The 840 Series Ventilator is intended to provide continuous ventilation to patients requiring respiratory support. The 840 Ventilator System with Expanded NeoMode Option is intended for patients with an Ideal Body Weight (IBW) as low as 0.3 kg. The 840 Series Ventilator with Expanded NeoMode Option is intended for use in hospitals and hospital-type facilities. It may be used during hospital and hospital-type facility transport provided that electrical power and compressed gas are supplied.
Device Story
The Puritan Bennett 840 Series Ventilator is a dual-microprocessor-based, touch-screen controlled critical care ventilator. It delivers precise, sensitive breaths to critically ill patients. The system consists of a Breath Delivery Unit (BDU), Graphical User Interface (GUI), and optional accessories (compressor, backup power, carts). It operates by connecting to electrical power and compressed gas in hospital settings. The device uses software to determine operational variables and settings based on patient IBW and breathing type; it prevents mismatching of patient size and circuit type. Healthcare providers use the GUI to monitor and adjust ventilation parameters. The device provides continuous respiratory support, potentially benefiting patients by maintaining adequate gas exchange. Modifications include software enhancements (communication, idle flow reduction, alarm defaults, apnea interval updates) and hardware updates (PCBAs, carts, backup power, O2 sensor, fuse).
Clinical Evidence
No clinical data. Substantial equivalence was demonstrated through non-clinical performance testing, including software verification and validation, electrical safety (IEC 60601-1), electromagnetic compatibility (IEC 60601-1-2), and usability testing for new hardware components.
Technological Characteristics
Dual-microprocessor-based critical care ventilator. Components: BDU, GUI, compressor, backup power, carts. Connectivity: Touch screen interface. Standards: IEC 60601-1 (electrical safety), IEC 60601-1-2 (EMC). Software: Major level of concern. Features: NeoMode option for IBW as low as 0.3 kg, tidal volume down to 2 mL.
Indications for Use
Indicated for neonate to adult patients (with Expanded NeoMode Option) or infant to adult patients (without option) requiring invasive or non-invasive continuous respiratory support. Minimum patient Ideal Body Weight (IBW) is 0.3 kg.
Regulatory Classification
Identification
A continuous ventilator (respirator) is a device intended to mechanically control or assist patient breathing by delivering a predetermined percentage of oxygen in the breathing gas. Adult, pediatric, and neonatal ventilators are included in this generic type of device.
Predicate Devices
Puritan Bennett™ 840 Ventilator System with Expanded NeoMode Option (K092847)
Reference Devices
Puritan Bennett™ 980 Series Ventilator System (K131252)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the seal of the U.S. Department of Health & Human Services. The seal features a stylized depiction of three human profiles facing right, arranged in a cascading manner. The profiles are rendered in a dark color, creating a silhouette effect. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the upper portion of the seal.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
March 29, 2016
Covidien LP Respiratory and Monitoring Solutions Kelsey Lee Regulatory Affairs Product Manager 6135 Gunbarrel Ave. Boulder, Colorado 80301
Re: K151252
> Trade/Device Name: Puritan Bennett™ 840 Series Ventilator System Regulation Number: 21 CFR 868.5895 Regulation Name: Continuous Ventilator Regulatory Class: Class II Product Code: CBK Dated: February 25, 2016 Received: February 26, 2016
Dear Ms. Lee,
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device
{1}------------------------------------------------
related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
*Tejashri Purohit-Sheth, M.D.*
Tejashri Purohit-Sheth, M.D. Clinical Deputy Director DAGRID/ODE/CDRH FOR
Erin I. Keith, M.S. Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K151252
Device Name
Puritan Bennett™ 840 Series Ventilator System
Indications for Use (Describe)
The 840 Series Ventilator is intended to provide continuous ventilation to patients requiring respiratory support. The 840 Ventilator System with Expanded NeoMode Option is intended for patients with an Ideal Body Weight (IBW) as low as 0.3 kg.
The 840 Series Ventilator with Expanded NeoMode Option is intended for use in hospital-type facilities. It may be used during hospital and hospital-type facility transport provided that electrical power and compressed gas are supplied.
