Evita 4 is a time cycled, constant volume, long term, intensive care ventilator for adults and children with a body weight of at least 3 kg. Intended Environment for the device: hospital use inner clinical transport
Device Story
Evita 4 is a microprocessor-controlled, time-cycled, constant-volume intensive care ventilator. It delivers gas via a high-pressure servo valve system. The device supports multiple ventilation modes including CMV/assist, SIMV, MMV, SB, CPAP, ASB/PSV, PCV+, APRV, Apnea Ventilation, and ILV. It is intended for use in hospital settings and during inner-clinical transport by trained clinicians. The device monitors patient respiratory parameters and displays curves and values to assist healthcare providers in managing patient ventilation. It provides features such as AutoFlow, leak compensation, and nebulization to support patient respiratory needs. The software version 3.0 integrates additional functions previously found in the Evita 2 dura, such as inspiration hold and pediatric nebulization, to enhance clinical utility.
Clinical Evidence
No clinical data provided; bench testing only.
Technological Characteristics
Microprocessor-controlled, time-cycled, constant-volume ventilator. Features high-pressure servo valve gas delivery. Software version 3.0. Supports multiple ventilation modes (CMV, SIMV, MMV, CPAP, etc.). Designed for hospital and clinical transport use. Connectivity and specific material standards not detailed.
Indications for Use
Indicated for adults and children with a body weight of at least 3 kg requiring long-term intensive care ventilation.
Regulatory Classification
Identification
A continuous ventilator (respirator) is a device intended to mechanically control or assist patient breathing by delivering a predetermined percentage of oxygen in the breathing gas. Adult, pediatric, and neonatal ventilators are included in this generic type of device.
{0}------------------------------------------------
**Dräger**
4 1888 SEP
file: 510kstate.RTF Date: June, 1998 Author: Frank Clanzett
# 510(k) SUMMARY Summary of Safety and Effectiveness
K980642
# APPLICANTS NAME AND ADDRESS:
Drager Inc. Critical Care Systems 3136 Quarry Road Telford , PA 18969
APPLICANTS TELEPHONE NUMBER: (215)-721-6917
APPLICANTS FACSIMILE NUMBER: (215)-721-6915
# APPLICANTS CONTACT PERSON:
Harald Kneuer Regulatory Affairs Manager
DATE THE SUMMARY WAS PREPARED: February, 1998
# DEVICE NAME:
Trade Name: Common Name: Classification Name: Evita 4 Evita 4 Ventilator, Continuous (per 21 CFR 868.5895)
# LEGALLY MARKETED DEVICE TO WHICH DRÄGER IS CLAIMING SUBSTANTIAL EQUIVALENCE:
Evita 4 - Manufactured by Dräger Medizintechnik GmbH, Lübeck, Germany and sold in the United States by Dräger, Inc.
Evita 2 dura- Manufactured by Dräger Medizintechnik GmbH, Lubeck, Germany and sold in the United States by Dräger, Inc.
Servo 300 - Sold in the United States by Siemens
{1}------------------------------------------------
# DESCRIPTION OF THE DEVICE:
The Evita 4 with the Software version 3.0 is a time cycled, constant volume ventilator The Evita + with the Oct.ware is microprocessor controlled and provides the following ventilation modes:
| - CMV/assist | Controlled Mandatory Ventilation<br>Controlled and assisted volume contant ventilation with the options:<br>- CPV Continuous Positive Pressure Ventilation<br>- PLV Pressure Limited Ventilation<br>- AutoFlow For automatic regulation of inspiratory flow<br>- IRV Inversed Ratio Ventilation |
|---------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| - SIMV | Synchronized Intermittent Mandatory Ventilation<br>Procedure for weaning patients off the ventilator after they have started spontaneous breathing. With the options:<br>- PLV<br>- AutoFlow |
| - MMV | Mandatory Minute Volume Ventilation<br>Spontaneous breathing with automatic adjustment of mandatory ventilation to the patients minute volume requirement. With the options:<br>- PLV<br>- AutoFlow |
| - SB | Spontaneous Breathing<br>Spontaneous breathing at ambient pressure |
| - CPAP | Continuous Positive Airway Pressure<br>Spontaneous breathing with positive airway pressure |
| - ASB/PSV | Assisted Spontaneous Breathing / Pressure Support Ventilation<br>Pressure supported spontaneous breathing |
| - PCV+ | Pressure Controlled Ventilation<br>Pressure controlled ventilation combined with free spontaneous breathing during the complete breathing cycle, and adjustable pressure increase to CPAP level |
| - APRV | Airway Pressure Release Ventilation<br>Spontaneous breathing on two pressure levels with long time ranges - independently adjustable |
| - Apnea Ventilation | For switching over automatically to volume controlled mandatory ventilation if breathing stops |
| - ILV | Independent Lung Ventilation<br>Separate, differentiated, synchronized ventilation with one Evita 4 for each lung |
{2}------------------------------------------------
# Dräger
# tensive Care Ventilator Data Comparison Sheet
| Specification | Evita 4 (K 961687) | Evita 2 dura (K 970165) | Servo 300 (K902859) | Evita 4 SW 3.0 |
|----------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------|---------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Manufacturer | Dräger | Dräger | Siemens | Dräger |
| Classification: Intensive Care<br>Ventilator | Yes | Yes | Yes | Yes |
| Operating Principle: Time<br>cycled, volume constant,<br>pressure controlled | Yes | Yes | Yes | Yes |
| Software controlled | Yes | Yes | Yes | Yes |
