MODIFICATION TO SERVO-I VENTILATOR SYSTEM, MODEL 64 87 800 E407E
K041223 · Maquet Critical Care AB · CBK · Jul 29, 2004 · Anesthesiology
Device Facts
Record ID
K041223
Device Name
MODIFICATION TO SERVO-I VENTILATOR SYSTEM, MODEL 64 87 800 E407E
Applicant
Maquet Critical Care AB
Product Code
CBK · Anesthesiology
Decision Date
Jul 29, 2004
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 868.5895
Device Class
Class 2
Attributes
Therapeutic, Pediatric
Indications for Use
The Servo-i Ventilator System is intended for treatment and monitoring of patients in the range of neonates, infants and adults with respiratory failure or respiratory insufficiency. Servo-i is a ventilator for use in healthcare facilities and for in-hospital transport.
Device Story
The NIV option is a software-controlled feature for the Servo-i ventilator system; enables non-invasive ventilation (NIV) mode. Device monitors patient respiratory status; microprocessor controls flow capabilities to meet patient needs in non-invasive applications. Includes independent alarm functions for excessive leakage and inability to maintain minimum PEEP. If leakage decreases, ventilation resumes automatically or via manual start. Used in healthcare facilities and for in-hospital transport by clinicians. Output provides respiratory support; assists in managing respiratory failure/insufficiency.
Clinical Evidence
No clinical data provided; substantial equivalence supported by bench testing of the software-controlled NIV feature and alarm functions.
Technological Characteristics
Software-controlled ventilation feature for Servo-i platform. Microprocessor-based control of flow for non-invasive application. Includes independent alarm functions for leakage detection and PEEP maintenance. Operates as an add-on software option for existing ventilator hardware.
Indications for Use
Indicated for neonates, infants, and adults with respiratory failure or respiratory insufficiency requiring treatment and monitoring. Used in healthcare facilities and for in-hospital transport. Infant weight range for NIV option is 3 kg to adult.
Regulatory Classification
Identification
A continuous ventilator (respirator) is a device intended to mechanically control or assist patient breathing by delivering a predetermined percentage of oxygen in the breathing gas. Adult, pediatric, and neonatal ventilators are included in this generic type of device.
K023289 — OPTION NIV, MODEL 8414115, UPGRADE KIT NIV, MODEL 8414072 · Drager Medizintechnik GmbH · Jan 15, 2003
K230173 — Servo-air Lite Ventilator System · Maquet Critical Care AB · Jul 6, 2023
K222644 — Nihon Kohden NKV-440 Ventilator System · Nihon Kohden Orangemed, Inc. · May 31, 2023
K103803 — HAMILTON-G5 · Hamilton Medical AG · Nov 23, 2011
K022674 — BIRD AVEA VENTILATOR · Bird Products Corp. · Nov 6, 2002
Submission Summary (Full Text)
{0}------------------------------------------------
JUL 2 9 2004
K041223
Section-Page
EVU-116506
Doc-ID
135 (140)
Issue no.
- 00
| | | :<br> | <br>1 |
|--|--|-------|-------|
| | | | |
Object/Subject NIV option for Servo-i
# 510 (k) Summary
as required by section 807.92(c)
### Subscribers Name & Address
Maquet Critical Care AB Röntgenvägen 2 SE-171 95 Solna, Sweden Tel: (011) 46 8 730 73 00 Fax: (011) 46 8 730 78 38
Contact Persons for this submission:
Mr. Anders Palm Tele. direct; (011) 46 8 730 74 88 Email: anders.palm@maquet.com
Document Type
Special 510(k)
USA Contact :
:
Mr. Jamie Yieh Manager, Regulatory Affairs Maquet Inc. 1140 Route 22 East, Suite 202 Bridgewater, NJ 08807 Phone: 908-947-2311 Fax: 908-947-2301 Cellphone; 908-227-8807 Email: jamie.yieh@maquet-inc.com
#### Trade Names
| Servo-i Ventilator System | article no.; 64 87 800 E407E | | |
|----------------------------------------------------------------|------------------------------|-------|-------------------|
| NIV option | article no.; 66 67 187 E407E | | |
| Device Classification | | | |
| Common Name | Classification<br>Number | Class | Regulation Number |
| Ventilator, Continuous (Respirator) | 73 CBK | II | 868.5895 |
| Predicate Device Identification | | | |
| Legally marketed devices to which equivalence is being claimed | | | 510(k) # |
| Servo-i Ventilatory System | | | K040221 |
| Evita 4, NIV option | | | K010093 |
{1}------------------------------------------------
| | Document Type<br>Special 510(k) | Section-Page | 136 (140) |
|------------------------|---------------------------------|--------------|-----------|
| Object/Subject | | Doc-ID | Issue no. |
| NIV option for Servo-i | | EVU-116506 | - 00 |
## Device Description
The ventilator is a platform with several selectable ventilation modes which monitor patients whom need respiratory assistance.
