K143035 · Flight Medical Innovations , Ltd. · CBK · Mar 30, 2016 · Anesthesiology
Device Facts
Record ID
K143035
Device Name
Flight 60 Ventilator
Applicant
Flight Medical Innovations , Ltd.
Product Code
CBK · Anesthesiology
Decision Date
Mar 30, 2016
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 868.5895
Device Class
Class 2
Attributes
Therapeutic, Pediatric
Indications for Use
The FLIGHT 60 Ventilator is intended to provide continuous or intermittent mechanical ventilation support for the care of individuals who require mechanical ventilation. Specifically, the FLIGHT 60 is applicable for adult and pediatric (i.e., infant, child and adolescent) patients, greater than or equal to 5kg (11 lbs). The FLIGHT 60 Ventilator is a restricted medical device intended for use by qualified, trained personnel under the direction of a physician; it is suitable for use in hospital, sub-acute, emergency room, and home care environments, as well as for transport and emergency response applications.
Device Story
Electrically powered, microprocessor-controlled ventilator; provides mechanical ventilation support (A/CMV, SIMV, Pressure Support, SPONT modes). Inputs: patient breathing signals (pressure/flow/time triggers). Transformation: internal blower-type compressor generates gas flow; exhalation valve with proportional solenoid provides PEEP; microprocessor manages ventilation parameters (volume/pressure/flow). Outputs: mechanical ventilation to patient; audible/visual alarms for limit violations. Used in hospital, sub-acute, emergency, home, and transport environments by trained personnel. Healthcare providers monitor via touch screen display; output assists in maintaining patient respiratory function. Benefits: provides life-sustaining respiratory support for patients ≥ 5kg.
Clinical Evidence
Bench testing only. Non-clinical side-by-side waveform performance testing compared the subject device to the predicate Flight 60 and Trilogy 100. Evaluated flow, pressure, and volume waveforms, ventilation control parameter accuracy, and patient trigger reliability/synchrony. Environmental testing (VOC, PM 2.5) and compliance with IEC 60601-1, ISO 80601-2-12, and IEC 60601-1-2 standards performed.
Technological Characteristics
Electrically powered, microprocessor-controlled ventilator. Features blower-type compressor, proportional solenoid exhalation valve, and internal Li-ion batteries. Dimensions: 295mm x 291mm x 249mm. User interface: touch screen. Connectivity: external power (100-240 VAC or 12-15 VDC). Standards: IEC 60601-1, ISO 80601-2-12, IEC 60601-1-2.
Indications for Use
Indicated for adult and pediatric patients (infant, child, adolescent) weighing ≥ 5kg requiring continuous or intermittent mechanical ventilation support. Restricted to use by qualified, trained personnel under physician direction in hospital, sub-acute, emergency, home, and transport settings.
Regulatory Classification
Identification
A continuous ventilator (respirator) is a device intended to mechanically control or assist patient breathing by delivering a predetermined percentage of oxygen in the breathing gas. Adult, pediatric, and neonatal ventilators are included in this generic type of device.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized symbol of a human figure, represented by three overlapping profiles facing to the right, with flowing lines suggesting movement or connection.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
March 30, 2016
Flight Medical Innovations Ltd. % Bosmat Friedman Regulatory Affairs Consultant MJ RAC 1208-12 Rockford Rd, Toronto. ON M2R 3A2 Canada
Re: K143035 Trade/Device Name: Flight 60® Ventilator Regulation Number: 21 CFR 868.5895 Regulation Name: Continuous Ventilator Regulatory Class: Class II Product Code: CBK. NOU Dated: March 1, 2016 Received: March 2, 2016
Dear Ms. Friedman,
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR
{1}------------------------------------------------
Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours.
Tejashri Purohit-Sheth, M.D.
