The MINISCAV™ waste gas evacuation apparatus is intended to remove patients' exhaled waste gases during procedures where analgesia is administered to a patient via inspiration of mixtures oxide and oxygen from a nitrous oxide / oxygen delivery device. Not intended for use with flammable anesthetic gases. This device is intended for professional use only in healthcare facilities, clinics, and physician and dentist offices.
Device Story
MINISCAV™ is a portable, diaphragm-based vacuum pump designed to evacuate exhaled waste gases from nitrous oxide/oxygen analgesia delivery systems. It connects via a standard 19mm flexible vacuum hose to the exhaust port of analgesia equipment. Operated by healthcare professionals in clinics and offices, the device provides a constant vacuum (≤90 mmHg) and flow rate (42 Lpm ± 5 Lpm) to ensure effective gas scavenging. By removing waste gases from the patient circuit, the device helps maintain a safe environment for both patients and staff. It is compatible with specific analgesia systems (e.g., Porter, Accutron, Nitronox). The device is not intended for use with flammable anesthetic gases.
Clinical Evidence
Bench testing only. Testing included electrical safety (AAMI/ANSI/ES60601-1), EMC (IEC 60601-1-2), durability (10,000 hours continuous operation), vacuum/flow performance (≤90 mmHg at 42 Lpm ± 5 Lpm), leakage testing, and oxygen-rich environment safety (IEC 60601-1 section 11.2.2). Compatibility was verified with Porter, Accutron, and Nitronox systems.
Technological Characteristics
Diaphragm-type vacuum pump; hermetically sealed; portable form factor. Connects via 19mm hose barb. Fixed vacuum output (≤90 mmHg) and flow rate (42 Lpm ± 5 Lpm). Electrical power source. Complies with IEC 60601-1 and IEC 60601-1-2 standards.
Indications for Use
Indicated for patients undergoing procedures requiring analgesia via nitrous oxide/oxygen delivery. Intended for professional use in healthcare facilities, clinics, and physician/dentist offices. Not for use with flammable anesthetic gases.
Regulatory Classification
Identification
A gas-scavenging apparatus is a device intended to collect excess anesthetic, analgesic, or trace gases or vapors from a patient's breathing system, ventilator, or extracorporeal pump-oxygenator, and to conduct these gases out of the area by means of an exhaust system.
Predicate Devices
G. Dundas, Active Waste Gas Scavenger System (K110930)
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized image of three human profiles facing right, stacked on top of each other, with flowing lines representing hair or movement.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
May 5, 2017
Sedation Systems LLC % Paul Dryden Consultant, Promedic LLC 2471 McMullen Booth Rd., Ste. 316 Clearwater, Florida 33759
Re: K160950
Trade/Device Name: MINISCAVTM Regulation Number: 21 CFR 868.5430 Regulation Name: Gas-Scavenging Apparatus Regulatory Class: Class II Product Code: CBN, BTA Dated: April 27, 2017 Received: May 1, 2017
Dear Paul Dryden:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices. good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{1}------------------------------------------------
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
# Michael J. Ryan -S
for Tina Kiang, Ph.D. Acting Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known)
#### K160950
Device Name
MINISCAVTM
Indications for Use (Describe)
The MINISCAV™ waste gas evacuation apparatus is intended to remove patients' exhaled waste gases during procedures where analgesia is administered to a patient via inspiration of mixtures oxide and oxygen from a nitrous oxide / oxygen delivery device. Not intended for use with flammable anesthetic gases.
This device is intended for professional use only in healthcare facilities, clinics, and physician and dentist offices.
Type of Use (Select one or both, as applicable)
XX Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
### PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.
#### FOR FDA USE ONLY
Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
FORM FDA 3881 (8/14)
Page 1 of 1
EF PSC Publishing Services (301) 443-6740
Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement on last page.
