K112754 · Versamed Medical Systems , Ltd. · CBK · Jun 15, 2012 · Anesthesiology
Device Facts
Record ID
K112754
Device Name
IVENT101
Applicant
Versamed Medical Systems , Ltd.
Product Code
CBK · Anesthesiology
Decision Date
Jun 15, 2012
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 868.5895
Device Class
Class 2
Attributes
Therapeutic, Pediatric
Indications for Use
The iVent101 ventilator [with single or dual limb configuration] is intended to provide continuous or intermittent ventilatory support for the care of individuals who require mechanical ventilation. Specifically, the ventilator is applicable for infants from 5 kg through adult patients who require invasive or non-invasive support via the following ventilatory modalities: - Assist/Control and SIMV with either Volume, Pressure Regulated Volume Control (PRVC) - CPAP with Pressure Support and CPAP / Pressure Support Volume Guaranteed (VG) - Adaptive Bi-Level for either NIV or invasive ventilation and Adaptive Bi-Level Volume Guaranteed [VG] The iVent101 ventilator is suitable for use in institution, home and portable settings
Device Story
Compact, portable, microprocessor-controlled ventilator; provides continuous/intermittent ventilatory support. Turbine-powered air source; internal battery; operates via AC/DC power. Inputs: internal flow and pressure sensors. Outputs: real-time clinical data (Tidal Volume, Rate, PIP, FiO2, Peak Flow, I:E, Minute Volume) displayed on LCD touch-screen. Operator (clinician/caregiver) performs all actions via touch-screen; adjusts alarm settings. Optional software displays waveforms, loops, trends, and logs. Supports single or dual limb patient circuits. Benefits: provides ventilatory support in diverse settings (institution, home, portable) with freedom from wall power/air.
Clinical Evidence
No clinical data; bench testing only.
Technological Characteristics
Microprocessor-controlled, turbine-powered ventilator. Features internal flow and pressure sensors. Supports single or dual limb patient circuits. LCD touch-screen interface. Power: AC/DC and integrated rechargeable battery. Software-driven ventilation modes including Assist/Control, SIMV, PRVC, CPAP/PSV, and Adaptive Bi-Level with Volume Guaranteed options.
Indications for Use
Indicated for infants (from 5 kg) through adult patients requiring invasive or non-invasive continuous or intermittent mechanical ventilatory support. Suitable for use in institutional, home, and portable settings.
Regulatory Classification
Identification
A continuous ventilator (respirator) is a device intended to mechanically control or assist patient breathing by delivering a predetermined percentage of oxygen in the breathing gas. Adult, pediatric, and neonatal ventilators are included in this generic type of device.
{1}------------------------------------------------
KII2754
JUN 15 2012 Page 1 of 3
## 510(k) Summary
In accordance with 21 CFR 807.92 the following summary of information is provided:
September 13, 2011 Date: GE Healthcare, VersaMed Medical Systems Submitter: Hasharon Industrial Park Ornat Bldg. p.o. box 5011 kadima, Israel 60920 Shlomi Deler - Regulatory Affairs Manager Primary Contact Person: GE Healthcare, VersaMed Medical Systems Tel : +972 98915407 Fax: +972 98911499 Secondary Contact Monica Morrison Person: GE Heathcare, Datex-Ohmeda, Inc. 3030 Ohmeda Drive Madison, WI 53718 Tel: 608-221-1551, Extension 3439 Fax: 608-646-7464 Email: Monica.Morrison@ge.com Pam Schaub US Agent: GE Healthcare 9900 w. innovation dr rp2138 wauwatosa, WI 53226 Phone: 414 7212584 ext Fax: 262 9971160 Email: pam.schaub@ge.com Device: Trade iVent101 Name: Common/Usual Portable Ventilator Name: Classification Names: 868-5895 , Continuous Ventilator 73 CBK, 73 NOU Product Code:
iVent101 [cleared under K092135] Predicate Device(s): Trilogy 100 [cleared under K083526]
{2}------------------------------------------------
### Page 2 of 3
Device The iVent101 is a compact, portable, microprocessor-controlled Description: ventilator intended to provide continuous or intermittent ventilatory support for the care of individuals who require mechanical ventilation.
> A turbine-powered air source and a rechargeable internal battery provide freedom from wall air and power outlets. Internal flow and pressure are read through low/pressure sensors. Clinical data [Tidal Volume, Rate, PIP, FiO2, Peak Flow, Inspiratory Time, I:E Minute Volume) are presented on machine screen.
