This product is intended for use as an in-line pressure relief device, designed to limit the system pressure within the limits of its stated technical specifications. This product is indicated for patients in a hospital for use by appropriately qualified healthcare professionals.
Device Story
F&P Optiflow Flow Diverter is an in-line pressure relief device for nasal high flow therapy; placed between flow meter and dryline. Device limits system pressure during therapy; prevents excessive pressure if downstream occlusion occurs. Operation: valve seals against body during normal flow; if relief pressure exceeded, shaft moves, opening seal to divert gas flow to atmosphere. Used in hospitals by qualified healthcare professionals. Output is mechanical pressure relief; protects patient from over-pressurization. Benefits include maintaining safe operating pressures during high-flow oxygen delivery.
Clinical Evidence
Bench testing only. Performance testing confirmed compliance with ISO 80601-2-74, ISO 80601-2-90, IEC 60601-1, and CGA V-5. Testing demonstrated that the device functions as intended to limit system pressure and does not raise new questions of safety or effectiveness compared to the predicate.
Technological Characteristics
Polycarbonate housing; mechanical spring-loaded pressure relief valve; DISS 1240 inlet; F&P female safety outlet. Flow range 5-70 L/min; max pressure 60 cmH2O. Non-sterile. Standards: ISO 80601-2-74, ISO 80601-2-90, IEC 60601-1, IEC 62366-1, CGA V-5.
Indications for Use
Indicated for patients in a hospital setting requiring in-line pressure relief during nasal high flow therapy, for use by qualified healthcare professionals.
Regulatory Classification
Identification
A nonrebreathing valve is a one-way valve that directs breathing gas flow to the patient and vents exhaled gases into the atmosphere.
{1}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
Fisher & Paykel Healthcare Ltd Reena Daken Regulatory Affairs Manager 15 Maurice Paykel Place, East Tamaki Auckland, 2013 New Zealand
Re: K234053
Trade/Device Name: F&P Optiflow Flow Diverter Regulation Number: 21 CFR 868.5870 Regulation Name: Nonrebreathing valve Regulatory Class: Class II Product Code: CBP Dated: December 21, 2023 Received: December 22, 2023
Dear Reena Daken:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
{2}------------------------------------------------
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Digitally signed by John S. John S. Bender -S Bender -- S Date: 2024.08.09
10:36:12 -04'00'
for Ethan Nyberg, Ph.D. Assistant Director DHT1C: Division of Anesthesia, Respiratory, and Sleep Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT, and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
{3}------------------------------------------------
Enclosure
{4}------------------------------------------------
# Indications for Use
510(k) Number (if known) K234053
Device Name F&P Optiflow Flow Diverter
Indications for Use (Describe)
This product is intended for use as an in-line pressure relief device, designed to limit the limits of its stated technical specifications.
This product is indicated for patients in a hospital for use by appropriately qualified healthcare professionals.
Type of Use (Select one or both, as applicable)
| Prescription Use (Part 21 CFR 801 Subpart D) | Over-The-Counter Use (21 CFR 801 Subpart C) |
|----------------------------------------------|---------------------------------------------|
|----------------------------------------------|---------------------------------------------|
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{5}------------------------------------------------
# 510(k) Summary
As Required by 21 CFR 807.92
# I. SUBMITTER
| Company Name and Address | Fisher & Paykel Healthcare Limited<br>15 Maurice Paykel Place<br>East Tamaki<br>Auckland 2013, New Zealand |
|---------------------------|------------------------------------------------------------------------------------------------------------|
| | Telephone: +64 9 574 0100 |
| Prepared and Submitted by | Danica Tung |
| | Regulatory Affairs Market Manager |
| Contact Person | Reena Daken |
| | Regulatory Affairs Manager |
| | Telephone: +64 9 574 0100 |
| | Email: reena.daken@fphcare.co.nz |
| Date Prepared | 8 July 2024 |
| II. DEVICE | |
| Name of Device | F&P Optiflow™ Flow Diverter |
| Common/Usual Name | Nonrebreathing Valve |
| Classification Name | Nonrebreathing Valve |
| Regulatory Class | Class II (21 CFR §868.5870) |
| Product Code | CBP |
### PREDICATE DEVICE III.
Predicate Device: OJR215 Pressure Relief Manifold, K173770
{6}------------------------------------------------
#### DEVICE DESCRIPTION IV.
The F&P Optiflow Flow Diverter is intended for use as an in-line pressure relief device, designed to limit the system pressure when used with the compatible F&P Optiflow Oxygen Kits and nasal interfaces.