{3}------------------------------------------------
Sedation Systems LLC 2471 McMullen Booth Rd., Suite 316 Clearwater, FL 33759
Tel – 888-959-5288
| Official Contact: | S. Michael Bender, Managing Member |
|-------------------------------------------------|------------------------------------------------------------------------------------------------------|
| Proprietary or Trade Name: | MINISCAV™ |
| Common/Usual Name: | Apparatus, Gas Scavenging |
| Classification Name /<br>Product Classification | Apparatus, Gas Scavenging<br>CBN, 21CFR 868.5430, Class II |
| Predicate Devices: | G. Dundas, Active Waste Gas Scavenger System, K11093<br>Medela, Dominant Flex Suction Pumps, K150134 |
#### Device Description:
The MINISCAV™ is a small, portable, vacuum pump that generates a constant vacuum that can be connected via a standard flexible vacuum hose to nitrous oxide / oxygen delivery equipment. The vacuum hose is connected to the "vacuum / suction" nipple on exemplary equipment.
#### Indications for Use:
The MINISCAV™ waste gas evacuation apparatus is intended to remove patients' exhaled waste gases during procedures where analgesia is administered to a patient via inspiration of mixtures of nitrous oxide and oxygen from a nitrous oxide / oxygen delivery device. Not intended for use with flammable anesthetic gases.
This device is intended for professional use only in healthcare facilities, clinics, and physician and dentist offices.
| Feature | Predicate<br>G. Dundas<br>Active Waste Gas<br>Scavenger System<br>K110930 | Predicate<br>Medela<br>Dominant Flex<br>Suction Pumps<br>K150134 | Proposed<br>MINISCAVTM |
|------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Product Classification | CBN<br>CFR 868.5430<br>Gas scavenging<br>apparatus<br>Class II | BTA<br>CFR 878.4780<br>Powered suction pump<br>Class II | CBN<br>CFR 868.5430<br>Gas scavenging apparatus<br>Class II |
| Indications for Use | The Waste Gas<br>Scavenger is designed<br>for use with vacuum<br>(suction) waste gas<br>disposal systems with<br>anesthesia machines<br>and heart/lung bypass | The Dominant Flex<br>Suction Pump is<br>indicated for vacuum<br>extraction, aesthetic<br>body contouring,<br>aspiration during<br>flexible endoscopy, and | The MINISCAVTM waste<br>gas evacuation apparatus<br>is intended to remove<br>patients' exhaled waste<br>gases during procedures<br>where analgesia is<br>administered to a patient |
#### Table 1 Substantial equivalence Comparison to Predicates
{4}------------------------------------------------
#### 510(k) Summary 5-May-17
Page 2 of 4
| Page 2 of 4 | | | |
|-----------------------------------------|-----------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | machines. | aspiration and removal<br>of surgical fluids, tissue<br>(including bone), gases,<br>bodily fluids or<br>infectious materials<br>from wounds or from a<br>patient's airway or<br>respiratory support<br>system either during<br>surgery or at the beside. | via inspiration of<br>mixtures of nitrous oxide<br>and oxygen from a<br>nitrous oxide / oxygen<br>delivery device. Not<br>intended for use with<br>flammable anesthetic<br>gases.<br>This device is intended<br>for professional use only<br>in healthcare facilities,<br>clinics, and physician and<br>dentist offices. |
| Environment of Use | Healthcare facilities,<br>clinics and physician<br>and dentist offices. | Healthcare facilities<br>and clinics | |
| Used with flammable<br>anesthetic gases | No | No | No |
| Requires a vacuum<br>source | Yes, connects to any<br>vacuum source<br>Central supply / wall /<br>Portable pump | Portable vacuum source | Portable vacuum source<br>to connect to a waste gas<br>scavenging apparatus to<br>supply the vacuum source |
| Means of connection | A flexible vacuum hose<br>from the device to the<br>vacuum source | Connects a flexible<br>hose from this pump to<br>any device requiring a<br>vacuum source, i.e., gas<br>scavenging apparatus | Connects a flexible hose<br>from this pump to a gas<br>scavenging apparatus |
| Fittings | 19 mm<br>Hose barb for vacuum | Hose barb for vacuum | 19 mm<br>Hose barb for vacuum |
| Vacuum pressures<br>Flow rate | Can accept a range of<br>applied vacuum as it no<br>range limit specified | Up to -700 mmHg | 90 mmHg<br>42 Lpm +/- 5 Lpm<br>Connects to a nitrous<br>oxide/ oxygen flowmeter<br>exhaust port |
| Technology | Passive device which<br>requires connection to a<br>vacuum source | Piston / cylinder design<br>Portable | Diaphragm design<br>Portable |
#### Substantial Equivalence Discussion -
Table 1 above compares the key features of the proposed MINISCA V™ with the identified predicate and reference to demonstrate that the proposed device can be found to be substantially equivalent.
