K190304 · Qingdao Kingon Medical Science and Technology Co. · CAW · Feb 29, 2020 · Anesthesiology
Device Facts
Record ID
K190304
Device Name
Kingon P2 Oxygen Concentrator
Applicant
Qingdao Kingon Medical Science and Technology Co.
Product Code
CAW · Anesthesiology
Decision Date
Feb 29, 2020
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 868.5440
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Kingon P2 is for prescription use by patients requiring high concentrations of oxygen on a supplemental basis. It is small, portable, and is capable of continuous use in the home, institutional, and travel / mobile environments.
Device Story
Kingon P2 is a portable oxygen concentrator for patients requiring supplemental oxygen. Device uses molecular sieve and pressure swing adsorption to extract oxygen from ambient air; nitrogen is adsorbed while oxygen is collected and delivered via nasal cannula. Device features pressure-sensing technology to trigger oxygen pulse delivery upon patient inhalation; regenerates oxygen pulse between breaths. Operates in home, institutional, and mobile environments; intended for prescription use. Output is pulse of oxygen (87-96% purity). Healthcare providers use device to provide supplemental oxygen therapy, potentially improving patient mobility and oxygenation. Device includes carry bag, AC power supply, and battery.
Clinical Evidence
Bench testing only. Device passed testing for electrical safety (IEC 60601-1, IEC 60601-1-11), oxygen concentrator performance (ISO 80601-2-69), oxygen-conserving equipment (ISO 80601-2-67), alarm systems (IEC 60601-1-8), EMC (IEC 60601-1-2, FCC Part 15), and battery safety (IEC 62133). Biocompatibility testing (ISO 10993) and gas pathway emissions testing (ISO 18562) also passed.
Technological Characteristics
Portable oxygen concentrator using pressure swing adsorption with molecular sieve. Dimensions: 6.30"H x 3.35"W x 8.70"L; Weight: 4.34 lbs. Power: 100-240 VAC, 50/60 Hz or 19 VDC. Connectivity: Standalone. Standards: IEC 60601-1, IEC 60601-1-11, ISO 80601-2-69, ISO 80601-2-67, IEC 60601-1-8, ISO 10993, ISO 18562.
Indications for Use
Indicated for adult patients requiring supplemental high-concentration oxygen. Suitable for home, institutional, and mobile/travel use.
Regulatory Classification
Identification
A portable oxygen generator is a device that is intended to release oxygen for respiratory therapy by means of either a chemical reaction or physical means (e.g., a molecular sieve).
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
Qingdao Kingon Medical Science and Technology Co. % Roman Huang General Manager Elliot Medical Solutions Parkland Drive Cleveland Heights, Ohio 44106
Re: K190304
Trade/Device Name: Kingon Portable Oxygen Concentrator P2 Regulation Number: 21 CFR 868.5440 Regulation Name: Portable Oxygen Generator Regulatory Class: Class II Product Code: CAW Dated: January 15, 2020 Received: January 24, 2020
### Dear Roman Huang:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part
{1}------------------------------------------------
801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Todd Courtney Assistant Director DHT1C: Division of ENT, Sleep Disordered Breathing, Respiratory and Anesthesia Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
Form Approved: OMB No. 0910-0120
Expiration Date: 06/30/2020
See PRA Statement below.
#### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
### Indications for Use
..............................................................................................................................................................................
网_皇冠 _ba
510(k) Number (if known)
#### K190304
#### Device Name
Kingon Portable Oxygen Concentrator P2
Indications for Use (Describe)
The Kingon P2 is for prescription use by patients requiring high concentrations of oxygen on a supplemental basis. It is small, portable, and is capable of continuous use in the home, institutional, and travel / mobile environments.
.
