30 lb / 13 kg and above) with respiratory insufficiency or respiratory failure, with or without obstructive sleep apnea. The device is for non-invasive use, or invasive use (with the use of the ResMed Leak Valve). Operation of the device includes both stationary, such as in hospital or home, or mobile, such as wheelchair usage.

## Device Story

Stellar 150 is a pressure-controlled ventilator for non-invasive or invasive respiratory support. Input: airway pressure via single-limb vented circuit. Operation: microprocessor-controlled blower generates required airway pressure; supports CPAP, Bi-level, and volume-assured pressure support (iVAPS) modes. Invasive use enabled by ResMed Leak Valve (includes leak port and anti-asphyxia valve). Used in hospital, home, or mobile settings (e.g., wheelchairs) by clinicians or patients. Output: delivered ventilation pressure. Healthcare providers use output to manage respiratory insufficiency/failure; benefits include improved patient breathing support and mobility.

## Clinical Evidence

Bench testing only. No clinical data. Performance testing of the ResMed Leak Valve conducted per ISO 5356-1 and ASTM F 1246-91. Side-by-side system verification testing demonstrated no significant difference in invasive therapy delivery compared to the predicate.

## Technological Characteristics

Pressure-controlled ventilator; microprocessor-controlled blower; single-limb vented circuit. Includes ResMed Leak Valve with integrated anti-asphyxia valve. Standards: ISO 5356-1 (conical connectors), ASTM F 1246-91 (electrically powered home care ventilators). Modes: CPAP, Bi-level, iVAPS. Stationary and mobile operation.

## Regulatory Identification

A continuous ventilator (respirator) is a device intended to mechanically control or assist patient breathing by delivering a predetermined percentage of oxygen in the breathing gas. Adult, pediatric, and neonatal ventilators are included in this generic type of device.

## Predicate Devices

- STELLAR 150 ([K103167](/device/K103167.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

## APR - 5 2012

| Date prepared          | February 6, 2012                                                  |
|------------------------|-------------------------------------------------------------------|
| Submitter              | Sandra Grünwald                                                   |
|                        | Manager Quality Management & Regulatory Affairs                   |
|                        | ResMed Germany, Inc.                                              |
| Official contact       | David D'Cruz                                                      |
|                        | V.P., Clinical & Regulatory Affairs                               |
|                        | ResMed Corp.                                                      |
|                        | 9001 Spectrum Center Blvd.,                                       |
|                        | San Diego CA 92123                                                |
|                        | USA                                                               |
|                        | Tel: +1 858-836-5984                                              |
|                        | Fax: +1 858-836-5522                                              |
| Proprietary/Trade name | Stellar 150                                                       |
| Common name            | Continuous ventilator                                             |
| Classification         | 21 CFR 868.5895                                                   |
|                        | Class II                                                          |
| Product code           | MNT                                                               |
|                        | Ventilator, continuous, minimal ventilatory support, facility use |
| Predicate Device       | STELLAR 150 (K103167)                                             |
| Reason for submission  | Change of Indication for Use                                      |

# 510(k) Summary – Stellar 150

i

{1}------------------------------------------------

#### Indications for Use

The device is intended to provide ventilation for non-dependent, spontaneously breathing adult and pediatric patients (30 lb / 13 kg and above) with respiratory insufficiency or respiratory failure, with or without obstructive sleep apnea. The device is for non-invasive use, or invasive use (with the use of the ResMed Leak Valve). Operation of the device includes both stationary, such as in hospital or home, or mobile, such as wheelchair usage.

#### Device Description

The Stellar 150 is a pressure controlled ventilator using a single limb vented circuit, product code MNT and is substantially equivalent to the already marketed Stellar 150 device (K103167). For both devices as they are substantial equivalent, it is essential that a microprocessor controlled blower generates the required airway pressure. CPAP and Bi-level modes are implemented. With ongoing technological progress, the device is further suitable to include a population above 30 lbs (13 kg) for CPAP and Bilevel modes. The device also includes a volume assured pressure support mode (iVAPS), indicated for patients above 66 lbs (30 kg). The Stellar 150 in combination with the new developed ResMed Leak Valve supports the invasive use therapy. The ResMed Leak Valve incorporates a leak port as well as an integrated anti-asphyxia valve.

#### Performance Data

Design and verification activities which were performed on the previously cleared Stellar 150 as a result of the risk analysis and product requirements remain still valid. All tests confirmed the product met the predetermined acceptance criteria.

