It is the expressed, intended use of the Fink Engineering SL8/DL8 and TL20 Hyperbaric Oxygen Treatment Facilities to provide therapy to those patients with selected medical conditions that have been determined to respond to the application of hyperbaric oxygen. As a Class II prescriptive device, it is further intended for physician involvement in their procurement and routine use.
Device Story
Multiplace hyperbaric chambers (SL8, DL8, TL20) provide hyperbaric oxygen therapy (HBO) at pressures up to 3.0 or 6.0 ATA. Chambers accommodate 8 to 20 patients. Input: pressurized air for chamber environment; 100% oxygen for patient delivery via hoods or BIBS. Operation: manual electropneumatic pressurization/depressurization; venturi-assisted overboard exhaust; whisper-quiet ventilation. Output: controlled hyperbaric environment for medical treatment. Used in clinical settings by physicians, nurses, and technicians. Features include rectangular design, walk-through sliding doors, and VDU display for monitoring pressure, temperature, humidity, O2, and CO2. Benefits: treats specific medical conditions (e.g., decompression sickness, wounds, burns) while reducing patient anxiety through hospital-room-like aesthetics.
Clinical Evidence
No clinical data provided. Substantial equivalence is based on design, intended use, and adherence to established safety standards for hyperbaric facilities.
Technological Characteristics
Rectangular steel pressure vessels fabricated per ASME Section VIII Div 1 and ASME PVHO-1. Operating pressure 3.0-6.0 ATA. Manual electropneumatic life support controls. Features: venturi-assisted overboard exhaust, pneumatic ventilation, digital depth measurement with analog backup, O2/CO2 gas analysis. Connectivity: internal/external PA system, sound-powered phone backup. Fire suppression per NFPA 99.
Indications for Use
Indicated for patients with air or gas embolism, carbon monoxide poisoning (with or without cyanide poisoning), clostridial myositis/myonecrosis, crush injury/compartment syndrome/acute traumatic ischemias, decompression sickness, problem wounds, exceptional anemia, necrotizing soft tissue infections, refractory osteomyelitis, delayed radiation injury, compromised skin grafts/flaps, thermal burns, and intracranial abscess.
Regulatory Classification
Identification
A hyperbaric chamber is a device that is intended to increase the environmental oxygen pressure to promote the movement of oxygen from the environment to a patient's tissue by means of pressurization that is greater than atmospheric pressure. This device does not include topical oxygen chambers for extremities (§ 878.5650).
Predicate Devices
Perry Sigma MP Multiplace Hyperbaric Chamber (K930748)
{0}------------------------------------------------
K0316449
DEC - 5 2003
# 510(k) NOTIFICATION SUMMARY (Per 21 CFR 807.92)
Prepared: 26 May 2003
TRADE NAME: SL8 Hyperbaric Oxygen Treatment Facility DL8 Hyperbaric Oxygen Treatment Facility TL20 Hyperbaric Oxygen Treatment Facility
COMMON NAME OF DEVICE: Multiplace Hyperbaric Chamber
CLASSIFICATION: 73 CBF, 21 CFR 868.5470
ESTABLISHMENT REGISTRATION NUMBER: Pending
### CLAIMED PREDICATE DEVICE(S):
Perry Sigma MP Multiplace Hyperbaric Chamber - K930748 Gulf Coast Multiplace Hyperbaric Chamber - K950957 Reimers T Class Hyperbaric Chamber - K954387
ADDRESS OF MANUFACTURER:
Fink Engineering PTY LTD 100 Talinga Road Cheltenham Victoria 3192 Australia Tele: 03-9585-2966
CONTACT PERSON: Eric Fink, Managing Director
#### EXECUTIVE SUMMARY
The Undersea and Hyperbaric Medical Society (UHMS) defines hyperbaric oxygen therapy The Ondersea and Hyperbart lessures higher than atmospheric in a hyperbaric chamber. as breating 100% offyger Fire Protection Association (NFPA), hyperbaric chambers are According to the National Fire Fress A (multi-occupant) and Class B (single occupant). The Classined into two outsgen Treatment Facilities are Class A multiplace hyperbaric chambers designed to treat up to 8 (SL/DL8) or 20 (TL20) patient at up to a maximum operating pressure of 3 Atmospheres Absolute (ATA) or 29.4 pounds per square maximum operating procears of 0 %... 8 %... 6 %. or 73.5 psig for the DL8 and TL20. Each incil gauge (psig) for the OLD and up tessurization gas and 100% oxygen as the hyperbaric treatment gas.