是 在
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|--|
|-------------------------------------------------|--|
X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# Section 4
# 510(k) Summary
# 1. Company making the submission:
Company Name - Qingdao Kingon Medical Science and Technology Co., Ltd Company Address - 24th East Building, No. 252 Yanhe Road, Tianhe Industrial Park, Huangdao, Qingdao, Shangdong, China Tel: 86-0532-58792324 Contact - Wu Xiao CEO and Chief Manager
# 2. U.S. Correspondent and Contact
Name: Elliot Medical Solutions Address: Cleveland Heights, Cleveland, Ohio Contact: Roman Huang General Manager Tel: 216-262-0962 Email: support@elliotmd.com
## 3. Submitted device
Device Name: Portable Oxygen Concentrator Trade name: Kingon Portable Oxygen Concentrator Models: P2 Common Name: Oxygen Concentrator Classification Name: Oxygen Concentrator, Portable Regulation Number: 868.5440 Regulatory Class: II Product Code: CAW Type of 510(k) submission: Traditional
- 4. Last Edited Date
February 28, 2020
# 5. Predicate Device.
| Manufacturer | Predicate<br>Device | 510(k)<br>number |
|--------------------------------------------------------------|---------------------------------------------|------------------|
| Philips<br>Respironics<br>SimplyGo<br>Oxygen<br>Concentrator | SimplyGo<br>Portable<br>Oxygen<br>Generator | K111885 |
{4}------------------------------------------------
### 6. Description.
The Kingon P2 oxygen concentrator includes the following items:
- One Kingon P2 oxygen concentrator ●
- One Carry bag ●
- One Nasal cannula ●
- One AC power supply ●
- Five Intake filters
- One Battery ●
- . One User manual
The Kingon P2 oxygen concentrator utilizes a molecular sieve and differential pressure swing adsorption to separate the gases in ambient air. In brief, after the device takes the room air, the molecular sieve packed in a sealed container can absorb nitrogen in the air, and the oxygen still exists in gaseous form that can be collected by specially designed pipelines and deliver to patients. When the environmental air pressure decreases, exhaust of the vacuum container occurs and nitrogen will be released from the molecular sieve. When the patient inhales, the device senses the pressure change and is triggered to release the oxygen pulse. In between breaths, the device regenerates an oxygen pulse and waits for the next inhalation breath before dispensing it. In this way, Kingon P2 device can concentrate the oxygen in the air to produce a pulse of oxygen between 87-96% in purity.
## 7. Indication for use.
The Kingon P2 is for prescription use by patients requiring high concentrations of oxygen on a supplemental basis. It is small, portable, and is capable of continuous use in the home, institutional, and travel / mobile environments.
| Element of<br>Comparison | Predicate Device | Subject Device | Comparison |
|-------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------|------------|
| Device name | Philips Respironics SimplyGo<br>Portable Oxygen Concentrator | Kingon P2 Portable<br>Oxygen Concentrator | |
| 510(k) | K111885 | N/A | |
| Indication for Use | The Portable Oxygen<br>Concentrator is for prescription<br>use by patients requiring high<br>concentrations of oxygen on a<br>supplemental basis. It is small,<br>portable and is capable of<br>continuous use in the home, | Same | Identical |
| | institutional, and travel/mobile<br>environments. | | |
| Environment of Use | Home, institutional, and<br>travel/mobile environments. | Same | Identical |
| Patient Population | Adult | Same | Identical |
| Single Patient, multi-<br>use | Yes | Yes | Identical |