> All the operator actions are performed on the LCD touch-screen on the front panel, allowing rapid control and continuous real-time monitoring of patient ventilation. Alarm settings are fully adjustable.
> Optional Waveform and Diagnostic Software package displays pressure and flow waveform data, loops, trends, and logged totals in a full array of time slices and presentation modes.
> The iVent101 can use external AC or DC power supply and contains an integrated battery.
> The iVent101 is equipped with two configurations, which differs in the patient circuit type that connects to the machine. One configuration is to be used with a standard one limb patient circuit, and the second configuration is to be used with a standard dual limb patient circuit. The two configurations are incorporating the same infrastructure and can be easily replaced by the user.
- The iVent101 ventilator [with single or dual limb configuration] is Intended Use: intended to provide continuous or intermittent ventilatory support for the care of individuals who require mechanical ventilation. Specifically, the ventilator is applicable for infants from 5 kg through adult patients who require invasive or non-invasive support via the following ventilatory modalities:
- Assist/Control and SIMV with either Volume. Pressure, or . Pressure Requlated Volume Control (PRVC)
- CPAP with Pressure Support and CPAP / Pressure Support . Volume Guaranteed [VG]
- . Adaptive Bi-Level for either NIV or invasive ventilation and Adaptive Bi-Level Volume Guaranteed [VG]
{3}------------------------------------------------
# Page 3 of 3
The iVent101 ventilator is suitable for use in institution, home and portable settings
The iVent101 has been updated from the predicate version Technology:
> K092135. There have been no changes to the general intended use of the device to provide continuous or intermittent ventilator support or to the fundamental scientific technology. The indications for use was updated to include two additional ventilation modes CPAP/PSV Volume Guaranteed [VG] and Adaptive Bi-Level Volume Guaranteed (VG).
> The software for the iVent101 in this submission has been updated to introduce minor software modifications and additional ventilation modes.
#### Determination of Summary of Non-Clinical Tests:
The iVent101 has been thoroughly tested through verification of Substantial Specifications and validation, including software validation. The Equivalence: following quality assurance measures were applied to the development of the system:
- 1. Risk Analysis
- 2. Requirements Reviews
- 3. Design Reviews
- 4. Testing on unit level (Module verification)
- 5. Integration testing (System verification)
- 6. Performance testing (Verification)
- 7. Safety testing (Verification)
- 8. Simulated use testing (Validation)
## Summary of Clinical Tests:
The subject of this premarket submission, the iVent101, did not require clinical studies to support substantial equivalence.
### Conclusion:
GE Healthcare, VersaMed Medical Systems considers the iVent101 to be as safe, as effective, and the performance to be substantially equivalent to the predicate devices.
{4}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for the United States Department of Health and Human Services. The logo features the department's name in a circular arrangement around a stylized emblem. The emblem consists of a stylized caduceus, a symbol often associated with healthcare, with three parallel lines forming the snake and staff.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
Mr. Shlomi Deler Regulatory Affairs Manager VersaMed Medical Systems Limited Hasharon Industrial Park Ornat Bldg. P.O. Box 5011 Kadima ISRAEL 60920
JUN 2 9 2012
un 2 3
US 28
Re: K112754
Trade/Device Name: iVent101 Regulation Number: 21 CFR 868.5895 Regulation Name: Continuous Ventilator Regulatory Class: II Product Code: CBK, NOU Dated: May 4, 2012 Received: May 7, 2012
Dear Mr. Deler:
This letter corrects our substantially equivalent letter of June 15, 2012.
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations. Title 21. Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{5}------------------------------------------------
Page 2- Mr. Deler
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to
http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers. International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
for
Anthony D. Watson, B.S., M.S., M.B.A. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
510(k) Number (if known):
iVent101 Device Name:
Indications for Use:
The iVent101 ventilator [with single or dual limb configuration] is intended to provide continuous or intermittent ventilatory support for the care of individuals who require mechanical ventilation. Specifically, the ventilator is applicable for infants from 5 kg through adult patients who require invasive or non-invasive support via the following ventilatory modalities:
- Assist/Control and SIMV with either Volume, Pressure Regulated Volume . Control (PRVC)
- CPAP with Pressure Support and CPAP / Pressure Support Volume Guaranteed (VG) ●
- Adaptive Bi-Level for either NIV or invasive ventilation and Adaptive Bi-Level Volume ● Guaranteed [VG]
The iVent101 ventilator is suitable for use in institution, home and portable settings
Prescription Use xxx (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use (Part 21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Schulten
Division Sign-Off) Offision of Anesthesiology, General Hospital nfection Control, Dental Devices
510(k) Number: k112754
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.