The F&P Flow Diverter is placed between the flow meter and the dryline during use. It is a multi-patient use prescription-only device, provided in a non-sterile state. It operates at flow rates between 5 to 70 L/min and is intended to be used by appropriately qualified healthcare professionals in hospitals.
When an anesthesia mask is applied over the Optiflow Switch interface, the respiratory gas no longer has a pathway to flow through the nasal interface to the patient. This will trigger respiratory gas flow to be diverted into the atmosphere.
#### INDICATIONS FOR USE V.
This product is intended for use as an in-line pressure relief device, designed to limit the system pressure within the limits of its stated technical specifications.
This product is indicated for patients in a hospital for use by appropriately qualified healthcare professionals.
{7}------------------------------------------------
### COMPARISON OF TECHNOLOGICAL CHARATCERISTICS WITH THE PREDICATE DEVICE VI.
| Design/Technological<br>Characteristics | Subject Device<br>(F&P Optiflow™ Flow Diverter) | Predicate Device<br>(OJR215 Pressure Relief Manifold – K173770) | Comments |
|------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------|
| Indications for use and intended use | | | |
| Indications for Use | This product is intended for use as an in-line<br>pressure relief device, designed to limit the system<br>pressure within the limits of its stated technical<br>specifications.<br>This product is indicated for patients in a hospital<br>for use by appropriately qualified healthcare<br>professionals. | The OJR215 Pressure Relief Manifold is designed<br>to protect neonates, infants and children from<br>excessive inspiratory pressure in the event of a<br>downstream occlusion occurring in continuous<br>flow breathing gas delivery systems via nasal<br>cannula. The device is intended for use with flow<br>rates greater than 0.5 L/min up to, and including,<br>36 L/min. The OJR215 is fitted upstream of the<br>patient. The OJR215 is single use only and is<br>prescription only. | Equivalent |
| Availability | Prescription use<br>(Part 21 CFR 801 Subpart D) | Prescription use<br>(Part 21 CFR 801 Subpart D) | Identical |
| Intended Use Environment | Hospital | Hospital/Institutional | Identical |
| Users | Qualified healthcare professionals | Qualified healthcare professionals | Identical |
| Operation and safety features | | | |
| Ambient Operating<br>Temperature | 18 to 26 °C (64.4 to 78.8°F) | 18 to 26 °C (64.4 to 78.8°F) | Identical |
| Shelf Life | Up to 7 years | 3 years | Different |
| Design/Technological<br>Characteristics | Subject Device<br>(F&P OptiflowTM Flow Diverter) | Predicate Device<br>(OJR215 Pressure Relief Manifold - K173770) | Comments |
| Storage Temperature | -10°C to +50 °C (14 to 122 °F) | -10°C to +50 °C (14 to 122 °F) | Identical |
| Sterility | Device not provided sterile. | Device not provided sterile. | Identical |
| Operating Pressure<br>- Maximum Pressure | Maximum system pressure of 60cmH2O at<br>maximum rated flow rate of 70L/min | Maximum relief pressure is 75 cm H2O at<br>maximum rated flow of 36 L/min. | Equivalent |
| Flow Range | 5-70L/min | 0.5-36 L/min | Equivalent |
| Application/Therapy | Nasal High Flow Therapy | Nasal High Flow Therapy | Identical |
| Function and Design | | | |
| Principles of Operation | During normal use, the valve seals against the<br>valve body to create a seal, allowing gas to flow<br>through the device to the patient. When the relief<br>pressure is exceeded, the shaft is moved, allowing<br>the seal to open and pressure is relieved, diverting<br>flow from the gas path to atmosphere. | During normal use, the valve seals against the<br>valve body to create a seal, allowing gas to flow<br>through the device to the patient. When the relief<br>pressure is exceeded, the shaft is moved, allowing<br>the seal to open and pressure is relieved, diverting<br>flow from the gas path to atmosphere. | Equivalent |