| Patient Interface | Standard Single Lumen Nasal<br>Cannula | same | Identical |
| Technology | Pressure Swing Adsorption with<br>molecular sieve | same | Identical |
| Dimensions | 3.6"H*8.3"w*9.4(Standard<br>Battery)<br>3.6"H*8.3"W*10.27"L(Extend<br>Battery) | 6.30"H*3.35"W*8.70"L | Similar |
| Weight | 5.0 lbs (with standard battery<br>installed)<br>6.0 lbs (with extended battery<br>installed) | 4.34lbs (±0.07Ibs with<br>standard battery) | Similar |
| Oxygen Concentration | At least 87% at all settings<br>(maximum of 96%) over<br>the rated environmental range. | Same | Similar |
| Equivalent Flow rates | 15-40 BMP increments of<br>5BPM. | 10-40BMP increments<br>of 5BPM. | Similar |
| Dose at Specified<br>Flow | 11mL per setting | 10.5mL per setting<br>(with 20BPM) | Identical |
| Filters | Input Filter; Patient Filter | same | Identical |
| User Interface | Buttons LCD Display | same | Identical |
| Electrical | 100-240 VAC; 50/60 Hz<br>19 VDC, 6.3 A | 100-240VAC ;50-60<br>Hz;<br>19 VDC + 5% 6A MAX | Similar |
| Acoustic Noise | 42 dBA typical at setting 2 and<br>20 BPM<br>48 dBA typical at setting 5 and<br>20 BPM<br>(when measured at 1 meter from<br>front of the device) | 49 dBA typical at<br>setting 2 and 20BPM<br>56 dBA typical at<br>setting 5 and 20 BPM<br>(when measured at 1<br>meter from front of<br>device) | Similar |
| Alarms | High Breath Rate Alarm | same | Similar |
| | Low Oxygen Concentration<br>Alarm | same | Similar |
| | Technical Fault Alarm | same | Similar |
| | Low Battery Alarm | same | Similar |
| Warm Up Indicator | same | | |
| No Flow Alarm | same | | |
| External Power Failure Alarm | same | | |
| Depleted Battery Alarm | same | | |
| Status Indicator | Tool icon | same | Similar |
| | Flow Setting | same | |
| | Battery Charge Level | same | |
| | Alarm Silence Symbol | same | |
| | Attention | same | |
| | Flow Control Setting | same | |
| Battery Duration | Up to 4.5 hours (Pulse setting of 2 at 20 BPM with Standard battery)<br>Up to 9 hours (Pulse setting of 2 at 20 BPM with Extended battery) | Up to 3.8 hours (Pulse setting of 1 at 20 BPM with Standard battery) | Similar |
| Operating Environment | Temperature: 41° F to 95° F (5° C to 35° C)<br>Relative humidity: 15% to 93%<br>Atmospheric Pressure: 700 hPa to 1010 hPa<br>Altitude: up to 10,000 ft (3048 m) | Temperature: 41 to 104°F (5 to 40°C)<br>Humidity: 10% to 90%, non-condensing<br>Altitude: 0 to 10,000 ft. (0 to 3048 meters) | Similar |
| Shipping Storage | -4° F to 140° F (-20° C to 60° C)<br>Relative humidity: up to 93%, non-condensing | Temperature: -4 to 158°F (-20 to 70°C)<br>Humidity: 5% to 90%, non-condensing<br>Store in a dry environment | Similar |
## 8. Similarities/Differences between the subject and predicate device.
{5}------------------------------------------------
{6}------------------------------------------------
## 9. Comparison of Technological Characteristics with Predicate Device.
Both the Kingon P2 oxygen concentrator and the predicate device (SimplyGo) have the same indication of use. Both of them have incorporated the same basic design and the same technological characteristics, and utilize the same operating principle. Both of them have been tested to the same electrical and electromagnetic safety standards for medical electrical equipment. And both of them are manufactured under a quality system. The differences between the subject device and predicate such as size, storage condition, operating condition, battery duration, alarm setting and panel indicator introduce risks mitigated by the performance testing provided in this submission.
