F&P OptiNIV Hospital Vented Full Face Mask Compatible with Single-limb Circuits – Size A (ONIV117A); F&P OptiNIV Hospital Vented Full Face Mask Compatible with Single-limb Circuits – Size B (ONIV117B); F&P OptiNIV Hospital Vented Full Face Mask Compatible with Single-limb Circuits – Size C (ONIV117C); F&P OptiNIV ONIV117-F Hospital Vented Full Face Mask with optional Expiratory Filter Compatible with Single-limb Circuits - Size A (ONIV117A-F); F&P OptiNIV ONIV117-F Hospital Vented Full

K252173 · Fisher & Paykel Healthcare, Ltd. · CBK · Sep 10, 2025 · Anesthesiology

Device Facts

Record IDK252173
Device NameF&P OptiNIV Hospital Vented Full Face Mask Compatible with Single-limb Circuits – Size A (ONIV117A); F&P OptiNIV Hospital Vented Full Face Mask Compatible with Single-limb Circuits – Size B (ONIV117B); F&P OptiNIV Hospital Vented Full Face Mask Compatible with Single-limb Circuits – Size C (ONIV117C); F&P OptiNIV ONIV117-F Hospital Vented Full Face Mask with optional Expiratory Filter Compatible with Single-limb Circuits - Size A (ONIV117A-F); F&P OptiNIV ONIV117-F Hospital Vented Full
ApplicantFisher & Paykel Healthcare, Ltd.
Product CodeCBK · Anesthesiology
Decision DateSep 10, 2025
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 868.5895
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The Fisher & Paykel Healthcare single-patient use masks are intended for use as an accessory to ventilators to enable non-invasive positive pressure ventilation (NPPV) therapy (CPAP or bi-level) to be delivered to spontaneously breathing adult patients (> 30 kg) with respiratory insufficiency or respiratory failure who have been prescribed NPPV. The masks are to be fitted and therapy maintained, by trained medical practitioners in a hospital/institutional environment with patient monitoring in place.

Device Story

Oro-nasal full face mask accessory for non-invasive positive pressure ventilation (NPPV) systems; delivers CPAP or bi-level therapy to spontaneously breathing adult patients. Device consists of plastic shell, soft seal for facial cushioning, and head straps. Connects to single-limb ventilator circuits via 22mm male ISO 5356-1 conical connectors. Used in hospital/institutional environments by trained medical practitioners. Provides interface for gas delivery; includes anti-asphyxiation valve. Benefits include respiratory support for patients with respiratory insufficiency or failure. Output is mechanical delivery of pressurized air/gas; healthcare providers monitor patient response and therapy parameters via the ventilator system.

Clinical Evidence

Bench testing only. Testing included ISO 5356-1 (connectors), ISO 17510 (masks), ISO 10993-1 (biocompatibility), and ISO 18562 (breathing gas pathway emissions). Performance testing confirmed pressure drop, resistance to flow, mask dead space, anti-asphyxiation valve function, venting leak (CO2 rebreathing), pressure waveform, humidity delivery, and product leak. Usability engineering performed per IEC 62366-1. Shelf-life and packaging integrity (ISTA 2A) also verified.

Technological Characteristics

Oro-nasal full face mask; plastic shell with soft seal; 22mm male ISO 5356-1 conical connector. Single-limb circuit compatible. Operating pressure range 4–30 cmH2O. Non-sterile. Materials evaluated per ISO 10993-1 and ISO 18562-1/2/3. Mechanical anti-asphyxiation valve. Available in three sizes (A, B, C).

Indications for Use

Indicated for spontaneously breathing adult patients (>30 kg) with respiratory insufficiency or respiratory failure requiring non-invasive positive pressure ventilation (CPAP or bi-level) in hospital/institutional settings.

Regulatory Classification

Identification

A continuous ventilator (respirator) is a device intended to mechanically control or assist patient breathing by delivering a predetermined percentage of oxygen in the breathing gas. Adult, pediatric, and neonatal ventilators are included in this generic type of device.

