Girgis Scope Video 200 Double Bladed Laryngoscope (GSV200.DBL)
Applicant
Girgis Scope, LLC
Product Code
CCW · Anesthesiology
Decision Date
Jun 22, 2017
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 868.5540
Device Class
Class 1
Indications for Use
The Girgis Scope Video 200 Double Bladed Laryngoscope (GSV200.DBL) is indicated to improve direct visualization of larynx and assist in both non-complicated and difficult intubation in situation of unpredictable airway condition.
Device Story
Device is a rigid, double-bladed video laryngoscope used by clinicians (anesthesiologists, ER physicians, EMS) to assist in tracheal intubation. It features a stationary handle with fiber-optic illumination and a movable blade equipped with a camera. The device provides both direct visualization (via traditional blade design) and indirect visualization (via the camera accessory) of the laryngeal anatomy. The movable blade, controlled by a levering system, articulates to create a pathway for endotracheal tube insertion. The camera output is displayed on a DVR monitor to assist in difficult intubation scenarios. The device is reusable, disassemblable for cleaning, and sterilized via pre-vacuum steam. It is intended to improve visibility of the vocal cords and epiglottis while reducing potential for teeth damage during intubation procedures.
Clinical Evidence
No clinical studies were conducted. Substantial equivalence is supported by bench testing, including electrical safety (IEC 60601-1, IEC 60335), electromagnetic compatibility (IEC 60601-1-2), biocompatibility (ISO 10993 series, ASTM F 756-08, ASTM F 619-03), and a simulated use/usability study involving anesthesiologists, ER physicians, and EMS personnel.
Technological Characteristics
Rigid laryngoscope with double-blade design; fiber-optic illumination; 3.7V Lithium-Ion rechargeable battery; lever-actuated flexible tip (70-degree range); ratchet spring holder with ball joint socket; camera accessory for indirect visualization. Materials tested per USP 24/NF 19 Class VI and ISO 10993. Sterilization: Pre-vacuum steam (132°C for 4 minutes).
Indications for Use
Indicated for patients requiring tracheal intubation, including those with non-complicated or difficult/unpredictable airway conditions, to improve direct visualization of the larynx.
Regulatory Classification
Identification
A rigid laryngoscope is a device used to examine and visualize a patient's upper airway and aid placement of a tracheal tube.
{1}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular border with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" arranged around it. Inside the circle is an abstract symbol resembling an eagle or a stylized human profile, composed of three curved lines.
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
June 22, 2017
Girgis Scope LLC % E. Smith Consultant Smith Associates 1468 Harwell Avenue Crofton. Marvland 21114
Re: K163412
Trade/Device Name: Girgis Scope Video 200 Double Bladed Laryngoscope (GSV200.DBL) Regulation Number: 21 CFR 868.5540 Regulation Name: Rigid Laryngoscope Regulatory Class: Class I Product Code: CCW Dated: May 23, 2017 Received: Mav 26, 2017
Dear E. Smith:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in
{2}------------------------------------------------
the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely,
Mark S. Fellman -S
for
Lori A. Wiggins, MPT, CLT Acting Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
# Indications for Use
510(k) Number (if known) K163412
#### Device Name
Girgis Scope Video 200 Double Bladed Laryngoscope (GSV200.DBL)
Indications for Use (Describe)
The Girgis Scope Video 200 Double Bladed Laryngoscope (GSV200.DBL) is indicated to improve direct visualization of larynx and assist in both non-complicated and difficult intubation of unpredictable airway condition.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|--|
|-------------------------------------------------|--|
X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
# CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
# *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
# 510(K) Summary
| Sponsor | |
|---------------------------|-----------------------------------------------------------------|
| Company Name: | Girgis Scope LLC |
| Company Address: | 2 Stonegate Drive<br>Monroe Township, NJ 08831 |
| Telephone: | 1-732-710-9724 |
| Fax: | 1-732-605-1631 |
| Contact Person: | Madgy Girgis, M.D. |
| Summary Preparation Date: | June 22, 2017 |
| Trade Name: | Girgis Scope Video 200 Doubled Bladed Laryngoscope (GSV200.DBL) |
| Common/Usual Name: | Video Laryngoscope |
| Classification Name: | Laryngoscope Rigid Accessory |
| Regulation Code: | 21 CFR 868.5540 |
| Product Code: | CCW |
| Device Class: | Class I |
| Panel: | Division of Anesthesiology |
Predicate Device
| Manufacturer | Brand Name | Classification |
|----------------------------------------|------------------------------------|----------------|
| Propper Manufacturing<br>Company, Inc. | Propper Flip-Tip™™<br>Laryngoscope | K915804 |
#### Device Description
GSV200.DBL is designed to assist in tracheal intubation procedure by direct visualization of laryngeal anatomy in closed position and use movable blade with camera as accessory from different angle in open position. The physical shape of GSV200.DBL is a modification of traditional Macintosh blade similar to Propper flip tip laryngoscope with added movable blade with camera as accessory to provide visualization to insert endotracheal tube.