#### Indications for Use -
While the indications for use are not identical to that of the predicate device (K110930), the subject device and Dundas predicate are intended to remove waste gases from anesthesia gas machine system..
Discussion - The MINISCAV™ is one part of a waste gas scavenging system. The parts include: (1) a scavenger to titrate exhaust gas flow from the patient circuit, which is connected to (2) a vacuum source. This device is intended to scavenge gas, which the same as the intended use of the predicate Dundas scavenger.
#### Technology, Construction, Performance -
{5}------------------------------------------------
The MINISCAVIM is a diaphragm hermetically sealed pump with a fixed vacuum flow rate. The inlet line is connected to a flowmeter exhaust port of the nitrous oxide / oxygen delivery equipment which has an exhaust port for scavenging gases. The Dundas scavenger (K110930) is a flowmeter that titrates exhaust flow with wall-supplied vacuum sources. While the Dundas scavenger does not supply the vacuum, both the MINISCAV and Dundas devices validate the ability to expel gases at specified vacuum pressures, mitigating risks associated with failure to vacuum. Therefore, the MINISCAV™ does not raise different concerns of safety or effectiveness for substantial equivalence.
The Medela Dominant Flex Suction Pump (K150134) is pump with an adjustable vacuum flow as high as -700 mmHg. The Medela includes indications to remove gases from respiratory support svstems, which the MINISCAV also performs by providing vacuum via a separate gas scavenger, Therefore, both devices are validated for vacuum specifications. Although different technologies (vacuum range -700mmHg compared to -90mmHg) exist, these differences do not raise different questions of safety and effectiveness.
#### Environment of Use -
The environments of use are similar to the predicate Medela Dominant Flex Suction Pump (K150134) and the Dundas Active Waste Gas Scavenger System (K110930).
Discussion - As the environments of use are similar to the predicate they should be considered substantially equivalent.
#### Non-Clinical Testing Summary -
We performed testing which evaluated:
- AAMI/ANSI/ES60601-1 for electrical safety ●
- IEC 60601-1-2 for EMC ●
- Durability ●
- o Continuous running demonstrated that the vacuum pump stays within its specifications for at least 10,000 hours
- Real-time testing of the complete unit supports that the device meets its O performance specifications after 2 years use
- . Vacuum and Flow testing
- Testing demonstrates that the device provides a constant vacuum of <=90 mmHg o at a constant Flow rate of 42 Lpm +/- 5 Lpm which has been deemed sufficient for scavenging waste gases from nitrous oxide / oxygen delivery equipment
- Determination of maximum length of exhaust tubing о
- Leakage
- Testing demonstrates that after 2 years the unit has no leaks and meets its O performance specifications
- Evaluation in an Oxygen Rich Environment per IEC 60601-1 section 11.2.2. .
- Risk Analysis of critical components, oxygen build-up in housing, and o construction and separation of the housing and components support compliance to IEC 60601-1 section 11.2.2, clause 1.b(3).
- . Compatibility
- o with Porter, Accutron, and Nitronox analgesia systems.
#### Discussion of Differences –
The differences between the proposed device and the predicate are:
{6}------------------------------------------------
- MINISCAV™ is intended to scavenge waste gases, which is similar to the Dundas ● Active Waste Gas Scavenger System (K110930).
- . While the intended uses are similar, the Dundas device (K110930) must be connected to a vacuum source (such as the MINISCAVTM) to operate.
- Limitation of vacuum. While the Medela Dominant Flex Suction Pump (K150134) has a ● higher vacuum, for connection to a waste gas scavenger, the vacuum pressure is much lower and within the performance specifications of the proposed device.
- o This lower vacuum range is appropriate for the intended use of scavenging gases.
These differences still allow one to find the subject device substantially equivalent to the predicate devices for the proposed indications for use.
### Substantial Equivalence Conclusion -
Based upon the presented information the sponsor has demonstrated through performance testing, design and features, and non-clinical testing that the proposed device and predicate device are substantially equivalent.
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.