The SL/DL8 and TL20 Hyperbaric Oxygen Treatment Facilities are intended to be procured The UEDEO and TEE Fry tribate a variety of medical conditions that respond to hyperbaric and used by pryololane to trout itst of medical conditions that have been identified for the appropriate primary or adjunctive use of hyperbaric oxygen. These approved conditions
{1}------------------------------------------------
include: air or gas embolism; carbon monoxide poisoning and smoke inhalation; clostridial include. and of gas embolism, oalbert may and partment syndrome and other acute traumatic ischemias; decompression sickness; enhanced healing of selected problem wounds; lightinas, decompreasion visime rotizing soft tissue infections; osteomyelitis (refractory); radiation tissue damage (osteoradionecrosis); compromised skin flaps and grafts; thermal radiation though abscess. Aggressive research into the beneficial effects of burns, and, measure when appropriately applied, will result in additional medical conditions being added to the list of indications by the UHMS.
The SL/DL8 and TL20 Hyperbaric Oxygen Treatment Facilities are designed and fabricated in accordance with the requirements of the ANSI/ASME Boiler and Pressure Vessel Code, in abool admod will, Division 1, Pressure Vessels; ANSI/ASME-PVHO-1 (American Society of Scoton Vill, Engineers-Pressure Vessels for Human Occupancy); and, NFPA 99, Health Care Facilities, Chapter 19 (Chapter 20, 2002 Edition), Hyperbaric Facilities.
The SL/DL8/TL20 rectangular hyperbaric chamber(s) include as a key component a new and very user-friendly rectangular chamber designed and manufactured specifically for the medical community to be used for hyperbaric oxygen treatment. The design of this unique ASME PVHO chamber has been driven by comments from hyperbaric physicians, technicians and nurses who require their equipment to be simple, easy to use and to simulate toomlount and harded inical conditions found elsewhere in their working environment.
The SL/DL8/TL20 facilities place major emphasis on patient comfort and user-friendly operator controls and incorporate several innovative and unique features including:
- A large, comfortable rectangular hyperbaric chamber that has been outfitted t . appear like any other clinical room in a hospital to reduce patient anxiety
- A rectangular medical lock with a unique locking system enabling single-handed . operation
- Large walk-trough rectangular sliding doors that fit flush with the floor so that . patients can be wheeled into the chamber without bumping over a door jam
- An easy to use control panel with large VDU display to monitor the treatment . profile including the internal chamber pressure, temperature, humidity, oxygen and carbon dioxide parameters
- An oxygen delivery system with venturi assisted overboard exhaust to eliminate . the possibility of "shrink wrapping" patients with their oxygen hoods
- Whisper quiet ventilation system to pneumatically control temperature, humidity . and upper oxygen limit without the need for internal electrical blowers
- Aesthetically pleasing external chamber cladding .