| Contoll Software: Version 3.0 | No (1.0) | Yes | N.A. | Yes |
| Gas delivery system: Fast<br>"High Pressure Servo Valve",<br>microprocessor controlled | Yes | Yes | Yes | Yes |
| Intended Use | Evita 4 is a time cycled, constant volume, long<br>term, intensive care<br>ventilator for adults and<br>children with a body<br>weight of at least 3 kg. | like Evita 4 (K 961687) | N.A. | Evita 4 is a time cycled,<br>constant volume, long<br>term, intensive care<br>ventilator for adults and<br>children with a body<br>weight of at least 3 kg |
| Specification | Evita 4 (K 961687) | Evita 2 dura (K 970165) | Servo 300 (K902859) | Evita 4 SW 3.0 |
| Modes of Ventilation: | - | - | - | identical with Evita 4<br>510(k) no: K 961687 |
| Ventilator Settings | - | - | - | identical with Evita 4<br>510(k) no: K 961687 |
| Flow Trigger Sensivity | - | - | - | identical with Evita 4<br>510(k) no: K 961687 |
| Parameter setting | - | - | - | identical with Evita 4<br>510(k) no: K 961687 |
| Monitors/Indicators | - | - | - | identical with Evita 4<br>510(k) no: K 961687 |
| Pressure Limitation | - | - | - | identical with Evita 4<br>510(k) no: K 961687 |
| Alarm Functions | - | - | - | identical with Evita 4<br>510(k) no: K 961687 |
| Displayed Values | - | - | - | identical with Evita 4<br>510(k) no: K 961687 |
| Displayed Curves | - | - | - | identical with Evita 4<br>510(k) no: K 961687 |
| Specification | Evita 4 (K 961687) | Evita 2 dura (K 970165) | Servo 300 (K902859) | Dräger<br>Evita 4 SW 3.0 |
| Special Functions | | | | |
| Intermittend PEEP | Yes | Yes | - | identical with Evita 2dura |
| Inspiration Hold | No | Yes | - | identical with Evita 2dura |
| Apnea Ventilation | Yes | Yes | - | identical with Evita 2dura |
| Nebulization | Yes | Yes | - | identical with Evita 2dura |
| Nebulization in pediatric mode | No | Yes | - | identical with Evita 2dura |
| Suction Help Mode | Yes | Yes | - | identical with Evita 2dura |
| Switch over function | Yes | Yes | - | identical with Evita 2dura |
| Leak-Compensation | No | Yes | - | identical with Evita 2dura |
| Minute Volume-Leak | No | Yes | - | identical with Evita 4<br>510(k) no. K 961687 |
| Physical Caracteristics | | | | |
{3}------------------------------------------------
4 of 6
#
------------------------------------------------------------------------------------------------------------------------------------------------------------------------------
{4}------------------------------------------------
{5}------------------------------------------------
# INTENDED USE OF THE DEVICE
Evita 4 (SW 3.0) is a time cycled, constant volume, long term, intensive care ventilator for adults and children with a body weight of at least 3kg.
Intended Environment for the device:
- user facilities …
- inner clinical transport -
SUMMARY OF THE TECHNOLOGICAL CHARACTERISTICS OF THE DEVICE COMPARED TO THE PREDICATE DEVICES:
The ventilator functions of the Evita 4 with the enhanced software version are substantially equivalent to the Evita 2 dura, the Evita 4 with the former software and the Servo 300 ventilators.
The Evita 4 with enhanced software integrates the same optional functions that are presently performed by the Evita 4 with the former software.
The device fulfils at least the same standards as the Evita 2 dura.
Therefore the device under review is substantial equivalent to the predicate devices concerning safety, efficiency and the intended use.
Flium
Frank Clanzett Regulatory Affairs Dräger Medizintechnik
{6}------------------------------------------------
Image /page/6/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized depiction of an eagle or bird with three curved lines representing its wings or feathers. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular fashion around the bird symbol.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
4 1998 SEb
Mr. Harald Kneuer Drager Inc. Critical Care Systems 3136 Quarry Road Telford, PA 18969
K980642 Re : Evita 4 Requlatory Class: II (two) Product Code: 73 CBK Dated: June 12, 1998 Received: June 15, 1998
Dear Mr. Kneuer:
We have reviewed your Section 510 (k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Requlations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Requlation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Druq Administration (FDA) will verify such assumptions. Failure to comply with the GMP requlation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.
{7}------------------------------------------------
## Page 2 - Mr. Harald Kneuer
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4648. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification"(21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597, or at its internet address "http://www.fda.gov/cdrh/dsma/dsmamain.html".
Sincerely yours,
Thomas J. Callehan
Thomas J. Callahan, Ph.D. Director Division of Cardiovascular, Respiratory, and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{8}------------------------------------------------
# INTENDED USE STATEMENT OF THE EVITA 4, SW 3.0
Fivita 4 is a time cycled, constant volume, long term, intensive care ventilator for adults and children with a body weight of at least 3 kg.
## Intended Environment for the device:
- hospital use
- inner clinical transport
Signature
Frank Clanzep (Regulatory Affairs)
1. September 1998
Mark Kramis
: Division Sign-Off) Division of Cardiovascular, Respiratory, and Neurological Devices 510(k) Number_K980(e4Z
\/ Prescription Use
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.