Tejashri Purohit-Sheth, M.D. Clinical Deputy Director DAGRID/ODE/CDRH FOR
Erin I. Keith, M.S. Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
## Indications for Use
510(k) Number (if known)
### K143035
Device Name FLIGHT60® Ventilator
### Indications for Use (Describe)
The FLIGHT 60 Ventilator is intended to provide continuous or intermittent mechanical ventilation support for the care of individuals who require mechanical ventilation. Specifically, the FLIGHT 60 is applicable for adult and pediatic (i.e., infant, child and adolescent) patients, greater than or equal to 5kg (11 1bs).
The FLIGHT 60 Ventilator is a restricted medical device intended for use by qualified, trained personnel under the direction of a physician; it is sutable for use in hospital, sub-acute, emergency room, and home care environments, as well as for transport and emergency response applications.
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW."
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(K) SUMMARY [as required by section 807.92(c)] FLIGHT 60 Ventilator
## 510(k) Number K143035
#### 1. SUBMITTER
## Applicant's Name:
Flight Medical Innovations Ltd. 7 Hatnufa Street Petach Tikva 4951025 Tel: (972) 3-6731660 Fax: (972) 3-6731690
## Company Contact:
Yael Spielmann, RA & QA Manager yaels@flight-medical.com
## Contact Person:
Bosmat Friedman Regulatory Affairs Consultant, MJ RAC 1208-12 Rockford Rd. Toronto, ON, M2R 3A2, Canada Phone: 647-975-3974; Fax: 647-427-1946; bosmat@pushmed.com
## Date Prepared:
March 23, 2016
#### 2. DEVICE
Trade Name: FLIGHT 60 Ventilator
Classification Name: Continuous Ventilator
## Classification:
Name: Continuous ventilator Product Code: CBK Regulation No: 868.5895 Class: 2
{4}------------------------------------------------
Classification Panel: Office of Device Evaluation (ODE) Division of Anesthesiology, General Hospital, Infection Control, and Dental Devices (DAGRID) Respiratory Devices Branch (RPDB)
and
Name: Continuous, Ventilator, Home Use Product Code: NOU Regulation No: 868.5895 Class: 2 Classification Panel: Office of Device Evaluation (ODE) Division of Anesthesiology, General Hospital, Infection Control, and Dental Devices (DAGRID) Respiratory Devices Branch (RPDB)
#### 3. PREDICATE DEVICES
### Main Predicate:
- Flight 60 Ventilator K130171 ●
### Secondary Predicates:
- Trilogy 100 K083526 ●
#### 4. DEVICE DESCRIPTION
The FLIGHT 60 Ventilator is an electrically powered, microprocessor controlled ventilator with the following types of ventilatory support: A/CMV Volume or Pressure Control, SIMV Volume or Pressure Control, Pressure Support & SPONT mode with Pressure Support. It can be pressure or time triggered; volume or pressure limited; time, pressure or flow cycled. Manual inflation is possible, and an emergency intake valve allows the patient to pull ambient air into the breathing circuit in the event of a complete loss of supply gas pressure. The FLIGHT 60 may be powered by external power (100 - 240 VACS or 12 - 15 VDC) or by its two internal Li Ion rechargeable batteries. The electrical system is comprised of three primary boards: the Main board (motherboard) which holds the majority of the electronics including the main CPU and the display CPU, the Power board, which holds the power subsystems, and internal communication functions, and the Communication board, which holds internal communication and external communication connectors. The main component of the pneumatic system is an electrically controlled compressor (pump). This compressor provides a compressed gas source so no external air compressor is needed. Additionally, the exhalation valve is activated by an electrically controlled proportional solenoid that provides a built in PEEP.
> Flight 60 Ventilator Traditional 510(k)
{5}------------------------------------------------
A comprehensive alarm system is built-in to alert the user to violations of set limits. The alarm system alerts the care giver by activating the audible alarm, screen display and the LED indicator.
The purpose of this 510(k) is to allow for the addition of a blower type compressor instead of the piston type compressor.