Specifications of the SL8 Hyperbaric Oxygen Treatment Facility are as follows:
| Design Code | ASME Section VIII, Div 1 and ASME PVHO-1 |
|-----------------------|------------------------------------------|
| Operating Pressure | 3.0 ATA |
| Operating Temperature | 62°F to 100°F |
| Design Pressure | 32.3 psi |
| Design Temperature | 62°F to 100°F |
| Design Life | 70,000 cycles (30 years) |
| Hydrostatic Pressure | 42.0 psi |
| Inspection Authority | ASME "U" Stamp |
| Weight | ~15,432 lbs |
| Dimensions | 8.1ftW X 7ftH X 10.8 ft L |
| Volume | 586 ft³ |
| Medical Lock | 13.8 in X 13.8 in X 19.7 in |
| Doorway Size | 32.2 in X 75.6 in |
{2}------------------------------------------------
| Lighting<br>13 in ID viewports<br>5.9 in ID viewport<br>Capacity<br>Fire Suppression<br>Finish<br>Life Support Controls | Four external dimmable lights<br>Two PVHO<br>One PVHO<br>Eight seated persons<br>IAW NFPA 99<br>Shotblasted and Painted<br>Manual electropneumatic pressurization<br>Manual electropneumatic depressurization |
|------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Environmental Control<br>Ventilation<br>BIBS with overboard dump<br>Hoods with overboard dump<br>Depth Measurement<br>Gas Analysis<br>Communications | Heating<br>Constant airflow<br>Four<br>Four<br>Digital with analog backup<br>Oxygen and carbon dioxide<br>Internal/external PA System<br>Sound powered phone backup |
| Entertainment<br>TV System | Individual four-channel selection for eight persons<br>External color with remote control<br>AM/FM Tuner/CD & DVD Player |
Specifications of the DL8 Hyperbaric Oxygen Treatment Facility are as follows:
| Design Code | ASME Section VIII, Div 1 and ASME PVHO-1 |
|---------------------------|------------------------------------------------------------------------------------|
| Operating Pressure | 6.0 ATA |
| Operating Temperature | 62°F to 100°F |
| Design Pressure | 80.0 psi |
| Design Temperature | 62°F to 100°F |
| Design Life | 70,000 cycles (30 years) |
| Hydrostatic Pressure | 104.0 psi |
| Inspection Authority | ASME "U" Stamp |
| Weight | ~26,400 lbs |
| Main Lock Dimensions | 8.1ftW X 7ftH X 11ftL |
| Entrance Lock Dimensions | 8.1ftW X 7ftH X 4.5ftL |
| Volume (ML) | 614.5 ft3 |
| Volume (EL) | 254.3 ft3 |
| Total Volume | 868.7 ft3 |
| Medical Lock | 13.8 in X 13.8 in X 19.7 in |
| Main Doorway Size | 39.4 in X 75.6 in |
| Lighting | Six external dimmable lights |
| 13.8 in ID viewports | Two PVHO in ML |
| 5.9 in ID viewports | One PVHO in each lock |
| Main Lock Capacity | Eight seated persons<br>Four wheelchairs<br>One hospital gurney |
| Fire Suppression | IAW NFPA 99 |
| Finish | Shotblasted and Painted |
| Life Support Controls | Manual electropneumatic pressurization<br>Manual electropneumatic depressurization |
| Environmental Control | Heating |
| Ventilation | Constant airflow |
| BIBS with overboard dump | Two |
| Hoods with overboard dump | Six |
| Depth Measurement | Digital with analog backup |
| Gas Analysis | Oxygen and carbon dioxide |
| Communications | Internal/external PA System<br>Sound powered phone backup |
{3}------------------------------------------------
| Entertainment | Individual four-channel selection for eight persons |
|--------------------|-----------------------------------------------------|
| TV System | External color with remote control |
| Compartment Relief | AM/FM Tuner/CD & DVD Player |
| Compartment Drain | ASME certified relief valve |
| | One manual drain valve in each lock |
Specifications of the TL20 Hyperbaric Oxygen Treatment Facility are as follows:
ASME Section VIII, Div 1 and ASME PVHO-1 Design Code Operating Pressure 6.0 ATA 62°F to 100°F Operating Temperature 80 psi Design Pressure 62°F to 100°F Design Temperature 70,000 cycles (30 years) Design Life 104.0 psi Hydrostatic Pressure ASME "U" Stamp Inspection Authority ~110,231 lbs Weight 10.3 ftW X 7ftH X 19.3 ftL Main Lock Dimensions 7.8 ftW X 7 ftH X 10.8ftL Entrance Lock Dimensions 7/8 ftW X 7 ftH X 10.8 ftL Inner Lock Dimensions 1536.2 ft3 Volume (ML) 529.7 ft3
480.3 ft3 Volume (EL) Volume (IL) 2546.2 H³ Total Volume 13.8 in X 13.8 in X 19.7 in Medical Lock 39.4 in X 75.6 in Main Doorway Size Fourteen external dimmable lights Lighting 23.6 in ID viewports Four PVHO Three PVHO 13 in ID viewports Two PVHO 6 in ID viewports Up to twenty persons Capacity Fire Suppression IAW NFPA 99 Shotblasted and Painted Finish Manual electropneumatic pressurization Life Support Controls Manual electropneumatic depressurization Environmental Control Heating Constant airflow Ventilation BIBS with overboard dump Four Twenty-two Hoods with overboard dump Digital with analog backup Depth Measurement Oxygen and carbon dioxide Gas Analysis Internal/external PA System Communications Sound powered phone backup Individual four-channel selection for eight persons Entertainment External color with remote control TV System AM/FM Tuner/CD & DVD Player
Fink Engineering has concluded that the general design approach, method of pressure control, and intended use of the SL&/DL8 and TL 20 Hyperbaric Oxygen Treatment Facilities are substantially equivalent to the Perry Baromedical Services Sigma MP Multiplace are ouberantially oque (KS30748), the Gulf Coast Multiplace Hyperbaric Chamber (K950957) and the Reimers Engineering T Class Multiplace Hyperbaric Chamber (K954387) and is proposing them as predicate devices for the SL8/DL8 and TL20 Hyperbaric Oxygen Treatment Facilities.
