INOGEN AT HOME OXYGEN CONCENTRATOR (AKA INOGEN AT HOME)
K132489 · Inogen, Inc. · CAW · Jun 20, 2014 · Anesthesiology
Device Facts
Record ID
K132489
Device Name
INOGEN AT HOME OXYGEN CONCENTRATOR (AKA INOGEN AT HOME)
Applicant
Inogen, Inc.
Product Code
CAW · Anesthesiology
Decision Date
Jun 20, 2014
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 868.5440
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Inogen At Home Oxygen Concentrator is used on a prescriptive basis by patients requiring supplemental oxygen. It supplies a high concentration of oxygen and is used with a nasal cannula to channel oxygen from the concentrator to the patient. The Inogen At Home Oxygen Concentrator may be used in a home or institution.
Device Story
Stationary oxygen concentrator; utilizes Pressure Swing Adsorption (PSA) to separate oxygen from ambient air; molecular sieve adsorbs nitrogen under pressure; enriched oxygen collected in accumulator; nitrogen exhausted during low-pressure cycle. System includes sieve beds, manifold, precision valves, sensors, and embedded software. Powered by AC mains. Used in home or institutional settings; operated by patients or caregivers. Provides continuous oxygen flow (1.0–5.0 LPM). Healthcare providers use output to manage supplemental oxygen therapy; benefits patients by providing high-concentration oxygen for respiratory support.
Clinical Evidence
Bench testing only. Conformance to EN ISO 8359:2012, ASTM F 1464:2005, and IEC 60601-1. Biocompatibility testing (particulate matter, Ozone, VOCs, CO2, CO). Performance testing included electrical safety, software functionality, fire mitigation, noise, environmental ranges (temperature/altitude), outlet pressure, flow indicator accuracy, and oxygen concentration.
Technological Characteristics
Stationary oxygen concentrator using Pressure Swing Adsorption (PSA). Powered by AC mains. Continuous flow 1.0-5.0 LPM. Includes sieve beds, manifold, valves, sensors, and embedded software. Conforms to EN ISO 8359:2012, ASTM F 1464:2005, and IEC 60601-1.
Indications for Use
Indicated for patients requiring supplemental oxygen on a prescriptive basis. Suitable for use in home or institutional settings. Administered via nasal cannula.
Regulatory Classification
Identification
A portable oxygen generator is a device that is intended to release oxygen for respiratory therapy by means of either a chemical reaction or physical means (e.g., a molecular sieve).
Inogen, Inc., Inogen One Oxygen Concentrator (K032818)
Submission Summary (Full Text)
{0}------------------------------------------------
# JUN 2 0 2014
## 5. 510(k) Summary
| 510(k) Owner/ Submitter: | Inogen, Inc.<br>326 Bollay Drive<br>Goleta, California 93117<br>P: (805) 562-0500<br>F: (805) 562-0516 |
|-------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------|
| Contact Person: | Mara Korsunsky<br>Inogen, Inc.<br>Director, Quality, Regulatory, and Compliance<br>P: (805) 562-0629<br>F: (805) 562-0516<br>E: mkorsunsky@inogen.net |
| Date Prepared: | August 5, 2013 |
| Trade Name: | Inogen At Home Oxygen Concentrator (AKA Inogen At Home/ Model GS-100) |
| Device Type:<br>(Common Name) | Oxygen Concentrator |
| Classification Name: | Portable Oxygen Generator |
| Classification Regulation: | 868.5440 |
| Class: | II |
| Panel: | Anesthesiology |
| Product Code: | CAW |
| Predicate Device(s): | Respironics, Inc., Respironics L4 Oxygen Concentrator (EverFlo),<br>K061261 |
| | Inogen, Inc., Inogen One Oxygen Concentrator, K032818 |
### Device Description:
The Inogen At Home is a stationary concentrator providing approximately 90% oxygen to patients on a continuous flow basis at a rate of 1.0 liter per minute (LPM) to 5.0 liters per minute (LPM), in 1.0 liter per minute (LPM) increments. Nasal cannula channels oxygen from device to patient; both concentrator
{1}------------------------------------------------
and nasal cannula are non-sterile. The device is capable of continuous use in the home or in an institutional setting.