Predicate Devices

Reference Devices

Related Devices

Submission Summary (Full Text)

{0} FDA U.S. FOOD &amp; DRUG ADMINISTRATION September 10, 2025 Fisher &amp; Paykel Healthcare Ltd Reena Daken Regulatory Affairs Manager 15 Maurice Paykel Place, East Tamaki Auckland, 2013 New Zealand Re: K252173 Trade/Device Name: F&amp;P OptiNIV Hospital Vented Full Face Mask Compatible with Single-limb Circuits - Size A (ONIV117A); F&amp;P OptiNIV Hospital Vented Full Face Mask Compatible with Single-limb Circuits - Size B (ONIV117B); F&amp;P OptiNIV Hospital Vented Full Face Mask Compatible with Single-limb Circuits - Size C (ONIV117C); F&amp;P OptiNIV ONIV117-F Hospital Vented Full Face Mask with optional Expiratory Filter Compatible with Single-limb Circuits - Size A (ONIV117A-F); F&amp;P OptiNIV ONIV117-F Hospital Vented Full Face Mask with optional Expiratory Filter Compatible with Single-limb Circuits - Size B (ONIV117B-F); F&amp;P OptiNIV ONIV117-F Hospital Vented Full Face Mask with optional Expiratory Filter Compatible with Single-limb Circuits - Size C (ONIV117C-F) Regulation Number: 21 CFR 868.5895 Regulation Name: Continuous ventilator Regulatory Class: Class II Product Code: CBK Dated: July 11, 2025 Received: July 11, 2025 Dear Reena Daken: We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and U.S. Food &amp; Drug Administration 10903 New Hampshire Avenue Silver Spring, MD 20993 www.fda.gov {1} K252173 - Reena Daken Page 2 adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download). Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181). Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050. All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems. {2} K252173 - Reena Daken Page 3 For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, Ethan L. Nyberg -S Ethan Nyberg, Ph.D. Assistant Director DHT1C: Division of Anesthesia, Respiratory, and Sleep Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT, and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {3} F&amp;P OptiNIV Hospital Vented Full Face Mask Compatible with Single-limb Circuits – Size A Page 8 of 30 | Indications for Use | | | | --- | --- | --- | | Please type in the marketing application/submission number, if it is known. This textbox will be left blank for original applications/submissions. | K252173 | ? | | Please provide the device trade name(s). | | ? | | F&P OptiNIV Hospital Vented Full Face Mask Compatible with Single-limb Circuits – Size A (ONIV117A); F&P OptiNIV Hospital Vented Full Face Mask Compatible with Single-limb Circuits – Size B (ONIV117B); F&P OptiNIV Hospital Vented Full Face Mask Compatible with Single-limb Circuits – Size C (ONIV117C); F&P OptiNIV ONIV117-F Hospital Vented Full Face Mask with optional Expiratory Filter Compatible with Single-limb Circuits - Size A (ONIV117A-F); F&P OptiNIV ONIV117-F Hospital Vented Full Face Mask with optional Expiratory Filter Compatible with Single-limb Circuits - Size B (ONIV117B-F); F&P OptiNIV ONIV117-F Hospital Vented Full Face Mask with optional Expiratory Filter Compatible with Single-limb Circuits - Size C (ONIV117C-F) | | | | Please provide your Indications for Use below. | | ? | | The Fisher & Paykel Healthcare single-patient use masks are intended for use as an accessory to ventilators to enable non-invasive positive pressure ventilation (NPPV) therapy (CPAP or bi-level) to be delivered to spontaneously breathing adult patients (> 30 kg) with respiratory insufficiency or respiratory failure who have been prescribed NPPV. The masks are to be fitted and therapy maintained, by trained medical practitioners in a hospital/institutional environment with patient monitoring in place. | | | | Please select the types of uses (select one or both, as applicable). | ☑ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C) | | {4} 510(k) Summary As Required by 21 CFR 807.92 # I. SUBMITTER Company Name and Address Fisher &amp; Paykel Healthcare Limited 15 Maurice Paykel Place East Tamaki Auckland 2013, New Zealand Telephone: +64 9 574 0100 Prepared and Submitted by Jung Yun Lee Regulatory Affairs Specialist Contact Person Reena Daken Regulatory Affairs Manager Telephone: +64 9 574 0100 Email: reena.daken@fphcare.co.nz Date Prepared 05 September 2025 # II. DEVICE Name of Device F&amp;P OptiNIV Vented Full Face Mask Common/Usual Name Full Face Mask Classification Name Ventilator, Continuous, Facility Use Regulatory Class Class II (21 CFR §868.5895) Product Code CBK # III. PREDICATE DEVICE Predicate device: F&amp;P Visairo NIV Mask, K203449 Reference device: F&amp;P Nivairo Mask, K191624 Page 1 of 5 {5} Page 2 of 5 ## IV. DEVICE DESCRIPTION The F&amp;P OptiNIV Vented Full Face Masks are oro-nasal full face masks that are intended for use as an accessory to deliver non-invasive positive pressure ventilation (NPPV) to a patient as part of a non-invasive ventilation system. The OptiNIV Vented Full Face Masks are prescription-only, provided in a non-sterile state. The F&amp;P OptiNIV Vented Full Face Mask consists of a plastic shell with a seal to allow cushioning onto the patient's face, and straps to enable the mask to be secured over the head. The mask fits under the patient's nose and covers the nares of the nose and the mouth, with a soft seal to maintain pressure inside the mask whilst minimizing discomfort to the user. The subject device is available in three sizes – A, B, and C to denote the seal sizes. ## V. INDICATIONS FOR USE The Fisher &amp; Paykel Healthcare single-patient use masks are intended for use as an accessory to ventilators to enable non-invasive positive pressure ventilation (NPPV) therapy (CPAP or bi-level) to be delivered to spontaneously breathing adult patients (&gt; 30 kg) with respiratory insufficiency or respiratory failure who have been prescribed NPPV. The masks are to be fitted and therapy maintained, by trained medical practitioners in a hospital/institutional environment with patient monitoring in place. {6} VI. COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE DEVICE | Design / Technological Characteristic | Subject Device F&P OptiSIV Vented Full Face Mask | Predicate Device F&P Visairo Mask (K203449) RT077 Model | Comments | | --- | --- | --- | --- | | Intended Use / Indications for Use | | | | | Indications for Use | The Fisher & Paykel Healthcare single patient use masks are intended for use as an accessory to ventilators to enable non-invasive positive pressure ventilation (NPPV) therapy (CPAP or bi-level) to be delivered to spontaneously breathing adult patients (>30 kg) with respiratory insufficiency or respiratory failure who have been prescribed NPPV. The masks are to be fitted and therapy maintained, by trained medical practitioners in a hospital/institutional environment with patient monitoring in place. | The Fisher & Paykel Healthcare single patient use masks are intended for use as an accessory to ventilators to enable non-invasive positive pressure ventilation (NPPV) therapy (CPAP or bi-level) to be delivered to spontaneously breathing adult patients (>30 kg) with respiratory insufficiency or respiratory failure who have been prescribed NPPV. The masks are to be fitted and therapy maintained, by trained medical practitioners in a hospital/institutional environment with patient monitoring in place. | Equivalent | | Availability | Prescription use (Part 21 CFR 801 Subpart D) | Prescription use (Part 21 CFR 801 Subpart D) | Equivalent | | Patient Population | Adult (>30 kg) | Adult (>30 kg) | Equivalent | | Operating Environment | Hospital/institutional environments | Hospital/institutional environments | Equivalent | | Reusability | Single use | Single use | Equivalent | | Technical Specifications and Features | | | | | Operating