← Product Code [CBI](/submissions/AN/subpart-f%E2%80%94therapeutic-devices/CBI) · K141888

# RUSCH ENDOBRONCHIAL TUBES (K141888)

_Teleflexmedical, Inc. · CBI · Apr 15, 2015 · Anesthesiology · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/AN/subpart-f%E2%80%94therapeutic-devices/CBI/K141888

## Device Facts

- **Applicant:** Teleflexmedical, Inc.
- **Product Code:** [CBI](/submissions/AN/subpart-f%E2%80%94therapeutic-devices/CBI.md)
- **Decision Date:** Apr 15, 2015
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 868.5740
- **Device Class:** Class 2
- **Review Panel:** Anesthesiology
- **Attributes:** Therapeutic

## Indications for Use

Rusch Endobronchial Tubes are used to isolate the left or right lung of a patient for surgery, one lung ventilation or one lung anesthesia.

## Device Story

Rusch Endobronchial Tube is a sterile, single-use PVC double-lumen tube; inserted into trachea via mouth to selectively ventilate one lung. Used in hospital OR and ICU settings by clinicians. Device features two cuffs and side arm assemblies for inflation; includes stylet; available in Robertshaw, Carlens, and White styles. Radiopaque design assists in placement. Clinicians use device to isolate lungs during surgery or anesthesia, facilitating one-lung ventilation. Benefits include controlled lung isolation for thoracic procedures.

## Clinical Evidence

Bench testing only. Performance testing included connector bonding strength, cuff resting diameter, tube collapse, cuff herniation, cuff burst evaluation, cuff bond strength, and side arm bonding strength. Biocompatibility testing (cytotoxicity, sensitization, intracutaneous activity) performed per ISO 10993-1.

## Technological Characteristics

Double-lumen tube; PVC construction; 26-41 French size range; 15 mm connector; radiopaque; cuffed. Compliant with ISO 5356 (connectors) and ISO 5361 (tracheal tubes). Sterile, single-use. No software or electronic components.

## Regulatory Identification

A tracheal/bronchial differential ventilation tube is a device used to isolate the left or the right lung of a patient for anesthesia or pulmonary function testing.

## Predicate Devices

- Silbroncho Double Lumen Tube ([K051522](/device/K051522.md))
- Well Lead Endobronchial Tubes ([K092886](/device/K092886.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, stacked on top of each other.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

April 15, 2015

Teleflex Medical, Inc. Lori Pfohl Senior Regulatory Affairs Specialist 2917 Weck Dr. Research Triangle Park, NC 27709

Re: K141888

Trade/Device Name: Rusch Endobronchial Tubes Regulation Number: 21 CFR 868.5740 Regulation Name: Tracheal/Bronchial Differential Ventilation Tube Regulatory Class: Class II Product Code: CBI Dated: March 13, 2015 Received: March 16, 2015

Dear Ms. Pfohl:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Page 2 - Ms. Pfohl

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

Tejashri Purohit-Sheth, M.D.

Tejashri Purohit-Sheth, M.D. Clinical Deputy Director DAGRID/ODE/CDRH FOR

Erin I. Keith, M.S. Division Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# Indications for Use

510(k) Number (if known)

K141888

Device Name Rusch Endobronchial Tubes

Indications for Use (Describe)

Rusch Endobronchial Tubes are used to isolate the left or right lung of a patient for surgery, one lung ventilation or one lung anesthesia.

Patient Population: Patients requiring one lung isolation

Environment of use: Hospitals - OR and ICU

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.

#### FOR FDA USE ONLY

Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)

This section applies only to requirements of the Paperwork Reduction Act of 1995.

### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number,"

Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below.

{3}------------------------------------------------

# 510(k) SUMMARY March 4, 2015

### Name, Address, Phone and Fax Number of Applicant

Teleflex Medical, Incorporated 2917 Weck Drive Research Triangle Park, NC 27709 USA Phone: 919-491-8960 Fax: 919-433-4996

### Contact Person

Lori Pfohl Senior Regulatory Affairs Specialist

### Device Name

Trade Name: Rusch Endobronchial Tubes

Common Name: Bronchial Tube

Classification Name: Tube, Trachael/Bronchial, Differential Ventilation (w/wo connector) (Class II per 21 CFR 868.5740, Product Code CBI)

### Predicate Devices

K051522 Silbroncho Double Lumen Tube K092886 Well Lead Endobronchial Tubes

### Device Description

The Rusch Endobronchial Tube is a sterile, single patient use PVC Double-Lumen Endobronchial Tube (also referred to as a DLT) that is inserted into the trachea via the mouth in order to selectively ventilate one lung. The Endobronchial tube is sold with or without accessories, contains a stylet and is available in Robertshaw, Carlens and White styles.

### Intended Use

Rusch Endobronchial Tubes are used to isolate the left or right lung of a patient for surgery, one lung ventilation or one lung anesthesia.

Patient Population: Patients requiring one lung isolation

Environment of use: Hospitals - OR and ICU

Teleflex Medical

{4}------------------------------------------------

# 510(k) SUMMARY March 4, 2015

## Contraindications

Left sides versions are contraindicated in patients with obstructions or stenosis in the left main bronchus. The right sided versions are contraindicated in patients with obstructions or stenosis in the right main bronchus. The versions with a carina hook are contraindicated for all procedures in the region of the carina.

