CLASSICSTAR NV SE AND NOVASTAR NV SE AND AAV FULL FACE MASKS
K081743 · Draeger Medical AG & Co. KG · CBK · Sep 22, 2008 · Anesthesiology
Device Facts
Record ID
K081743
Device Name
CLASSICSTAR NV SE AND NOVASTAR NV SE AND AAV FULL FACE MASKS
Applicant
Draeger Medical AG & Co. KG
Product Code
CBK · Anesthesiology
Decision Date
Sep 22, 2008
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 868.5895
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
ClassicStar™, NV, Full Face Mask, SE (with standard elbow), size S-L: The mask ClassicStar noninvasive ventilation (NV) with standard elbow (SE) is intended to provide a patient interface for application of noninvasive ventilation. The mask is to he used as an accessory to ventilators that have adequate alarms and safety systems for ventilator failure and which are intended to administer CPAP or positive pressure ventilation for treatment of respiratory failure or respiratory insufficiency. It is intended for use on adult patients (>30 kg/66 lbs), who are appropriate candidates for noninvasive ventilation in the hospital or institutional environment. The mask is disposable and for single patient use. NovaStar™, NV, Full Face Mask, SE (with standard elbow), size S-L: The mask NovaStar noninvasive ventilation (NV) with standard elbow (SE) is intended to provide a patient interface for application of noninvasive ventilation. The mask is to be used as an accessory to ventilators that have adequate alarms and safety systems for ventilator failure. The full face mask NovaStar, noninvasive ventilation (NV), with anti-asphyxia valve (AAV) is intended to be used with positive airway pressure devices, operating at or above 3 mbar (3 cmH2O). The mask contains exhalation ports and does not require the use of a separate exhalation device. It is intended for use on adult patients (>30 kg), who are appropriate candidates for noninvasive ventilation in the hospital, institutional and in the home environments. The mask can be used multiple times on multiple patients. Reuse, however, is limited up to 5 times.
Device Story
Full-face masks providing patient interface for noninvasive ventilation; covers mouth and nose. ClassicStar (disposable) and NovaStar (reusable up to 5 times) variants available with standard elbow (SE) or anti-asphyxia valve (AAV). SE models require ventilators with integrated alarm/safety systems; AAV models include exhalation ports and anti-asphyxia valve. Used in hospital, institutional, or home environments by clinicians or patients. Masks connect to positive pressure ventilation devices to treat respiratory failure/insufficiency. Output is delivered air/pressure to patient airway; facilitates respiratory support and patient benefit through noninvasive ventilation.
Clinical Evidence
No clinical data provided; bench testing only.
Technological Characteristics
Full-face mask interface; available in sizes S, M, L. Materials include headgear and mask components. SE models feature standard elbow; AAV models feature integrated anti-asphyxia valve and exhalation ports. Reusable models (NovaStar) validated for up to 5 uses. No electronic components, software, or energy sources.
Indications for Use
Indicated for adult patients (>30 kg/66 lbs) requiring noninvasive ventilation (CPAP or positive pressure) for respiratory failure or insufficiency. SE models require ventilators with adequate alarms/safety systems; AAV models require positive airway pressure devices operating at ≥3 mbar. Used in hospital, institutional, or home (AAV only) settings.
Regulatory Classification
Identification
A continuous ventilator (respirator) is a device intended to mechanically control or assist patient breathing by delivering a predetermined percentage of oxygen in the breathing gas. Adult, pediatric, and neonatal ventilators are included in this generic type of device.
{0}------------------------------------------------
# 510(k) Summary
## Summary of Safety and Effectiveness
#### Applicants Name and Address:
Dräger Medical AG & Co. KG Moislinger Allee 53-55 23542 Lübeck Germany
Establishment Registration Number: 9611500
#### Contact Person:
Dr. Karin Luebbers Senior Manager Regulatory Affairs
Tel. No.: + 49 (451) 882-5367 Fax No .: + 49 (451) 882-7-5367
#### Applicants US Contact Person
Ms. Jovce Kilrov VP Processes, Quality & Regulatory Affairs
Tel. No .: (215) 660-2626 Fax No .: (267) 885-9989
#### Date submission was prepared:
2008-09-09
#### Device Name:
ClassicStar™, NV, Full Face Mask, SE Trade Name: Common Name: Full Face Mask, Noninvasive ventilation Trade Name: NovaStar™, NV, Full Face Mask, SE Common Name: Full Face Mask, Noninvasive ventilation NovaStar™, NV, Full Face Mask, AAV Trade Name: Common Name: Full Face Mask, Noninvasive ventilation
{1}------------------------------------------------
### Noninvasive Ventilation Masks
#### Classification:
| Class II | | |
|----------------|--------------------------------------|--------------|
| Regulation No. | Device | Product Code |
| 868.5895 | Ventilator, continuous, facility use | 73CBK |
Legally Marketed Device to which Substantial Equivalence is claimed:
| 510(k) number | Trade name | Company |
|---------------|-------------------------------------------------|------------------|
| K023135 | Image3 SE Disposable Full Face Mask | Respironics Inc. |
| K063806 | SleepNet MoJoTM -NV Full Face Mask<br>nonvented | SleepNet Corp. |
| K060273 | SleepNet MoJoTM Full Face Mask | SleepNet Corp. |
### Device Description:
Within the medical device family "Noninvasive Ventilation Masks" are devices to provide a patient interface for the application of noninvasive ventilation.