Pressure Range | 4 – 30 cmH₂O | 4 – 40 cmH₂O | Equivalent | | Compatibility with F&P Systems | F&P 850 (K110019) F&P 950 (K220703) | F&P 850™ (K110019) F&P 950™ (K220703) | Equivalent | | Interface Connections | ISO 5356-1 Conical Connectors 22mm male ISO medical taper | ISO 5356-1 Conical Connectors 22mm male ISO medical taper | Equivalent | Page 3 of 5 {7} | Design / Technological Characteristic | Subject Device F&P OptiNIV Vented Full Face Mask | Predicate Device F&P Visairo Mask (K203449) RT077 Model | Comments | | --- | --- | --- | --- | | Mask Dead Space | < 200 cm³ | < 200 cm³ | Equivalent | | Anti-Asphyxiation Valve Operation | • Open to Atmospheric Pressure 0.41 cmH₂O • Closed to Atmospheric Pressure 0.92 cmH₂O | (F&P Visairo Mask RT075 model) • Open to Atmospheric Pressure 0.41 cmH₂O • Closed to Atmospheric Pressure 0.92 cmH₂O | Equivalent | | Resistance to Flow through mask | • 0.23 cmH₂O @ 50 L/min • 0.61 cmH₂O @ 100 L/min | • 0.26 cmH₂O @ 50 L/min • 0.63 cmH₂O @ 100 L/min | Similar | | Breathing Circuit | Single limb | Single limb | Equivalent | | Sterility | Device not provided sterile | Device not provided sterile | Equivalent | | Maximum duration of use | Mask 14 days Filter 48 hours | 14 days | Similar | | Shelf life | 3 years | 3 years | Equivalent | | Sizes | Available in three sizes – A, B, and C | Available in three sizes – A, B, and C | Equivalent | Page 4 of 5 {8} Page 5 of 5 # VII. PERFORMANCE DATA ## Summary of non-clinical tests The F&amp;P OptiNIV Vented Full Face Mask range has been tested to applicable requirements and standards which have all passed: - ISO 5356-1:2015 Anaesthetic and respiratory equipment — Conical connectors Part 1: Cones and sockets - ISO 17510:2020 Medical devices — Sleep apnoea breathing therapy — Masks and application accessories - Pressure drop/resistance to flow testing - Mask deadspace testing - Pressure-flow curve testing - Anti-asphyxiation valve function testing - ISTA 2A:2011 Procedure 2A: Packaged-products weighing 150 lb (68 kg) or less. Basic Requirements: atmospheric conditioning, compression, fixed displacement or random vibration, and shock vibration - IEC 62366-1:2015/AMD 1:2020 Medical devices – Part 1: Application of Usability Engineering to Medical Devices – Amendment 1 and FDA Guidance “Applying Human Factors and Usability Engineering to Medical Devices” - ISO 10993-1:2018 Biological evaluation of medical devices Part 1: Evaluation and testing within a risk management process - ISO 18562-1:2024 Biocompatibility evaluation of breathing gas pathways in healthcare applications Part 1: Evaluation and testing within a risk management process - ISO 18562-2:2024 Biocompatibility evaluation of breathing gas pathways in healthcare applications – Part 2: Tests for emissions of particular matter - ISO 18562-3:2024 Biocompatibility evaluation of breathing gas pathways in healthcare applications – Part 3: Tests for emissions of volatile organic substances Additional performance testing has also been completed to confirm the safety and effectiveness of the F&amp;P OptiNIV Vented Full Face Mask range which have all passed: - Shelf-life testing – all relevant functional testing was performed after simulated shelf-life of three years - Venting leak testing to address CO2 rebreathing - Pressure waveform testing - Humidity delivery testing - Product leak testing # VIII. CONCLUSIONS The F&amp;P OptiNIV Vented Full Face Mask range is substantially equivalent to the predicate device, F&amp;P Visario NIV Masks (K203449), based on patient population, intended uses, comparison of the general technological characteristics, and performance. In addition, the conclusions drawn from the non-clinical tests demonstrate that the subject device is substantially equivalent to the legally marketed predicate device.
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...