#### INDICATIONS FOR USE ನ.
The FLIGHT 60 Ventilator is intended to provide continuous or intermittent mechanical ventilation support for the care of individuals who require mechanical ventilation. Specifically, the FLIGHT 60 is applicable for adult and pediatric (i.e., infant, child and adolescent) patients, greater than or equal to 5kg (11 lbs). The FLIGHT 60 Ventilator is a restricted medical device intended for use by qualified, trained personnel under the direction of a physician; it is suitable for use in hospital, sub-acute, emergency room, and home care environments, as well as for transport and emergency response applications.
#### 6. COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE DEVICE
The suggested modification to the Flight 60 has been tested and validated. No new issues of safety or effectiveness have been identified as a result of this change. A comparison of FLIGHT 60's technological characteristics with the predicate devices is presented in the table below:
{6}------------------------------------------------
| DEVICE<br>NAME<br>SUBJECT | FLIGHT 60 (REVISED) | FLIGHT 60 | TRILOGY 100 |
|---------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| MANUFACTURER | Flight Medical Innovations Ltd. | Flight Medical Innovations Ltd. | Respironics Inc. |
| 510(K) NUMBER | | K130171 | K083526 |
| PRODUCT CODE | CBK, NOU | CBK, NOU | CBK |
| INTENDED USE | The FLIGHT 60 Ventilator is<br>intended to provide continuous or<br>intermittent mechanical ventilation<br>support for the care of individuals<br>who require mechanical ventilation.<br>Specifically, the FLIGHT 60 is<br>applicable for adult and pediatric<br>(i.e., infant, child and adolescent)<br>patients, greater than or equal to<br>5kg (11 lbs). | The FLIGHT 60 Ventilator is<br>intended to provide continuous or<br>intermittent mechanical<br>ventilation support for the care of<br>individuals who require<br>mechanical ventilation. Specifically,<br>the FLIGHT 60 is applicable for<br>adult and pediatric (i.e., infant,<br>child and adolescent) patients,<br>greater than or equal to 5kg (11<br>lbs). | The Philips Respironics Trilogy<br>100 system provides continuous<br>or intermittent ventilatory<br>support for the care of individuals<br>who require mechanical<br>ventilation. Trilogy 100 is<br>intended for pediatric through<br>adult patients weighing at least 5<br>kg (11 lbs). |
| PHYSICAL<br>DIMENSIONS | 11.641 x 11.457 x 9.803<br>(295mm x 291 mm x 249 mm) | 11.641 x 11.457 x 9.803<br>(295mm x 291 mm x 249 mm) | 16.68cm x 28.45cm x 23.52cm |
| USER INTERFACE | Touch screen with alpha numeric<br>data | Touch screen with alpha numeric<br>data | Control buttons, display screen |
| CONTROL BUTTONS | Touch screen function & buttons | Touch screen function & buttons | Buttons control panel |
| MECHANICAL<br>SYSTEM | Blower type compressor | Piston type compressor | Blower type compressor |
| POWER SUPPLY AC | 100-240V<br>50-60Hz | 100-240V<br>50-60Hz | 100-240V<br>50-60Hz |
| EXTERNAL DC | 12-30 V | 12-30 V | 12/14.4 V |
| INTERNAL BATTERY | Li- ion battery, 12V 5.2Ah | Li- ion battery, 12V 5.2Ah | Li- ion battery, 14.4V 4176mAh |
| BACK-UP BATTERY | Li-ion, 14.8V 2.6Ah | Li-ion, 14.8V 2.6Ah | Li- ion battery, 14.4V 4176mAh |
| | | | |
| AVAILABLE MODES | Volume/pressure/PRVC control<br>ACMV<br>Volume/pressure/PRVC control<br>SIMV + pressure support<br>SPONT + pressure support<br>Bilevel, MVG, VtG | Volume/pressure/PRVC control<br>ACMV<br>Volume/pressure/PRVC control<br>SIMV + pressure support<br>SPONT + pressure support<br>Bilevel, MVG, VtG | Volume/pressure control ACMV,<br>Volume/pressure control SIMV +<br>pressure support<br>CPAP (bilevel)<br>Bilevel, ST AVAPS, AVAPS |