### Substantial Equivalence Comparison to Predicates

| Features                            | Teleflex Medical<br>Endobronchial Tube<br>(proposed)                                                                                                                                                                                                                                                                                                    | Fuji Systems Silcobronch<br>(K051522) | Well Lead<br>EndobronchialTube<br>(K092886) |
|-------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------|---------------------------------------------|
| Classification Name                 | Tube, Trachael/Bronchial, Differential<br>Ventilation (w/wo connector)                                                                                                                                                                                                                                                                                  | Same                                  | Same                                        |
| Product Code                        | CBI                                                                                                                                                                                                                                                                                                                                                     | Same                                  | Same                                        |
| Regulation Number                   | 868.5740                                                                                                                                                                                                                                                                                                                                                | Same                                  | Same                                        |
| Indications for Use                 | Rusch Endobronchial Tubes are used to<br>isolate the left or right lung of a patient<br>for surgery, one lung ventilation or one<br>lung anesthesia                                                                                                                                                                                                     | Same                                  | Same                                        |
| Environment of Use                  | Hospital - OR and ICU                                                                                                                                                                                                                                                                                                                                   | Same                                  | Not stated                                  |
| Patient Population                  | Patients requiring one lung isolation                                                                                                                                                                                                                                                                                                                   | Same                                  | Not stated                                  |
| Contraindications                   | Left sides versions are contraindicated in<br>patients with obstructions or stenosis in<br>the left main bronchus. The right sided<br>versions are contraindicated in patients<br>with obstructions or stenosis in the right<br>main bronchus. The versions with a<br>carina hook are contraindicated for all<br>procedures in the region of the carina | Not Stated                            | Not Stated                                  |
| Design Features                     | Double lumen shaft, 2 cuffs, Stylet                                                                                                                                                                                                                                                                                                                     | Same                                  | Same                                        |
| Single Use                          | Yes                                                                                                                                                                                                                                                                                                                                                     | Same                                  | Same                                        |
| Size Range                          | 26-41 French                                                                                                                                                                                                                                                                                                                                            | 33-39 French                          | 28-41 French                                |
| Cuffed                              | Yes                                                                                                                                                                                                                                                                                                                                                     | Same                                  | Same                                        |
| Radiopaque                          | Yes                                                                                                                                                                                                                                                                                                                                                     | No                                    | Yes                                         |
| Connection to<br>ventilation source | 15 mm connector                                                                                                                                                                                                                                                                                                                                         | Same                                  | Same                                        |
| DLT Materials                       | PVC                                                                                                                                                                                                                                                                                                                                                     | Silicone                              | PVC                                         |

The proposed device is substantially equivalent to the predicate devices:

- Indications for Use-The indications for use are identical for the proposed device when 9 compared to the predicates. Each device is used to isolate the left or right lung of a patient for surgery, one lung ventilation or one lung anesthesia.
- Technology and construction The design, fabrication, shape, size, etc. are equivalent to the . predicates. The designs of the propose and predicate devices consist a double lumen tube with

{5}------------------------------------------------

## 510(k) SUMMARY March 4, 2015

cuffs on each lumen, and corresponding side arm assemblies to inflate the cuffs.

- Environment of use Identical to predicate. Hospital OR and ICU .
- Patient Population Identical to predicate. Patients requiring one lung isolation _
- Materials -All patient contacting materials are in compliance with ISO 10993-1. Testing 0 included cytotoxicity, sensitization and intracutaneous activity.

### Comparison to Predicate Device:

The proposed Endobronchial tubes are substantially equivalent to the predicate devices with respect to indications for use, technology and construction. The differences between the predicate and the proposed devices are minor and any risks have been mitigated through testing. Additionally, the proposed device is made with the same materials as the K092886 predicate.

## Non-clinical Comparative Performance Testing

A brief summary of tests relied upon to demonstrate substantial equivalence to the predicate can be found in the table below:

| Test                          | Reference to<br>Standard (if<br>applicable) | Principle of Test                                                                                                                                                                                                                                                                                                                                                |
|-------------------------------|---------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Connector<br>bonding strength | ISO 5356                                    | The security of the attachment of the connector to the<br>Endobronchial tube is tested by applying an axial separation<br>force to the connector                                                                                                                                                                                                                 |
| Cuff resting<br>diameter      | ISO 5361                                    | The resting diameter of the cuff is measured when the cuff<br>is inflated to a reference pressure which is intended to<br>remove creases but minimize stretching of its walls                                                                                                                                                                                    |
| Tube collapse                 | ISO 5361                                    | The patency of the Endobronchial tube airway lumen is<br>tested by passing a steel ball through the Endobronchial<br>tube lumen with the cuff inflated within a transparent tube                                                                                                                                                                                 |
| Cuff herniation               | ISO 5361                                    | The tendency of the cuff to herniate beyond the plane<br>perpendicular to the long axis of the tube at the nearest<br>edge of the bevel is tested by applying an axial force with<br>the cuff inflated within a transparent tube. A cuff which<br>protrudes excessively at its patient end may partially or<br>completely occlude the orifice at the patient end |
| Cuff Burst<br>Evaluation      | N/A                                         | The cuff restrained burst test is designed to ensure the cuff<br>will not burst or rupture when inflated inside the trachea                                                                                                                                                                                                                                      |
| Cuff Bond<br>Strength         | N/A                                         | To evaluate the strength needed to separate the cuff from<br>the tube                                                                                                                                                                                                                                                                                            |
| Side arm<br>bonding strength  | N/A                                         | To evaluate the retention force of the inflation line<br>connection to the Endobronchial tube                                                                                                                                                                                                                                                                    |

### Substantial Equivalence Conclusion

The Rusch Endobronchial Tubes have the same indications for use, patient population and technology of construction as the predicate devices. Performance test results demonstrate that the proposed device meets its intended use. It is for these reasons that the proposed device can be found substantially equivalent.

---

**Source:** [https://fda.innolitics.com/submissions/AN/subpart-f%E2%80%94therapeutic-devices/CBI/K141888](https://fda.innolitics.com/submissions/AN/subpart-f%E2%80%94therapeutic-devices/CBI/K141888)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