Performance testing on the Stellar 150 device was performed and cleared in the previous submission of the Stellar 150 (K103167). Within this submission performance testing on the ResMed Leak Valve was accomplished. The ResMed Leak Valve performance tests comprise functionality tests according to:

- ISO 5356-1 Anaesthetic and respiratory equipment Conical connectors Part 1: Cones and o sockets
- ASTM F 1246-91 Specification for Electrically Powered Home Care Ventilators, Part 1 -O Positive-Pressure Ventilators and Ventilator Circuits

In addition to the performance testing of the new ResMed Leak Valve, system verification testing was performed as well as bench testing. The side-by-side testing demonstrated that there is no significant difference in delivering invasive therapy using the new ResMed Leak Valve (leak port with integrated anti-asphyxia valve) and therefore the Stellar 150 is Substantially Equivalent to the predicate device.

#### Substantial Equivalence

The Stellar 150 is substantial equivalent to the previously cleared predicate device Stellar 150 (K103167).

- Similar indication for use A
- Same operating principle A
- A Same technologies
- A Same manufacturing process

The extension of the indication for use can be realized by combining the Stellar 150 with the new developed ResMed Leak Valve. Especially the ResMed Leak Valve with its integrated anti-asphyxia valve (AAV) ensures that safety and effectiveness of the invasive use therapy is not affected adversely. Performance and side-by-side testing demonstrated that there is no significant difference in auversely. Forromanoo and the new ResMed Leak Valve (leak port with integrated anti-asphyxia valve).

{2}------------------------------------------------

As the operating principle, the technology and the manufacturing process of the Stellar 150 device are identical to the Stellar 150 (K103167) design and verification activities which were performed on the previously cleared Stellar 150 as a result of the risk analysis and product requirements remain still valid. All tests confirmed the product met the predetermined acceptance criteria.

Thus the Stellar 150 with the modified indication for use is substantial equivalent to the previous cleared Stellar 150 device (K103167).

#### Conclusion

The indication for use is similar to the predicated device, whereas the technological characteristics, principles of operation and manufacturing process remain unchanged. Performance data of the new ResMed Leak Valve and side-by-side testing demonstrate that the new combination consisting of the ResMed Leak Valve and the Stellar 150 is as safe and effective as the Stellar 150 device (K103167). The Stellar 150 with modified indication for use is therefore substantially equivalent to the Stellar 150 device (K103167).

{3}------------------------------------------------

#### DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/3/Picture/1 description: The image shows the seal of the U.S. Department of Health and Human Services. The seal features a stylized caduceus, a symbol often associated with medicine and healthcare, consisting of a staff with two snakes coiled around it. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" is arranged in a circular pattern around the caduceus.

#### Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002

ResMed Germany, Inc. C/O Mr. David D'Cruz Vice President, Clinical & Regulatory Affairs ResMed Corporation 9001 Spectrum Center Blvd. San Diego, California 92123

Re: K113640

Trade/Device Name: Stellar 150 Regulation Number: 21 CFR 868.5895 Regulation Name: Continuous Ventilator Regulatory Class: II Product Code: MNT Dated: March 14, 2012 Received: March 19, 2012

#### Dear Mr. D'Cruz:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Eederal Register.

APR - 5 2012

{4}------------------------------------------------

### Page 2 - Mr. D'Cruz

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If vou desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices /ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.

Sincerely yours,

hh for

Anthony D. Watson, B.S., M.S., M.B.A. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

{5}------------------------------------------------

### Indications for Use

510(k) Number (if known):

K113640

Device Name:

Stellar 150

#### Indications for Use:

The device is intended to provide ventilation for non-dependent, spontaneously breathing adult and pediatric patients (30 lb / 13 kg and above) with respiratory insufficiency or respiratory failure, with or without obstructive sleep apnea. The device is for non-invasive use, or invasive use (with with of milliour. Operation of the device includes both stationary, such as in hospital or home, or mobile, such as wheelchair usage.

| Prescription Use            | X | AND/OR | Over-The-Counter Use        |
|-----------------------------|---|--------|-----------------------------|
| (Part 21 CFR 801 Subpart D) |   |        | (Part 21 CFR 807 Subpart C) |

(PLEASE DO NOT WRITE BELOW THIS LINE – CONTINUE ON ANOTHER PAGE IF NEEDED)

| Concurrence of CDRH; Office of Device Evaluation (ODE)                            | Page 1 of 1      |        |         |
|-----------------------------------------------------------------------------------|------------------|--------|---------|
| (Division Sign-Off)                                                               |                  |        |         |
| Division of Anesthesiology, General Hospital<br>Infection Control, Dental Devices | K113640          |        |         |
| 510(k) Number:                                                                    | February 6, 2012 | C89440 | Page 20 |

---

**Source:** [https://fda.innolitics.com/submissions/AN/subpart-f%E2%80%94therapeutic-devices/MNT/K113640](https://fda.innolitics.com/submissions/AN/subpart-f%E2%80%94therapeutic-devices/MNT/K113640)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