| MANDATORY<br>BREATH CONTROLS | Volume/pressure control<br>Flow/Pressure/time trigger | Volume/pressure control<br>Flow/Pressure/time trigger | Volume/pressure control<br>Flow trigger |
| SPONTANEOUS<br>BREATH SUPPORT | Pressure support<br>Cycle variables: flow or time | Pressure support<br>Cycle variables: flow or time | Pressure support<br>Cycle variables: flow |
| BASIC PRESSURE<br>CONTROL | Pressure targeted waveform | Pressure targeted waveform | Pressure targeted waveform |
| BASIC FLOW<br>CONTROL | Descending/Square flow waveform | Descending/Square flow waveform | Ramp (=Descending )/Square<br>Flow Pattern |
| ALARMS | AC Disconnection, Low Battery,<br>Empty Battery, Patient<br>Disconnection | AC Disconnection, Low Battery,<br>Empty Battery, Patient<br>Disconnection | AC Power Disconnect, Low<br>Battery, Battery Depleted, Circuit<br>Disconnect |
| Non-invasive<br>positive<br>pressure<br>ventilation<br>(NPPV) | Yes | Yes | Yes |
| Reusable<br>exhalation valve | Yes | Yes | Yes |
{7}------------------------------------------------
As can be seen in the table above, the Revised Flight 60 has the same technological characteristics as the predicate Flight 60, with the exception of its compressor The same as Trilogy 100's compressor. Altogether, the technological characteristics of the Revised Flight 60 are substantially equivalent to the predicate Flight 60 or the Trilogy 100 (compressor type only). Thus, the comparison of the Revised Flight 60 to its predicate devices do not raise new safety and effectiveness concerns.
{8}------------------------------------------------
#### 7. PERFORMANCE DATA
Design and verification activities were performed on the FLIGHT 60 as a result of the risk analysis and product requirements. Verification of compliance with recognized standards has been made to support safe use of the device for its intended use and in its intended environment. Specifically, FLIGHT 60 was designed and tested in accordance with the applicable requirements in relevant FDA guidance documents and international standards including:
- FDA Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices (May 11, 2005)
- . IEC 60601-1: Medical Electrical Equipment, Part 1: General Requirements for Basic Safety and Essential Performance (2005), including US national deviations
- . ISO 80601-2-12: Medical Electrical Equipment - Part 2-12: Particular Requirements for Basic Safety and Essential Performance of Critical Care Ventilators (2011); and
- . IEC 60601-1-2: Medical Electrical Equipment - Part 1-2: Electromagnetic Compatibility - Requirements and Tests (2007):
- o ESD contact discharge and air discharge test levels were replaced by 8 kV and 15 kV, respectively.
- Radiated immunity test level was 30 V/m. o
- Magnetic field immunity test level was 30 A/m o
Flight 60 was subjected to environmental tests and was tested for Volatile Organic Compounds (VOC) by EPA test TO-15 and for particles by EPA's PM 2.5. All tests confirmed the product met the predetermined acceptance criteria. Testing also included non-clinical side-by-side waveform performance test, in which the waveform characteristics of Flight 60 were compared to those of Trilogy 100 (K083526) and to Flight 60 (K130171). Characteristics tested included flow, pressure and volume waveforms, ventilation control parameter accuracy and patient trigger reliability and synchrony. The comparison of the recorded waveforms supports the claim that FLIGHT 60 is substantially equivalent to the predicate devices.
#### 8. CONCLUSION
Flight Medical Innovations Ltd. believes that, based on the information provided in this submission, the FLIGHT 60 Ventilator is substantially equivalent to its predicate device without raising any new safety and effectiveness concerns.
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.