{4}------------------------------------------------
#### Intended Use:
It is the expressed, intended use of the Fink Engineering SL8/DL8 and TL20 Hyperbaric Oxygen Treatment Facilities to provide therapy to those patients with selected medical Oxygen Trouthons that have been determined to respond to the application of hyperbaric oxygen. As a Class II prescriptive device, it is further intended for physician involvement in their procurement and routine use.
The UHMS is the professional medical organization chartered with setting the standards of care defining the appropriate use of hyperbaric oxygen. More specifically, the UHMS our cloning the appropriations that have been clearly established as appropriate primary or adjunctive use of hyperbaric oxygen (HBO). The disorders on the list have been primary of dajaridated and verified through extensive data collection. It should be noted that the list is dynamic. Based on the strength of the scientific data, disorders are both added and removed from the list, depending on the outcomes of scientific pursuit.
The conditions listed as appropriate for the use of HBO in the current edition of the Hyperbaric Oxygen Therapy Committee Report (1999) are as follows:
- Air or gas embolism 1.
- Carbon monoxide poisoning and carbon monoxide poisoning complicated by 2. cvanide poisoning
- Clostridial myositis and myonecrosis ന്
- Crush injury, compartment syndrome, and other acute traumatic ischemias 4.
- Decompression sickness ട്.
- Enhanced of selected problem wounds .
- Exceptional blood loss anemia 7.
- Necrotizing soft tissue infections 8.
- Osteomyelitis (refractory) ഗ്
- 10. Delayed radiation injury (soft tissue and bony necrosis)
- 11. Skin grafts and flaps (compromised)
- 12. Thermal burns
- 13. Intracranial abscess
{5}------------------------------------------------
Image /page/5/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized graphic of an eagle or bird-like symbol with three curved lines representing its wings or body.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
DEC - 5 2003
Fink Engineering Pty. Ltd. c/o Mr. W. T. Workman Workman Hyperbaric Services. Inc. 18111 Copper Ridge Dr. San Antonio, TX 78259
Re: K031649
Trade/Device Name: Multiplace Hyperbaric Oxygen Treatment Chamber, Models SL8/DL8/TL20 Regulation Number: 868.5470 Regulation Name: Hyperbaric Chamber Regulatory Class: II Product Code: CBF Dated: September 30, 2003 Received: October 1, 2003
Dear Mr. Workman:
. . . .
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{6}------------------------------------------------
Page 2 - Mr. W. T. Workman
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (301) 594-4646. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html
Sincerely yours,
Susan Runnd
Chiu Lin, Ph.D. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{7}------------------------------------------------
## Indications for Use
510(k) Number (if known): K031649
Device Name: Fink Engineering PTY LTD, SL8/DL8 and TL20 Hyperbaric Treatment Facilities
Indications For Use:
- 1. Air or gas embolism
- 2. Carbon monoxide poisoning and carbon monoxide poisoning complicated by cyanide poisoning
- 3. Clostridial myositis and myonecrosis
- 4. Crush injury, compartment syndrome, and other acute traumatic ischemias
- 5. Decompression sickness
- 6. Enhanced healing of selected problem wounds
- 7. Exceptional anemia
- 8. Necrotizing soft tissue infections
- 9. Osteomyelitis (refractory)
- 10. Delayed radiation injury (soft tissue and bony necrosis)
- 11. Skin grafts and flaps (compromised)
- 12. Thermal burns
- 13. Intracranial abscess
X Prescription Use _ (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IN NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
2. Intochu
Page 1 of
**510(k) Number:** k031649
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.