The masks ClassicStar and NovaStar are Full-Face masks which cover the mouth and the nose and are available with a standard elbow (SE) or an anti-asphyxia valve (AAV). Masks with a standard elbow (SE) may only be used on ventilation devices, which incorporate adequate alarm and safety systems for ventilation failure.
A mask with an anti-asphyxia valve (AAV) incorporates the anti-asphyxia valve in the mask elbow.
The ClassicStar Masks are disposable, while the NovaStar Masks are reusable up to 5 times for multiple patients. Further differences are the headgears, which are similar in materials but differ in means of connection to the mask. The headgears are generally provided with the masks, for the reusable masks further headgears are available as optional accessory.
All noninvasive ventilation masks are available in three different sizes (S, M, L),
#### Intended Use:
ClassicStar™, NV, Full Face Mask, SE (with standard elbow), size S-L:
The mask ClassicStar noninvasive ventilation (NV) with standard elbow (SE) is intended to provide a patient interface for application of noninvasive ventilation. The mask is to he used as an accessory to ventilators that have adequate alarms and safety systems for ventilator failure and which are intended to administer CPAP or positive pressure ventilation for treatment of respiratory failure or respiratory insufficiency. It is intended for use on adult patients (>30 kg/66 lbs), who are appropriate candidates for noninvasive ventilation in the hospital or institutional environment. The mask is disposable and for single patient use. NovaStar™, NV, Full Face Mask, SE (with standard elbow), size S-L:
The mask NovaStar noninvasive ventilation (NV) with standard elbow (SE) is intended to provide a patient interface for application of noninvasive ventilation. The mask is to be used as an accessory to ventilators that have adequate alarms and safety systems for ventilator
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the seal of the Department of Health & Human Services (HHS) of the United States. The seal features a stylized eagle with its wings spread, symbolizing protection and service. Encircling the eagle is the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA", indicating the department's name and national affiliation.
SEP 2 2 2008
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Ms. Karin Lübbers Senior Manger Regulatory Affairs Draeger Medical AG & Co. KG 53/55 Moislinger Allee Luebeck GERMANY 23542
Re: K081743
Trade/Device Name: ClassicStar™, NV, Full Face Mask, SE NovaStar™, NV, Full Face Mask, SE NovaStarTM, NV, Full Face Mask, AAV Regulation Number: 21 CFR 868.5895 Regulation Name: Continuous Ventilator Regulatory Class: II Product Code: CBK Dated: September 10, 2008 Received: September 15, 2008
Dear Ms. Lübbers:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{3}------------------------------------------------
Page 2 - Ms. Lübbers
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801). please contact the Center for Devices and Radiological Health's (CDRH's) Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometric's (OSB's) Division of Postmarket Surveillance at 240-276-3474. For questions regarding the reporting of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at 240-276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
ThAnnuele Yendms foell
Chiu Lin, Ph.D. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
# Indications for Use
510(k) Number (if known): K
Device Name:
ClassicStar ™, NV, Full Face Mask, SE Star TM, NV, Full Face Mask, SE
NovaStar TM, NV, Full Face Mask, AAV
Indications for Use:
The mask ClassicStar noninvasive ventilation (NV) with standard elbow (SE) and the mask NovaStar noninvasive ventilation (N); with standard elbow (SE) are intended to provide a patient interface for application of noninvasive ventilation. The masks are to be used as an accessory to ventilators that have adequate alarms and safety systems for ventilator failure and which are intended to administer CPAP or positive pressure ventilation for treatment of respiratory failure or respiratory insufficiency. The masks are intended for use on adult patients (>30 kg/66 lbs), who are appropriate candidates for noninvasive ventilation in the hospital or institutional environment.
The mask ClassicStar NV, with standard elbow is disposable and for single patient use.
The mask NovaStar NV, with standard elbow can be used multiple times on multiple patients. Reuse, however, is limited up to 5 times.
The full face mask NovaStar, noninvasive ventilation (NV), with anti-asphyxia valve (AAV) is intended to be used with positive airway pressure devices, operating at or above 3 mbar (3 cmH2O). The mask contains exhalation ports and does not require the use of a separate exhalation device. It is intended for use on adult patients (>30 kg), who are appropriate candidates for noninvasive ventilation in the hospital, institutional and in the home environments. The mask can be used multiple times on multiple patients. Reuse, however, is limited up to 5 times.
Prescription Use
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
M. The
(Division Sign-Off) Division of Anesthesiology, General Hospital Infection Control, Dental Devices
Page *1* of *1*
Ko81743 510(k) Number: -
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.