#### 10. Safety and Performance Data, Biocompatibility Data
{7}------------------------------------------------
| Bench Tests | Standards | Results | Report File<br>No. |
|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------|---------|----------------------|
| General requirements for basic safety and<br>essential performance | IEC 60601-1 | Pass | GZME18010<br>0001801 |
| General requirements for basic safety and<br>essential performance - Collateral Standard:<br>Requirements for medical electrical<br>equipment and medical electrical systems<br>used in the home healthcare environment | IEC 60601-1-11 | Pass | GZME18010<br>0001802 |
| Medical electrical equipment part 2:<br>particular requirements for the basic safety<br>and essential performance of oxygen<br>concentrator equipment | ISO 80601-2-69 | Pass | GZME18010<br>0001803 |
| ISO 80601-2-67: medical electrical<br>equipment, part 2-67: particular requirements<br>for basic safety and essential performance of<br>oxygen-conserving equipment | ISO 80601-2-67 | Pass | GZME18010<br>0001804 |
| IEC 60601-1-8: Medical electrical<br>equipment, General requirements for basic<br>safety and essential performance - collateral<br>standard: General requirements, tests and<br>guidance for alarm systems in medical<br>electrical equipment and medical electrical<br>system | IEC 60601-1-8 | Pass | GZME18010<br>0001805 |
| EMC TEST | IEC 60601-1-2 | Pass | GZME18010<br>0001901 |
| Radiated & Conducted emissions test (FCC) | 47 CFR Part 15,<br>subpart B | Pass | GZME18010<br>0001902 |
| Secondary cells and batteries containing<br>alkaline or other non-acid electrolytes -<br>safety requirements for portable sealed<br>secondary cells, and for batteries made from<br>them, for use in portable applications | IEC 62133,<br>ST/SG/AC.10/11/<br>Rev.6/Section<br>38.3 | Pass | SHES180500<br>540701 |
| Electromagnetic Immunity from RFID reader | AIM standard<br>7351731 | Pass | 1909ESU013<br>-U1 |
Safety and performance data of the Kingon P2 device are listed in Table as below.
Biocompatibility data are listed in Table as below:
{8}------------------------------------------------
| Biocompatibility<br>Tests | Standards | Results | Report File name<br>and Number |
|------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------|---------|--------------------------------|
| In vitro<br>cytotoxicity test of<br>shell and<br>accessories | ISO 10993-5:2009 Test<br>Method MTT Method<br>MEM with 10% FBS<br>extract | Pass | SDWH-<br>M201800623-1 |
| Skin sensitization<br>test of shell and<br>accessories | ISO 10993-10: 2010 Test<br>Methods Guinea Pig<br>Maximization Test 0.9%<br>Sodium Chloride Injection<br>Extract | Pass | SDWH-<br>M201800623-2 |
| Skin sensitization<br>test of shell and<br>accessories | ISO 10993-10: 2010 Test<br>Methods Guinea Pig<br>Maximization Test Sesame<br>oil extract | Pass | SDWH-<br>M201800623-3 |
| Skin irritation test<br>of shell and<br>accessories | ISO 10993-10: 2010 Test<br>Methods 0.9% Sodium<br>Chloride Injection Extract | Pass | SDWH-<br>M201800623-4 |
| Skin irritation test<br>of shell and<br>accessories | ISO 10993-10: 2010 Test<br>Methods Sesame oil<br>extract | Pass | SDWH-<br>M201800623-5 |
| Tests | Standards | Results | Report File name<br>and Number |
| Emission of<br>VOCs and<br>aldehydes | ISO 18562-3 | Pass | 18914-N01 |
| Emissions of<br>particulate matter.<br>carbon dioxide,<br>carbon monoxide<br>and ozone | ISO 18562-2 | Pass | 18914-N02 |
| Biological<br>evaluation of<br>medical devices -<br>part 17:<br>establishment of<br>allowable limits | ISO 10993-17<br>ISO 18562-1 | Pass | 18914-N03 |
{9}------------------------------------------------
| for leachable<br>substances | | |
|-----------------------------|--|--|
| | | |
#### 10. Conclusion.
The Kingon P2 oxygen concentrator has similar intended use, principle of operation, and similar technological characteristics as the predicate device identified. Performance testing contained in this submission demonstrates the minor differences in technological characteristics between the subject device and the predicate do not raise different questions of safety and effectiveness. Thus, in accordance with the Federal Food, Drug and Cosmetic Act, 21 CFR Part807 and based on the information provided in this premarket notification, we conclude that Kingon P2 oxygen concentrator is substantially equivalent to predicate device.
End.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.