## Summary of technological characteristics of modified Device and Predicate Device: NIV option for Non -Invasive ventilation
The Non-Invasive Ventilation Option is a software controlled feature available on the Servo-i ventilator. The microprocessor control of this feature allows a larger range of flow capabilities designed to meet patient needs safely in a non-invasive application.
To detect disconnect or excessive leakage there are two independent alarm functions:
- an alarm will be given if the leakage is excessive. .
- if the minimum PEEP will be unable to maintain (as in predicate device Servo-i) .
If leakage decreases the ventilation will resume automatically (or by manual start as in predicate Servo-i).
### Intended Use of the Device:
The intended use(s) and indications of the Servo-i application, as described in its labelling, are the same as the intended uses and indications for the unmodified Servo-i, except for the intended population for Infant using the NIV option where the lower range is changed from 0,5 Kg to 3 KE:
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is circular and contains the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the top half of the circle. Inside the circle is a stylized image of an eagle or bird with three curved lines representing its wings or body.
#### Public Health Service
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
# JUL 2 9 2004
Maquet Critical Care AB C/O Mr. Jamie Yieh Manager, Regulatory Affairs Maquet, Incorporated 1140 Route 22 East, Suite 202 Bridgewater, New Jersey 08807
Re: K041223
Trade/Device Name: Modification to Servo-1 Ventilator System, Model 64 87 800 Regulation Number: 868.5895 Regulation Name: Continuous Ventilator Regulatory Class: II Product Code: CBK Dated: May 10, 2004 Received: May 10, 2004
Dear Mr. Yieh:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting (1 wr ), ice can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{3}------------------------------------------------
Page 2 - Mr. Yieh
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice and if the (2 - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), nl Jase contact the Office of Compliance at (301) 594-4646. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html
Sincerely yours,
Chiu Lin, Ph.D.
Chiu Lin, Ph.D. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
| MAQUET | | |
|---------------------------------|--------------------------|-----------|
| Document Type<br>Special 510(k) | Section-Page<br>28 (140) | |
| Object/Subject | Doc-ID | Issue no. |
| NIV option for Servo-i | EVU-116506 | - 00 |
# Indications for Use
510(k) Number (if known):
Device Name: Servo-i Ventilator System
#### Indications For Use:
INGICALONS I OF OSC.
The Servo' Ventilator System is intended for treatment and monitoring of patients in the range of I ne or re Fentually with respiratory failure or respiratory insufficiency. Servo' is a ventilator neonales, thanks and datas will respiratery gain. I o realthcare facilities and for in-hospital transport.
| Prescription Use | X |
|-----------------------------|---|
| (Part 21 CFR 801 Subpart D) | |
AND/OR
| Over-The-Counter Use | |
|------------------------|--|
| (21 CFR 807 Subpart C) | |
(Please do not WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Ouu Sulton
Page 1 of _
(Division Sign-Qff)
Division of Anesthesiology, General Hospital,
Infection Control, Dental Devices
**510(k) Number:** *_*_
spital,
Panel 1
/
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.