Girgis Scope Video DBL is designed to use two power sources, one in the stationary handle
{5}------------------------------------------------
for fiber optic illumination of the airway and the second in the DVR monitor to illuminate the Larynx for the video camera. Battery chargers, cables to computer and cable to connect to TV screen are accessories included in the kit. The GSV200.DBL is provided non-sterile and is reusable and can be disassembled to provide cleaning and disinfection.
# Indications for Use Statement
The Girgis Scope Video 200 Double Bladed Laryngoscope (GSV200.DBL) is indicated to improve direct visualization of larynx and assist in both non-complicated and difficult intubation in situation of unpredictable airway condition.
| Parameters | Girgis Scope LLC<br>(K163412) | Propper Manufacturing<br>Company, Inc.<br>(K915804) | Comments |
|---------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------|
| Indications for Use | The Girgis Scope Video<br>200 Double Bladed<br>Laryngoscope<br>(GSV200.DBL) is<br>indicated to improve<br>direct visualization of<br>larynx and assist in<br>both non-complicated<br>and difficult intubation<br>in situation of<br>unpredictable airway<br>condition. | Propper Flip Tip™<br>laryngoscope is<br>indicated to improve<br>direct visualization of<br>larynx and assist in<br>both non-complicated<br>and difficult intubation<br>in situation of<br>unpredictable airway<br>condition. | Identical |
| Type of Device | GVS200.DBL is a<br>modification of the<br>traditional MacIntosh<br>laryngoscope.<br>In addition to proper<br>design movable blade<br>and ratchet spring<br>holder are attached. | Propper Flip-Tip™<br>Laryngoscope is a<br>modification of the<br>traditional MacIntosh<br>laryngoscope. | Similar - Propper offers<br>a flip tip at the distal<br>end of the<br>laryngoscope blade and<br>the GSV200.DBL offers<br>a second blade. |
| Design | Anatomically designed<br>to improve visibility of<br>the vocal cords and<br>reduce teeth damage | Anatomically designed<br>to improve visibility of<br>the vocal cords and<br>reduce teeth damage | Identical |
| Laryngoscope Flexi-Tip<br>and Lever Control | Flexi-tip is adjustable to<br>70 degrees using<br>levering movable<br>handle to provide a<br>precise control to<br>expose the epiglottis. | Flexi-tip is adjustable to<br>70 degrees using<br>levering movable<br>handle to provide a<br>precise control to<br>expose the epiglottis. | Identical |
# Predicate Product Comparison Table
{6}------------------------------------------------
| Fiber Optic Intubating<br>Laryngoscope | Stationary handle for<br>fiber optic illumination<br>of the mouth | Stationary handle for<br>fiber optic illumination<br>of the mouth | Identical |
|----------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------|
| Power Supply and Light<br>Source | Stationary<br>laryngoscope handle<br>containing a Lithium<br>Ion rechargeable<br>battery (3.7V).<br>GSV200.DBL utilizes a<br>rechargeable battery to<br>power light bulb and<br>through their 5mm<br>fibro-optic bundle to<br>provide bright white<br>illumination. | Stationary<br>laryngoscope handle<br>containing battery.<br>Propper utilizes AA and<br>C Batteries to power<br>light bulb and through<br>their 5mm fibro-optic<br>bundle to provide<br>bright while<br>illumination. | Predicate uses C<br>batteries and the<br>GSV200.DBL utilizes a<br>rechargeable battery |
| Sterilization | Pre-vacuum sterilizer<br>132 degrees for 4<br>minutes | Pre-vacuum sterilizer<br>132 degrees for 4<br>minutes | Identical |
| Laryngoscope Blades | Reusable | Reusable | Identical |
#### Discussion of Technological Characteristics
- 1. Propper Flip-Tip™ Laryngoscope blade (K915804) and the GBV200.DBL Laryngoscope both have a flexible tip that is adjustable to 70 degrees using a levering system which includes a movable handle, tip and spring mechanism and both offer fiber optic illumination.