The Inogen At Home can be powered by mains AC power found in the home or in an institutional setting. The Inogen At Home includes a detachable AC power cord to make the electrical connection to the AC mains.
The Inogen At Home utilizes Pressure Swing Adsorption (PSA) technology to produce oxygen. Oxygen is separated from ambient air under pressure by molecular characteristics and affinity for adsorbent material. The molecular sieve preferentially adsorbs nitrogen, allowing enriched oxygen to collect in an accumulator; the process then swings to low pressure to desorb the nitrogen, exhausting it back into the environment. A series of sieve beds, manifold with precision valves, sensors, and embedded software are used collectively to govern system functionality.
#### Intended Use/ Indications for Use:
The Inogen At Home Oxygen Concentrator is used on a prescriptive basis by patients requiring supplemental oxygen. It supplies a high concentration of oxygen and is used with a nasal cannula to channel oxygen from the concentrator to the patient. The Inogen At Home Oxygen Concentrator may be used in a home or institution.
#### Technological Characteristics:
The Inogen At Home has a similar indications statement and technological characteristics (materials, design, and energy source) as the predicate devices. It raises neither new nor different questions of safety and effectiveness compared to the predicate devices; it is substantially equivalent.
#### Non-clinical Performance Data:
The Inogen At Home conforms with EN ISO 8359:2012, Oxygen Concentrators for Medical Use - Safety Requirements (ISO 8359:1996, Second edition, Amendment 1), ASTM F 1464:2005, Oxygen Concentrators for Domiciliary Use, IEC 60601-1, Medical Electrical Equipment -- Part 1: General requirements for safety, 1988; Amendment1, 1991-11, Amendment2, 1995.
Biocompatability testing included the evaluation of particulate matter, Ozone, Volatile Organic Compounds, Carbon Dioxide, and Carbon Monoxide. Performance testing included the assessment of electrical safety/ compatability, software functionality, fire mitigation and noise, temperature and altitude ranges, outlet pressure, flow indicator accuracy, and oxygen concentration.
This supports a determination of substantial equivalence.
#### Conclusion:
The Inogen At Home is substantially equivalent to the predicate devices.
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with its wings spread, and three bars extending from the eagle's body. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the eagle.
#### Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
June 20, 2014
Inogen, Inc. Mara Korsunsky Director, Quality, Regulatory, and Compliance 326 Bollay Drive Goleta, California 93117
Re: K132489
> Trade/Device Name: Inogen At Home Oxygen Concentrator Regulation Number: 21 CFR 868.5440 Regulation Name: Portable oxygen generator Regulatory Class: II Product Code: CAW Dated: May 21, 2014 Received: May 22, 2014
Dear Ms. Korsunsky:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{3}------------------------------------------------
Page 2 - Ms. Korsunsky
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers. International and Consumer Assistance at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely vours.
Tejashri Purohit-Sheth, M.D. Tejashri Purofist-Sheth, M.D. Clinical Deputy Director DAGRID/ODE/CDRH FOR
Erin I. Keith, M.S. Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
### Indications for Use
510(k) Number (if known): ≤132489
Device Name: Inogen At Home Oxygen Concentrator
Indications for Use:
The Inogen At Home Oxygen Concentrator is used on a prescriptive basis by patients requiring supplemental oxygen. It supplies a high concentration of oxygen and is used with a nasal cannula to channel oxygen from the concentrator to the patient. The Inogen At Home Oxygen Concentrator may be used in a home or institution.
× Prescription Use (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use
(21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
:
・・
Concurrence of Center for Devices and Radiological Health (CDRH)
Anya C. Harry -S 2014.06.19 13:43:53 -04'00'
Page 1 of
Page 15 of 184
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.