X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(K) SUMMARY
In accordance with the Food and Drug Administration Rule to implement provisions of the Safe Medical Devices Act of 1990 and in conformance with 21 CFR 807.92, this information serves as a 510k Summary for the use of the Puritan Bennett™ 840 Series Ventilator System
| Submitted By: | Covidien<br>6135 Gunbarrel Avenue<br>Boulder, CO 80301 |
|-----------------------------------|---------------------------------------------------------------------------------|
| Date: | March 22, 2016 |
| Contact Person: | Kelsey Lee<br>Regulatory Affairs Product Manager<br>(303) 305-2760 |
| Proprietary Name: | Puritan Bennett™ 840 Series Ventilator System |
| Common Name: | Continuous Ventilator |
| Device Classification Regulation: | 21 CFR 868.5895 – Class II |
| Device Product Code & Panel: | CBK |
| Predicate Devices: | Puritan Bennett™ 840 Ventilator System with<br>Expanded NeoMode Option, K092847 |
| Reference Device: | Puritan Bennett™ 980 Series Ventilator System,<br>K131252 |
{4}------------------------------------------------
## Device Description
The PB 840 Ventilator consists of a Breath Delivery Unit (BDU, a Graphical User Interface (GUI) and a number of optional accessories, including a Compressor, a Back-Up Power Source (BPS), and three cart options. Depending on the patient ideal body weight (IBW), the appropriate patient circuit is attached to the PB 840 Ventilator System and patient. When the system is operational and connected to the appropriate utilities, the ventilator system delivers sensitive, precise breaths to critically ill patients
The Puritan Bennett (PB) 840 Ventilator System is a dual-microprocessor-based, touch screen controlled, critical care ventilator intended to provide continuous ventilation for neonate to adult patients (with expanded NeoMode Option) or infant to adult patients (with no Option) who require either invasive ventilation or non-invasive ventilation.
The PB 840 Ventilator System includes software that is intended for patients with Ideal Body Weight (IBW) as low as 0.3 kg and provides the user with tidal volume to 2 mL. The ventilator determines values for operational variables and allowable settings based on breathing types and IBW. The software controls prevent inadvertent mismatching of patient size and breathing circuit type.
## Indications for Use/Intended Use
The 840 Series Ventilator is intended to provide continuous ventilation to patients requiring respiratory support. The 840 Ventilator System with Expanded NeoMode Option is intended for patients with an Ideal Body Weight (IBW) as low as 0.3 kg.
The 840 Series Ventilator with Expanded NeoMode Option is intended for use in hospitals and hospital-type facilities. It may be used during hospital and hospital-type facility transport provided that electrical power and compressed gas are supplied.
### Technological Characteristics Comparison
The subject Puritan Bennett™ (PB) 840 Series Ventilator System is considered a modification of the predicate PB840 (K092847).
The technological differences between the subject and predicate PB840 are limited to the following:
- · Software Enhancements
- Communication Enhancements —
- Reduction of idle flow for Neonates —
- -Update to alarm defaults for CPAP
- Removal of redundant code
{5}------------------------------------------------
- Update apnea interval ।
- Hardware Enhancements
- New BD and GUI PCBAs due to obsolescence
- A new Pole Cart and new Compressor Cart
- Optional four-hour 803 Backup Power Source (BPS)
- New O2 Sensor Manufacturer
- Fuse Upgrade on the 9.4" GUI Backlight Inverter PCBA
The subject and predicate devices have the same technological aspects:
- Intended use ●
- Intended Population ●
- . Principles of Operation
- Basic ventilator design
- . Manufacturing and Packaging
All other PB840 Series Ventilator System requirements and specifications remain the same
The reference device, PB980, has similar intended use, population, principles of operation and overall design. The reference device has the same oxygen sensor, alarm default for CPAP, high pressure settings and idle flow during "Idle Mode" (referred to as "Standby Mode" in the reference device).
### Substantial Equivalence - Non-Clinical Evidence
The following performance data were provided in support of substantial equivalence determination:
# Software Verification and Validation Testing
Software verification and validation testing were conducted and documentation was provided as recommended by FDA's Guidance for Industry and FDA Staff, "Guidance for the Content of Premarket Submission for Software Contained Medical Devices". The software for this device was considered as a "major" level of concern, since a failure or latent flaw in the software could result in serious injury or death to the patient.
# Electrical Safety and electromagnetic compatibility (EMC)
{6}------------------------------------------------
Electrical safety and EMC testing were conducted on the subject device since it required EMC enhancements based on the two new carts introduced. In addition, the new 803 Backup Power Source (BPS) for the PB840 required repeating EMC testing. The system complies with IEC 60601-1:1988 + A1:1991 + A2:1995 standard for electrical safety and IEC 60601-1-2:2014 standard for EMC. Additionally, the subject device complies with all known national differences.
### Usability Testing
Usability testing was performed to ensure the new Pole Cart and Compressor Cart meet the user requirements and can be used as intended.
#### Substantial Equivalence - Clinical Evidence
N/A - Clinical evidence was not necessary to show substantial equivalence
#### Substantial Equivalence - Conclusions
Since the predicate Puritan Bennett™ (PB) 840 Ventilator with Neomode Option was cleared, this device has over time underwent both software and hardware modifications. As part of the change control process, the subject device has undergone the appropriate verification, validation and safety testing, all of which confirms that the device meets its design, performance, and safety specifications; all testing is described within the body of this submission. The performance data demonstrates that the subject device can be considered substantially equivalent to the predicate device that is currently marketed for the same intended use.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.