- 2. Both devices are anatomically designed to improve visibility of the epiglottis and reduce teeth damage. Both devices provide a laryngoscope stationary handle containing batteries, Propper utilizes C Batteries and GBV200.DBL a rechargeable battery, to power light bulb through their fibro-optic bundle which provides bright white illumination.
- Both systems have a stationary handle and in addition to the Propper design the 4. GSV200.DBL uses a ratchet spring holder with a ball joint socket that locks the movable blade in closed position while introducing the two blades in patient mouth and unlocks to open position in step by step progression.
- In addition to the Propper laryngoscope design, GSV200.DBL uses the movable blade, which 5. is attached to the movable handle and articulates with stationary blade so movement in the movable handle is transmitted to the movable blade to:
- A. To create a pathway to introduce the endotracheal tube with ease
- B. The movable blade carries camera arrangement as accessory for indirect visualization of laryngeal anatomy in unpredictable difficult intubation. The camera angle will be looking upward and forward to show the most anterior position of larynx in difficult intubation
- 5. Both offer a reusable Flexi-Tip style laryngoscope blade, both recommend pre-vacuum sterilization at 132°C for 4 minutes.
{7}------------------------------------------------
### Non-Clinical Performance Data
- IEC 60335-1 Corrigendum 1 Household and similar electrical appliances Safety Part ● 1: General Requirements
- . IEC 60335-2-29 Safety of household and similar electrical appliances - Safety - Part 2-29 Particular requirements for battery charges
- IEC 60950-1 Information technology equipment – Safety – Part 1: General requirements
- IEC 60601-1 + CORR 1 (2006) + CORR 2 (2007) + A1 (2012) Medical electrical equipment . Part 1: General requirements for basic safety and essential performance
- IEC 60601-1-2:2007 Medical electrical equipment Part 1-2: Collateral standard: Electromagnetic compatibility - Requirements and tests
- . USP 24, NF 19 Biological Test for Plastics, Class VI 70 degrees C
- . AAMI/ANSI ST79 Comprehensive Guide to Steam Sterilization and Sterility Assurance in Health Care Facilities
- . ISO 10933-5 Biological Evaluation of Medical Devices – Part 5: Tests for in vitro Cytotoxicity L929 MEM Elution Test
- ISO 10993-10 Biological Evaluation of Medical Devices – Part 10: Test for Irritation and Skin Sensitization Kligman Maximization Test
- ISO 10993-10 Biological Evaluation of Medical Devices – Part 10: Test for Irritation and Skin Sensitization Intracutaneous Injection Test
- ISO 10993-11 Biological Evaluation of Medical Devices – Part 11: Tests for Systemic Toxicity
- ASTM F 756-08 Standard Practice for Assessment of Hemolytic Properties of Materials
- ASTN F 619-03 Standard Practice for Extraction of Medical Plastics
- A Simulated Use/Usability study was conducted using Anesthesiologists, Emergency Room Physicians, and Emergency Medical Service (EMS).
- . Software Validation per Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices
# Clinical Study
No clinical studies were conducted.
# Conclusion
The Girgis Scope Video 200 Double Bladed Laryngoscope model (GSV200.DBL) has the same indications for use, flexible tip design, fiber optic bundle for lighting, laryngoscope blade design, laryngoscope handle design, same methods of sterilization, and similar nonclinical testing and are battery operated as the Propper Flip-Tip Laryngoscope (K915804). The DVR accessory improves visualization of the epiglottis and the usability study demonstrated ease-of-use of the system. Based on the similarities of the two laryngoscope systems no different questions of safety and effectiveness have been raised and we conclude that our laryngoscope system is substantially equivalent to the predicate laryngoscope system.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.