| Inlet Connection | DISS 1240 | Inlet barb | Different |
| Outlet Connection | F&P female safety connection | 22mm female taper connection as per ISO 5356-1 | Different |
| Housing | The body and enclosure is moulded and includes<br>the inlet and the outlet connection and is<br>polycarbonate. | The manifold housing is ABS, and is moulded to<br>include the gas inlet barb, the 22mm chamber inlet<br>connection and the 15mm oxygen sensor port. | Equivalent |
| Design/Technological<br>Characteristics | Subject Device<br>(F&P OptiflowTM Flow Diverter) | Predicate Device<br>(OJR215 Pressure Relief Manifold – K173770) | Comments |
| Shaft | The shaft connects to the 2-piece valve seal | The manifold plunger shaft is a 2-piece valve seal. | |
| Plunger Seal | The valve seal rests on the valve body during<br>normal use and then accurately relieves pressure by<br>opening when the relief pressure is reached. | The seal on the manifold plunger seals the<br>manifold valve during normal use and then<br>accurately relieves pressure by opening when the<br>relief pressure is reached. | Equivalent |
| Spring Pressure Relief Valve | The shaft is moved by the valve seal that moves<br>when the set pressure is exceeded. | The shaft is moved by a spring pressure relief valve<br>that moves when the set pressure is exceeded. | |
| Manifold Shroud | The rear and front caps prevent any modification<br>by the user to the pressure relief that has already<br>been set during production. | The purpose of the manifold shroud is to prevent<br>access to the pre-set operating pressures and<br>tampering. | Equivalent |
| Accessories | | | |
| Breathing Circuit | F&P AA404 Optiflow Oxygen Kit<br>Clearance status: a modified version of AA403<br>cleared under K211096 | F&P RT330 OptiflowTM Junior Optiflow Tubing<br>Kit<br>Clearance status: cleared (K173770) | Equivalent |
| Patient Interface | Nasal cannula | Nasal cannula | Equivalent |
{8}------------------------------------------------
{9}------------------------------------------------
{10}------------------------------------------------
#### VII. PERFORMANCE DATA
## Summary of Non-Clinical Tests
Performance testing of the Flow Diverter was completed and confirms the subject device does not raise new questions of safety and effectiveness. The testing provided demonstrates substantial equivalence of the Flow Diverter to the predicate device.
The Flow Diverter has been tested to the applicable requirements of the following standards:
- ISO 80601-2-74:2017 Part 2-74 Particular requirements for basic safety and essential ● performance of respiratory humidifying equipment
- CGA V-5: 2019 Standard for Diameter Index Safety System (Noninterchangeable Low ● Pressure Connections for Medical Gas Applications)
- IEC 60601-1 Edition 3.2 (2020) Medical electrical equipment Part 1: General ● requirements for basic safety and essential performance
- . ISO 17664-2:2021 Processing of health care products - Information to be provided by medical device manufacturer for the processing of medical devices. Part 2: Non-critical medical devices
- ISO 80601-2-90:2021 Medical electrical equipment - Part 2-90: Particular requirements for basic safety and essential performance of respiratory high-flow therapy equipment
- IEC 62366-1:2015 + A1:2020 Medical devices Part 1: Application of usability ● engineering to medical devices
#### CONCLUSION VIII.
The Flow Diverter is substantially equivalent to the predicate based on intended use, technological characteristics, and performance. In addition, the conclusions drawn from the non-clinical tests demonstrate that the device is substantially equivalent to the legally marketed predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.